GILSON Pipetman Classic Single Channel Manual Air Displacement
Adjustable Volume Pipette
Product Information
Specifications
Each pipette is inspected and validated by qualified technicians in
accordance with the Gilson Quality system. Gilson declares that its
manufactured pipettes comply with the requirements of the ISO 8655 standard by
type testing. The adjustment is carried out under strictly defined and
monitored conditions (ISO 8655-6).
Description
The PIPETMAN Classic is a high-quality pipette designed for accurate and
precise liquid handling. It features a thumbwheel for volume adjustment, a
color-coded push button, a tip ejector button, a connecting nut, a tip holder,
and a body or handle clip tip ejector. The pipette is GLP compliant and comes
with a dual-position adapter for P2 and P10 models, a safety bag, and a
Certificate of Conformity.
Parts Checklist
- PIPETMAN Classic
- PIPETMAN Classic Quick Guide
- Dual-position adapter (for P2 and P10 models only; refer to Fitting the Tips on page 5) Safety bag
- Certificate of Conformity (including barcode sticker)
Serial Number
The serial number is engraved on the body of the pipette. It provides unique identification of your pipette and the date of manufacture. For more details, refer to PIPETMAN Classic User’s Guide lT801120. The barcode on the box and the Certificate of Conformity provide traceability of your pipette. The updated line of PIPETMAN Classic is identified with an underlined serial number starting from QG.
Setting the Volume
The volume of liquid to be aspirated is set using the volumeter. The dials
are colored either black or red to indicate the position of the decimal point,
depending on the model. The volume is set by turning the thumbwheel or the
push button. The push button makes it easier and faster to set volumes,
especially when wearing gloves. The thumbwheel may be turned to slowly reach
the required setting. Dial colors by model
Model | Color of Volumeter Numbers | Increment |
---|---|---|
P2 (P/N F144801) WITH D10 OR DL10 TIPS | Black | 0.01 ml |
P10 (P/N F144802) WITH D10 OR DL10 TIPS | Black | 0.1 ml |
P20 (P/N F123600) WITH D200 TIPS | Black | 2 ml |
P100 (P/N F123615) WITH D200 TIPS | Black | 10 ml |
P200 (P/N F123601) WITH D200 TIPS | Black | 20 ml |
P1000 (P/N F123602) WITH D1000 TIPS | Black | 100 ml |
P5000* (P/N F123603) WITH D5000 TIPS | Red | 500 ml |
P10mL* (P/N F161201) WITH D10ML TIPS | Red | 1000 ml |
Maximum Permissible Errors
The data provided in the following tables conform to the ISO 8655-2
standard.
GILSON | ISO 8655 | Pipette Model | Volume (μl) | Error (± μl) |
---|---|---|---|---|
GILSON | ISO 8655 | P2 (P/N F144801) WITH D10 OR DL10 TIPS | 0.2 | 0.012 |
GILSON | ISO 8655 | P10 (P/N F144802) WITH D10 OR DL10 TIPS | 1 | 0.012 |
FAQ
Can I lubricate the pipette?
No, lubricating the pipette will cancel the warranty.
Introduction
PIPETMAN Classic is a fully adjustable, air-displacement pipette that is used with disposable tips. Eight single-channel models cover a volume range from 0.2 μL to 10 mL for many applications.
Parts Checklist
- PIPETMAN Classic
- PIPETMAN Classic Quick Guide
- Dual-position adapter (for P2 and P10 models only; refer to Fitting the Tips on page 5)
- Safety bag
- Certificate of Conformity (including barcode sticker)
GLP Compliant
The serial number is engraved on the body of the pipette. It provides unique identification of your pipette and the date of manufacture. For more details, refer to PIPETMAN Classic User’s Guide LT801120. Ex: QG10369 The barcode on the box and the Certificate of Conformity provide traceability of your pipette.
Figure1 Serial number location
NOTE
The updated line of PIPETMAN Classic is identified with an underlined
serial number starting from QG.
Description
Figure 2 PIPETMAN Classic single-channel model
Specifications
PIPETMAN Classic is a high-quality pipette that offers excellent accuracy and precision. The data provided in the Maximum Permissible Errors table on page 4 were obtained using PIPETMAN® DIAMOND Tips. These values are guaranteed only when genuine PIPETMAN DIAMOND Tips are used. Each pipette is inspected and validated by qualified technicians in accordance with the Gilson Quality System. Gilson declares that its manufactured pipettes comply with the requirements of the ISO 8655 standard by type testing. The adjustment is carried out under strictly defined and monitored conditions (ISO 8655-6).
Maximum Permissible Errors
The data provided in the following tables conform to the ISO 8655-2 Standard.
VOlUME
(µl)
| GILSON| ISO 8655
---|---|---
SYSTEMATIC ERROR (µL)| RANDOM ERROR (µL)| SYSTEMATIC ERROR (µL)| RANDOM ERROR
(µL)
P2 (P/N F144801) WITH D10 OR DL10 TIPS
0.2| ± 0.024| £ 0.012| ± 0.08| £ 0.04
0.5| ± 0.025| £ 0.012| ± 0.08| £ 0.04
1| ± 0.025| £ 0.012| ± 0.08| £ 0.04
2| ± 0.030| £ 0.014| ± 0.08| £ 0.04
P10 (P/N F144802) WITH D10 OR DL10 TIPS
1| ± 0.025| £ 0.012| ± 0.12| £ 0.08
5| ± 0.075| £ 0.030| ± 0.12| £ 0.08
10| ± 0.100| £ 0.040| ± 0.12| £ 0.08
P20 (P/N F123600) WITH D200 TIPS
2| ± 0.10| £ 0.030| ± 0.2| £ 0.1
5| ± 0.10| £ 0.040| ± 0.2| £ 0.1
10| ± 0.10| £ 0.050| ± 0.2| £ 0.1
20| ± 0.20| £ 0.060| ± 0.2| £ 0.1
P100 (P/N F123615) WITH D200 TIPS
10| ± 0.35| £ 0.10| ± 0.8| £ 0.3
50| ± 0.40| £ 0.12| ± 0.8| £ 0.3
100| ± 0.80| £ 0.15| ± 0.8| £ 0.3
P200 (P/N F123601) WITH D200 TIPS
20| ± 0.50| £ 0.20| ± 1.6| £ 0.6
100| ± 0.80| £ 0.25| ± 1.6| £ 0.6
200| ± 1.60| £ 0.30| ± 1.6| £ 0.6
P1000 (P/N F123602) WITH D1000 TIPS
100| ± 3.0| £ 0.6| ± 8.0| £ 3.0
500| ± 4.0| £ 1.0| ± 8.0| £ 3.0
1000| ± 8.0| £ 1.5| ± 8.0| £ 3.0
*P5000 (P/N F123603)* WITH D5000 TIPS
500| ± 12| £ 3| ± 40| £ 15.0
2500| ± 15| £ 5| ± 40| £ 15.0
5000| ± 30| £ 8| ± 40| £ 15.0
P10mL (P/N F161201) WITH D10ML TIPS
1000| ± 30| £ 6| ± 60| £ 30.0
2000| ± 30| £ 6| ± 60| £ 30.0
5000| ± 40| £ 10| ± 60| £ 30.0
10000| ± 60| £ 16| ± 60| £ 30.0
*P5000 and P10mL do not have tip ejectors
Setting the Volume
The volume of liquid to be aspirated is set using the volumeter. The dials are colored either black or red to indicate the position of the decimal point, depending on the model (refer to Figure 3).
Figure 3Dial colors by model
The volume is set by turning the thumbwheel or the push button. The push
button makes it easier and faster to set volumes, especially when wearing
gloves. The thumbwheel may be turned to slowly reach the required setting.
To achieve maximum accuracy when setting the volume:
- When decreasing the volume setting, slowly reach the required setting, making sure to not pass the setting.
- When increasing the volume setting, pass the required value by 1/3 of a turn and then slowly decrease to reach the volume, making sure to not pass the setting.
Model | COlOR OF VOlUMETER NUMBERs |
---|---|
BlACK | RED |
P2 | µl |
P10 to P20 | µl |
P100–P200 | µl |
P1000–P5000 | 0.01 ml |
P10ml | ml |
Fitting the Tips
To fit a new PIPETMAN DIAMOND Tip, push the tip holder into the tip using a slight twisting motion to ensure a firm and airtight seal.
NOTE
For the P2 and P10 models, a dual-position adapter (plastic) is required to
fit DL10 tips (long tips) or D10 tips (short tips). The metallic rod of the
tip ejector is shaped so that the adapter may be clipped to it in either
position
P2 and P10 models are delivered with the adapter in place ready to use DL10 tips. When D10 tips are used, the adapter must be repositioned in the shorter slot as follows:
- Pull the adapter down from the metallic rod.
- Turn the adapter 180°.
- Refit the adapter so the end of the metallic rod engages the shorter slot of the adapter.
Figure 4 Dual-position adapter for P2 and P10
NOTE
Plastic tip ejector can be used and ordered separately, (refer to PIPETMAN
Classic User’s Guide LT801120).
Pre-Rinse the Tips
Some liquids (e.g., protein-containing solutions and organic solvents) can leave a film of liquid on the inside wall of the tip. Pre-rinse the tip to minimize any errors that may be related to this phenomenon. Pre-rinsing consists of aspirating the first volume of liquid and then dispensing it back to waste. Subsequent volumes you pipette will have levels of accuracy and precision within specifications.
- Aspirate
Press the push button to the first stop (this corresponds to the set volume of liquid).
- Hold the pipette vertically and immerse the tip in the liquid (refer to PIPETMAN Classic User’s Guide LT801120 Guidelines for Good Pipetting).
- Release the push button slowly and smoothly (to the top position) to aspirate the set volume of liquid.
- Wait one second. The time depends on the model (refer to the table). Then, withdraw the pipette tip from the liquid. You may wipe any droplets away from the outside of the tip using a medical wipe; however, if you do so, take care to avoid touching the tip’s orifice.
Figure 5 Pipetting motion – aspirate and dispense
Dispense
- Place the end of the tip against the inside wall of the recipient vessel (at an angle of 10° to 40°).
- Press the push button slowly and smoothly to the first stop.
- Wait for at least a second, then press the push button to the second stop to expel any residual liquid from the tip. Keep the push-button pressed fully down and (while removing the pipette) draw the tip along the inside surface of the vessel.
- Release the push button smoothly. Eject the tip by pressing firmly on the tip of the ejector button.
Maintenance
Routine maintenance will help keep your pipette in good condition, ensuring a continued high level of performance. Maintenance is limited to cleaning or autoclaving the parts specified in the Cleaning and Decontamination chapter of the PIPETMAN Classic User’s Guide (LT801120).
NOTE
PIPETMAN P2 and P10 should not be disassembled; you may only replace the
push button, tip ejector, and dual-position adapter. With these pipettes, if
the tip holder is damaged, the piston may also be damaged. After replacing any
parts you should verify the performance of your pipette following the
verification procedure available on the Gilson website
(www.gilson.com). If the pipette needs to be
readjusted, please contact your local authorized Gilson Service Center.
Warranty
Gilson warrants this pipette against defects in material under normal use and
service for a period of three years from the date of purchase.
This warranty shall not apply to pipettes that are subject to abnormal use
and/or improper or inadequate maintenance (contrary to the recommendations
given in the user’s guide), including, but not limited to pipettes that have
been subjected to physical damage, improper handling, or spillage or exposure
to any corrosive environment. This warranty shall also be void in the event
pipettes are altered or modified by any party other than Gilson or its
designates. Gilson’s sole liability under this warranty shall be limited to,
at Gilson’s sole option, repair or replacement of any defective components of
pipettes or refund of the purchase price paid for such pipettes. Routine
cleaning, control and recalibration are not covered under the warranty. The
replacement of wearing parts such as seals, O-rings, broken pistons assembly,
and broken tip holders are not covered under the warranty.
NOTICE
Yearly routine maintenance is highly recommended to keep your pipette in good
condition, ensuring a continued high level of performance.
Web Resources
- Visit Gilson’s website at www.gilson.com for:
- Service contact information
- PIPETMAN Classic User’s Guide LT801120 pdf
- Information about accessories, spare parts, and related products
- Additional product information and special offers
www.gilson.com/contactus LT801623/C | ©2019 Gilson, Inc. All rights reserved.
References
- Manual-Hub.com - Free PDF manuals!
- Gilson | Home | Liquid Handling, Purification and Extraction Solutions
- Manual-Hub.com – Free PDF manuals!
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>