TRANSTEK TMB-1776-BZY Blood Pressure Monitor User Manual

June 17, 2024
TRANSTEK

TRANSTEK TMB-1776-BZY Blood Pressure Monitor

Product Information: Blood Pressure Monitor Model TMB-1776-BZY

Specifications:

  • Model: TMB-1776-BZY
  • Manufacturer: Guangdong Transtek Medical Electronics Co.,Ltd.
  • Address: Zone A, No.105, Dongli Road, Torch Development District, 528437 Zhongshan, Guangdong, China
  • Measurement Method: Oscillometric Measuring
  • Features: Blood pressure measurement, pulse rate measurement, result storage
  • Warranty: 2 years

User Manual
INTRODUCTION
The TRANSTEK Blood Pressure Monitor TMB-1776-BZY is a device that allows you to measure your blood pressure and pulse rate. This manual provides important safety information and step-by-step instructions for using the product.

General Description
The TMB-1776-BZY blood pressure monitor features accurate readings equivalent to those obtained by a trained observer using the cuff and stethoscope auscultation method. It also has a result storage function.

Indications for Use
The blood pressure monitor is intended for adult indoor use only.

Measurement Principle
The TMB-1776-BZY uses the Oscillometric Measuring method to detect blood pressure. It establishes a zero pressure equivalent to atmospheric pressure before each measurement. The arm cuff is then inflated while the unit detects pressure oscillations generated by beat-to-beat pulsatile. These oscillations are used to determine the systolic and diastolic pressure as well as the pulse rate.

Receiving and Inspecting your Monitor
Before using the device, check that the packaging has not been tampered with and ensure that all contents are present. Make sure there is no visible damage to the device or accessories and remove all packaging material. If you have any doubts, do not use the device and contact your retailer or the specified Customer Services address.

BEFORE YOU START
Before using the blood pressure monitor, there are a few things you need to do:

Choice of Power Supply
The TMB-1776-BZY can be powered by batteries or by pairing it with a smart device. Choose the power supply option that suits your preference.


Installing and Replacing the Batteries
If you choose to use batteries, follow the instructions in the manual to install or replace them.

Setting Date and Time
Set the date and time on the blood pressure monitor according to the instructions provided.

Select the User ID
If applicable, select the user ID on the monitor to ensure accurate record keeping.

Pair a smart device with the monitor
If you choose to pair the blood pressure monitor with a smart device, follow the instructions provided to establish the connection.

MEASUREMENT
To measure your blood pressure using the TMB-1776-BZY, follow these steps:

Applying the cuff
Wrap the arm cuff around your upper arm, following the instructions provided. Ensure a snug fit but not too tight.

Start the Measurement
Press the start button on the monitor to begin the blood pressure measurement. Remain still and avoid talking or moving during the measurement process.

DATA MANAGEMENT
The TMB-1776-BZY allows you to manage your measurement records.
Follow these instructions:

Recall the Records
Use the appropriate buttons on the monitor to recall your previous measurement records.

Delete the Records
If desired, delete specific records from the memory using the provided options on the monitor.

INFORMATION FOR USER
Here are some tips and information for accurate blood pressure measurement:

Tips for measurement
Follow the recommended guidelines for accurate blood pressure measurement, such as sitting in a comfortable position with your feet flat on the ground and keeping your arm supported at heart level.

Maintenances
Refer to the manual for maintenance instructions to ensure the longevity and accuracy of the blood pressure monitor.

ABOUT BLOOD PRESSURE
Here are some frequently asked questions about blood pressure:

What are systolic pressure and diastolic pressure?
Systolic pressure is the highest pressure in your arteries when your heart beats, while diastolic pressure is the lowest pressure in your arteries when your heart rests between beats.

What is the standard blood pressure classification?
The standard blood pressure classification includes categories like normal, elevated, hypertension stage 1, hypertension stage 2, and hypertensive crisis. Consult with your healthcare professional for more information.

Irregular pulse rate detector
The blood pressure monitor is equipped with an irregular pulse rate detector to detect irregularities in your pulse rate. Consult with your healthcare professional if you receive irregular pulse rate readings.

Why does my blood pressure fluctuate throughout the day?
Blood pressure can fluctuate due to various factors such as physical activity, stress, and time of day. It is normal for blood pressure to vary throughout the day.

Why do I get a different blood pressure at home compared to the hospital?
Home blood pressure measurements may differ from hospital measurements due to various factors such as white coat syndrome, different measurement techniques, and environmental differences. Consult with your healthcare professional for accurate interpretation.

Is the result the same if measuring on the right arm?
For accurate comparison, it is recommended to measure blood pressure on the same arm consistently. However, consulting with your healthcare professional is advised for personalized
guidance.

FAQ

  1. Q: How long is the warranty period for the TMB-1776-BZY blood pressure monitor?
    A: The warranty period for the TMB-1776-BZY blood pressure monitor is two years.

  2. Q: Can I use the blood pressure monitor outdoors?
    A: No, the blood pressure monitor is intended for indoor use only.

  3. Q: How do I pair the monitor with a smart device?
    A: Follow the instructions provided in the manual to establish a connection between the blood pressure monitor and your smart device.

  4. Q: Can I delete specific measurement records from the memory?
    A: Yes, you can delete specific records from the memory using the options provided on the monitor.

  5. Q: What should I do if I receive irregular pulse rate readings?
    A: Consult with your healthcare professional if you receive irregular pulse rate readings for further evaluation and guidance.

Thank you very much for selecting TRANSTEK Blood Pressure Monitor TMB-1776-BZY.
Please read the user manual carefully and thoroughtly so as to ensure the safe usage of this product, Keep the manual well for further reference in case you have problems.

INTRODUCTION

General Description
Thank you for selecting TRANSTEK blood pressure monitor (TMB-1776-BZY). The monitor features blood pressure measurement, pulse rate measurement and the result storage. The warranty period is two years. Readings taken by the TMB-1776-BZY are equivalent to those obtained by a trained observer using the cuff and stethoscope auscultation method. This manual contains important safety and care information, and provides step by step instructions for using the product. Read the manual thoroughly before using the product.

Features:

  • 80mm×58.5mm Digital LCD display
  • Maximum 199 records per user
  • Measuring-during-inflation technology

Indications for Use
This Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate with arm circumference ranging from 22cm to 32cm (about 8¾˝-12½˝) or 22cm to 42cm (about 8¾˝-16½˝). It is intended for adult indoor use only.

Measurement Principle
This product uses the Oscillometric Measuring method to detect blood pressure. Before every measurement, the unit establishes a “zero pressure” equivalent to the atmospheric pressure. Then it starts inflating the arm cuff, meanwhile, the unit detects pressure oscillations generated by beat-to-beat pulsatile, which is used to determine the systolic and diastolic pressure, and also pulse rate.

Receiving and Inspecting your Monitor
Check that the device packaging has not been tampered with and make sure that all contents are present. Before use, ensure that there is no visible damage to the device or accessories and that all packaging material has been removed. If you have any doubts, do not use the device and contact your retailer or the specified Customer Services address.

Safety Information

The signs below might be in the user manual, labeling or other component. They are the requirement of standard and using.

Precaution

  • This device is intended for indoor, home use and is not intended for self-use in public areas.

  • This device is portable, but it is not intended for use during patient transport.

  • This device is not suitable for continuous monitoring during medical emergencies or operations.

  • This device is intended for non-invasive measuring and monitoring of arterial blood pressure.
    It is not intended for use on extremities other than the arm, or for any purpose other than obtaining a blood pressure measurement.

  • This device is for adults. Do not use this device on neonates or infants. Do not use it on children and adolescents unless otherwise instructed by a medical professional.

  • Consult with your physician before using this monitor if you suffer from the following conditions: common arrhythmias such as premature ventricular beats or atrial fibrillation;
    peripheral arterial disease; pregnancy; preeclampsia; implantation with electrical devices; undergoing intravascular therapy; arteriovenous shunt or mastectomy. Please note that any of these conditions may affect measurement readings, in addition to patient motion, trembling or shivering.

  • Do not use this device for diagnosis or treatment of any health problem or disease. Contact your physician if you have or suspect any medical problem. Do not change your medications without the advice of your physician or health care professional.

  • If you are taking medication, consult your physician to determine the proper time to measure your blood pressure.

  • This device may be used only for the intended use described in this manual, the manufacturer shall have no liability for any incidental, consequential, or special damages caused by misuse or abuse.

  • Please use the device under the environment which is provided in the user manual. Otherwise, the performance and lifetime of the device will be impacted and reduced.

  • The device may require up to 30 minutes to warm up / cool down from the minimum/ maximum storage temperature before it is ready for use.

  • The blood pressure monitor, its adapter, and the cuff are suitable for use within the patient environment.

  • Do not wash the cuff in a washing machine or dishwasher !

  • The device contains sensitive electronic components. To avoid measurement errors, avoid taking blood pressure measurements near a strong electromagnetic field radiated interference signal or electrical fast transient/burst signal.

  • Wireless communication equipment, such as wireless home network devices, mobile phones, cordless telephones and their base stations, walkie-talkies may cause interference that may affect the accuracy of measurements. A minimum distance of 1 foot (30 cm) should be kept from such devices during a measurement.

  • Blood Pressure Monitor is intended for use by medical staffs and lay persons, and patient is also an intended user or operater.

Caution

  • Do not attempt to repair the unit yourself if it malfunctions. Only have repairs carried out byauthorized service centers.
  • It is recommended that the performance should be checked after repair, maintenance, and every two years of use, by retesting the requirements in limits of the error of the cuff pressure indication and air leakage (testing at least at 50 mmHg and 200 mmHg). Please contact manufacturer or distributor for authorized service personnel.
  • Store your device, cuff and adapter in a clean and dry place, protect it against extreme moisture, heat, lint, dust and direct sunlight. Never place any heavy objects on it.
  • Make sure the rubber tube of the cuff is not squeezed, stretched, or kinked during storage.
  • Dispose of accessories, detachable parts, and the device according to the local guidelines.

Warning

  • Do not apply the cuff on an arm that has an intravenous drip or a blood transfusion attached.
  • Do not kink, fold, stretch, compress, or otherwise deform the tube during measuring, as the cuff pressure might continuously increase, which could prevent blood flow and result injury.
  • Taking blood pressure measurements too frequently could disrupt blood circulation and cause injuries.
  • Do not apply cuff to areas on patient where skin is delicate or damaged. Check cuff site frequently for irritation.
  • Do not place the cuff on the arm of a person whose arteries or veins are undergoing medical treatment, i.e. intra-vascular access or intra-vascular therapy or an arteriovenous (A-V) shunt, which could disrupt blood circulation and cause injuries.
  • Do not place the cuff on the arm on the same side of a mastectomy (especially when lymph nodes have been removed). it is recommended to take measurements on the unaffected side.
  • Do not wrap the cuff on the same arm to which another monitoring device is applied. One or both devices could temporarily stop functioning if you try to use them at the same time.
  • Please check (for example, by observation of the limb concerned) that the operation of the device does not result in prolonged impairment of patient blood circulation.
  • On the rare occasion of a fault causing the cuff to remain fully inflated during measurement, loosen and remove the cuff immediately. Prolonged high pressure applied to the arm (cuff pressure > 300 mmHg or constant pressure >15 mmHg for more than 3 minutes) might lead to bruising and discolored skin.
  • Do not use this device with high-frequency (HF) surgical equipment at the same time.
  • This device is not used in conjunction with oxygen rich environments, not intended for use with flammable anaesthetics, not intended for use in conjunction with flammable agents.
  • Do not touch output of the batteries/adapter and the user simultaneously.
  • The power cord is considered the disconnect device for isolating this equipment from supply mains. Do not position the equipment so that it is difficult to reach or disconnect.
  • Do not use this device if you are allergic to polyester, nylon, or plastic.
  • Only use accessories approved by manufacturer. Using unapproved accessories might cause damage to the unit and injure users.
  • If you experience discomfort during a measurement, such as pain in the arm or other complaints, press the Power button immediately to release the air from the cuff.
  • Do not use the device while under maintenance, or being serviced.
  • The air tube poses a risk of strangulation. Furthermore, the small parts of product and batteries present a choking hazard if swallowed. They should therefore always be kept away from infants/children.
  • Sensor degradation or looseness may reduce performance of device or cause other problems.

Notice

  • You can use this device to take your own measurement, no third-party operator is required.
  • Adapter is specified as a part of ME EQUIPMENT.
  • At the request of authorized service personnel, circuit diagrams, component part lists, descriptions, and calibration procedures will be made available by the manufacturer or distributor.
  • The expected lifetime of the cuff may vary by the frequency of washing, skin condition, and storage state.
  • Please report to the manufacturer and the competent authority of the Member State / the FDA in which you are established about any serious incident that has occurred in relation to this device.

LCD display signal

Monitor Components

List

  1. Blood Pressure Monitor (TMB-1776-BZY)

  2. Cuff (Type BF applied part)
    Upper arm cuff:22-32cm or Upper arm cuff:22-42cm.

  3. 4×AAA batteries

  4. User manual

  5. Type-C cable& Adapter

BEFORE YOU START

Choice of Power Supply

  1. Battery powered mode: 6VDC 4×AAA batteries
  2. AC adaptor powered mode: 5V 1A

(Please use the AC adapter which is authorized by the manufacturer!)
Please unplug the adaptor to depart from the using utility power.

CAUTION
In order to get the best effect and protect your monitor,please use the right batteries and special power adaptor which complies with local safety standard.

Installing and Replacing the Batteries

  • Open the battery cover.
  • Install the batteries by matching the correct polarity, as shown.
  • Replace the battery cover.

Replace the batteries whenever the below happen bAt The shows.
The display is dim.
The display does not light up.

CAUTION

  • New and used batteries, or different types of batteries shall not be used together.
  • Remove batteries if the device is not likely to be used for some time. Do not heat or deform the batteries, or dispose of them in fire. Batteries should not be disposed of with household waste. Please check with your local authority for battery recycling advice.

Setting Date and Time.
It is important to set the clock before using your blood pressure monitor, so that a time stamp can be assigned to each record that is stored in the memory. (The setting range of the year :2023—2053; Time format:24 H/12 H)

  1. When the monitor is off, hold pressing “USER” button for about 3 seconds, it will display the bluetooth symbol first. No operation within 60s or Press the “START/STOP” button, it will skip Bluetooth pairing and enter the [YEAR] setting.

  2. Press “STRAT/STOP” button to confirm [YEAR]. then the [date format] will flash. Press the “USER” or “MEM” button to switch the date format between [month/day] and [day/month].

  3. Repeat the same steps to set the [MONTH] and [DAY].

  4. Repeat the same steps to set the [time format] between [24 H] and [12 H].

  5. Repeat the same steps to set the [HOUR] and [MINUTE].

  6. After setting the [MINUTE], the LCD will display all the settings you have done, and then display “do nE” , several seconds the monitor will turn off automatically.

Pair a smart device with the monitor
You are the intended operator of this blood pressure monitor. You can measure your blood pressure and then save and send measurement data to a smart device (such as smartphone or tablet) with Bluetooth wireless connectivity.

  1. Turn on Bluetooth and the app on your smart device. Make sure both are ON when pair-up is proceeding.
  2. When the monitor is off, press and hold the “START/STOP” button to start pair-up, the bluetooth symbol will flash.
  3. If successful, the bluetooth symbol will not flash any more and the monitor will display “do nE”, and then it will shut off after about several seconds automatically.

Note:

  1. The date and time on your monitor will automatically be set after paired with your smart device successfully.
  2. The device can also enter the Bluetooth pairing automatically when the battery is installed for the first time.
  3. If unsuccessful within 60 seconds, it is judged timeout and the monitor will shut off.

Specifications for Bluetooth Transmission

Bluetooth| Throughput| 2.5K-5K
Latency| 50ms
Error Rate| 0 when≤-70dBm
Operating Frequency| 2402-2480MHz
Transmission Power| see test report
Transmission Distance| 10m
Note:

  1. The necessary Quality of Service (QoS) is fully considered here for wirelessly enabled functions.
  2. Interference may occur in the vicinity of equipment marked with the following symbol . And TMB-1776-BZY may interfere the vicinal electrical equipment.
  3.  Keep the monitor at least 20 centimeters away from the human body (especially the head) when data transmission is proceeding after measurement.
  4. To enable the data transmission function, this device shall be paired to an appropriate BT mobile terminal.

Warning
About a wireless communication interference
The monitor operates in the unlicensed ISM band at 2.4 GHz. In case it is used around the other wireless devices including microwave and wireless LAN, which operate at the same frequency band as the monitor, there is a possibility that interference occurs between the monitor and such other devices. If such interference occurs, please stop the operation of other devices or relocate the monitor before using it or do not use it around the other wireless devices.

List of compatible devices:
For iOS devices:
The operating system must be iOS 13.0 or more.

For Android devices:
The operating system must be Android 5.0 or more.

Select the User ID
There are 3 user ID available. The user and , each with 199 memory spaces, are designed for 2 different people to save the measured values separately. The user , no memory space, is reserved for guest.

  1. When the monitor is off, press the “USER” button to display the current user ID and flash. Press the “USER” button to switch the user ID between the user , and .     
  2. Press the “START/STOP” button to confirm the selected user ID, the User ID will not flash any more and the monitor will enter the measurement automatically.

Applying the cuff

Only use a cuff that has been approved by the manufacturer for this device model. Before use, please confirm if it fits your arm circumference.

  1. Remove all jewelry, such as watches and bracelets from your left arm. Note: If your doctor has diagnosed you with poor circulation in your left arm, use your right arm.

  2. Roll or push up your sleeve to expose the skin. Make sure your sleeve is not too tight.

  3. Hold your arm with your palm facing up and tie the cuff on your upper arm, align the Artery indicator with the main Artery (on the inside of your arm).
    Note: Locate the main Artery by pressing with 2 fingers approximately 2 cm above the bend of your elbow on the inside of your left arm. Identify where the pulse can be felt the strongest, that is your main Artery!

  4. Make sure the bottom edge of the arm cuff 2 to 3 cm above the inside elbow. Then wrap the cuff securely.
    Note: The cuff should be snug but not too tight.
    You should be able to insert one finger between the cuff and your arm.

  5. Sit upright in a comfortable chair with your back against the backrest of the chair. Keep your feet flat and your legs uncrossed.
    Place your arm resting comfortably on a flat table. The cuff worn on your arm should be placed at the same level as your right atrium of the heart.

  6. Take 5-6 deep breaths and let’s start measuring!

Helpful tips:

  • Take the measurement in a silent room.
  • Rest for 5 minutes before a measurement.
  • Wait at least 3 minutes before another measurement.
  • This allows your blood circulation to recover.
  • Be relaxed and do not move and talk during the measurement procedure.
  • For a meaningful comparison, try to measure under similar conditions. For example, take daily measurements at approximately the same time, on the same arm, or as directed by a physician.

MEASUREMENT

Start the Measurement

  1. When the monitor is off, press “START/STOP” button to turn on the monitor, and it will finish the whole measurement.
    (Take user 1 for example.)
    Cuff wrap detection Inflating and measuring. Display and save the measurement result. Note:

    • Any time, to stop the measurement, press the “START/STOP” button.
    • If you don’t pair with the device or don’t keep the app ON, the bluetooth symbol will flash during the measurement.
  2. Press the “START/STOP” button to turn off the monitor, or it will shut off after about 1 minute.

  3. If your monitor is already paired with your smart device and both Bluetooth and app are ON, when the measurement completed, the measurement result will start transmitting. (only user 1 and 2 available) If successful, the symbol “ ” will disappear first , after sevral seconds, “ ” will also disappear, and then the monitor will turn off automatically. If unsuccessful within 60 seconds, it is judged timeout and the monitor will shut off. In the case of a data transmission failure, up to 199 measurements are saved on the device and will be sent when a successful connection is achieved.

  4. About the irregular pulse rate and excessive body motion during the measurement.

    • During a measurement, If an irregular pulse rate is detected, the symbol will display in the measurement result. See page 25 for more information.
    • During a measurement, when the excessive body motion, especially of the arm the cuff is worn is detected, the symbol will flash about 5 seconds and detect again. If it is no longer detected, the symbol will disappear; If still detected, the symbol will final display in the measurement result.

Note
The measured blood pressure reading may not be accurate if this symbol is displayed.

DATA MANAGEMENT

Recall the Records

  1. When the monitor is off, press “MEM” button to show the average value of the latest three records.If the records are less than 3 groups, it will display the latest record instead.
  2. Press the “USER” or “MEM” button to get the record you want.
    The date and time of the record will be shown alternately.

Note

  • If there is untransmitted data, the symbol and will display on the record.
  • The latest record (1) is shown first. Each new measurement is assigned to the first (1) record. All other records are pushed back one digit (e.g., 2 becomes 3, and so on), and the last record (199) will be dropped from the list.
  • When you pass the limit, every time you recall the records, the monitor will display “FULL” along with the group number “199”.

Delete the Records
Delete the records by following the steps below.

Delete one record (Example shown below for User 1)

  1. Enter the memory recall mode as described in the previous section “Recall the records”, find out the record you want to delete.
  2. Press and hold the “START/STOP” button for about 5 seconds, the LCD will display “dELE yES” with the “yES” blinking.
  3. Short press the “START/STOP” button again to confirm the deletion, the LCD will display “do nE” and then the previous record will be displayed.

Note
Before you confirm the deletion, you could press the “USER” or “MEM” button to switch the LCD display from “dELE yES” to “dELE no”. Then press the “START/STOP” button, you could stop clearing the memory.

Delete all records (Example shown below for User 1)

  1. Enter the memory recall mode as described in the previous section “Recall the records”.

  2. Press and hold the “USER” and “MEM” button for about 5 seconds, the LCD will display “dELE ALL” with the “ALL” blinking.

  3. Short press the “START/STOP” button to confirm the deletion, the LCD will display “do nE”. Note
    Before you confirm the deletion, you could press the “USER” or “MEM” button to switch the LCD display from “dELE ALL” to “dELE no”. Then press the “START/STOP” button, you could stop clearing the memory.

  4. Once deleted, your readings cannot be restored. The LCD will display “–” like the following picture.
    Press the “START/STOP” button to turn off the monitor, otherwise it will power off automatically after about 1 minute.

INFORMATION FOR USER

Tips for Measurement
Measurements may be inaccurate if taken in the following circumstances.

Maintenance

In order to get the best performance, please follow the instructions below.

  1. Cleaning Process:
    Step 1: Make sure to switch off and unplug the device prior to cleaning.
    Step 2: Use a soft cloth wetted with soapy water to clean the cuff first, and then use a soft cloth wetted with clear water to remove residual soap until there is no visible residual contaminants. Attention shall be paid to avoid liquid invasion into the cuff.
    Step 3: Use a dry soft cloth to wipe the cuff, in order to remove residual moisture.
    Step 4: Dry the cuff at a well-ventilated place after cleaning.

  2. Disinfection Process:
    Step 1: Make sure to switch off and unplug the device prior to disinfection.
    Step 2: Use a soft cloth wetted with 70% isopropanol to disinfect the cuff for about 3 minutes. Attention shall be paid to avoid liquid invasion into the cuff.
    Step 3: Use a clean dry cloth or towel to wipe off the disinfectant until there is no visible residue.
    Step 4: Dry the cuff at a well-ventilated place after disinfection.

Suggestion:
Frequency of Cleaning and Disinfection:
For single patient multiple use, it’s recommended to clean the device surface once a month or whenever it’s necessary. For multiple patient multiple use, it’s recommended to clean the device every time before and after usage. Maintenance procedures shall be taken as per instruction.

ABOUT BLOOD PRESSURE

What are systolic pressure and diastolic pressure?
When ventricles contract and pump blood out of the heart, the blood pressure reaches its maximum value in the cycle, which is called systolic pressure. When the ventricles relax, the blood pressure reaches its minimum value in the cycle, which is called diastolic pressure.

What is the standard blood pressure classification?
The blood pressure classification published by World Health Organization (WHO) and International Society of Hypertension (ISH) in 1999 is as follows:

CAUTION
Only a physician can tell your normal BP range. Please contact a physician if your measuring result falls out of the range. Please note that only a physician can tell whether your blood pressure value has reached a dangerous point.

Level Blood Pressure (mm Hg)| Optimal| Normal| High-normal| Mild| Moderate| Severe
---|---|---|---|---|---|---
SYS| <120| 120-129| 130-139| 140-159| 160-179| ≥180
DIA| <80| 80-84| 85-89| 90-99| 100-109| ≥110

Irregular Pulse Rate Detector
An irregular pulse rate will be detected if there is an irregular pulse rhythm while measuring systolic and diastolic blood pressure. When measurements were performed, the monitor will record all pulse intervals and calculate the average. If two or more pulse intervals were recorded, and the difference between each interval and the average is larger than ±25% of the average; or if four or more pulse intervals were recorded, and the difference between each interval and the average is larger than ±15% of the average value, the irregular pulse symbol will be displayed along with measurement results.

CAUTION
The appearance of the IPR icon indicates that a pulse irregularity consistent with an irregular pulse rate was detected during measurement. Usually this is NOT a cause for concern. However, if the symbol appears often, we recommend you seek medical advice. Please note that the irregular pulse rate detector only serves as a non-medical feature, and the results cannot be used directly for clinical judgement. Please seek medical advice from professionals before making any medical decisions.

Why does my blood pressure fluctuate throughout the day?

  1. Individual blood pressure varies multiple times everyday. It is also affected by the way you tie your cuff and your measurement position, so please take the measurement under the same conditions.
  2. If the person takes medicine, the pressure will vary more.
  3. Wait at least 3 minutes for another measurement.

What you need to pay attention to when you measure your blood pressure at home:
If the cuff is tied properly. If the cuff is too tight or too loose. If the cuff is tied on the upper arm. If you feel anxious. Taking 2-3 deep breaths before beginning will be better for measuring.
Advice: Relax yourself for 4-5 minutes until you calm down.

Why do I get a different blood pressure at home compared to the hospital?
The blood pressure is different even throughout the day due to weather, emotion, exercise etc, Also, there is the “white coat” effect, which means blood pressure usually increases in clinical settings.

Is the result the same if measuring on the right arm?
It is ok for both arms, but there will be some different results for different people. We suggest you measure the same arm every time.

TROUBLE SHOOTING

If any abnormality arises during use, please check the following points:

PROBLEM SYMPTOM CHECK THIS REMEDY
No power Display can not light up. Batteries are depleted. Replace

with new batteries.
Batteries are inserted incorrectly.| Insert the batteries correctly.
Adapter is inserted incorrectly.| Insert the AC adapter correctly.
High Battery| bAt H shows| The battery is too high.| Replace with new batteries.
Low Battery| bAt Lo & shows| The battery is too low.| Replace with new batteries.
Error message| E 1 shows| The cuff is not wrapped or wrapped incorrectly, or the cuff air plug is loose.| Refasten the cuff and insert air tube plug correctly then measure again.
E 2 or shows| Excessive body motion (such as shaking of the arm with the cuff on) or weak pulse is detected.| Relax for 5 minutes. and then keep still, measure again.
E 3 shows| Pulse is not detected during measuring.| Loosen the clothing on the arm and measure again.
E 4 shows| The measurement failed.| Relax for 5 minutes and measure again.
EEx shows| Hardware error (X can be some digital symbol, such as1, 2, etc.)| Turn off monitor and measure again. If EEx still appears on the display, please contact the retailer or our customer service.
USb Err shows| The voltage of the adapter is too high or too low.| Replace with the authorized adapter.
Warning message| out shows| Out of measurement range| Relax for a moment and then measure again. If the problem persists, contact your physician.

NOTE: If the product still does not work, contact Customer Service. Under no circumstance should you disassemble or attempt to repair the unit by yourself.

SPECIFICATIONS

WARNING: No modification of this equipment is allowed.

Authorized Component
Please use the authorized adapter.

Adapter

  • Type: BLJ06L050100U-U
  • Input: 100-240V, 50-60Hz, 0.2A max
  • Output: 5V 1000 mA

Contact Information
For more information about our products, please visit www.transtekcorp.com.you can get customer service, usual problems and customer download, transtek will serve you anytime.
Manufactured by: Guangdong Transtek Medical Electronics Co., Ltd.
Company: Guangdong Transtek Medical Electronics Co., Ltd.
Address: Zone A, No.105 ,Dongli Road, Torch Development District, 528437 Zhongshan, Guangdong, China

EMC GUIDANCE

The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments.

Essential performance:
Accuracy of measuring blood pressure and pulse rate

Measurement Range| Systolic pressure: 60-230 mmHg Diastolic pressure: 40-130 mmHg Pulse: 40-199 beats/minute
---|---
Rated Cuff Pressure| 0-299 mmHg (0-39.9 kPa)
Accuracy| Static Pressure: 5℃-40℃ within ±3mmHg Pulse value: ±5%
Clinical validation: Mean difference within ±5mmHg; Standard deviation ≤8mmHg

The Basis Safety of the Blood Pressure Monitor (TMB-1776-BZY) is as following:
Deviation from normal operation that poses an unacceptable risk to the patient or operator.
Warning: Don’t be near the active HF surgical equipment and the RF shielded room of an ME system for magnetic resonance imaging, where the intensity of EM disturbances is high.

  • Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
  • Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
  • Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the equipment including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.

Technical description

  1. All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFORMANCE with regard to electromagnetic disturbances for the expected lifetime.
  2. Guidance and manufacturer’s declaration-electromagnetic emissions and Immunity.

Guidance and manufacturer’s declaration – electromagnetic emissions

Emissions test| Compliance
RF emissions CISPR 11| Group 1
RF emissions CISPR 11| Class [ B ]
Harmonic emissions IEC 61000-3-2| Class A
Voltage fluctuations / flicker emissions IEC 61000-3-3| Comply
Guidance and manufacturer’s declaration – electromagnetic Immunity

Immunity Test| IEC 60601-1-2
Test level| Compliance level
Electrostatic discharge (ESD) IEC 61000-4-2| ±8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 kV air| ±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast transient/burst IEC 61000-4-4| ±2 kV for power supply lines ±1 kV signal input/output 100 kHz repetition frequency| ±2 kV for power supply lines Not Applicable 100 kHz repetition frequency
Surge IEC61000-4-5| ±0.5 kV, ±1 kV differential mode ±0.5 kV, ±1 kV, ±2 kV common mode| ±0.5 kV, ±1 kV differential mode Not Applicable
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11| 0% UT; 0,5 cycle. At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°. 0% UT; 1 cycle and 70% UT; 25/30 cycles; Single phase: at 0°. 0% UT; 250 / 300 cycle| 0% UT; 0,5 cycle. At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°. 0% UT; 1 cycle and 70% UT; 25/30 cycles; Single phase: at 0°. 0% UT; 250 / 300 cycle
Power frequency magnetic field IEC 61000-4-8| 30 A/m
50 Hz / 60 Hz| 30 A/m
50 Hz / 60 Hz
Conduced RF IEC61000-4-6| 3 V
0,15 MHz – 80 MHz 6 V in ISM and amateur radio bands between 0,15 MHz and 80 MHz 80% AM at 1 kHz| 3 V
0,15 MHz – 80 MHz 6 V in ISM and amateur radio bands between 0,15 MHz and 80 MHz 80% AM at 1 kHz
Radiated RF IEC61000-4-3| 10 V/m
80 MHz – 2,7 GHz 80% AM at 1 kHz| 10 V/m
80 MHz – 2,7 GHz 80% AM at 1 kHz
NOTE UT is the a.c. mains voltage prior to application of the test level.

FCC Statement

FCC ID: OU9TMB1776BZY
This device complies with Part 15 of the FCC Rules.

Operation is subject to the following two conditions:

  1. This device may not cause harmful interference, and
  2. this device must accept any interference received, including interference that may cause undesired operation.

FCC Regulatory Compliance
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

  • Reorient or relocate the receiving antenna.
  • Increase the separation between the equipment and receiver.
  • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
  • Consult the dealer or an experienced radio/TV technician for help.

RF Exposure Compliance
This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter.

Guangdong Transtek Medical Electronics Co., Ltd.
Zone A, No.105 , Dongli Road, Torch Development District, 528437 Zhongshan, Guangdong, China