INFICON ELT3000 E-Check Lithium Ion Battery Leak Detector Instruction Manual

June 1, 2024
INFICON

INFICON ELT3000 E-Check Lithium Ion Battery Leak Detector

INFICON-ELT3000-E-Check-Lithium-Ion-Battery-Leak-Detector-product-
image

Product Information

Specifications

  • Catalog No.: 600-105
  • Model: ELT3000 / ELT3000 PLUS
  • Manufacturer: INFICON GmbH
  • Address: Bonner Strasse 498, 50968 Cologne, Germany
  • Software Version: lime10en1-08-(2401)

Description
The calibration leak is designed for use with ELT3000 / ELT3000 PLUS devices to ensure accurate calibration for leak detection purposes.

Function
The calibration leak helps in calibrating the leak detection equipment to maintain accurate and reliable performance.

Technical Data
Mechanical Data:
Ambient Conditions:
Maintenance
Regular maintenance of the calibration leak is essential to ensure proper functioning and accurate calibration results.

Decommissioning
Proper procedures must be followed for decommissioning the calibration leak, including sending it for filling or disposal.

Product Usage Instructions

Safety
It is important to follow all safety instructions provided in the manual to prevent any hazards or injuries.

Intended Use
The calibration leak is intended for use with ELT3000 / ELT3000 PLUS devices by technically qualified personnel and instructed users.

FAQ

  1. Q: How often should the calibration leak be checked?
    A: It is recommended to check the calibration leak periodically as per the manufacturer’s guidelines to ensure accurate calibration.

  2. Q: Can the calibration leak be used with other devices?
    A: The calibration leak is specifically designed for use with ELT3000 / ELT3000 PLUS devices and may not be compatible with other equipment.

About this manual

This document applies to the calibration leak indicated on the title page. Product names may occur in the document, which are added for identification purposes only and belong to the respective owner of the rights.

Warnings
DANGER
Imminent hazard resulting in death or serious injuries

WARNING
Hazardous situation resulting in potential death or serious injuries

CAUTION
Hazardous situation resulting in minor injuries

NOTICE
Hazardous situation resulting in damage to property or the environment

Target groups
This instruction manual is intended for the user of the calibration leak, technically qualified personnel and instructed personnel.

Safety

Intended use

  • Only use the calibration leak as intended, as described in the instruction manual, to avoid hazards due to misuse.
  • The calibration leak is used to calibrate the battery leak detector. This usually takes place once a day. Before first use, the calibration leak must be removed from the packaging and stored for 24 hours with the opening facing downwards. The calibration leak, opening downwards, is placed in the test chamber of the battery leak detector and evacuated. The escaping solvent in gaseous form is detected. It can be used in roofed and closed rooms. 12 months after manufacture, the calibration leak must no longer be used. It must then be sent in for a refill see “Decommissioning [} 14]”. Storage is in a well-ventilated place. A closed container is not suitable.
  • Adhere to the restrictions of use, see “Technical data [} 12]”.

Incorrect usage

Avoid the following unintended uses:

  • Operation of the calibration leak by untrained personnel
  • Use outside the technical specifications, see “Technical Specifications”
  • Use of the calibration leak for detectable defects
  • Operation at too high ambient temperature
  • Using the device outside of the specific area
  • Exceeding permissible ambient conditions for calibration leaksExceeding permissible environmental conditions for the calibration leaks
  • Storage in a poorly ventilated place or closed container
  • Use or storage near sources of ignition or heat
  • Use the calibration leak to calibrate other devices with an internal ignition source.
  • Abuse for purposes of intoxication
  • Open the calibration leak
  • Refilling by the customer
  • Use with chemicals/liquids other than DMC
  • Rücksendung ohne geeignete Transportmittel/Verpackung/Warnaufkleber

Duties of the user

  • Read, observe, and follow the information in this instruction manual and in the work instructions provided by the owner. This concerns in particular the safety and warning instructions.
  • Always observe the complete operating instructions for all work.
  • If you have any questions about operation or maintenance that are not answered in this manual, contact customer service.

Requirements for the user
The following notes are for companies or any person who is responsible for the safety and effective use of the calibration leak by the user, employees or third parties.

Safety-conscious operation

  • Only operate the calibration leak if it is in a technically perfect condition and shows no signs of damage such as escaping solvents.
  • Only operate the calibration leak as intended, in a safety and risk conscious manner, and in accordance with this instruction manual.
  • Adhere to the following regulations and observe their compliance:
    • Intended use
    • General applicable safety and accident prevention regulations
    • International, national and local standards and guidelines
    • Additional device-related provisions and regulations
  • Keep this instruction manual available on site.

Personnel qualifications

  • Only allow instructed personnel to work with the calibration leak leak. The instructed personnel must have received training in handling the calibration leak.
  • Make sure that authorized personnel have read and understood the instruction manual and all other applicable documents.

Scope of delivery, transport, storage

Scope of delivery

Item Quantity
E-Check 1
Instruction manual 1
Calibrated Leak Certificate 1
Safety data sheet 1
Return material 1

Transport

WARNING
Risk of explosion due to leaked solvent
For transport, the calibration leak is wrapped in a fleece and sealed in foil packages. If liquid is visible inside the film packaging or the fleece is damp, there is a risk of fire or explosion in the vicinity of ignition sources due to the leaked solvent

  • Remove the foil packaging of the calibration leak only if there is no liquid in the packaging or the fleece is damp.
  • If liquid escapes from the calibration leak, do not unpack it, avoid any contact and dispose of it properly without delay.
  • Ensure good ventilation

NOTICE
Damage due to improper packaging
Hazard to health and the environment

  • Use the supplied packaging materials (2 x plastic bags and fleece) for returning and disposing of the calibration leak.
  • Use a new carton for return and disposal.
  • When returning and disposing of the product, observe the applicable dangerous goods regulations for road, air and sea transport, as well as the associated packaging regulations and labels!

Storage
Damage caused by improper storage

  • Remove the calibration leak from the packaging and store the calibration leak for at least 24 hours with the opening/flange facing downwards before use.
  • Store the calibrated leak in a well-ventilated place so that the print on the top is always legible.

Description

WARNING
Risk of explosion due to escaping solvent
If the calibration leak is used to calibrate equipment with internal ignition sources (e.g. valves, electric motors, Pirani vacuum gauges, etc.), an ignition source may cause an explosion.
Solvents can escape through a damaged membrane and lead to an explosive concentration in the air.

  • No smoking.
  • Ensure adequate ventilation.
  • Keep ignition sources away from the calibrated leak.
  • Properly dispose of the calibrated leak if liquid escapes.
  • Avoid any contact with the solvent.
  • Observe the manufacturer’s safety data sheets and follow the applicable work instructions.
  • Do not expose the calibration leak to temperatures higher than 40 °C.
  • Operate the calibrated leak only in the INFICON battery leak detector ELT3000/ELT3000 PLUS / ELT3000 PLUS.

WARNING
Danger due to electrical charge
Static charge can cause a sudden discharge, possibly with sparking, when handling the calibration leak.

  • Wear ESD clothing and shoes

CAUTION
Risk of injury from falling
Always place the calibration leak on a level surface and secure it against falling down.

CAUTION
Risk of injury through inhalation of solvent vapors
Inhalation of solvent vapors from the calibration leak may cause nausea and intoxication up to unconsciousness and respiratory arrest.

  • Avoid inhalation of solvent vapors.
  • Do not use the solvent to become intoxicated.
  • Avoid any skin contact with the solvent.
  • Ensure that there is sufficient ventilation.

INFICON-ELT3000-E-Check-Lithium-Ion-Battery-Leak-
Detector-\(8\)

Position Name
1 E-Check (top view)
2 Sticker on the E-check
3 Shelf life notice
4 Danger notice
5 Nameplate

INFICON-ELT3000-E-Check-Lithium-Ion-Battery-Leak-
Detector-\(9\)

Position Name
1 E-check (view from the bottom)
2 Flange with leakage element
3 Support surface of the E-Check in the test chamber

The different screw height prevent the closing of the outlet opening.

NOTICE
Leaking solvent damages the calibration leak.

  • Do not loosen the screws on the flange.
  • Do not remove the flange with leakage element.

Function
The DMC solvent in the calibration leak constantly escapes via a membrane, in very small quantities, in gaseous form. To check the detection function, place the calibration leak in the measuring chamber with the outlet opening facing downwards. The membrane is protected by a grid. The calibrated leak is used to calibrate the ELT3000 battery leak detector. The central flange contains the diaphragm and is sealingly connected to the housing by means of a copper ring.

  • Note that several measuring cycles are necessary until the actual leakage rate can be determined.

Refilling and recertification is only possible through INFICON, see “Maintenance [ 13]”.

Technical data

Mechanical data

Material Stainless steel
Dimensions (Ø x H) 150 mm x 25 mm
Weight (filled) < 1250 g

Ambient conditions

Temperature range (°C) 15 °C to 35 °C
Relative humidity (%) 80 % at 30 °C, linear decrease to 50 % at 40 °C
Height above sea level (m) 2000 m
Permissible storage temperature 0 °C to 40 °C

Maintenance
WARNING
Danger from improper filling.

  • The calibration leak may only be refilled by INFICON customer service.

For refilling and recertification, contact INFICON customer service.

Decommissioning

The owner can dispose of the device or it can be sent to INFICON. The device consists of materials that can be recycled. This option should be exercised to prevent waste and also to protect the environment. During disposal, observe the environmental and safety regulations of your country.
INFICON-ELT3000-E-Check-Lithium-Ion-Battery-Leak-
Detector-\(10\)The E-Check leak must not be disposed of with household waste.

Send in calibration leak for filling or disposal
Injury and environmental damage from spilled DMC
Injury and environmental damage from spilled DMC
Residual amounts of DMC must remain in the calibration leak and must not be drained under any circumstances.

WARNING
Danger due to harmful substances
Contaminated devices could endanger health. The contamination declaration serves to protect all persons who come into contact with the device. Devices sent in without a return number and completed contamination declaration will be returned to the sender by the manufacturer.
Complete the contamination declaration in full except for item 3.

  1. When returning the calibration leak, observe the applicable dangerous goods regulations for road, air and sea transport.
  2. Contact the manufacturer and send in a completed declaration of contamination before return shipment.
    • You will then receive a return number and the shipping address.
  3. Use suitable packaging for the return, see the following “Scope of delivery, transport, storage [ 7]”.
  4. Before shipping the instrument, attach a copy of the completed contamination declaration to the outside of the package.

For contamination declaration see below.

Packing the calibration leak
Avoidance of damage, bodily injury and environmental damage

  • Wrap the calibration leak in the fleece, place the calibration leak with fleece in the first bag and seal it. Then place it in the second bag and seal this bag as well.
  • Pack the calibration leak in a suitable box, if possible as for the original packaging.

Declaration of Contamination

The service, repair, and/or disposal of vacuum equipment and components will only be carried out if a correctly completed declaration has been submitted. Non-completion will result in delay This declaration may only be completed (in block letters) and signed by authorized and qualified staff.

INFICON-ELT3000-E-Check-Lithium-Ion-Battery-Leak-
Detector-\(11\)

www.inficon.com
reachus@inficon.com
Due to our continuing program of product improvements, specifications are subject to change without notice
The trademarks mentioned in this document are held by the companies that produce them.

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals