INFICON SL3000EX Sniffer Leak Detectors Instruction Manual
- June 3, 2024
- INFICON
Table of Contents
Translation of the original operating instructions
Sniffer line
SL3000EX
Catalog No. 525-008, 525-009
From software version — linj10en1-03-(2111)
INFICON GmbH
Bonner Strasse 498
50968 Cologne, Germany
About this manual
These operating instructions describe the setup and maintenance of the
SL3000EX sniffer line.
Product names may occur in the document, which are added for identification
purposes only and belong to the respective owner of the rights.
Target groups
This instruction manual is intended for operators and technically qualified personnel with experience in leak detection technology and the integration of leak detectors in leak detection systems. In addition, the installation and use of the device require knowledge of electronic interfaces.
Warnings
DANGER Imminent hazard resulting in death or serious injuries
WARNING Hazardous situation resulting in potential death or serious
injuries
CAUTION Hazardous situation resulting in minor injuries
NOTICE Hazardous situation resulting in damage to property or the
environment
Safety
Intended use
The SL3000EX sniffer line is used to locate gas leaks on test objects in conjunction with the Ecotec E3000 leak device in Zone 2 hazardous areas.
Incorrect usage
- Extension of the cable between interface box and Ecotec E3000.
- Installation of the Interface-Box or Ecotec E3000 in explosive areas.
- Use with devices other than Ecotec E3000
- Use in radioactive areas
- Use of accessories or spare parts, which are not included in this instruction manual
- Use outside the technical specifications, see “Technical Specifications”
- Testing of wet or damp test objects
- Pumping off solids · Suctioning of liquids into the device
- Using the device with detectable defects or defective power switch
- Use of the device in ATEX Zone 0 and ATEX Zone 1.
Duties of the operator
- Read, observe, and follow the information in this manual and in the work instructions provided by the owner. This concerns in particular the safety and warning instructions.
- Always observe the complete operating instructions for all work.
- If you have any questions about operation or maintenance that are not answered in this manual, contact customer service.
Owner requirements
The following notes are for companies or any person who is responsible for the safety and effective use of the product by the user, employees or third parties.
Safety-conscious operation
- Operate the device only if it is in perfect technical condition and has no damage.
- Only operate the device properly in accordance with this instruction manual, in a safety and risk conscious manner.
- Adhere to the following regulations and observe their compliance:
- Intended use
- Universally valid safety and accident prevention regulations
- International, national and local standards and guidelines
- Additional device-related provisions and regulations
- Only use original parts or parts approved by the manufacturer.
- Keep this instruction manual available on site.
Personnel qualifications
- Only instructed personnel should be permitted to work with and on the device. The instructed personnel must have received training on the device.
- Make sure that authorized personnel have read and understood the instruction manual and all other applicable documents.
Scope of delivery
Item | Quantity |
---|
SL3000EX 5m, including 385mm flexible sniffer tip
or
SL3000EX 10m, including 600mm flexible sniffer tip| 1
Operating manual| 1
Description
SL3000EX
The SL3000EX sniffer line is used to locate gas leaks on test objects in
conjunction with the Ecotec E3000 leak detector.
The sniffer line consists of a multi-function cable, a sniffer handle, a
sniffer tip and an interface box.
Construction
CAUTION
Danger of blockage due to liquids
Liquid may destroy the device.
► Do not draw in any liquids (e.g. water or adhesives).
1 | ZERO push button | 2 | Push button (individual assignment) |
---|
There are two push buttons on the sniffer handle. Pressing the left button sets the background display to ZERO. For more information on the functions of these buttons, refer to the operating instructions of the leak detectors used.
SL3000EX with flexible sniffer tip
1 | Sinter filter (bronze Poral filter) | 5 | Brass capillary filter |
---|---|---|---|
2 | O ring | 6 | Metal filter |
3 | Sniffer tip | 7 | Cotton filter |
4 | Small O-ring | 8 | Replacement brass capillary filter |
A water protection tip must not be used.
Technical data
Name plates SL3000EX
Fig. 1: Name plates SL3000EX
1 | Interface box nameplate | 2 | Nameplate sniffer line SL3000EX |
---|
EX certification (intrinsic safety)
EX marking
Sniffer line| Ex ic IIB T4 EPL Gc
II 3G
---|---
EX marking
Interface-Box| [Ex ic Gc] IIB
Um: 250 V, Po: 28 mW, lo:16 mW, Uo: 7.14V
Table 1: EX marking
Mechanical data
SL3000EX
Gas flow| (160 sccm)
Connection on the device| Via separate socket on the device
SL3000EX
Acknowledge faults via buttons on the sniffer probe| Yes
End standby via a button on the sniffer probe| Yes
Acknowledge calibration via a button on the sniffer
probe| Yes
Electrical data
SL3000EX
Operating voltage| 5V
Mains frequency| DC voltage
Power consumption| 20 VA
Ambient conditions
SL3000EX
Max. height above sea level| 2000 m
Maximum relative humidity| 75% to 31 °C and 50% above 31 °C
Max. storage temperature| -20 °C to +60 °C
Permissible ambient temperature (during operation)| 5 °C to 40 °C
Installation
DANGER
Explosion hazard from wrong or missing grounding
Wrong or missing grounding can cause sparks. This can ignite the surrounding
potentially explosive atmosphere.
► The device may only be operated with the protective earth wire connected.
• Observe the grounding requirements for type of protection Ex i according to
IEC/EN 60079-14.
► Connect a protective earth wire from the interface box to the
protective earth of the main supply system.
► Ensure that the external grounding connections are exist, are in good
condition, and are not damaged or corroded.
DANGER
Explosion hazard due to gas accumulation in the Ecotec E3000
If the unit fans fail, gas may accumulate in the Ecotec E3000.
► Remove the test probe from the leakage area in case of a rough leakage.
NOTICE
The cable for the protective earth wire connection is not part of the scope of
delivery and must be provided by the user. The cross-section of the protective
earth wire must be at least 4 mm² or 12 AWG.
For connection to the leak detector Ecotec E3000, the sniffer line is connected to the connection socket provided on the front of the device for this purpose. The sniffer line communicates with the leak detector via multi function cable cable and is supplied with power via the data cable.
Fig. 2: The grounding strap (here yellow-green) must be attached to the protective earth connection of the interface box!
- The leak detector is supplied by an electric circuit that meets the requirements for “Limited-energy circuits” of EN 61010-1.
- The protective conductor of the interface box is connected.
► Connect the sniffer line to the leak detector.
The sniffer line is now ready for operation.
Application area
Maintenance
Maintenance sniffer tip
Supplementary pack set
1 | O-rings (5x) | 4 | Replacement brass capillary filter (5x) |
---|---|---|---|
2 | Sinter filter (bronze poral filter, 5x) | 5 | Cotton filter (50x) |
3 | Small O-rings (5x) | 6 | Needle tip tool |
Change cotton filters in the sniffer tip
- The cotton filters in the sniffer tip are dirty (brass capillary filters).
- The capillary filter is not removed.
1. Carefully remove both cotton beads from the front using the pointed end of the needle tip tool.
Note: The metal filter behind it is maintenance-free and must not be damaged or moved out of position by the needle point tool. This would impair the effectiveness of the metal filter. For the position of the metal filter, see also “SL3000EX [►} 8]”.
2. Using the blunt end of the needle tip tool, stuff two clean cotton filters into the capillary filter.
Capillary filter change with flexible sniffer tip
1 | Small O-ring | 2 | Brass capillary filter |
---|
- A damaged brass capillary filter should be replaced with another brass capillary filter. A capillary filter made of plastic must not be used.
1. Unscrew the capillary filter.
2. Take the new brass capillary filter and slide an intact O-ring over the thread for sealing. The capillary filter already has the necessary filters installed.
3. Screw the new capillary filter onto the sniffer tip hand-tight.
NOTICE
Perform a leakage resistance check “Checking connection between capillary
filter and ground [} 16]” after each capillary filter change.
Capillary filter change with rigid sniffer tip
- A damaged brass capillary filter should be replaced with another brass capillary filter. A capillary filter made of plastic must not be used.
1. Unscrew the capillary filter.
2. Screw a new brass capillary filter onto the sniffer tip hand tight. The capillary filter already has the necessary filters installed. An O-ring for sealing is not required.
NOTICE
Perform a leakage resistance check “Checking connection between capillary
filter and ground [►} 16]” after each capillary filter change.
Check / replace sinter filter and O-ring
1 | Sinter filter (bronze Poral filter) | 3 | Connection adapter |
---|---|---|---|
2 | O ring | 4 | Mounting screws (2x) |
- The leak detector is switched off and the sniffer line is removed.
- The sniffer tube is removed from the sniffer handle.
1. Loosen the two fastening screws.
2. Remove the O-ring and the sinter filter.
3. Visually inspect the sinter filter and the O-ring for dirt and damage.
• If you notice any damage or contamination, replace the O-ring with the sinter filter.
4. Reinsert the O-ring with the sintered filter into the sniffer tip.
5. Screw the sniffer tube back on to the sniffer handle.
• The sniffer line is ready to be used again.
NOTICE
Perform a leakage resistance check “Checking connection between capillary
filter and ground [►} 16]” after each sintered filter change.
Checking connection between capillary filter and ground
DANGER
Danger of deflagration or explosion
If there is poor conduction between the filter tip and the interface box,
sparking may cause deflagration or explosion.
► Check the leakage resistance each time the filter tip is changed.
► To ensure EX protection, the leakage resistance must be < 50 Ohm.
► In case of service, send the sniffer line and the interface box to INFICON
Service.
Fig. 4: Checking connection between capillary filter and ground
Maintenance plan SL3000EX
Maintenance required 1| Assembly| Material| Material no:| Operating hours|
Repair level
---|---|---|---|---|---
| | | | 2000| 5000| 10000|
Check sintered filter, replace if necessary| Sniffer tip| Sinter filter-F
0-ring
(5pcs)| 20003500| X| | | I
Warning W35- “Flow too low” Exchange filter| Sniffer tip| Felt disc for
capillary
filter (50pcs)| 200001116| dependin
g on
environm
ent| | | I
Check capillary filter, replace if necessary| Sniffer tip| Capillary filter
plastic| 20003501| | | X| I
Rubberized sniffer probe, replace if necessary| Sniffer probe| Sniffer probe,
rubberized| 581-200| | depending on usage| | III
| | (EX)| | | | |
EX protection| EX Box| Flame Arrester| 200010894| | 3 years/or after EX-
incident| | Ill
Zener diode| EX Box| Zener diode| 200010895| | 3 years/or after fuse failure|
| Ill
Repair Level I: Customer
Repair Level II: Customer with technical training by INFICON
Repair Level III: INFICON Service Engineer
Decommissioning
WARNING
Danger due to harmful substances
Contaminated devices could endanger health. The contamination declaration
serves to protect all persons who come into contact with the device. ► Fill in
the declaration of contamination completely.
-
Contact the manufacturer and send in a completed declaration of contamination before return shipment.
⇒ You will then receive a return number. -
Use the original packaging when returning.
-
Before sending the device, attach a copy of the completed contamination declaration. See below.
Declaration of Contamination
Appendix
Accessories and spare parts
Sniffer tips SL3000EX|
---|---
FT385EX, 385 mm length, flexible| 521-041
FT600EX, 600 mm length, flexible| 521-040
FT120EX, 120 mm length, flexible| 521-042
ST120EX, 120 mm length, rigid| 521-043
Sniffer line SL3000EX|
SL3000EX 5m, including 385mm flexible sniffer tip| 525-009
SL3000EX 10m, including 600mm flexible sniffer tip| 525-008
CE Declaration of Conformity
We — INFICON GmbH – herewith declare that the products defined below meet the
basic requirements regarding safety and health and relevant provisions of the
relevant EU Directives by design, type and the versions which are brought into
circulation by us. This declaration of conformity is issued under the sole
responsibility of INFICON GmbH.
In case of any products changes made without our approval, this declaration
will be void.
Designation of the product:
Sniffer line
Models: SL3000EX
€x) 113G, Ex ic IIB T4 Ge
Catalogue numbers:
525-008
525-009
The products meet the requirements of the following Directives:
- Directive 2014/34/EU (ATEX)
- Directive 2014/30/EU (EMC)
- Directive 2011/65/EU (RoHS)
Applied harmonized standards:
-
EN 61326-1:2013
Class A according to EN 55011 -
ENIEC 60079-0:2018
-
EN 60079-1171 :2012
-
ENIEC 63000:2018
Cologne, February 19, 2021
| Cologne, February 19, 2021
---|---
INFICON GmbH
Bonner Strasse 498
D-50968 Cologne
Tel.: +49 (0)221 56788-0
Fax: +49 (0)221 56788-90
www.inficon.com
E-mail: leakdetection@inficon.com
inspired by visions. Proven by success.
www.inficon.com
reachus@inficon.com
Due to our continuing program of product improvements, specifications are
subject to change without notice.
The trademarks mentioned in this document are held by the companies that
produce them.