W HLER CM220 CO Detector Instruction Manual

June 16, 2024
W HLER

W HLER CM220 CO Detector

General Information

Operation Manual Information

This operation manual allows you to safely work
with the Wohler CM 220. Please keep this manual for your information.
The Wohler CM 220 should be used by trained professionals for its intended use only.
Liability is void for any damages caused by not following this manual.

Notes

WARINING!
Not following this warning can cause injury or death.

ATTENTION!
Not following this note can cause permanent damage to the analyzer.

NOTE!
Useful information

Intended Use

The Wohler CM 220 is a digital pen type CO meter that can be used for domestic purposes to check furnaces, room heaters, central heating furnaces, water heaters etc. as well as for commercial purposes, e.g. the ventilation control of indoor parking garages.
The electrochemical sensor giantizes an exact measurement even if other gases interfere..

Scope of Supply

Meter Components
Wohler CM 220 CO-Meter

3 x AAA batteries
Carrying pouch

Information on disposal

Electronic equipment does not belong into domestic waste, but must be disposed in accordance with the applicable statutory provisions.
You may hand in any defective batteries taken out of the unit to our company as well as to recycling places of public disposal systems or to selling points of new batteries or storage batteries.

Address

Wohler Technik GmbH
Wohler-Platz
33181 Bad Wunnen berg
Tel.: + 49 2953 73-100
Fax: + 49 2953 73-250
E-Mail: [email protected]

Technical Data

Description Value
Range 0 – 999 ppm
Accuracy ± 20 % at 0 – 100 ppm ± 15 % at 100–500 ppm (at 20 ± 5°C, 50 ± 20%

RH)
Resolution| 1 ppm
Power Supply| 3 AAA batteries
Response time| within 60 s
Operating humidity| 5 – 99% RH (not condensing)
Storage condition| – 20 – + 60°C
Operation temperature| 0 – 45°C
Meter Size| 175 x 47 x 28 mm

CO Levels and its effects

CO-Concentration Symptoms
0-1 ppm Normal background levels
9 ppm Maximum indoor air quality level
200 ppm Mild headache, fatigue, nausea and dizziness within 2- 3 hours
400 ppm Frontal headache, life threatening after 3 hours
800 ppm Dizziness, nausea, convulsions, life threatening after 2- 3 hours
1600 ppm Nausea within 20 minutes, death within 2-3 hours

Design and Function

Display

  1. Aktueller CO-Wert
  2. maximizer CO Wert
  3. Batterie symbol: Batterie bald wechseln
    Display
Keys
  1. Turning the backlight on and off
    Switch between temperature units C and F
    Reset the max. CO value

  2. Switching the gate on and off. Setting alarm value 1 + 2: Turn on calibration mode

  3. Keys

Operation

Power on/off

Press the power-key to turn on and turn off the meter. When the meter is powered on, the preset alarm value and current air temperature will be blinking on the LCD. The meter is now running a self-test and will finish in about 15 seconds with a short beep. It will then enter the measuring mode.

WARNING!
It is recommended to turn on the meter in an area free of CO since high CO level may cause failure in passing Self-Test.

Switch temperature units

Users can select °C or °F as the temperature unit by pressing the backlight key for more than 1 second during the self-test.

Alarm setting

The alarm value of the meter is preset at 25 ppm to give warnings of possible danger. Users can adjust the value for different needs or standard by turning off the meter, then press and hold the power-key until the value options appear in the display. There are 10, 25, 30, 35, 45, 50, 70 and 100 ppm displaying in cycles. Release the button to select the preferred value. Then the meter will return to Self-Test mode in some seconds.

Measurement

The meter measures the CO concentration at the sensor and displays the reading in ppm. It also indicates the maximum value (at the lower side of the display) of all readings since it is powered on.
Users can reset the maximum value by pressing the backlight-key for two seconds in the measuring mode.

Backlight

The meter features backlight function for using in dark areas. Press the backlight to activate and switch off the backlight.

Low battery indicator

The low battery indicator appears in the display when the battery voltage gets low. The measured values may be more inexact now. In this case you have to change the batteries.

Auto power off

The meter turns itself off automatically after 15 minutes non-operation. (It will not auto power off during the calibration).

0 ppm calibration

Calibrate the meter in an area free of CO. To enter the calibration mode, turn off the meter and press the power-key and the backlight-key simultaneously for two seconds. „ppm“ will blink in the upper display and the calibration value „0“ at the lower side.
Now the meter is running the 0 ppm calibration as a default.
Now the meter calibrates automatically. The LCD is displaying a blinking ppm value and it stops blinking when the calibration is completed. This takes about 10 minutes.

NOTE!
Do not be alarmed if high ppm or minus value shows up when entering the calibration mode. It will go down and approach to 0 in seconds.

Troubleshooting

Problem Solution
The meter is switched on but there is no display. Make sure that the

batteries are in good contact and correct polarity.
Replace a new battery and try again.
Calibration failure| If the low battery indicator was shown before the calibration, replace a new battery.
E 2| The value is underflow.
E 3| The value is overflow.
E 31 und E 33| Return the meter to the dealer for repairing.
E 35| a) Self-Test failure: Turn on the meter again in other area free of CO. b) Sensor failure: Return the meter to Wohler for repairing.

Warranty and Service

Warranty

Each Wohler CM 220 will be tested in all functions  and will leave our factory only after extensive quality control testing.
If used properly, the warranty period for the Wohler CM 220 will be 12 month from the date of sale.
Not covered by the warranty are the batteries and the costs for transport and package.
Service by non authorized personnel or making modifications to the meter voids any warranty.

Service

We see SERVICE as a very important element in our business. That is why we are still available to you even after the guarantee period has expired.

  • An immediate repair will be carried out if you bring your meter to one of our approved service centers listed below.
  • You can obtain immediate help from our engineers by telephone.

Declaration of conformity

The product:

Name: CO-Meter
Model: Wohler CM 220

has been tested in accordance to essential protection requirements of Council Directive 2014/30/EU and found the test results indeed meet the limitation of the relevant test standards listed below:

EN 61326-1 (2013)
(CISPR11, IEC/EN 61000-4-2 (2009)/-3 (2006+A2 :2010)/-8(2010)

Bad Wunnenberg, 21.09.2022
Signature

Johannes Lottering, Geschäftsführer/Managing DirectorLogo

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

W HLER User Manuals

Related Manuals