INFICON ECOTEC E3000 Leak Detector Instruction Manual
- June 16, 2024
- INFICON
Table of Contents
INFICON ECOTEC E3000 Leak Detector
Product Information
Specifications
- Order no.: 530-001, 530-002 530-103, 530-104
- Software version: V3.32
About this Manual
This manual provides operating instructions for the Leak Detector. It includes information on the target group, applicable documents, presentation of information, and safety guidelines.
1.1 Target Group
The target group for this manual includes individuals who will be operating and using the Leak Detector.
1.2 Other Applicable Documents
There may be other documents that are relevant to the operation and use of the Leak Detector. Please refer to those documents as well.
1.3 Presentation of Information
The manual provides important information and warnings. Pay close attention to the warnings provided to ensure safe and proper use of the product.
1.3.1 Warnings
Warnings are provided throughout the manual to highlight potential dangers or hazards associated with the use of the Leak Detector. It is important to read and understand these warnings before operating the product.
Safety
2.1 Intended Use
The Leak Detector is intended for detecting leaks in various applications. It is important to use the product only for its intended purpose.
2.2 Owner Requirements
As an owner of the Leak Detector, it is your responsibility to ensure that the product is used and maintained properly. Follow all guidelines and recommendations provided in this manual.
2.3 Operator Requirements
Operators using the Leak Detector should be trained on its proper use and safety precautions. They should also be familiar with the information provided in this manual.
2.4 Dangers
The manual highlights potential dangers associated with the use of the Leak Detector. It is important to be aware of these dangers and take appropriate safety measures to avoid accidents or injuries.
Shipment, Transport, Storage
This section provides information on how to handle and store the Leak Detector during shipment and transport. Follow the guidelines provided to ensure the product’s safety and integrity.
Description
4.1 Function and Setup of the Device
This section explains the function and setup process of the Leak Detector. It provides step-by-step instructions on how to set up and operate the device effectively.
4.2 Main Unit
The main unit of the Leak Detector houses the core components and controls. It is important to familiarize yourself with the main unit and its features before using the product.
4.3 Sniffer Line
The sniffer line is a crucial component of the Leak Detector that is used for detecting leaks. This section provides information on how to properly connect and use the sniffer line.
4.4 Technical Data
The technical data section provides detailed specifications and information about the Leak Detector. It includes data such as dimensions, weight, power requirements, and other technical details.
FAQs
- Q: Can I use the Leak Detector for detecting leaks in gas pipelines?
- A: Yes, the Leak Detector can be used for detecting leaks in gas pipelines. However, please refer to the specific guidelines and safety precautions provided in the manual.
- Q: What should I do if the device malfunctions?
- A: If the device malfunctions, refer to the troubleshooting section in the manual. If the issue persists, contact the authorized service center for assistance.
- Q: Can I use the Leak Detector in hazardous environments?
- A: The suitability of the Leak Detector for hazardous environments depends on its specific certifications and ratings. Refer to the manual and consult with relevant authorities to determine its compatibility with hazardous environments.
Translation of the original operating instructions
Ecotec® E3000
Leak detector
Order no.: 530-001, 530-002 530-103, 530-104
from software version: V3.32
kina22en1-25, 2102
Reprint, translation and duplication need to be approved in writing by INFICON GmbH.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
2
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
4
Table of Contents
About this manual
This document applies to the software version stated on the cover page.
Documents for other software versions are available from our sales department.
1.1 Target group
These operating instructions are intended for the owner of the leak detection
unit E3000 and for technically qualified personnel with experience in leak
detection technology and integration of leak detection devices in leak
detection systems.
1.2 Other applicable documents
ECO-Check installation instructions, document no. liqa10 Interface
description, document no. kins22
1.3 Presentation of information
1.3.1 Warnings
Imminent threat of danger resulting in death or severe injuries
Dangerous situation potentially resulting in death or severe injuries
Dangerous situation resulting in minor injuries
Dangerous situation resulting in damage to property or the environment
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
1 About this manual
5
6
1 About this manual
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
2 Safety
2.1 Intended use
The device is a leak detector for sniffer leak detection. With the device you
locate and quantify leaks on test objects. The device sniffs for light gases,
refrigerants and natural gases.
The test objects must contain the gas under overpressure. The outsides of the
test objects are checked for escaping gas with a sniffer line (sniffing
method).
The sniffer line is available as an accessory (see “10.1 Accessories”, page
81).
You must install, operate and service the device only in compliance with these
operating instructions.
Adhere to the restrictions of use (see “4.4 Technical data”, page 14).
Unauthorized use
Do not suction in liquids with the device.
Never hold the sniffer probe into liquids but sniff only for gases. To sniff
test objects that have low surface moisture, such as condensation moisture,
use the water protection tip.
Use the device only outside of explosive areas.
2.2 Owner requirements
Safety-conscious operation
Operate the device only in technically perfect working order. Operate the
device exclusively as specified, in a safety-conscious and hazard-conscious
manner and in compliance with these operating instructions.
Fulfill the following regulations and monitor their compliance: Intended use
Generally applicable safety and accident prevention regulations
International, national and local standards and guidelines Additional
provisions and regulations that are specific to the unit
Use only original parts or parts approved by the manufacturer. Keep these
operating instructions available at the equipment location. Do not touch any
live parts with the sniffer tip. Sniffing unspecified and aggressive gases.
Personnel qualifications
Allow only qualified technical staff to work with and on the device. The
qualified technical staff must have received training on the device.
Allow personnel in training to work with and on the device only under the
supervision of trained qualified technical staff.
Make sure that the authorized personnel have read and understood the operating
instructions and all other applicable documents (see “1.2 Other applicable
documents”, page 5), especially the information on safety, maintenance and
repairs, before starting work.
Define responsibilities, authorizations and supervision of personnel.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Safety
7
2.3 Operator requirements
Read, observe and follow the information in these operating instructions and
the working instructions created by the owner, especially the safety
instructions and warnings.
Carry out any work only based on the complete operating instructions.
If you have questions regarding operation or maintenance that you cannot find
answers for in these instructions, please contact the INFICON customer
service.
2.4 Dangers
The unit was built according to the state of the art and the recognized safety regulations. Nevertheless, improper use can result in danger to life and limb of the operator or other persons and damage to the unit and other property.
Danger due to use in EX/ ATEX areas
If the workplace is classified as a hazardous area according to Ex/ATEX
regulations it is mandatory to apply Ex/ATEX certified equipment.
Ecotec E3000 including INFICON sniffer lines are not Ex/ATEX certified.
In no risk areas and non-classified environments Ecotec E3000 can be safely
used.
The major risk associated with the leak detector is always the atmosphere
around and inside the instrument enclosure and not the gas aspired by the
sniffer into the interior of the detector.
Dangers due to electric power
The device is operated with electrical voltages of up to 236 V. Touching parts
where electric voltage is applied can result in death.
Disconnect the device from the power supply prior to any installation and
maintenance work. Make sure that the electric power supply cannot be
reconnected unauthorized.
Touching live parts with the sniffer probe can result in death.
Before starting the leak test, disconnect electrically operated test objects
from the power supply. Make sure that the electric power supply cannot be
reconnected unauthorized.
The device contains electric components that can be damaged from high electric
voltage.
Before connecting the device to the power supply, make sure that the supply
voltage specified on the device is the same as the local power supply.
Dangers due to liquids and chemical substances
Liquids and chemical substances can damage the unit.
Adhere to the restrictions of use (see “4.4 Technical data”, page 14).
Do not suction in liquids with the device.
Never try to find toxic, caustic, microbiological, explosive, radioactive or
other harmful substances with the device.
You must not smoke near the device. Do not subject the device to open fire and
avoid sparking.
Dangers due to strong light irradiation
Exposure of the eyes to LED light can lead to lasting eye damage.
Do not look into the LEDs of the sniffer handle from a short distance or for a
longer period of time.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
8
2 Safety
3 Shipment, transport, storage
Scope of delivery Transport
Table 1: Scope of delivery
Article Ecotec E3000 (main unit) Power supply cable, 3 m length Fuses
Replacement air filter 8 mm Allen wrench 19 mm box wrench Operating
instructions Interface description
Quantity 1 1
30 1 1 1 1 1
Check the scope of delivery of the product for completeness after receipt. The following must be ordered separately: sniffer lines with the desired length, ECO-Check calibrated leak, for device version E3000RC: display and connecting cable. Accessory list: see “10.1 Accessories”, page 81
Storage
Damage from transport Transport in unsuitable packaging material can damage
the unit. Parts inside the device can be damaged during transportation without
transport restraint. Store the original packaging. Transport the device only
in the original packaging. Screw the transport restraint into the device floor
before transportation, see
“5.1 Setup”, page 17.
Store the device adhering to the specifications, see “4.4 Technical data”,
page 14.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
3 Shipment, transport, storage
9
10
3
Shipment, transport, storage
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
4 Description
4.1 Function and setup of the device
The Ecotec E3000 is comprised of main unit and sniffer line. The Ecotec E3000
can verify and quantify gases sucked in by the sniffer line with the help of a
selective mass spectrometer. Working in the Ecotec E3000: a Quadrupol mass
spectrometer as a detection system a high vacuum pump system an inlet system
for the gas flow electrical and electronic sub-assemblies for electrical
supply and signal processing The mass spectrometer works under high vacuum,
i.e. the pressure in the mass spectrometer must always be below 10-4 mbar.
This vacuum is created by the turbo molecular pump with the help of a
diaphragm pump.
4.2 Main unit
The main unit will be called “device” only in the following so long as this
does not falsify the meaning.
a
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
b e
Fig. 1: Front view
1 Display 2 Handles and ventilation openings 3 Speaker
b
dc
4 ECO-Check calibrated leak 5 Lemo plug-in connector for sniffer line
4 Description
11
g
Headphone
I/O Port
RS-232
530-001 Ecotec E3000
90001054694
Sicherungen / Fuses 2 x 4 AT
ab c
Fig. 2: Rear view
1 Headphone connection, 3.5 mm jack 2 Inputs/outputs (I/O port) 3 RS-232
connection 4 Mains plug
de
f
5 Fuses behind cover 6 Power supply 7 Rating plate
1 Headphone connection, 3.5 mm jack In order to better hear signals in a loud
environment, you can connect head phones.
2 Inputs/outputs (I/O port) The I/O port enables communication with a PLC.
Some functions of the Ecotec E3000 can be controlled from the outside and
measurement results and device states of the Ecotec E3000 can be transmitted
to the outside.
Relay changeover contacts allow monitoring of the trigger values and the
operating status of the Ecotec E3000. Please refer to the “Interface
description Ecotec E3000” (doc. no. kins22e1) for further information.
3 RS-232 connection A PC can read out all data and measurement results of the
device and control the device via the RS-232 connection. Please refer to the
“Interface description Ecotec E3000” (doc. no. kins22e1) for further
information.
4 Mains plug The mains plug serves to switch the device on and off.
5 Fuses behind cover For information on replacing the fuses, see “8.2.3
Replacing mains fuses”, page 76.
6 Power supply For information on the power supply, see “4.4 Technical data”,
page 14, as well as the labeling on the power supply (rating plate).
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
12
Description
7 Rating plate The rating plate contains the supply voltage specification and other information with which the device can be clearly identified.
Fig. 3: Rating plate
1 Supply voltage 2 Serial number 3 Production date
4.3 Sniffer line
You need a sniffer line to operate the device. Sniffer lines are available in
four lengths: 3 m, 5 m, 10 m and 15 m.
The sniffer line is made of a flexible tube (multi-function cable), a handle
with controls (sniffer handle) and a sniffer probe.
There is a special sniffer line for robotic applications (see “10.1
Accessories”, page 81).
Sniffer probe
There are rigid and flexible sniffer probes with different lengths.
Sniffer handle display and functions
The display on the sniffer handle shows current information on the measuring process. You can operate functions frequently used during measurement with both buttons. LEDs inserted in the handle illuminate the test area.
c
a
d
b
e
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Fig. 4: Handle: display and functions
1 Display 2 ZERO adjustment 3 LEDs
4 Speaker (on the rear) 5 I·Guide operation
4 Description
13
4.4
If the measuring value limit is exceeded, the display switches from green to red. You can additionally set one speaker in the handle to output a signal and the LEDs in the handle to flash or shine with maximum brightness, see “6.3 Settings before measurements”, page 28.
Technical data
Table 2: Technical data
Mechanical data Dimensions (W H D)
610 mm 370 mm 265 mm
Weight
34 kg
Ambient conditions
Permissible ambient temperature (during 10 °C to 45 °C operation)
Permissible storage temperature
-20 °C to 60 °C
Max. relative humidity up to 31 °C
80 %
Max. relative humidity from 31 °C to 40 °C Linearly dropping from 80 % to 50 %
Max. relative humidity above 40 °C
50 %
Pollution degree
II
Max. altitude above sea level
2000 m
Electrical data
Supply voltages and frequencies
100 … 120 V, 50 / 60 Hz
207 … 236 V, 50 / 60 Hz
Power consumption
200 VA
Protection class
IP 20
Overvoltage category Mains fuse
II 2 4 A slow-blow
Power supply cable
2.5 m
Noise level
< 54 dBA
Physical data
Minimum detectable leak rate
R134a
0.05 g/a (0.002 oz/yr)
R600a Helium
0.05 g/a (0.002 oz/yr) < 1 10-6 mbar l/s
Measurement range
6 decades
Detectable masses
2 to 200 amu
Mass spectrometer
Quadrupol mass spectrometer
Ion source
2 cathodes
Time constant of the leak rate signal
< 1s
Gas flow through the capillary
120 to 200 sccm
Measured at 1 atm (1013 mbar) at sea level.
The flow rate changes with geographical
height and barometric pressure.
Time until ready for operation
< 2 min
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
14
4 Description
Table 2: Technical data (cont.)
Response time 3 m sniffer line 5 m sniffer line 10 m sniffer line 15 m sniffer
line
Table 3: Factory settings
Alarm profile Alarm delay Number of measuring points (I·Guide) Functions
Internal Recording output Baud rate and blank flange Pressure unit Flow rate
Lower limit Upper limit Sensitivity test Gas, definition Device speaker Handle
speaker Selecting a cathode I·Guide Calibration (cal), internal Contrast
Volume Minimum volume Leak rate, selected, highest Leak rate filter Menu PIN
Measuring time (I·Guide) Measuring mass Peak value ECO-Check Relay outputs
RS-232 protocol Sniffer probe, light Sniffer probe, filter, maintenance
Recorder, gas Scaling of the recorder Language SPS outputs and inputs
0.7 sec. 0.9 sec. 1.4 sec. 3.0 sec.
Trigger alarm Deactivated 4 Activated Activated Auto 9600 CR+LF mbar
100 sccm 250 sccm Activated Gas 1, gas 2, gas 3, gas 4, gas 5, gas 6 Activated
Trigger value A Deactivated Activated Not inverted, setting 30 2 2 Automatic
I·Filter Deactivated, 0000 1 second 69 Deactivated, 5 seconds Activated See
interface description ASCII Activated, Level 4 100 hours Auto Logarithmic
English See interface description
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
4 Description
15
Table 3: Factory settings (cont.)
Control location Search threshold Trigger value, sum (I·Guide) Trigger and
unit Idle time (I·Guide) Maintenance, sniffer probe filter Zero time Zero key
sniffer line ZERO key main unit
Local and RS-232 90 % 10 g/a 4 g/a 3 seconds 100 hours 5 seconds Activated Activated
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
16
4 Description5
Installation
5.1 Setup
Danger due to moisture and electricity Moisture getting into the device can
lead to personal injury from electric shocks and to material damage from short
circuits. Operate the Ecotec E3000 only in a dry environment. Operate the
Ecotec E3000 away from sources of liquid and moisture.
Danger due to dropping heavy loads The device is heavy and can damage persons
and items through tilting or dropping. Place the device only on a sufficiently
sturdy surface.
Material damage due to vibration. Parts of the measurement technology rotate
and must not be shaken. The parts continue to rotate for several minutes after
the device is shut down. Place the device only on a sturdy, vibration-free
surface. The device must not be shaken during operation and at least five
minutes after being
switched off.
Material damage due to an overheated device The device heats up during
operation and can overheat without sufficient ventilation. Please note the
specifications, see page 14. Ensure sufficient ventilation especially on the
ventilation openings on the left and
right of the device: free space on the side at least 20 cm, in front and in
the rear at least 10 cm. Keep heat sources away from the device. Do not
subject the device to direct sunlight.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
5 Installation
17
Material damage due to transport restraint that has not been removed The
transport restraint blocks the mechanical system in the device. Remove the
transport restraint before the start-up.
The transport restraint is on the bottom of the Ecotec E3000 and consists of a
yellow star screw.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Fig. 5: Unscrew the yellow transport restraint before the start-up
5.2 Connecting the sniffer line
Material damage due to a missing sniffer line The device must not be operated
without a connected sniffer line in order to avoid overpressure in pump and
measurement system. Connect the sniffer line before you start up the device.
Do not replace the sniffer line while the device is in operation.
Align the red marking on the sniffer line plug with the red marking on the
socket. Push the sniffer line plug into the socket on the device until it
locks into place. Pull the grooved ring on the plug to release it. The ring
opens the lock and you can pull out the plug.
5.2.1
Replacing the capillary filter of the sniffer probe
The metal capillary filter is the standard filter. With the plastic capillary
filter there is less danger of scratching the surfaces to be sniffed. The
water conservation tip is used if there is a danger of suctioning in liquids.
18
5 Installation
a bc
d
Fig. 6: Capillary filter
1 Sniffer probe end 2 Plugwasher 3 Metal capillary filter d Plastic capillary
filter
5.2.1.1
Replacing the plastic capillary filter with the metal capillary filter
1 Turn off the Ecotec E3000. 2 Unscrew the plastic capillary filter. 3 Insert
the plugwasher, see fig. 6 on page 19. 4 Screw in the metal capillary filter
on the sniffer probe end. 5 Calibrate the Ecotec E3000, see “6.4.2
Calibration”, page 38.
5.2.1.2
Replacing the metal capillary filter with the plastic capillary filter
If you want to switch from a capillary filter made of metal to a capillary
filter made of plastic, you have to the remove the plugwasher. The plugwasher
sits on the steel capillary in the sniffer probe.
1 Turn off the Ecotec E3000.
2 Unscrew the filter.
3 Unscrew the two cross-head screws in the sniffer probe flange and take out
the sniffer probe.
4 Push the capillary out of the plastic sheath a bit from the top with a
narrow pin or a thin needle (about 0.5 mm). Make sure that the sinter filter
in the sniffer probe flange does not get lost in doing so
5 Remove the plugwasher from the sniffer probe.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Fig. 7: Pushing out the capillary 6 Put the sinter filter back in and tighten the sniffer probe on the handle.
5 Installation
19
7 Screw the plastic capillary filter on the sniffer probe. 8 Calibrate the Ecotec E3000, see “6.4.2 Calibration”, page 38.
5.2.1.3
Water protectiontip mounting and remove
With the help of the water protection tip, you can check test objects with low
surface moisture, e.g. condensation moisture, for tightness.
Short circuit hazard Sucked in liquid can destroy the device. Do not suction
in liquids with the device.
The water protection tip is screwed onto the sniffer probe end like the metal
capillary filter. The small plugwasher must also be placed below the end, see
“5.2.1.2 Replacing the metal capillary filter with the plastic capillary
filter”, page 19.
Fig. 8: Tightening the water conservation sniffer probe
For renewed mounting of the plastic capillary filter, see “5.2.1.1 Replacing
the plastic capillary filter with the metal capillary filter”, page 19.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
20
5 Installation
5.2.2
Attaching the sniffer lines holder
A holder is available for the sniffer probe. The holder can be installed on
the right or the left of the device.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Fig. 9: Sniffer line holder mounting The holder has two hooks that are
inserted into two slits on the front panel of the device. The holder is pulled
to the front panel of the device with a magnet on its rear.
5.3 Connecting the ECO-Check calibrated leak
A built-in calibrated leak (ECO-Check) and different external calibrated leaks
are available as accessories available for the Ecotec E3000, see “10.1
Accessories”, page 81. Please refer to the ECO-Check installation manual on
how to connect the ECO-Check.
5.4 Connecting the external display unit to the Ecotec E3000RC
Connect the external display unit and the Ecotec E3000RC with the appropriate
connecting cable. Secure the plug on the socket by tightening the screws.
Fig. 10: The Ecotec E3000RC with external display unit as a table device (left), for rack mounting (right)
5 Installation
21
5.5 Connecting to the power supply system
The supply voltage of the Ecotec E3000 is specified on the labeling of the
mains plug (rating plate). The Ecotec E3000 cannot be switched for other
supply voltages.
Danger from electric shock Improperly earthed or protected products may be
dangerous to life in case of a fault. The use of the device is not permitted
without a connected protective conductor. Only use the included 3-wire power
cable. Make sure that the mains plug is always accessible.
Connect the device to the electric power supply with the supplied power cable.
5.5.1
Connecting to a PC
The connection is made with a commercially available 9-pin SUB-D plug. Please
refer to the “Interface description Ecotec E3000” (doc. no. kins22e1) for
further information on data exchange.
5.5.2
Connecting with a PLC
The connection is made with a commercially available 25-pin SUB-D plug. Please
refer to the “Interface description Ecotec E3000” (doc. no. kins22e1) for
further information on data exchange.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
22
5 Installation
6 Operation
6.1 Start-up
Connect a sniffer line and switch on the device with the mains plug. The
Ecotec E3000 starts a self-test lasting several minutes. The display shows the
headline “Running up” and the individual steps of the self-test.
Start screen in EN
Fig. 11: The device runs up
After running up, the Ecotec E3000 already measures the gas concentration in
the environment. There is no separate start function. But you have to still
calibrate the device and make different settings for your intended
measurement.
If the ECO-Check calibrated leak is not in the Ecotec E3000, an acoustic
warning signal and the warning 71 are output during the first start-up.
To stop the alarm quickly, press the key on the bottom right (labeled “OK”).
If you are working without the ECO-Check, you should deactivate the alarm
permanently, see “ECOCheck”, page 34.
6.2 Operating the device
6.2.1
Display and keys
All settings are made with the eight keys to the left and the right of the
display. The function allocations of the keys change depending on the current
operating step. The function allocation is shown directly next to the key
which allows fast and purposeful operation after a short learning period.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Fig. 12: Start display after the self-test
6 Operation
23
6.2.1.1 Recurring function symbols
The keys are always assigned with the following functions and labeled with the
shown symbols.
Setting the volume for speakers and head phones. Set volume: The set volume is
displayed on the bottom edge of the display. Value range: 0 (off ) to 10
(max.)
Back
Calling the main menu. Calling a window again that was closed with .
Calling up calibration.
The lower edge of the display shows “Zero” if a zero point has been set since
the device start-up. Calling up information: software version, operating
hours, serial number, date & time, alarm profile. Back to last menu level.
Navigating in a list box.
Pressing the key assigns the allocation “0” to that key and “1” to an adjacent
key. The same setting option exists for the numbers “2/3”, “4/5”, “6/7”, and
“8/ 9”. Closing the window and calling up the measurement view. Back to the
window with Measurement with I·Guide: calling up the list of I·Guide
programs.
Calling up help for the current function.
Confirming an entry or selection.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
24
6 Operation
6.2.2
Elements of the measurement view
The measured leak rates are shown numerically and with logarithmically divided
bar graph displays.
The other elements of the measurement view are shown in the following figure.
a bcd
ef
g
h
i
Fig. 13: Elements of the measurement view
1 Gas
6 Leak rate unit
2 Bell: search threshold exceeded; flashing bell: 7 Marking arrow: marks the measurement
trigger value exceeded
displayed on the sniffer handle
3 Search threshold (broken line)
8 Bar graph, logarithmic
4 Trigger value
9 Status bar: symbols and text overlay provide
5 Numerical leak rate display
information about the device status
The two center keys on the left side of the display can be used to adjust the volume of the alarm signal at any time. If one of the two keys is pressed, the device emits a sound with the selected volume through the speaker and shows the setting with a bar graph in the status line. The set value is also the first entry in the status line on the bottom of the display and applies only to the speaker of the main unit. To adjust different alarm profiles, see page 29.
Menu key The key on the bottom left of the display has two functions:
Calling the main menu. Returning you to the last window that was closed with .
Calibration key (CAL) The key on the top right next to the display can be used to start a calibration of the Ecotec E3000 with an external test leak at any time. Additional information on completing an external calibration, see “6.4.2.2 External calibration with external calibrated leak”, page 40.
ZERO key Pressing the ZERO key briefly stores the currently displayed leak rate as zero point for all selected refrigerants. Pressing the ZERO key for more than 2 seconds switches off the ZERO function. The display ZERO disappears from the status line in this case. For more information on the ZERO function, see “Zero”, page 31.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
6
Operation
25
Information key
Pressing the information key i (to the bottom right of the display) shows
information on
the status of the Ecotec E3000. Details: see “6.4.6.1 Calling up information
on the measurement”, page 46.
Status line Status information is displayed in the lower line of the
measurement window. The set volume for the alarm is displayed in the left of
the line.
If the small speaker flashes, it indicates that the device speaker is switched
off. If the number flashes, it indicates that the alarm delay is switched on,
see “Alarm delay”, page 29
A small black triangle with an exclamation point next to it can indicate an
active warning.
If the ZERO function is activated, the word “ZERO” follows in the status line.
If the first cathode (filament A) of the mass spectrometer has been consumed
and the Ecotec E3000 switches automatically to the second cathode (filament
B), the display will show “Fil. B” in the status line.
If you working with an activated IGS, “IGS” will be shown in the status line.
6.2.3
Operating elements and display on the sniffer handle
The display of the sniffer handle shows the most important information for the
current measurement. Measurements can be controlled with the two keys.
a
b
2.0E-7
c HE
d
e
Fig. 14: Operating elements and display on the sniffer handle
1 Leak rate as bar graph 2 Leak rate, numerical. Unit, as shown on the main
unit 3 Measured gas 4 Left key, assigned “Zero” e Right key, different
assignments
The measured leak rate is shown as an increasing or decreasing bar. The second
line shows the leak rate as a numerical figure (with the same measurement unit
as in the main display). The third line shows the abbreviation for the
measured gas.
Depending on the measurement, the display can also show something else, e.g.
“Error” or the number of a warning.
If you search for several gases simultaneously, you can switch between
individual measurement results with the right key. The right key can also be
used to confirm messages or states during a measurement cycle.
You can trigger the ZERO function with the left key, see “Zero”, page 31.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
26
6 Operation
The key can be deactivated to prevent accidental triggering: Press the key until you hear a signal. The key is reactivated by pressing it longer.
6.2.4
Special features of the Ecotec E3000RC
The Ecotec E3000RC has a connection panel for the external display unit
instead of the built-in display. Two LEDs (to the left of the plug) provide
information on the status of the Ecotec E3000RC, even if the external display
unit is not connected.
The green LED shows that the Ecotec E3000RC is switched on. It has a steady
green light if an external display is connected and flashes if no external
display can be detected.
The red LED flashes in case of an error message; a steady red light indicates
a warning.
Fig. 15: Connection panel with LEDs
If no display unit is connected, you can confirm error messages and warnings
by pressing both keys on the sniffer line simultaneously.
The external display unit has four keys:
The menu key opens the main menu. The current measured background value is set
as zero point with the ZERO key, see
“Zero”, page 31. The START / STOP keys do not have any function (the external
display unit can also be
used with other leak detectors made by INFICON that require these keys).
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
6 Operation
27
6.3 Settings before measurements
Before the first measurements, you should make different device settings in
the following menus: Miscellaneous Display Audio Vacuum & access control
Interfaces (for control via interfaces and if you use an ECO-Check) You can
access the menus via the main menu
6.3.1 Miscellaneous
Language
Time & Date Sniffer light Pressure unit Leak rate filter
You can select one of the following languages:
English (factory setting) German French Italian Portuguese Spanish Japanese
(Katakana) Chinese (Mandarin, simplified Chinese)
To set the language to English temporarily, press keys two and six during the
run-up of the Ecotec. Call up the language setting after the run-up and set
the required language.
First page: internal date with the format DD.MM.YYYY Second page (press key on
the bottom right ) time with the format SS:MM.
Activate/deactivate light Adjust brightness from 1 (min) to 6 (max)
atm Torr Pa mbar
Auto Fixed I·Filter
The I·Filter is an intelligent filter algorithm that delivers the best results
regarding interference suppression and stability of the leak rate signal. It
was especially developed for the use in the Ecotec E3000.
Only in cases in which the older Ecotec II model was replaced with an Ecotec
E3000 and the Ecotec E3000 is used in a fixed test device can it be necessary
to select the older filter settings “Auto” or “Fixed”.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
28
6 Operation
Alarm delay Wake up
With very unstable surface conditions, it can be expedient to emit an acoustic
alarm only if the trigger value is exceeded for a specified period of time. If
the function is activated, the number indicating the volume of the device
speaker flashes in the status bar.
Setting range: 0 to 9.9 seconds, in tenths of a second
Although the alarm delay is activated thedevice possibly does not warn when it
finds a leak. So that every test successes do the following.
Please note the measurement display in the device and the handle.
Deactivate the function when surface conditions are stable again.
If the Ecotec E3000 is in idle state (sleep), a time can be set for it to
start automatically. This way you can set the Ecotec up to run through its
warm-up phase already before the shift starts.
You can set a separate wake-up time exact to the minute for each day of the
week.
To deactivate the wake up again, set the wake-up time to 00:00.
6.3.2 Audio settings
Beep sound Device speaker Audio sniffer
Alarm profile
You can switch off the signals that indicate the completion of specific
functions.
You can switch off the speaker installed in the main unit. This does not
influence the speaker connection.
You can choose whether the speaker in the sniffer handle should signal the
exceedance of search threshold or trigger value. You can also switch off the
speaker completely.
You can assign one of three alarm profiles to the device speaker: Pinpoint
Setpoint Trigger/alarm
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
6 Operation
29
Table 4: Features of the alarm profiles Alarm profile pinpoint
Search value exceeded Trigger value exceeded Acoustic tracing of the
measurement result
Comment
< 1/10 trigger value: low frequency >1/10 trigger value up to 10 trigger
value: rising frequency > 10 trigger value: high frequency Recommended, for
precise leak localization
Alarm profile setpoint
Acoustic signal low frequency
Acoustic signal high frequency
Alarm profile trigger/alarm Acoustic signal low frequency Two-tone signal
Three different two-tone
signals available for selection.
This allows identification with a signal for devices that are used together.
Volume
You can adjust the volume that cannot be set lower with the plus and minus
keys next to the display. This way you can prevent the acoustic signals from
being switched off accidentally during the measurement.
The adjustment applies to the speaker in the main unit and the head phones.
You can also set the current volume of speaker and head phones here.
Setting range: 0 to 15
Hearing damage from excessively loud signals The volume of signals can exceed 85 dB(A). Keep away from the device if the volume has been set to high. Wear ear protection, when needed.
6.3.3
Display settings
In the menu “Settings > Display”, you can set the display details of the
device display and the display in the sniffer handle.
Contrast
Increase or decrease the display contrast using the arrow buttons. Holding the
button scrolls through the values. The new setting takes immediate effect in
the display.
To adjust the contrast to the currently displayed menu, select “Automatic”. To
display a dark background and the script in bright characters, select “Invert
display”. If the display is no longer legible because it is too dark or too
bright, you can reset the setting as follows:
1 Switch the device off and back on.
2 Press keys 3 and 7 during the run-up until the display can be recognized
again.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
30
6 Operation
Peak hold
3 Call up the window for the contrast setting and confirm the new value. The
device will otherwise use the old, unrecognizable setting after the next
start-up.
The factory setting with black font on light background can be inverted. This
causes the display to emit less light overall.
You can set if and how long the highest measured value is to be displayed
additionally below the current leak rate.
Setting range: 0 to 20 seconds.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
a
Fig. 16: Measurement view with maximum value display 1 Maximum value
6.3.3.1
Gas display handle
In the menu “Settings > Display > Gas display handle” you can define which gas
is shown in the display in the sniffer handle.
Automatic
The display always shows the gas that the most of is measured currently. If a trigger value is exceeded, this gas will be displayed.
Manual
You can switch between gases with the right key on the handle.
Auto with hold
You can switch to another gas with the right key on the handle. After the hold time has elapsed, the display again shows the gas that the most of is measured currently.
Hold time
The hold time can be set to 5, 10, 15, or 20 seconds.
6.3.4 Vacuum & access control
Zero
The gas concentration contained in the measurement environment can be set as
zero point for measurement (background suppression). The function has the
brief name “Zero”.
If the gas concentration drips after setting the zero point, a negative
measured value should be displayed. In order to avoid that, the zero point is
revised downward if the measured value is negative for the duration of the
“Zero time”, see below.
The zero point is not automatically revised up (wards). It is therefore
important to regularly reset the zero point.
The zero point can be set with the left key on the handle and with the key
“Zero” in the measurement view.
6 Operation
31
Zero time Flow limits
You can activate or deactivate the keys here in this menu. Deactivation
prevents that the function is triggered inadvertently and that an incorrect
absolute measured value is thus displayed.
The key on the sniffer line handle can also be activated or deactivated by
pressing the key longer.
Zero time is the time during which the leak rate must be negative so that the
zero point can be revised downward automatically. The best setting depends on
your measurement conditions (sampling speed, gas background, test specimen).
Setting range: 1 to 9.9 s
To detect a leak on the 160-sccm capillary, an upper limit value is set. If
the value is exceeded, the system issues the warning “Capillary broken”. With
longer exceedance, device components are also switched off to protect them.
To detect a clogging of the 160-sccm capillary, a lower limit value is set. If
the value is fallen short of, the system issues warning “Changed flow!”. With
a strong lower deviation, the error message “Flow through capillary too low”
is issued.
Possible settings range from 160 to 999 sccm or 0 to 160 sccm.
The closer the lower limit value is set to the actual flow rate, the more
sensitive the Ecotec E3000 reacts to a beginning clogging of the filter and
the sniffer line.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Sensitivity
Fig. 17: Setting flow limits
The flow through the sniffer line depends on the atmospheric pressure of the
environment. If you operate the device at great height, the flow rate through
the sniffer line can drop considerably, approx. 20 % per 1000 m height. Reset
the flow limits accordingly in this case.
The sensitivity check in the device ensures that the sensitivity of the Ecotec
E3000 is always sufficient. The entire gas flow from sniffer probe to sensor
is monitored and the software simultaneously checks whether the Ecotec E3000
can determine the correct signal strength with it. The monitoring ensures that
the Ecotec E3000 does not become insensitive and that leaks are not detected
without the user noticing. If the sensitivity is reduced, the error message
“Sensitivity too low” is issued. Recalibration can restore sensitivity in this
case, see “6.4.2 Calibration”, page 38. The error message is repeated every 15
seconds until calibration is started.
32
6 Operation
Monitoring check
Fig. 18: Monitoring check
We urgently recommend leaving the monitoring switched on at all times. It
should be deactivated only for measurements in an argon-free environment
because monitoring requires the argon signal.
Cathode automatically
When the function is activated, each time the E3000 is switched on, the other Cathode switched. This ensures that both cathodes are used equally.
Calibration
In the window “Calibration” you can activate or deactivate the external calibration. In this way, you prevent an external calibration that was previously carried out inadvertently is overwritten. More about calibration, see “6.4.2 Calibration”, page 38.
Changing the menu PIN
You can protect access to the settings with a PIN.
To prevent entry errors, you have to enter the PIN twice. After confirmation
with “OK”, the main menu is displayed and the PIN is effective immediately.
To cancel protection again, enter “0000” as the new PIN (factory setting).
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Fig. 19: Defining the menu PIN
6.3.5
Interfaces
In “Settings > Interfaces”, you enter the settings for the interfaces and for the ECO-Check. Please refer to the interface description (kins22e1) for detailed information on interfaces.
Control location
Local RS-232 Local and RS-232
Local: Only measurements can be read out via the RS-232 interface. They are
not available to control the device.
6 Operation
33
Recorder outputs > Recorder output scale
Recorder outputs > Recorder output gas PLC Setup > Define PLC inputs (outputs)
RS-232 settings
RS-232 protocol
ECO-Check
RS-232: The Ecotec E3000 is controlled almost exclusively via the interface.
The display serves only for a visual check. Some settings can be changed on
the device. Please use the protection via access PIN if all functions on the
device are to be inaccessible, see “Changing the menu PIN”, page 33. Local and
RS-232: The Ecotec E3000 can be controlled via interface and entries on the
device.
Linear Logarithmic The output is on channel 1 (pin 1 of the I/O connection).
Gas 1 to 4 Auto
You will assign different commands to the pins of the I/O connection here.
Baud rate 1200/2400/4800/9600/19200/38400 End sign LF/CR/CR + LF
LD ASCII Diagnosis Printer Auto Printer Manual
If you do not use an ECO-Check calibrated leak, you should choose “Deactivate”
here. Otherwise, warning 71 “No communication with ECO-Check” will be issued
at every start of the Ecotec E3000. If you use an ECO-Check calibrated leak,
you can set up a warning for the time of use here: 14/30/60/90 days.
6.4 Settings for the measurements
With delivery, the data are programmed for the following gases and the
measurement results are shown below each other in the display: R134a R32 R600a
He
The gas selection can be changed at any time.
6.4.1
Selecting the gas, changing gas parameters, activating measurement
You can access the settings via the main menu Select “Measuring parameters”
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
34
6 Operation
Fig. 20: Gas information for the measurement
The display shows:
four gases the respective mass position the respective trigger value the
addition “Deactivated” if the associated gas is currently not searched for,
see below.
You can also define your own gas with the key “Gas def.”, see “6.4.5 Setting a
user-defined gas”, page 44.
Press the key to the right of the gas whose parameters you want to change. The
window “Measuring param. gas …” will open.
Fig. 21: Changing the settings for gas 1
With the keys up and down you select a setting. Pressing the key “change” on
the bottom right opens the associated settings menu.
Gas
The gas to search for is displayed. You can open the gas library with “change” and choose
from approx. 100 gases there.
User-defined gases are displayed at the end of the list.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
6 Operation
35
Status Trigger and unit
Fig. 22: The gas library
“Status” shows whether the search for this gas is activated or deactivated.
You can change the setting at any time. If the search for a gas is
deactivated, it will also be displayed in the window “Measuring parameters”.
In the measurement window, the display of the measurement result for a
deactivated gas is dropped and the display is clearer.
In the window “Trigger & unit”, you can set the trigger value with the keys on
the left and select the unit with the keys on the right.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Search & Limit (search threshold and display limit)
Fig. 23: Setting trigger value and unit
The following table shows the adjustable units and the associated limits for the trigger value.
Table 5: Trigger value according to unit
Unit g/a oz/yr ppm mbar l/s Pa m3/s atm cc/s Torr l/s sft³/yr
Lower trigger value limit Upper trigger value limit
0.1
1000
0.004
100
1 2 10-7 2 10-8 2 10-7 2 10-7 2 10-4
999999 9.9 10-2 9.9 10-3 9.9 10-2 9.9 10-2 9.9 10+1
In the window “Search threshold display limit”, you can set the search threshold with the keys on the left and select the factor for the lower display limit with the keys on the right.
36
6 Operation
The search threshold is a percentage of the trigger value and serves as an
additional warning level. Smaller leaks that are even below the trigger value
can also be reported, when needed, see “Alarm profile”, page 29.
The absolute value of the search threshold is calculated by the device and
displayed.
Internal calibration Mass and position
Fig. 24: Setting display limit and search threshold
With the function “Display limit”, you can hide measurement results that are
below the expected leak rate. This makes the measurement view clearer
especially in view of the measurement bar because smaller measurement results
are hidden.
You can define the lower display limit as a multiple of the lowest measurable
leak rate (1 , 2 , 5 , 10 , 20 , 50 , 100 ).
You can deactivate internal calibration. If it is deactivated, only the more
precise external calibration can be made for the gas, see “6.4.2 Calibration”,
page 38.
Internal calibration is precluded from the start if the mass position of a gas
is outside the range from 40 to 105 amu.
With the selection of a gas from the gas library, a standard mass position is
automatically selected for gas to be measured. If the device could respond to
other substances in the working environment for the leak test, we recommend
selecting another mass position for the verification of the required gas. You
can find a list of all possible gases with their normal and alternative mass
positions in the Appendix, see page 82.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Cal. factor
Fig. 25: Selection of another mass position
Below the selected mass is the information whether it is the preferred mass or
not. In addition, the molecular mass of the gas as well as the height of the
peak in relation to the highest peak for this gas are displayed. The standard
factor is a measure of the sensitivity of the device for the gas at the set
mass position.
The calibration factor is displayed in this line.
6 Operation
37
Last cal. Cal. method
Based on this line, you can check when the last calibration took place. The line shows whether calibration was external or internal.
6.4.2
Calibration
The Ecotec E3000 can be calibrated most conveniently by adding an ECO-Check
calibrated leak. The ECO-Check can be integrated in the front panel of the
device or set up at the testing position. It compensates temperature
fluctuation and thus enables the accuracy required for calibration.
The ECO-Check calibrated leak includes R134a. It can be used for the
calibration of gases with a mass position from 40 to 105 amu because the
Ecotec E3000 converts the calibration result for the measurement of these
gases.
You get the most accurate calibration with the external calibrated leak. The
calibrated leaks are valid for one gas each and temperature-insensitive.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
38
6 Operation
When to calibrate?
The device should be calibrated daily and after every operator change.
Calibration is additionally required after the following events:
Sniffer line replacement Sniffer line probe replacement Switch between gases
(if you calibrate with an external calibrated leak) Filter replacement Prompt
for calibration by the system
6.4.2.1 Internal calibration with ECO-Check
Incorrect calibration because of operating temperature that is too low
Calibrating the device in the cold state can deliver incorrect measurement
results. The device must have been switched on for at least 60 minutes before
calibration for
hydrogen measurement. The device must have been switched on for at least 20
minutes before calibration for
the measurement of all gases.
The ECO-Check must be installed; refer to installation manual of the ECO-
Check.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Fig. 26: Built-in ECO-Check calibrated leak
If a gas measurement cannot be calibrated with the ECO-Check because the gas
mass position is outside of 40 to 105 amu, the message “Int. calibration
impossible” is displayed for this gas subsequent to calibration.
If a gas was blocked for internal calibration in the menu “Settings gas”, the
message “Gas deactivated” (see “Internal calibration”, page 37) is displayed.
The Ecotec E3000 notices if you insert the sniffer tip in the opening of the
calibrated leak and starts calibration automatically. Messages subsequently
guide you through the calibration process.
A warning is displayed if the device has not be switched on for 20 minutes
yet. Simply confirm the warning and then continue with calibration if you know
that the device has
6 Operation
39
reached operating temperature because it was switched off only briefly before
calibration. Otherwise remove the sniffer probe again and start calibration
later.
After measuring and a short calculation period, the results of calibration are
shown in the display. The old and the new calibration factor as well as the
old and the new relative peak position are displayed.
In order to prevent that an earlier external and thus more accurate
calibration is accidentally overwritten, you have press the key “Confirm new
values” to complete the calibration.
1 Switch to measurement view.
2 Insert the sniffer tip in the opening of the ECO-Check until you feed some
resistance.
3 Press the right key on the sniffer handle as soon as the display shows the
line “Calibration: Press right key”.
4 Remove the sniffer tip from the reference leak if the display shows the line
“Remove sniffer from cal-opening”.
5 Confirm the new values with the key on the bottom right.
Check calibration (Test function)
If you insert the sniffer tip in the opening of the ECO-Check during
measurement operation, an examination of the calibration is started
automatically (test function). The device checks the measured value of the
ECO-Check while the sniffer probe is inside the opening of the calibrated
leak. The operator is subsequently prompted to remove the sniffer tip from the
opening of the calibrated leak.
For gases released for internal calibration, either “Test o.k.” or
“Recalibration required!” will appear. For gases not activated for internal
calibration, “Gas deactivated” will be displayed. For gases that cannot be
calibrated with the ECO-Check because of very high or very low mass positions,
“Gas spec. test not possible” is shown.
Press the key “OK” or the right key on the handle to return to measurement
operation.
6.4.2.2 External calibration with external calibrated leak
For external calibration of the Ecotec E3000, we recommend calibrated leaks
with leak rates > 2 g/a. If background concentrations in the testing
environment are clearly increased, a calibrated leak with a higher leak rate
is required.
External calibration is a semi-automatic process. Text messages in the display
guide you through the calibration process. Calibration can be stopped at any
time with the key “Cancel”.
Incorrect calibration because of operating temperature that is too low
Calibrating the device in the cold state can deliver incorrect measurement
results.
The device must have been switched on for at least 60 minutes before
calibration for hydrogen measurement.
The device must have been switched on for at least 20 minutes before
calibration for the measurement of all gases.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
40
6 Operation
A warning is displayed if the device has not be switched on for 20 minutes
yet. Simply confirm the warning and then continue with calibration if you know
that the device has reached operating temperature because it was switched off
only briefly before calibration. Otherwise start calibration again later.
The gas measurement to be calibrated is generally activated. If you want to
calibrate a deactivated measurement, activate the gas in the menu “Measuring
parameters”.
After measuring and a short calculation period, the results of calibration are
shown in the display. The old and the new calibration factor as well as the
old and the new relative peak position are displayed.
1 Switch to measurement view.
2 Press the key “Cal”. The list of gases currently set for measurement is
displayed (up to four gases).
3 Select the gas that the measurement is to be calibrated for.
4 Check whether the gas and the displayed leak rate match the data of the
calibrated leak. If the leak rate does not match, select “Leak rate change”
and correct the value.
5 Select “Start”.
6 Hold the sniffer tip into the center of calibrated leak opening and follow
the instructions in the display. If you have to wait until the air signal has
stabilized then this can take up to 30 seconds with a helium or hydrogen
calibration.
7 Confirm the new values with the key on the bottom right.
6.4.3
Gas equivalent to helium and hydrogen, settings for diluted gas
If you are looking for helium or hydrogen, you can have also have the
determined leak rate for the gas equivalent displayed, e.g. as R134a.
If you have an equivalent set, all displays will show the original gas
followed by the equivalent in brackets. Example: He (R134a)
Proceed as follows to set a gas equivalent:
1 Select helium or hydrogen from the gas library.
2 Select the line “Status” in the window “Settings gas …” and press “Change”.
3 Select “Equivalent name” in the window that opens. The gas library will be
displayed again.
4 Select the equivalent name and confirm with “OK”.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
6 Operation
41
Fig. 27: Example for a measurement view with helium as refrigerant equivalent
You can also take a different pressure and/or different concentration between
original gas and gas equivalent into consideration.
An internal conversion of the measurement result thus allows the Ecotec E3000
to approximate the result of a leak pre-check close to the result of a main
leak examination.
Enter the settings for concentration and pressure in the window “Settings gas
… > Status > Change > Equivalent settings”.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Switching off the gas equivalent function
Fig. 28: Setting the parameters for gas equivalent and diluted gas
In this window, you can enter the helium or hydrogen filling pressure. You can
enter the pressure for the equivalent gas next to that.
The bottom right shows the correction factor of helium/Hydrogen based on the
gas equivalent. If a set of parameters that exceeds the limits of the Ecotec
E3000 is entered, the correction factor is displayed with inverted colors. In
this case, adjust the parameter until the correction factor display returns to
normal.
On the bottom left, you can enter the value for the gas concentration. If you
are looking for diluted helium or hydrogen, you can take the dilution into
consideration in this setting. The leak rate for the undiluted gas is then
displayed as the measured value.
Select “OK” when all parameters are set correctly.
Please note: If you are working with diluted gas, you have to select the
original gas as gas equivalent, i.e. gas and gas equivalent are identical.
To switch off the gas equivalent setting, select the last entry from the gas
library (Settings gas … > Status > Change > Equivalent name).
42
6 Operatio
Fig. 29: The last entry switches off the gas equivalent function
6.4.4
Suppressing interfering gases (Sophisticated interfering gas suppression –
IGS)
IGS causes the findings of the interfering gases cyclopentane, isopentane as
well as any mixture thereof to be suppressed during the search for refrigerant
R600a. Up to an interfering gas concentration of 50 g/a, the error is only 1 %
at most.
If IGS is activated, only one additional gas can be entered in the list of
sample gases. If more than two gases are activated for R600a when IGS is
activated, the additional gases (starting with the highest gas number from 1
to 4) will automatically be deactivated so that there are only two gases left.
If R600a is measured with IGS and R134a is set as second gas, select mass
position 83 for R134a because malfunctions between propellant and R134a will
occur otherwise.
IGS requires little maintenance. If repeated false alarms occur during the
sniffing with IGS, calibration must be set to interfering gases, see below.
Activating IGS
1 Select R600a as the gas to be searched for: “Measuring parameters > Gas… >
Gas > Change > R600a > OK”.
2 Switch to mass setting in the window “Measuring param. gas …”.
3 Browse the available mass positions until “IGS” is displayed in the line for
the preferred mass.
4 Confirm with “OK”.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
IGS alignment
Fig. 30: Selection of the IGS mode for R600a
The previous, actual calibration is carried out with an external test leak as
it is with other gases. For the additional IGS alignment, you need a
cyclopentane test leak and an isopentane test leak, available as “Calibration
set for IGS mode”.
6 Operation
43
The Ecotec E3000 recognizes errors you make in the gas sequence during the
alignment and notifies you with a flashing gas display.
1 Activate IGS, see above.
2 Select “Cal” in the measurement window.
3 Select the gas R600a in the window “Select gas”. The entry should be
expanded with IGS.
4 Select “IGS alignment” in the window “Start external calibration” and follow
the instructions on the display.
5 Confirm the calibration with “OK”.
6.4.5
Setting a user-defined gas
You can save the settings for six individual gases. In the main menu, select
“Measuring parameters > Gas def.”, then one entry and
“Change”.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
User library no. Name
Massposition
Normalisation factor
Fig. 31: Settings for a user-defined gas
You can switch between settings with the arrow keys and open a settings window
by pressing the “change” key.
Shows the number of the gas definition to be edited currently.
You have to enter a name for the gas to be defined.
The buttons are assigned with letters and you can assign names with six
characters with them. Once you have entered the sixth character, you can leave
the window by selecting “OK”.
The mass determines the position of the peak at which the user-defined gas is
measured. The Ecotec E3000 can verify masses from 2 to 200 amu.
The buttons are with assigned with numbers and you can enter masses between 2
and 200.
The standard factor is used to convert the current supplied by the sensor into
a leak rate signal. If you are setting a user-defined gas, the Ecotec E3000
should subsequently be calibrated with an external test leak, if possible. If
calibration is successful, you will not modify the standard factor. If
calibration fails and the error message “Calibration factor too high” is
displayed, the standard factor must be reduced by decade, e.g. from 1.0E+08 to
1.0E+07. If the error message “Calibration factor too low” is displayed,
increase the standard factor by a decade, e.g. from 1.0E+08 to 1.0E+09. Repeat
this process until Ecotec E3000 calibration is completed successfully.
44
6 Operation
Molecule mass
Enter the molecular mass of the gas to be measured with the arrow keys (usually included in the data sheet for the gas).
6.4.6 Measuring
Electric shock hazard
Electrical voltages can be transmitted via sniffer tip and can cause damage to
property or personal injury.
Do not touch live parts with the sniffer probe.
Disconnect electrically operated test objects from the mains before starting
the leak test and secure them against being restarted without authorization.
Danger of eye damage LEDs generate a bundled light that can damage eyes. Do not look into the LEDs from a short distance or for a longer period of time.
Electric shock hazard
Sucked in liquids can trigger short circuits and cause damage to property or
personal injury.
Do not suction in liquids with the device.
Use the water conservation tip in humid environments, see “5.2.1.3 Water
protectiontip mounting and remove”, page 20.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Requirements
Material damage due to a missing sniffer line The device must not be operated
without a connected sniffer line in order to avoid overpressure in pump and
measurement system. Connect the sniffer line before you start up the device.
Do not replace the sniffer line while the device is in operation.
Measurement requires: A sniffer line connected to the main unit. The device
has run up and warmed up, see “6.1 Start-up”, page 23. The device is
calibrated, see “6.4.2 Calibration”, page 38.
6 Operation
45
You have made the device settings requires for measurement, see “6.3 Settings
before measurements”, page 28.
You have made the measurement settings requires for measurement, see “6.4
Settings for the measurements”, page 34.
Measuring position and speed
Hold the sniffer probe to the possible leak as close as possible. The tip may even touch the test specimen. If a weldseam or the like is to be tested, the tip should be moved along the path at a speed of less than 10 cm/s. Please also note the minimum measurement time for the search for helium, see “Special features with helium”, page 56.
Test sequence
1 Keep the sniffer probe away from possible gas sources and press the left key
on the sniffer handle (ZERO).
2 Sniff the test specimen.
If there is a leak, it will be reported in the displays, with flashing LEDs in
the sniffer handle and – depending on your settings – also with an acoustic
signal.
Because of the high measuring sensitivity of the device and because
interfering gases can falsify the measurement result, you should repeat the
measurement if a leak has been reported. Remember to suppress the background
again beforehand (press the left key on the sniffer handle).
Measuring with unstable surface conditions
With very unstable surface conditions, it can be expedient to emit an alarm only if the trigger value is exceeded for a specified period of time, see “Alarm delay”, page 29.
6.4.6.1
Calling up information on the measurement
Press the key to obtain information on the current measurement:
Software version Operating hours Serial number Date and time Alarm profile
Selected gases with mass position and trigger value. Gases that are set but
that are not
currently sought for are marked.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Fig. 32: Information on the measurement If there is an error message or a warning then this will be displayed instead of the gases.
46
6 Operation
Fig. 33: Information page with warning
6.4.7
Measuring with I·Guide
I·Guide was developed to support the operator in the application of the
correct technique for leak tests.
With an I·Guide program, a time sequence and a repetition rate are predefined
for sniffing for one or two gases. The following details can be set:
one or two gases trigger values for the gases number of measuring points
measurement time per measuring point idle time between measurements
(transition to next measuring point) maximum permissible overall leak rate for
the part to be tested
The right key on the sniffer handle must be pressed in order to confirm the
measurement at an individual measuring point. Confirmation can alternatively
take place via the interface programming.
You can set ten I·Guide programs.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Fig. 34: Measurement view in an I·Guide program
I·Guide as a timer signal
You can also do without the calculation of an overall leak rate. The I·Guide program then serves merely as a default for time-controlled measuring. Set the number of measuring points to zero for this.
Capturing the results of a long measurement series with I·Guide
You can combine the leak rates of a maximum of 99 measurements with I·Guide. Set the number of measuring points to 99 for that. If you subsequently press the right key on the sniffer handle for two seconds during the measurement, a result window with the individual measurements and the overall leak rate will be displayed. After the 98th measuring point, the result is displayed automatically.
6 Operation
47
6.4.7.1
Setting the I·Guide program
1 Select “Main menu > Settings > Setup I·Guide”. 2 To activate I·Guide select
“Activate”. 3 To enable confirmation with the right key on the sniffer handle,
select “Key on”. Control
is otherwise possible only via interface. 4 Select one of the 10 programs with
the arrow keys and then press “Change”.
Name
Gas A
Gas B Glob. Trigger A/ Glob. Trigger B
Number of measuring points Measuring time
Wait time
Fig. 35: List of I·Guide programs 5 Select the setting you wan to change with
the arrow keys and press “Change”. 6 Adjust the setting as described in the
following are and confirm with “OK”.
Name of the program. A name comprises six characters. After you have entered
the last character, you can confirm with “OK”.
One of the four gases can be selected for the search, see “6.4.1 Selecting the
gas, changing gas parameters, activating measurement”, page 34.
See above. If no second gas is to be searched for, set this to “0”.
The maximum permissible leak rate for the addition of all measurements of an
I·Guide program can be set here. The trigger value for a single measurement
complies with the trigger value originally set for a gas.
0 to 99.
1 to 25 seconds. You must not set the measurement time to be shorter than the
response time of the device, see “Table 2: Technical data” on page 14.
You can set a time between 0.1 and 25 seconds for the transition to the next
measuring point.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
48
6 Operation
Fig. 36: Settings for an I·Guide program
6.4.7.2 Starting the I·Guide program
Messages in the display of the main unit, messages in the display of the handle and acoustic signals guide you through the program.
1 If due, calibrate the device externally. During the work with I·Guide, the device can be calibrated internally only.
2 Activate I·Guide in the I·Guide menu, see above. A message shows the gas that is being searched for. It is the gas from the first activated I·Guide program. Confirm with “OK”.
3 Switch to the main menu. Measurement will start immediately.
4 Follow the messages.
Table 6: User prompts I·Guide program
Process
Message in main device display
Idle time to
Tip to point …
move to the
measuring point
Prompt to confirm the position.
Confirm point …
Measuring
Measure point …
Measurement Tip to point … time elapsed
Message in handle display
For pos. …
OK? Pos. … Confirm with the right key if the sniffer probe is at the measuring
point. Measure pos. … For pos. …
Acoustic Acoustic
signal main signal
unit
handle
Ticking
Short signal Short signal
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
6 Operation
49
Table 6: User prompts I·Guide program
Cycle completed Cycle number of measurement Measured gas
Long signal Long signal
Measured gas
Sum of measured leak rates
Single leak rates and sum of measured leak rates
“OK!” with a total leak rate below the limit value
“OK!” with a total leak rate below “Error!” with a total leak rate above
the limit value
the limit value or if the trigger
“Leak in the cycle!” with a total leak rate above the limit value or if the
value was exceeded in a single measurement.
trigger value was exceeded in a
single measurement.
If you are searching for two gases: If you are searching for two gases:
you can switch between the
the display switches automatically
displays for the two gases with the between the displays for the two
keys “A” and “B”.
gases.
Constant signal in case of an error
“Please wait” or “Wait” is displayed if you want to start the next measurement
by pressing the button before the idle time has elapsed.
During a cycle, you can use the key to move back one measuring point.
The key takes you back to the beginning of the cycle.
Fig. 37: Display after a successful measurement
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Changing the I·Guide program
Fig. 38: Search for gas R134a with I·Guide program 1: The trigger value and
the overall leak rate were exceeded at measuring points 1 and 3 The next
measurement cycle starts if you confirm with “OK” or press the right key on
the sniffer handle.
The new I·Guide program must be set and activated in the I·Guide settings, see
above.
50
6 Operation
Resetting the I·Guide cycle counter
Open the list of I·Guide programs in the main menu with and select one.
Measurement will start immediately.
In the window “Setup I·Guide”, you can reset the I·Guide cycle counter
manually with the key “Reset counter”. The cycle counter is always set to zero
when the device is switched off.
Fig. 39: Cycle counter reset
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
6 Operation
51
Calling up information on the I·Guide program
Press the key to obtain information on the current measurement:
Software version Operating hours Serial number Date and time Alarm profile
Information on the current I·Guide program
If two gases were selected for the I·Guide program, gas type A/B and leak rate
sums A/B are displayed alternately.
Fig. 40: I·Guide information page
If there is an error message or a warning then this will be displayed instead
of the I·Guide program information.
6.5
Idle state (sleep)
The idle state is a meaningful alternative to switching the device off only if
you allow the device to run up automatically with the wake-up function, see
“Wake up”, page 29.
If you press the SLEEP key in the main menu of the device, the device goes
into idle state. The mass spectrometer is switched off and the pumps are
stopped.
“Sleep” is replaced with “START” In the main menu and the Ecotec E3000 is run
up again with “START” or the wake-up function.
The electrical components cannot maintain operating temperature in the idle
state. After a restart, precise measurements are therefore possible only after
the warm-up time, see “6.4.2 Calibration”, page 38.
6.6 Service
The service menu is password-protected. Settings in the service menu can be
made only after a special INFICON service training.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
52
6 Operation
6.7
Calling up information about the device
You can call up all information about the device by selecting “Info” in the main menu. The information is spread out over nine pages. The key on the bottom right switches you to the next page and the key on the bottom left takes you back one page. The page number is displayed in the upper right corner.
Table 7: Device information
Menu item
Format
Description
Page 1: General data
Backing pressure mbar
Flow
sccm
Flow through the sniffer line
Total pressure
mbar
Pressure in the mass spectrometer
Time since power on min
Operating hours h
Serial number
9000 XXX XXXX
Software version x.xx.xx
Electronic system °C / °F temp.
Temperature of CPU board
TSP temperature °C / °F
Temperature of mass spectrometer
Calibrated leak temperature
°C / °F
Temperature of ECO-Check
Page 2: Turbo molecular pump data
State
On / Off / Run-up
Curr. error code
Speed
Hz
Current
A
Voltage
V
Driving power
W
Operating hours of h the TMP
Operating hours of the turbo molecular pump
Operating Hours TC h
Operating hours of the frequency converter for TMP
Run-up time
s
Software version XXXXXX
Page 3: Transpector data
Configuration
Box version
x.xx
Control SW version x.xx
Measure SW version x.xx
Filament
A or B/A or B Set cathode / active cathode
Power on time
h
Emission on time A h
Emission on time B h
Serial number box
Serial number sensor
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
6 Operation
53
Table 7: Device information (cont.)
Menu item
Format
Page 4: ECO-Check Data
Description
Gas
Rxxx
Gas of internal leak
Leak rate nom. / at T x.x g/a / x.x g/a Nominal leak rate / leak rate at current temperature
Version / Checks.
x.x / hexadecimal Software version with checksum code
Serial no.
9000 XXX XXXX
Serial no. reservoir 9000 XXX XXXX
Filling date
DD.MM.YYYY
Expiration date Gain / Offset
DD.MM.YYYY
Test leak temperature
°C / °F
Status of light barrier
Parameters for the temperature measurement in the ECO-Check
Page 5: Sniffer line information
Type Software version
SL3000 / system x.x
Length
3 m / 5 m / 10 m / 15 m
Serial no.
9000 XXX XXXX
Key left Key right
Background light Green/red
Background light
Bar graph Flow (calibration)
sccm
Motion
Page 6: I/O port data
Recorder A
V
Recorder B
V
Sleep Zero
Low / High Low / High
Gas a/b/select Input reserved
Low / Low / Low High / High / High
Low / High
Leak/Ready/Error Relay “Leak”
Low / Low / Low High / High / High
Low / High
Relay “Leak”
Low / High
Output reserved Low / High Page 7: Analog data
AIN3 sniffer length V
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
54
6 Operation
Table 7: Device information (cont.)
Menu item AIN4 +5V II leak
Format V
Description
AIN5 +24V III ext. V
AIN6 +5V I sniffer V AIN8 -15V MC50 V
AIN9 +15V MC50 V
AIN10 +24V MC50 V AIN11 +24V I TSP V
AIN12 +24V II TMP V
Page 8: Analog data (2)
AIN0
V
AIN0 offset
V
Foreline pressure AIN1
mbar V
Flow
sccm
AIN2
V
Page 9: Info RS-232
Ecotec E3000 Sniffer
ASCII string
Sniffer Ecotec E3000
ASCII string
Command sent from main device to sniffer line Command sent from sniffer line to main device
The information in the first two lines can be switched with the keys “Sniffer” and “Leak”:
Ecotec E3000 Leak ASCII string
Command sent from main device to test leak
Leak Ecotec E3000 ASCII string
Command sent from test leak to main device
Host Ecotec E3000
ASCII string
Command sent from mainframe to Ecotec E3000
Ecotec E3000 Host ASCII string
Command sent from Ecotec E3000 to mainframe
Page 10: Info field bus
IC1000
24V supply SW-Version
SW-Version BL
BM1000
Serial number Bus type
SW-Version Module
Serial number Module
Address act. Value
Page 11: Info field bus BM 1000
Baud rate Exception code
Error counters
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
6 Operation
55
Table 7: Device information (cont.)
Menu item
Format State
Station name
IP address IP subnet mask
Gateway IP adress
DHCP enabled
Description
6.8 Special features of individual gases
R134a: Influence of cyclopentane and R245fa
R600a: Influence of cyclopentane and isopentane
Special features with helium
If you are sniffing for R134a, the presence of cyclopentane and R245fa can lead to incorrect measurement results. Search for R134a with alternative mass position 83, if cyclopentane and R245fa could be sniffed. Setting another mass, see “Mass and position”, page 37.
If you are sniffing for R600a, the presence of cyclopentane and isopentane can lead to incorrect measurement results. Search for R600a with the IGS mass position, if cyclopentane and isopentane could be sniffed. Setting the IGS- mass position, see “6.4.4 Suppressing interfering gases (Sophisticated interfering gas suppression – IGS)”, page 43.
If you are sniffing for helium, the Ecotec E3000 needs longer for an analysis than with refrigerants. Therefore stick to the following times during which you will not move the sniffer tip.
Table 8: Minimum measurement times for helium
Length of sniffer line
3 m 5 m 10 m 15 m
Minimum measurement time 2.2 s 2.5 s 3.3 s 4.5 s
The smallest detectable leak rate of the Ecotec E3000 for helium is 1 10-6
mbar l/s (higher than for refrigerant).
You can use a PRO-Check calibrated leak for the internal calibration of
helium. Because the PRO-Check calibrated leak does not fit into the opening on
the front panel of the Ecotec E3000, you have to connect it with a SUB D
cable, see installation manual of the ECOCheck.
If you are working with diluted helium, you can have the determined leak rate
also displayed as the gas-equivalent leak rate. For additional details, see
“6.4.3 Gas equivalent to helium and hydrogen, settings for diluted gas”, page
41.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
56
6 Operation
Special features with hydrogen / forming gas
If you are sniffing for hydrogen / forming gas, the Ecotec E3000 needs longer for an analysis than with refrigerants. Therefore stick to the following minimum measurement times.
Table 9: Minimum measurement times for hydrogen
Length of sniffer line
3 m 5 m 10 m 15 m
Minimum measurement time 2.7 s 3.0 s 3.8 s 5.0 s
If you work with hydrogen (forming gas), you can have the determined leak rate
also displayed as gas-equivalent leak rate, see “6.4.3 Gas equivalent to
helium and hydrogen, settings for diluted gas”, page 41.
If you want to verify hydrogen, the warm-up phase of the device must be
extended to 1 hour before the first calibration.
The smallest detectable leak rate of the Ecotec E3000 for hydrogen is 1 10-6
mbar l/s (higher than for refrigerant).
You can use a PRO-Check calibrated leak for the internal calibration of
hydrogen / forming gas. Because the PRO-Check calibrated leak does not fit
into the opening on the front panel of the Ecotec E3000, you have to connect
it with a SUB D cable, see installation manual of the ECO-Check.
Methane
Methane cannot be calibrated with the built-in ECO-Check because methane can
be detected only on mass 15 (which is outside the range for the permissible
internal calibration from 40 to 105).
Therefore use the external calibrated leak “TL4-6 for methane” for
calibration.
6.9 Switching off the device
Damage caused by rotating parts
The tubo molecular pump needs 5 min to stop.
Let the turbo molecular pump stop before all maintenance work or before you
move the device.
You can switch off the Ecotec E3000 at any time with the mains plug (position
“0”). It takes a few minutes for the turbo molecular pump to come to a
standstill. The Ecotec E3000 must not be moved during that time.
The parameter set in the Ecotec E3000 are saved. After the start-up, the
Ecotec E3000 returns to the same state that it was in before it was switched
off.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
6 Operation
57
58
6 Operation
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
7 Warnings and error messages
During operation, the display shows information that supports you in the
operation of the Ecotec E3000. Measurements are displayed along with current
device states, operating instructions as well as warnings and error messages.
The Ecotec E3000 is equipped with extensive self-diagnosis functions. If the
electronic system detects a faulty state, the device will show this insofar as
possible on the display and will interrupt operation, if necessary.
Error messages
Errors are events that the Ecotec E3000 cannot remedy itself and that force an
interruption of operation. The error message consists of a number and a
description.
After you have removed the cause of the error, start operation again with the
restart key.
Warnings
Warnings warn of device states that can impair the accuracy of measurements. Operation of the device is not interrupted.
Confirm acknowledgment of the warning with the OK key or the right key on the sniffer handle.
The following table shows all warnings and error messages. It lists possible causes for the malfunction and directions for fault clearance.
Please note that work marked with an asterisk must be carried out only by service staff that is authorized by INFICON.
Table 10: Warning and malfunction messages
No. Message
Possible error sources
Fault clearance
E1 Input voltage 24 V on the MC50 is too low
Fuse F1 on the wiring board is blown. Replace the fuse.*
CPU board MC50 is defective.
Contact the INFICON customer service.
E2 Input voltage 24 V on the Transpector is too low
Fuse F2 on the wiring board is blown. Replace the fuse.*
The Transpector is defective.
Contact the INFICON customer service.
E3 Input voltage 24 V on the Fuse F3 on the wiring board is blown. Replace the fuse.*
frequency converter is too low The turbo molecular pump is
Contact the INFICON customer service.
defective.
W4 Voltage 24 V on OPTION output is too low
Fuse F4 on the wiring board is blown. Replace the fuse.*
Power input through external wiring is Check the wiring. too high.
W5 Voltage 5 V on sniffer line is Fuse F5 on the wiring board is blown. Replace the fuse.*
too low
The sniffer line is defective.
Replace the sniffer line.
W6 Voltage 5 V on the ECO-Check The electronic system of the ECO-
is too low
Check calibrated leak is defective.
Replace the ECO-Check calibrated leak, see installation manual of the ECO- Check calibrated leak.
CPU board MC50 is defective.
Contact the INFICON customer service.
E7 Input voltage -15 V on the CPU board MC50 is defective. MC50 is too low
Contact the INFICON customer service.
E8 Input voltage 15 V on the MC50 is too low
CPU board MC50 is defective.
Contact the INFICON customer service.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
7
Warnings and error messages
59
Table 10: Warning and malfunction messages (cont.)
No. Message
Possible error sources
W12 Turbo pump frequency
The humidity is too high.
during run-up not reached or
TMP current too high!
Fault clearance
If W12 is displayed with very high humidity or after very long idle periods,
allow the device to warm up longer. If the problem still exists, restart the
device.
Contact the INFICON customer service if the problem still exists.
The turbo molecular pump is defective.
Contact the INFICON customer service.
W14 Maintenance interval for filter The change interval for the sniffer
in sniffer probe elapsed
probe air filter is reached.
Replace the filter, see page 77. Confirm the work, see page 69.
W16 Maintenance interval for turbo molecular pump elapsed
The change interval for the operating Replace the reservoir, see page 74. fluid reservoir of the turbo molecular Confirm the work, see page 69. pump is reached.
W17 Maintenance interval for diaphragm pump elapsed
Diaphragm pump maintenance is due. Replace the membranes of the diaphragm pump!* Confirm the work, see page 69.
W18 Maintenance interval for main The change interval for the main air Clean or replace the filter, see page 73.
air filter elapsed
filter is reached.
Confirm the work, see page 69.
E19 Temperature on CPU board The ambient temperature is too low / Please note the ambient conditions, see
MC50 too low(<-21°)
temperature sensor is defective
page 14. Contact the INFICON customer
service.
E20 Temperature on CPU board The ambient temperature is too high Please note the ambient conditions, see
MC50 too high (>60°)
page 14. Contact the INFICON customer
service.
One fan failed.
Check the air stream through the ventilation openings on both sides of the housing (left inlet, right outlet).
The main air filter is soiled
Clean or replace the filter, see page 73. Confirm the work, see page 69.
E22 TMP frequency is too low!
The sniffer line is incorrectly connected.
Check the connection.
The turbo molecular pump is defective.
Contact the INFICON customer service.
E23 Turbo pump frequency too The turbo molecular pump is
high!
defective.
Contact the INFICON customer service.
W24 Voltage 24 V of the external display unit is too low
Only for Ecotec E3000RC: The fuse on Replace the fuse.* driver board “External operation of the device” is blown.
The RC control panel is drawing too Contact the INFICON customer service. much power.
E25 Remove the sniffer probe from the opening of the calibrated leak!
The sniffer probe is in the calibration opening of the ECO-Check.
The light barrier of the ECO-Check is soiled.
Remove the sniffer probe.
Blow out the calibration opening with fresh air or clean it with a cotton
cloth.
W28 Real-time clock was reset! Please enter date and time.
CPU-board MC50 was replaced.
The battery on CPU-board MC50 is defective.
Enter the date and the time, see page 28. Contact the INFICON customer service.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
60
7 Warnings and error messages
Table 10: Warning and malfunction messages (cont.)
No. Message
Possible error sources
Fault clearance
W29 Voltage 24 V on the audio output is too low!
Fuse F6 on the wiring board is blown. Replace the fuse.*
The speaker is defective.
Replace the speaker.*
E30 Sensitivity too low
The sensor in the Transpector is defective.
Contact the INFICON customer service.
W31 Factor K1 outside of the range Other interfering gases, not
(0.9 to 1.1)!
cyclopentane or isopentane, were
detected, e.g. alcohols.
Recalibrate the IGS, see page 43.
W34 Changed flow!
The sniffer line is incorrectly
(The flow rate has been
connected.
reduced by more than 30 % The sniffer line filters are clogged. since the last calibration. The
warning goes out if the flow
change is less than 20 %
again.)
Check the connection.
Replace the sniffer line filters, see page 71. Confirm the work, see page 69.
Recalibrate the device, see page 38.
W35 Flow through capillary is too The lower limit value is set too high. Lower the lower flow rate limit value, see
low
page 32.
(The lower limit value is fallen The capillary filter in the sniffer line is Replace the capillary filter, see page 77.
short of longer than 2 s.)
clogged.
Confirm the work, see page 69.
Recalibrate the device, see page 38.
The sinter filter in the sniffer handle is Replace the sinter filter, see page 78.
clogged.
Confirm the work, see page 69.
Recalibrate the device, see page 38.
The capillary is clogged.
Replace the sniffer line. Recalibrate the device, see page 38.
Replace the multi-function cable of the sniffer line. * Recalibrate the device, see page 38.
The internal filters in the main unit are Replace the internal filters (three pieces).*
clogged
Confirm the work in the service menu.
Recalibrate the device, see page 38.
E36 Flow through capillary too The sniffer line is incorrectly
Check the connection.
high.
connected.
(The upper flow rate limit The upper limit value is set too low. value is exceeded longer than
2 s.)
The capillary is broken or leaking.
Set the upper limit value for the flow higher, see page 32.
Replace the sniffer line.
Recalibrate the device, see page 38.
Replace the multi-function cable of the sniffer line.Replace the multi- function cable of the sniffer line.* Recalibrate the device, see page 38.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
7 Warnings and error messages
61
Table 10: Warning and malfunction messages (cont.)
No. Message
Possible error sources
Fault clearance
E37 Capillary broken! (>10 s)
The upper limit value is set too low. Set the upper limit value for the flow
(The upper limit value is
higher, see page 32.
exceeded longer than 10 s. The capillary is broken or leaking. The emission is switched off to protect the cathodes.)
Replace the sniffer line. Recalibrate the device, see page 38.
Replace the multi-function cable of the
sniffer line.Replace the multi-function
cable of the sniffer line.*
Recalibrate the device, see page 38.
E38 Capillary broken! (>60 s)
The upper limit value is set too low. Set the upper limit value for the flow
(The upper limit value is
higher, see page 32.
exceeded longer than 60 s. The capillary is broken or leaking. The pumps are switched off to protect the cathodes.)
Replace the sniffer line. Recalibrate the device, see page 38.
Replace the multi-function cable of the
sniffer line.Replace the multi-function
cable of the sniffer line.*
Recalibrate the device, see page 38.
E39 Emission failed
If the device was switched off for a Confirm the error message and restart
(The emission on both
longer period of time, this error can the device. If the problem still exists:
cathodes cannot be switched occur during the first 10 minutes after Contact the INFICON customer service.
on.)
start-up.
Both cathodes are defective.
The Transpector is defective.
E40 Emission failed
If the device was switched off for a Confirm the error message and restart
(The emission failed during longer period of time, this error can the device. If the problem still exists:
operation.)
occur during the first 10 minutes after Contact the INFICON customer service.
start-up.
The input pressure is too high.
E41 No communication with Transpector!
The Transpector is defective.
The software cannot establish the connection with the Transpector.
Check the connection between Transpector and wiring board.*
Contact the INFICON customer service.
The Transpector is defective.
Contact the INFICON customer service.
E42 Transpector temperature > 70 °C or < 0 °C!
The main air filter is soiled.
Clean or replace the filter, see page 73. Confirm the work, see page 69.
The ambient temperature is too high. Please note the ambient conditions, see The ambient temperature is too low. page 14.
E43 Transpector limit value exceeded!
Internal Transpector data error
Contact the INFICON customer service.
E44 Transpector communication Internal Transpector data error error!
Contact the INFICON customer service.
E45 Transpector hardware error! Internal Transpector data error
Contact the INFICON customer service.
E46 Transpector hardware warning!
Internal Transpector data error
Contact the INFICON customer service.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
62
7 Warnings and error messages
Table 10: Warning and malfunction messages (cont.)
No. Message
Possible error sources
Fault clearance
E47 Transpector overpressure!
If the device was switched off for a Confirm the error message and restart longer period of time, this error can the device. If the problem still exists: occur during the first 10 minutes after Contact the INFICON customer service. start-up.
The sniffer line is not connected.
Connect the sniffer line and confirm the error message. Restart the device.
E48 Emission failed
The input pressure is too high.
(The emission failed during operation.)
The Transpector is defective.
Confirm the error message and restart the device. If the problem still exists: Contact the INFICON customer service.
W49 No emission with first cathode Start-up of the Emission failed. The device has switched to the second cathode.
You can continue to measure but need to have the cathodes checked.
E50 Control of turbo pump or electronic system!
There is an error in the control of the Contact the INFICON customer service. turbo molecular pump.
E51 No communication with the Wiring board defective
turbo controller!
CPU board MC50 defective
Contact the INFICON customer service.
W57 Too many EEPROM write cycles!
Constant change of gas configuration Use the alternative commands. (e.g.
or I*Guide program via interface.
BEF749 instead of BEF750)
W58 Measuring parameters
The gas library was changed by a
inconsistent! Please check! software update.
Reselect the gases to be measured. (The affected gas number is displayed.)
W59 Overflow of EEPROM parameter queue!
EEPROM defective.
Contact the INFICON customer service.
W60 All EEPROM parameters lost! A new wiring board was installed. Please check your settings!
All settings are reset to factory settings. Make your settings again.
If the message occurs constantly
Contact the INFICON customer service.
during running up, the EEPROM on the
wiring board is defective.
W61 0 EEPROM parameters initialized! Please check the settings!
A software update introduced new parameters. The new parameters are listed in the warning.
Confirm the warning.
Check whether the factory settings comply with the new parameters of your
application.
If the message occurs constantly
Contact the INFICON customer service.
during running up, the EEPROM on the
wiring board is defective.
W62 0 EEPROM parameters lost! Changed parameters were detected Check the settings of the listed
Please check the settings! during the running up.
parameters.
The affected parameters are listed in the warning.
If the message occurs constantly
Contact the INFICON customer service.
during running up, the EEPROM on the
wiring board is defective.
W63 TSP parameters inconsistent! The Transpector was replaced.
Please check!
The wiring board was replaced.
The EEPROM on the wiring board is defective.
Check the Transpector parameters. Contact the INFICON customer service.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
7 Warnings and error messages
63
Table 10: Warning and malfunction messages (cont.)
No. Message
Possible error sources
Fault clearance
W64 There are warnings pending! Confirmed but not yet valid warnings Remedy the cause of the warning.
are repeated every two hours or at every new start-up.
Deactivate the warning reminder, see “Fig. 46:Maintenance plan”, page 70.
W65 TSP serial number inconsistent! Please check!
The Transpector was replaced. The wiring board was replaced.
Contact the INFICON customer service.
The EEPROM on the wiring board is defective.
W66 ECO-Check brand-new!
A new ECO-Check calibrated leak was Enter the serial number and the code of
connected.
the calibrated leak, refer to the
installation manual of the calibrated leak.
W67 ECO-Check expires on DD.MM.YYYY!
The validity of the ECO-Check gas store expires in three months.
Order an ECO-Check gas store.
W68 ECO-Check expired!
The validity of the ECO-Check gas Replace the ECO-Check gas store, see store is expired (2 years operation or installation manual of the ECO-Check. older than 3 years).
W70 All EEPROM parameters of the The EEPROM in the ECO-Check
Replace the ECO-Check calibrated leak,
ECO-Check lost!
calibrated leak is empty or defective. see installation manual of the ECO- Check.
W71 No communication with ECO- The ECO-Check calibrated leak cannot Check the connection to the ECO-Check
Check!
be addressed by the main unit.
calibrated leak.
If the problem still exists: Contact the
INFICON customer service.
No ECO-Check calibrated leak connected.
Connect an ECO-Check calibrated leak. If you do not want to connect the ECOCheck calibrated leak, deactivate the warning, see page 34.
W72 No communication with sniffer line!
The sniffer line cannot be addressed by the main unit.
Check the connection of sniffer line to main unit (separate and reconnect; try another sniffer line, if possible). If the problem still exists, contact the INFICON customer service.
E73 Unsuitable sniffer line!
The SL3000XL of Protec P3000 was connected accidentally.
Connect the correct sniffer line.
W74 IC1000 SW version is obsolete! An SW update of the IC1000 is required Contact the INFICON service. for the full range of functions.
W75 No communication with IC1000!
The communication existed, but failed.
Check the connection to IC1000.
W77 Changed calibration factor! The calibration has changed by more Recalibrate the device, see page 38. (Possible message while the than 15 % since the last calibration calibration is being tested.) changed.
W78 Signal difference between test leak and air too small. (1.25 with R134a)
The calibrated leak is too small. The calibrated leak is defective.
The background signal during calibration is too high.
Check the leak rate of the calibrated leak or use a calibrated leak with a
higher leak rate.
Check the background signal by switching off the ZERO function (press the ZERO
key longer than 2 s).
The air signal was not steady during Recalibrate the device, see page 38. calibration (confirmed too early).
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
64
7 Warnings and error messages
Table 10: Warning and malfunction messages (cont.)
No. Message W79 Factor outside the range!
Possible error sources
Fault clearance
An invalid factor was detected during Repeat the IGS-calibration, see “IGS
IGS calibration.
alignment”, page 43.
If the problem still exists: Contact the INFICON customer service.
W80 Cathode switched over!
The device was switched over to
Recalibrate the device, see page 38.
another cathode without calibration.
W81 Calibration factor too small! During the internal calibration, the Check the entry for the leak rate, see calibration factor was determined to “6.4.2 Calibration”, page 38. be < 0.1 or < 0.01 during the external calibration.
It was incorrectly calibrated.
Recalibrate the device, see page 38.
The leak rate of the calibrated leak is incorrect (especially during external calibration).
Replace the external calibrated leak.
W82 Calibration factor too large!
During the internal calibration, the calibration factor was determined to be
10 or > 99.9 during the external calibration.
Check the entry for the leak rate, see “6.4.2 Calibration”, page 38.
It was incorrectly calibrated.
Recalibrate the device, see page 38.
The leak rate of the calibrated leak is incorrect (especially during external calibration).
Replace the external calibrated leak.
The Transpector sensor is not sensitive. Contact the INFICON customer service.
W83 Baseline offset out of range! E3000 was not in use for a long time. Let the device run longer.
E84 Test leak signal too small
The calibrated leak is too small.
Use a calibrated leak with a higher leak
(Flow of test leak absolutely
rate.
too small, approx. 1 g/a with The calibrated leak is defective.
Check the leak rate of the calibrated leak.
R134a)
The Transpector sensitivity is too low. Contact the INFICON customer service.
The gas flow to the sensor is blocked, the flow divider is clogged.
W85 Problem during peak finding! The Transpector cannot find the peak. Replace the calibrated leak. If the error
Calibration impossible.
still exists contact the INFICON customer
service.
W86 Internal calibration impossible
You are trying to calibrate while the Wait for calibration until the device has
device is not ready to measure.
run up.
W87 Unsupported gas in the ECO- EEPROM in ECO-Check not
Check!
programmed.
Call up the measurement view with the menu.
Use another ECO-Check.
The ECO-Check is defective.
E90 Overspeed (of TMP, E001)
The rotational speed of the turbo molecular pump is too high.
Contact the INFICON customer service.
E91 Overvoltage (on TMP, E002) Error in the electric power supply of Contact the INFICON customer service. the turbo molecular pump
E92 Run-up time error (E006)
The turbo molecular pump does not Contact the INFICON customer service. run up correctly.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
7 Warnings and error messages
65
Table 10: Warning and malfunction messages (cont.)
No. Message
Possible error sources
Fault clearance
E93 Connection electronic drive The connection between the control Contact the INFICON customer service. unit – pump defective (E008) and the turbo molecular pump is faulty.
E94 Error in the TC control unit (E015)
Defective turbo molecular pump control.
Switch the device off. Wait until the pump has stopped (> 5 min). Switch the device back on.
Contact the INFICON customer service.
E95 Electronic drive unit does not Error in the communication between Contact the INFICON customer service.
recognize pump. (E021)
turbo molecular pump and control.
E96 Error in the temperature control of the TC (E025)
The main air filter is soiled.
Clean or replace the filter, see page 73. Confirm the work, see page 69.
The ambient temperature is too high. Please note the ambient conditions, see page 14.
E97 Error of the temperature sensor in the TC (E026)
The temperature sensor is defective. Contact the INFICON customer service.
E98 Error in the motor stage or the There is an error in the motor stages or Contact the INFICON customer service.
control (E037)
the control.
E99 Error on the power supply (F007)
The supply voltage has failed.
Contact the INFICON customer service.
- The work may be carried out only by service personnel who have been authorized by INFICON.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
66
7 Warnings and error messages
8 Maintenance
8.1 Calling up and managing maintenance information
In the main menu you can call up information regarding the last operational
states of the device and its maintenance by selecting “History & maintenance”.
Here you will find especially the list of the last error messages. You will
also determine which maintenance work has to be confirmed and confirm
maintenance.
Error history
Fig. 41: Selection options in “History & maintenance”
The error history includes the errors and warnings that occurred during the
operation of the Ecotec E3000. Date and time are shown followed by an
identification for the error or the warning (E… with errors and W… with
warnings), with a brief description of the error or the warning.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Calibration history
Fig. 42: List with errors and warnings
To have the full warning or error message displayed, select the appropriate
line and press the key with the magnifying glass. Detailed information for
possible causes is also specified for many messages.
This list shows all calibrations that were carried out during the operation of
the Ecotec E3000. Information provided includes
Date and time Type of calibration (internal or external) Number of gases (only
with external calibration) Calibration factor
8
Maintenance
67
Maintenance intervals
Fig. 43: Calibration list
To have the complete calibration information displayed, select the appropriate
line and press the key with the magnifying glass. The calibration information
includes:
Calibration mode (internal or external); for internal: gas in calibrated leak
Gas
Internal calibration: one or more gas numbers External calibration: gas
number, mass position, gas Date and time of calibration Number of operating
hours at the time of calibration Calibration factor Peak position (if
different from mass position) Flow through sniffer line at the time of
calibration Cathode that was in use at the time of calibration (A/B) Size of
the calibrated leak that is used (external calibrated leak for external
calibration and ECO-Check for internal calibration) Argon flow and mass
deviation at the time of calibration Flow of the gas that was calibrated and
background signal
To call up the elapsed operating hours of the device since the start-up,
select “Maintenance intervals”. The information does not apply to the sniffer
line because different lines could have been used.
The remaining operating hours up until the next maintenance work are displayed
below.
The times below “Next maintenance for …” are based on confirmations that you
enter after maintenance work, see “Confirm maintenance”, page 69.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
68
8 Maintenance
Maintenance history
The maintenance history shows all maintenance work that was confirmed within the scope of the work. Shown are date and time of completed works, the number of operating hours of the device at the time of the maintenance work and the type of work performed. To have the complete entry displayed, select the appropriate line and press the key with the magnifying glass.
Confirm maintenance
Fig. 44: Entries in a maintenance list
The detailed view also shows maintenance work that you as operator cannot
confirm. In order to explain and confirm this work, you have to have special
technical knowledge and access to the “Service” menu.
In the menu “Confirm maintenance”, confirm the replacement of the lubricant
(operating fluid) reservoir and the main air filter replacement.
Select “Lubricant reservoir” or “Main air filter” and then press “OK”. You are
then prompted to state whether you want to confirm the maintenance, i.e. make
an entry in the maintenance list.
The maintenance intervals for the operating fluid reservoir and the air filter
of the main unit are specified and the system will remind you when the
interval has elapsed.
Sniffer line filter Because the device can be operated with different sniffer
lines, any replacement of the filter in the sniffer probe is not subject to
the maintenance schedule. Instead, you can enter a period in “Sniffer filter”
after which the device will ask you to replace the filter.
Setting range: 10 to 1000 hours and infinite (). Set if you do want to
suppress the request for replacement.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Fig. 45: Confirming maintenance work
8 Maintenance
69
Maintenance plan If you select “Maintenance Plan”, you have the option to deactivate the plan and thus the reminder messages. Under the menu item Maintenance Plan the warning W80 “Cathode switched!” can be deactivated independently from the other reminders.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Select filament (cathode)
Replacing ECO-Check
Fig. 46: Maintenance plan
Warning reminder
If the maintenance plan is activated but no maintenance work is confirmed, the
warning “Reminder of pending maintenance work” will be displayed every two
hours. You can deactivate the repeated display of this warning in “Warning
reminder”.
In this window, you can switch from cathode B back to cathode A if the device
has independently selected cathode B. The setting can only be made in the idle
state and with a stopped turbo molecular pump.
Select the cathode and confirm with “OK”.
If an ECO-Check calibrated leak is connected or the gas store of the ECO-Check
us replaced, you have to enter the serial number as well as the identification
with the calibration data. The installation of the ECO-Check in the Ecotec
E3000 as well as the gas store replacement are described in the ECO-Check
manual.
The ECO-Check must be connected to the device. You must write down the serial
number and the identification on the gas store or you can refer to the
supplied certificate for that information.
The appropriate entry window opens with “Replacing the ECO-Check”.
Fig. 47: Enter serial number and identification of the gas store.
The Ecotec E3000 checks the entered number. It will report “Invalid entry” if
the ECOCheck cannot be identified with the number.
70
8 Maintenance
8.2 Maintenance work
Failure to perform the maintenance work specified in the maintenance schedule
will void the warranty.
Danger to life due to electric shock Inside the device there are voltages of
considerable magnitude. Touching parts where electric voltage is applied can
result in death. Disconnect the device from the power supply prior to any
maintenance work. Make
sure that the electric power supply cannot be reconnected unauthorized.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Material damage from rotating parts
The turbo molecular pump requires 5 minutes to power down.
Allow the turbo molecular pump to power down before any maintenance work is
performed or before moving the device.
You will need the following tool for maintenance:
2 screwdrivers, size 2 Box wrench, 19 mm (included in the delivery scope) Allen wrench, 8 mm (included in the delivery scope) Allen wrench, 3 mm (not included in the delivery scope) Tweezers Table 11: Maintenance schedule
Maintenance
Material description Parts number
Check the sinter filter of the sniffer probe and replace, if necessary.
Sinter filter for sniffer probe
200 03 500
Replace the filter inserts of the Felt for capillary filter capillary filters and the water conservation tip.
200 001 116
Clean or replace the main air Air filter Ecotec E3000
filter in the floor of the
(104 x 154 mm; 5 pcs.)
housing.
200 001 552
Check the internal filter and replace, if necessary (three pcs.).
Internal filter
200 03 679
Replace operating fluid reservoir of the turbo molecular pump.
Operating fluid reservoir 200 003 801
The date on the packaging is the latest possible installation date.
Operating hours Period Maintenance
500 2000 10,000
level
X1
I
X
I
X
I
X
II
3 years II
8 Maintenance
71
Table 11: Maintenance schedule (cont.)
Replace the membranes of the Wear parts set for
diaphragm pump.
diaphragm pump
Replace gas store no later than Replacement gas store after 2 years of operation. Maximum durability, storage plus operation: 3 years
200 03 504 531-010
X
III
2 years
1 Heavier soiling of the measurement environment can also require earlier
replacement.
Explanation of maintenance levels:
Maintenance level I: Customer without any technical training Maintenance level
II: Customer with technical and INFICON training Maintenance level III:
INFICON Service
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
72
8 Maintenance
8.2.1
Replacing the air filter of the main unit
The air filter is inside a duct that is accessible from the bottom of the
device. The duct is closed with a cover plate. The cover plate is held in
place with a 3-mm hexagon socket screw.
Danger to life due to electric shock
Inside the device there are voltages of considerable magnitude. Touching parts
where electric voltage is applied can result in death.
Disconnect the device from the power supply prior to any maintenance work.
Make sure that the electric power supply cannot be reconnected unauthorized.
Material damage from rotating parts The turbo molecular pump requires 5
minutes to power down. Allow the turbo molecular pump to power down before any
maintenance work is
performed or before moving the device.
1 Remove the sniffer line and the ECO-Check from the main unit. 2 Place the
main unit with the front panel on a soft surface. 3 Loosen the screw of the
cover plate until you can rotate the cover plate to the side.
Fig. 48: Loosening the air filter cover 4 Pull out the air filter and replace it with a new one.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Fig. 49: Air filter in the air filter duct
8 Maintenance
73
5 Retighten the cover plate in front of the duct. 6 Put the device back on its feet and connect sniffer line and ECO-Check, if needed. 7 Confirm the work, see “Confirm maintenance”, page 69.
8.2.2
Replacing the operating fluid reservoir
The operating fluid reservoir supplies the turbo molecular pump with
lubricant. It consists of a plastic container with a soaked liner and 8 soaked
sticks (Porex rods). Plastic container and Porex rods are below the turbo
molecular pump and accessible from the bottom of the Ecotec E3000.
The bore for the operating fluid reservoir is closed with an aluminum plug and
a plastic screw.
The operating fluid reservoir has a limited service life and storage period;
see the maintenance schedule.
Danger to life due to electric shock
Inside the device there are voltages of considerable magnitude. Touching parts
where electric voltage is applied can result in death.
Disconnect the device from the power supply prior to any maintenance work.
Make sure that the electric power supply cannot be reconnected unauthorized.
Danger of poisoning The operating fluid reservoir can contain toxic substances
from the pumped medium. Wear appropriate protective clothing, when needed.
Dispose of the operating fluid reservoir as stipulated by local regulations. A
data safety sheet for the lubricants is available upon request.
1 Remove the sniffer line and the ECO-Check from the main unit. 2 Place the
main unit with the front panel on a soft surface. 3 Unscrew the plastic screw
with a 19 mm box wrench.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
74
8 Maintenance
Fig. 50: Fastener of the operating fluid reservoir
4 Lever out the aluminum plug with one or two narrow screw drivers.
5 Hook something into the center bore of the plastic container and pull out
the plastic container.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Fig. 51: Opened operating fluid reservoir 6 Pull the eight Porex rods out of the front side of the bore with tweezers.
Fig. 52: Porex rods of the operating fluid reservoir 7 Insert the new Porex rods with tweezers.
8 Maintenance
75
8 Place the plastic container in the bore with the soaked liner first and
close it with the aluminum plug.
9 Tighten the plastic screw again. Ensure that the o-ring is placed correctly
in the groove of the plastic screw and that the bore is lined up properly.
10 Confirm the work, see “Confirm maintenance”, page 69.
8.2.3
Replacing mains fuses
The fuses are behind a flap near the mains plug. They are in two slots. The
mains fuses are available with order no. 200 000 914. Two identical fuses must
be used at all times.
Danger to life due to electric shock Inside the device there are voltages of
considerable magnitude. Touching parts where electric voltage is applied can
result in death. Disconnect the device from the power supply prior to any
maintenance work. Make
sure that the electric power supply cannot be reconnected unauthorized.
1 Lever out the cover of the mains plug to the right with a screwdriver.
Fig. 53: Levering out cover to the right 2 Pull out the two slots and replace the fuses.
Translation of the original operating instructions Ecotec E3000, kina22en1-25, 2102
Fig. 54: Slot with fuse 3 Push the slots back in. Ensure that the arrows point up. 4 Close the flap.
76
8 Maintenance
8.2.4
Replacing
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>