GILSON PIPETMAN P2 Metal Ejector User Guide
- June 15, 2024
- Gilson
Table of Contents
- Introduction
- Parts Checklist
- Description
- Specifications
- Fitting the Tips
- Fitting the Tips on P2 and P10 Models
- Multichannel models
- Ejector Spacer for PIPETMAN® 10 µL Multichannel Models
- Pre-Wetting the Tips
- Aspirating
- Dispensing
- Maintenance
- Cleaning and Decontamination
- Warranty
- Web Resources
- Documents / Resources
Quick Guide
Introduction
PIPETMAN®, recognized as the pipetting standard, is a fully adjustable, air-
displacement pipette that uses disposable tips. PIPETMAN combines its
legendary accuracy, precision, and durability with easy pipetting and
handling.
PIPETMAN offers a large range of accurate and precise pipettes with optimized
pipetting and purge forces for hours of pipetting.
Eight single channel models cover a volume range from 0.2 µL to 10 mL.
Eight multichannel models cover a volume range from 1 µL to 300 µL.
Parts Checklist
After unpacking the pipette, verify that the following items were included and
are undamaged:
Single models
- PIPETMAN®
- Quick Guide
- Safety bag
- Certificate of Conformity (including barcode sticker)
- Dual-position adapter (for P2 and P10 models only)
Multichannel models
- PIPETMAN®
- Quick Guide
- Safety bag
- Certificate of Conformity (including barcode sticker)
- Ejector spacers for D10 tips (for Px10 models only)
Good Laboratory Practice (GLP) Compliance
The serial number is engraved on the body of the pipette. It provides unique
identification of your pipette and the manufacturing date. The barcode on the box and the
Certificate of Conformity provides traceability of your pipette.
Description
Figure 2
PIPETMAN® multichannel and single channel model components
Specifications
PIPETMAN® is a high-quality pipette that offers excellent accuracy and
precision. The data provided in the Maximum Permissible Errors table on page 4
was obtained using PIPETMAN® DIAMOND Tips.
These values are guaranteed only when genuine PIPETMAN DIAMOND Tips are used.
NOTICE
With a precise pipetting technique the P2 model may be used to aspirate
volumes as low as 0.1 µL and the P10 model as low as 0.5 µL.
Maximum Permissible Errors
The data in the following table conforms to the ISO 8655-2 standard.CV
means Coefficient of Variation
P10 model can be used up to 0.5 μL with a precise pipetting technique.
P5000 and P10mL do not have tip ejectors.
Gilson maximum permissible errors are only guaranteed when PIPETMAN® pipettes
are used with the recommended PIPETMAN® DIAMOND Tips. Setting the Volume
The volume of liquid to be aspirated is set using the volume display. The dial
colors are either black or red to indicate the position of the decimal point,
depending on the model (refer to Figure 3). Figure 3
Dial colors by model
The volume is set by turning the thumbwheel or the push button (refer to
Figure 4). The push button makes it easier and faster to set volumes,
especially when wearing gloves.
The thumbwheel may be turned using only one hand to slowly reach the required
setting.
To achieve maximum accuracy when setting the volume, proceed as follows:
When decreasing the volume setting, slowly reach the required setting,making
sure to not overshoot the mark.
Figure 4
Location of push button and thumbwheel
When increasing the volume setting, pass the required value by 1/3 of a turn
and then slowly decrease to reach the volume, making sure to not overshoot the
mark.
MODEL | COLOR OF VOLUMETER NUMBERS |
---|---|
BLACK | RED |
P2 | µL |
P10 – P20 – Px10 – Px20 | µL |
P100 – P200 – Px200 – Px300 | µL |
P1000 | 0.01 mL |
P5000 | 0.01 mL |
P10mL | mL |
Fitting the Tips
Single Channel Models
To fit a new PIPETMAN DIAMOND Tip, push the tip holder into the tip using a
slight twisting motion to ensure a firm, airtight seal.
PIPETMAN DIAMOND Tip Compatibility for single channel variable volume models
Figure 5
PIPETMAN® DIAMOND Tip compatibility chart for single channel models
NOTICE
D5000 and D10mL PIPETMAN DIAMOND tips do not have any filter. P5000 and
P10mL models are sold with a bag of 10 filters. For more details, refer to
PIPETMAN User’s Guide LT801122.
Fitting the Tips on P2 and P10 Models
NOTE
For the P2 and P10 models, a dualposition adapter (plastic) is required to fit
DL10 tips (long tips) or D10 tips (short tips). The metallic rod of the tip
ejector is shaped so the adapter may be clipped to it in either position.
P2 and P10 models are delivered with the adapter in place, ready-to-use DL10
tips. If D10 tips are used, the adapter must be repositioned in the shorter
slot as follows:
- Pull the adapter down from the metallic rod.
- Turn the adapter 180°.
- Refit the adapter so the end of the metallic rod engages the shorter slot of the adapter.
Figure 6
Dual-position adapter for P2 and P10 with stainless steel tip ejector
NOTE
The dual position is autoclavable.
Multichannel models
PIPETMAN® DIAMOND Tips are best fitted with the ROCKY RACK technique, invented
by Gilson, available only in our TIPACK and TOWERPACK.
PIPETMAN® DIAMOND Tip Compatibility for Multichannel Models
P8x10, P12x10 | D10*, DL10, DF10, DFL10 | |
---|---|---|
P8x20, P12x20 | DL10, DFL10, DF30, D200 | |
P8x200, P12x200 | D200, D300, DF200, DF300 | |
P8x300, P12x300 | D200, D300, DF200, DF300 |
- Using the broad ejector spacer D10, you can adapt a D10 tip (see above). Figure 7
ROCKY RACK technique
Ejector Spacer for PIPETMAN® 10 µL Multichannel Models
According to the tip used, D10 or DL10, you may have to exchange the ejector spacer. The broadly spaced one is for D10, and the narrowly spaced one is for DL10.
- Remove the tip ejector, keeping both ejector locks depressed. Pull the tip ejector down.
- Gently press the tabs from the ejector spacer and remove it from the tip ejector.
- Insert the alternative ejector spacer and click it to the tip ejector.
- To refit the tip ejector, gently re-insert the tip ejector vertically into the rails of the ejector support.
Figure 8
Ejector spacer location
Pre-Wetting the Tips
Pre-wetting the tips before pipetting helps prepare the tips for the best
pipetting performance. Ideally, the prewet includes both immersing the tip in
the liquid and performing one pipetting step.
Pre-wetting the tips helps ensure that volumes your pipette’s volume will
achieve accuracy and precision within specifications.
Aspirating
- Press the push button to the first stop (this corresponds to the set volume of liquid).
- Hold the pipette vertically and immerse the tip in the liquid (refer to PIPETMAN® User’s Guide LT801122 available on gilson.com).
- Release the push button slowly and smoothly (to the top position) to aspirate the set volume of liquid.
- Wait one second, and then withdraw the pipette tip from the liquid. You may wipe any droplets away from the outside of the tip using a medical wipe; however, if you do so, take care to avoid touching the tip’s orifice.
Figure 9
Pipetting motion – aspirate and dispense
Dispensing
- Place the end of the tip against the inside wall of the recipient vessel (at an angle of 10° to 40°).
- Press the push button slowly and smoothly to the first stop.
- Wait for at least one second, then press the push button to the second stop to expel any residual liquid from the tip. Keep the push button pressed fully down and (while removing the pipette) draw the tip along the inside surface of the vessel.
- Release the push button smoothly. Eject the tip by pressing firmly on the tip ejector button.
NOTICE
It is recommended to regularly verify the volume displayed on the dial while
pipetting.
Maintenance
Routine maintenance will help keep your pipette in good condition, ensuring a
continued high level of performance.
Maintenance is limited to cleaning or autoclaving the parts specified in the
PIPETMAN® User’s Guide LT801122, or replacing the push button, connecting nut,
tip ejector, tip holder, seal, and O-ring.
NOTICE
Gilson recommends maintenance and calibration at least annually, more
frequently as needed, depending on use.
NOTICE
PIPETMAN® P2 and P10 should not be disassembled. You may only replace the push
button, tip ejector, and its dual-position adapter. With these pipettes, if
the tip holder is damaged, the piston may also be damaged. After replacing any
parts, you should verify
the performance of your pipette following the verification procedure available
on the Gilson website (www.gilson.com). If the pipette needs to be readjusted,
please contact your local authorized Gilson Service Center.
NOTICE
Decreased pipetting forces are due to the design of the piston assembly, which
includes a very high-quality lubricant (part number 5440011070). The use of
any other lubricant voids the warranty of this pipette.
Cleaning and Decontamination
PIPETMAN is designed so the parts normally in contact with liquid contaminants
can easily be cleaned and decontaminated.
For further information, please refer to PIPETMAN User’s Guide LT801122
available on gilson.com.
Warranty
Gilson warrants this pipette against defects in material under normal use and
service for a period of three years for both single channel and multichannel
models from the date of purchase.
This warranty shall not apply to pipettes which are subject to abnormal use
and/or improper or inadequate maintenance (contrary to the recommendations
given in the user’s guide), including, but not limited to pipettes that have
been subjected to physical damage, improper handling, or spillage or exposure
to any corrosive environment. This warranty shall also be void in the event
pipettes are altered or modified by any party other than Gilson or its
designates. Gilson’s sole liability under this warranty shall be limited to,
at Gilson’s sole option, repair or replacement of any defective components of
pipettes or refund of the purchase price paid for such pipettes. Routine
cleaning, control and recalibration are not covered under the warranty. The
replacement of wearing parts such as seals, O-rings, broken pistons assembly,
and broken tip holders are not covered under the warranty.
NOTICE
Yearly routine maintenance is highly recommended to keep your pipette in good
condition, ensuring a continued high level of performance.=
THE FOREGOING WARRANTY IS EXCLUSIVE AND GILSON HEREBY DISCLAIMS ALL OTHER
WARRANTIES, WHETHER EXPRESS OR IMPLIED, INCLUDING ANY WARRANTIES OF
MERCHANTABILITY AND ANY WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE, UNDER
NO CIRCUMSTANCES SHALL GILSON BE LIABLE FOR ANY CONSEQUENTIAL, PUNITIVE,
INDIRECT OR INCIDENTAL DAMAGES ARISING OUT OF ANY BREACH OF ANY EXPRESS OR
IMPLIED WARRANTY.
Web Resources
Visit Gilson’s website at www.gilson.com for:
- Service contact information
- PIPETMAN® User’s Guide LT801122 pdf
- Information about accessories, spare parts, and related products
- Additional product information and special offers
NOTES
gilson.com/contactus
LT801624/E | ©2023 Gilson, Inc.
All rights reserved.
QUICK GUIDE | PIPETMAN®
Documents / Resources
|
GILSON PIPETMAN P2 Metal
Ejector
[pdf] User Guide
PIPETMAN P2 Metal Ejector, PIPETMAN P2, Metal Ejector, Ejector
---|---
References
- Gilson | Home | Liquid Handling, Purification and Extraction Solutions USA
- Gilson | Home | Liquid Handling, Purification and Extraction Solutions
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>