NANOCOLOR® VARIO C2
NANOCOLOR ® VARIO 4
Water Analysis
C2 Digital Block Heater
Handback Manual
General information and introduction
The following operating instruction includes information femmages resulting
from errors or omissions in this operating manual. The manufacturer reserves
the right to make improvements to this manual and the products described
herein at any time without prior notice or obligation. Revised editions of
this manual are available on the manufacturer’s website.
The heating blocks NANOCOLOR ® VARIO C2, VARIO C2 M and VARIO 4 from
MACHEREY‑NAGEL are devices for stationary use in the laboratory. They are used
for sample preparation using colorimetric test kits or the models NANOCOLOR®
VARIO C2, VARIO C2 M and VARIO 4. The manufacturer is not responsible for
direct, indirect, accidental or consequential d and NANOCOLOR ® digestion
chemicals from MACHEREY‑NAGEL. Standard parameters for routine digestions are
pre‑programmed in the heating blocks and help the user to avoid errors. The
touch screen and the intuitive menu navigation allow a fast, safe and
comfortable workflow with the devices and allow the opera‑ tion without
complex training.
CE Marking
The CE marking declares that the product complies with the following European
Community harmonization legislation:
European Directive 2011/65/EU on the restriction of the use of certain
hazardous substances in electrical and electronic equipment (RoHS)
European Directive 2012/19/EU on waste electrical and electronic equipment
(WEEE)
European Directive 2014/30/EU on the harmonization of the laws of the member
states relating to electromagnetic compatibility (EMC)
European Directive 2014/35/EU of electrical equipment designed for use with
certain voltage limits (LVD)
Type label
info@mn-net.com · www.mn-net.com
Figure 1: Type label of the heating blocks NANOCOLOR ® VARIO C2, VARIO C2
M and VARIO 4
Explanation of symbols
The device type plate and these operating instructions contain the symbols or terms listed below, which have the following meaning:
Term/Symbol | Meaning |
---|---|
NANOCOLOR’® VARIO 4 | Device type designation |
SN | Serial number of the device |
110/230 V, —50/60 Hz | Power supply |
300/550 VA | Power consumption |
Attention (observe documentation)! Observe the safety instructions in the |
operating instruction of the device
| According to 2012/19/EU, it is prohibited to dispose of the device through
public waste disposal systems. Note also the information in the “Disposal”
section.
| The CE symbol indicates fulfillment of the applicable harmonization
legislation of the European Community.
| Identification USB interface
MACHEREY-NAGEL| Identification of the manufacturer
| This symbol indicates that there is a risk of electric shock.
| This symbol indicates that the chemicals used have a corrosive effect.
Observe the safety measures in the laboratory and wear the suitable protective
equipment. Observe the instructions in the current safety data sheets (SDS) of
the products used.
| This symbol indicates that there may be a danger by using flammable
substances.
| This symbol indicates that the designated area can become hot and must
not be touched without suitable protective measures.
| Explanations to the text. Tips and tricks for better handling.
Safety instructions
BE SURE TO READ THE FOLLOWING SAFETY INSTRUCTIONS CAREFULLY BEFORE USING
THE DEVICE.
Failure to follow these instructions may result in serious injury to the
operator, malfunction or damage to the equipment.
Keep this operating instruction in a safe place for future reference.
Follow the safety notes and instructions in the operating instruction and
observe the stickers and notices on the device.
Do not work on internal parts of the unit. Non‑compliance will invalidate any
warranty claim.
Use of the hazard warnings:
DANGER
Indicates an imminent or potentially hazardous situation which, if not
avoided, will result in death or serious injury.
WARNING
Indicates an imminent or potentially hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result
in minor or moderate injury.
NOTICE
Indicates a situation which, if not avoided, may result in malfunction or
damage
Chemical safety
WARNING
Possible dangers of contact with chemical substances.
Working with samples, reagents and corresponding accessories is associated
with dangers.
Wear suitable protective equipment when working with the cuvettes. Observe the
safety data sheet (SDS) of the test kit used
During normal operation of this device, it may be necessary to use chemicals
that are hazardous to your health or biologically harmful samples.
Before handling these substances, read all hazard statements and safety
information printed on the containers with the original solutions and in the
safety data sheets. All used solutions must be disposed of in accordance with
the national laws and regulations. The type of protective equipment must be
selected according to the concentration and amount of the hazardous substance
present at the corresponding workplace.
Technical description
The heating blocks NANOCOLOR® VARIO C2, VARIO C2 M, and VARIO 4 offer space for the sample preparation of cuvettes with 16 mm outer diameter and in the case of the VARIO C2 M for cuvettes with 16 mm and 22 mm outer diameter. Six temperatures (70 °C, 100 °C, 120 °C, 148 °C, 150 °C and 160 °C) and five heating times (10 min, 30 min, 60 min, 120 min and continuous) are pre‑programmed. The aluminum block of the unit is heated to the set temperature by a heating cartridge. The temperature is automatically maintained for the set heating time after it has been reached. After the time has expired, the heating process is automatically interrupted and an acoustic signal is given.
7.1 Technical Data
Type: | Heating block for chemical-analytical digestion |
---|---|
Bores: | 2 x 12 with 16 mm OD (VARIO 4) |
8 with 16 mm OD + 2 with 22 mm OD (VARIO C2 M) 12 with 16 mm OD (VARIO C2)
Display:| Backlit, multicolor LCD touchscreen
Operations:| Icon-based menu guidance via touch screen
Temperatures:| 6 preprogrammed temperatures 70 °C/ 100 °C/ 120 °C/148 °C/ 150
°C/160 °C, 6 free memory locations for individual temperature settings
Temperature range:| 40 °C-160 °C (1 °C increments)
Temperature stability:| ± 1 °C (according to DIN, EN, ISO and EPA methods)
Warm up time:| From 20 °C to 160 °C within 10 min (at 20 °C)
Heating times:| 5 preprogrammed heating times, 10 min/30 min /60 min /120 min/
cont. 7 free memory locations for individual heating times
Time range:| 0 h:01 min-9 h:59 min (1 min increments)
Safety:| Replaceable safety covers as contact protection, lockable protective
lids, overheating protection
Interfaces:| Bidirectional RS-232, USB A (function), USB B (host)
Internal quality control (IOC):| With NANOCOLOR® T-Set (REF 919917) and
NANOCOLOR® USB T-Set (REF 919921) optional fully automatic calibration and
generation of a test certificate for device control and monitoring
Languages:| DE/EN/FR/ES/HU/PL/CZ/TR/DK
Update:| via USB-Stick
Operating range:| 10 °C-40 °C; max. 80% relative humidity (non-condensing), up
to 3000 m
Power supply:| 110V/230V, –50/60 Hz
Power consumptions:| 300/550 W (VARIO 4) 150/300W (VARIO C2, VARIO C2 A/I)
Dimensions:| 290 mm x 287 mm x 146 mm (VARIO 4) 169 mm x 282 mm x 146 mm
(VARIO C2, VARIO C2 M)
Weight:| ca. 3.2 kg (VARIO 4) ca. 2.0 kg (VARIO C2, VARIO C2 A4)
Warranty:| 2 Years
CE certified:| Yes
Declaration of conformity:| On request
Overvoltage category:| II
Degree of pollution:| 2
Setting up and connecting the device
WARNING
Risk of burns: The safety cover(s) must be installed to prevent personal
injury. The safety cover prevents burns. Always keep the protective cover(s)
closed during operation.
DANGER
Risk of electric shock: The safety cover(s) must be installed to prevent
personal injury.
The safety cover prevents damage to the power cord. Always keep the protective
cover(s) closed during operation.
CAUTION
Risk of injury due to the danger of tipping over due to slipping. Place the
device on a flat surface. Do not stack the appliance. Set up the unit in a
laboratory or similar environment (recommended location: fume hood).
NOTICE
Only place the device in a designated location. Make sure that the surface is
dry, clean, level and horizontal to avoid overheating the unit. If possible,
set up the unit on a flame‑retardant surface (e.g. laboratory bench).
NOTICE
Damage due to condensation and exceeding the temperature limits:
The device is designed for indoor use only. Operate the device in a clean and
dry environment.
WARNING
Take care not to overload the mains socket. There is a risk of overload and
fire. Make sure that the power cord is not damaged. Check the suitability of
the power source used for the equipment.
WARNING
Risk of injury: Familiarize yourself with the device before working with it
and read this document carefully. Do not use the device unless you have
received instruction in its use.
NOTICE
Defects at the power supply and housing can lead to malfunction of the device.
If the device shows an apparent breakage of the housing or a damaged power
supply, it must be taken out of operation.
8.1 Scope of delivery
- Check the delivery for completeness. If parts of the delivery are missing, contact MACHEREY‑NAGEL or your local distributor directly.
- Heating block NANOCOLOR® VARIO C2 / VARIO C2 M / VARIO 4
- Power cable
- Fuse with socket, Size: 5x20mm T3,15AH (time lagging, high breaking capacity)
- 1 × protective cover (VARIO C2 and VARIO C2 M only)
- 2 × protective cover (VARIO 4 only)
- USB data cable (A/B)
- Software DVD
- Quick start Guide
8.2 Connecting the device
NOTICE
Use only the approved power cord.
The separately packed fuse of the heating block is marked by a sticker in the
delivery box.
- Remove the fuse and insert it into the socket on the back of the heating block. The insertion direction is determined by the locking device. (Figure 2)
- Place the supplied protective hood(s) on the device.
- Connect the supplied power cord to the unit and then to the power outlet.
8.3 Transport
To transport the device, the delivery box with its protective interior is best
suited. Therefore, keep the packaging. If you no longer have the packaging in
hand, use a suitable cardboard box as outer packaging and soft padding
material so that the device cannot slip during transport.
Device views
- Display / Touchscreen
- Safety cover as contact protection
- Foldaway protective cover
- USB‑A interface
- USB‑B interface
- RS‑232 interface
- I / O switch
- Fuses
- Device type label
- Power connection
Initial operation
10.1 Turning on the device
Connect the heating block to the power supply and switch it on with the main I
/ O switch (Figure 4, No. 7 ) located on the back of the device. The
device will start up and the manufacturer’s logo will be displayed. After a
few seconds the start screen will appear.
NOTICE
****By pressing , the info screen will be displayed. The serial number,
installed software‑ as well as bootloader versions will be displayed.
NOTICE
The device must always be set up in such a way that the on / off switch can be
operated at any time.
10.2 Switching off
NOTICE
If you are not going to use the device for a longer period of time, take it
out of operation and disconnect it from the power supply.
To switch off the device, tilt the I / O switch on the back of the device
(Figure 3, No. 7 ) to the right (O).
Operation and user guidance
NOTICE
Do not operate the touch screen with sharp or pointed objects, as this will
damage the screen.
The heating block has an illuminated, color touch screen display (Figure 4,
No. 1 ). The display is operated by touching individual fields.
Familiarize yourself with the operation of the touch screen display by
touching individual buttons with your finger or a special stylus. Touching
buttons selects the corresponding function.
Program selection
On the Home screen (Figure 4), select the desired temperature or time by
pressing or several times. By setting the values on the right side of
the display, the right half of the heating block is controlled. The same
applies accordingly to the left side (VARIO 4 only).
Call a pre‑programmed temperature. The last used temperature is always
displayed first. The keys work in jog‑roll mode. Only programmed temperatures
from the smallest to the largest are displayed.
Call a pre‑programmed time. The last used time is always displayed first.
The keys operate in jog‑roll mode. Only programmed times from the shortest to
the longest are displayed. Press ****to start the digestion (see Performing a
digestion, page 48).
12.1 Settings
Press on the Home screen to access the settings (Figure 5). Press to
return to the home screen. Pressing **** in the respective settings menus will
return to the Settings overview menu without saving the selection.
12.1.1 Language
Pressing takes you to the language selection (Figure 6). Select the
desired language and confirm with .
12.1.2 Graphic mode
Pressing displays a window for switching on (icon ) and off (icon
) the graphic mode. Confirm the desired setting with .
When the graphics mode is active, the temperature curve appears on the display
(see Temperature monitoring, page 51).
12.1.3 Programming the temperature
Press **** to access the selection of programmed temperatures(Figure 8).
Select a free memory location ( ) and enter the desired temperature in the
range 40 °C– 160 °C (Figure 9). Confirm the temperature with . Temperatures programmed by the user can be deleted or edited in the
same way. Preprogrammed temperatures cannot be changed.
12.1.4 Programming heating times
Press **** to access the heating time settings (Figure 10).
Select a free memory location ( ) and enter the desired heating time in the range 0h:01 min‑9h:59 min (Figure 11). Confirm the heating time entry with . Heating times programmed by the user can be deleted or edited in the same way. Preprogrammed heating times cannot be changed.
12.1.5 Date/Time
Pressing takes you to the date and time settings (Figure 12). Enter the
time in the form hh:mm:ss. Select the date field by touching it. Enter the
date in the form dd:mm‑20yy and confirm with .
12.1.6 Protective cover warning
Pressing displays a window for switching the protective cover warning on
(icon ) and off (icon ****Select the desired setting and confirm with .
If the protective cover warning is activated and the cover is not closed, an
icon appears in the start screen during measurement and an entry is generated
in the error log.
NOTICE
**** It is recommended to activate the protective cover warning and close the
cover.
12.1.7 Display settings
Pressing takes you to the display settings (Figure 13). By pressing the
arrow keys and , you can adjust the brightness , contrast and
the saturation of the display to the desired conditions. Select the
desired settings and confirm them with ****.
12.1.8 Signal tone
Pressing takes you to the signal tone settings. Use the arrow keys
and to adjust the volume. The setting is indicated in the status line
by the icon . Select the desired settings and confirm with .
When the signal tone is activated, different device states (e.g. finished
digestion) are indicated acoustically.
12.1.9 T‑Set
Press to access test equipment monitoring with the T‑Set. Here you can
independently carry out quality controls with the NANOCOLOR ® USB T‑Set (REF
919921) or NANOCOLOR ®
T‑Set (REF 919917). For further details, refer to chapter Quality Control,
page 53.
12.1.10 System settings
Pressing takes you to the menu for restoring the factory settings. Press
again to reset the factory defaults for the following settings:
User‑defined temperatures
User‑defined heating times
Display settings
Sound settings
Select the desired setting that you want to reset to factory defaults and
confirm with .
The service menu (icon ) is secured by a service code. The use of this
menu is reserved exclusively for service personnel for service work.
12.1.11 Error log
Press ** to enter the error log menu. A window appears with a list of the
errors that have occurred.
All system messages that have occurred are logged. The messages can be
filtered by the categories Notes/Warnings and Error messages by clicking on
the respective icon.
12.1.12 WAV‑Player**
To use the “WAV File Player”, contact MACHEREY‑NAGEL.
Performing a digestion
13.1 Sample preparation
WARNING
** Risk of chemical burn: Test kits with corrosive and dangerous
substances. Wear suitable protective equipment when working with the cuvettes.
Refer to the Safety Data Sheet (SDS) of the test kit used.
WARNING
** Cuvettes can overheat! Risk of burns, chemical burns and cuts. Check the
digestion conditions. Check the notes and instructions in the operating manual
of the test kit used. Pay particular attention to the specified digestion
temperatures and digestion times.
WARNING
Damaged cuvettes can break during the digestion process. Do not use
cuvettes that are damaged or were dropped. Make sure that the cuvettes are not
damaged before inserting them into the heating block.
Prepare the sample according to the instructions in the package insert and the
pictogram instructions (Figure 14). HKS 61
13.2 Insert cuvettes
NOTICE
**** Contaminated cuvettes can damage the heating block. Clean the outside of
the cuvette before inserting it into the bores of the heating block.
WARNING
**** The heating block becomes very hot, there is a risk of burns. Do not
insert your fingers into the bores of the heating block. Close the protective
lid during operation and keep the lid closed during the entire heating and
cooling phase.
Prepare the cuvettes as described in the instruction manual of the test kit
and insert them into the provided bores of the heating block (Figure 15).
Close the protective lid(s) of the heating block.
13.3 Perform digestion
WARNING
The protective cover(s) must be installed and closed to prevent injury.
WARNING
Risk of inhalation of irritant vapors. During digestion in an open system
(e.g. digestion with top‑mounted cooler) harmful vapors may escape. When
performing a digestion with open cuvettes (digestion with top‑mounted cooler),
the heating block must be operated in a fume hood.
WARNING
**** Risk of chemical burns: If a cuvette breaks, make sure that the liquid
does not come into contact with the skin. If necessary, use an exhaust system
to remove chemical vapors.
Press to start the selected program. The selected temperature will flash
on the display during the heating phase. When the desired program temperature
has been reached, a signal tone sounds and the selected time runs out. The
colon between hours and minutes flashes in the display while the heating time
is running.
During the heating phase, the current temperature can be displayed by pressing
the key.
NOTICE
A running program can be aborted at any time. A program abort is
initiated by pressing the key. STOP appears on the display. Within 5 s,
the abort must be confirmed by pressing the ** key again, otherwise the
program will continue to run normally.
13.4 Temperature monitoring**
When Graphic mode is activated (see chapter Graphic mode, page 44), the start
screen changes to Graphic mode (Figure 16) a few seconds after a program is
started and displays the temperature curve of the current heating
process.
The temperature range can be enlarged or reduced by tapping the graphic field. Pressing the button takes you back to the program selection. With the heating block NANOCOLOR ® VARIO 4 it is possible to switch between the two sides of the heating block in the graphic mode (see chapter Graphic mode, page 44) by pressing the key. If the display is not touched for some time in the program selection, the view automatically changes back to the graphic mode.
13.5 Remove cuvettes
WARNING
Danger of burns. The sample cuvettes are hot. Wear suitable protective
equipment **(e.g. thermally insulating gloves). Remove the vials only when
the temperature is below 80 °C.
WARNING
Risk of injury: If the cuvettes cool down quickly, there is a risk that
they will break and cause cuts. Do not cool hot vials with cold water unless
specifically required. Allow the vials to cool in the air or in the heating
block.
WARNING
Dispose of chemicals, waste and used cuvettes according to national,
regional and **local regulations.
When a program is finished, a signal tone is emitted (see chapter Signal tone,
page 48) and the digestion is finished. “END” appears in the display. Open the
protective hoods and remove the vials from the heating block (Figure 17).
After pressing one of the keys ****the last selected program is
recalled.
NOTICE
**** The internal fan of the device runs parallel to the sample digestion
and, if necessary, even beyond.
14. Quality Control
Each institution must establish its own QC policy.
To check the performance of the device, we recommend measuring the temperature
with a thermocouple. The temperature accuracy can be checked with the
NANOCOLOR ® USB T Set (REF 919921) or NANOCOLOR ® T Set (REF 919917) provided
by MACHEREY‑NAGEL. These temperature probes are suitable for checking the
device in case of a deviation and for adjusting the device.
Pressing in Settings **takes you to test equipment monitoring with (USB)
T‑Set.
Press the Test key to open the test menu. In the test menu all programmed
temperatures are checked. In this menu the calibration of the device is not
changed.
Press the ** button to open the calibration menu (Figure 18). In the
calibration menu, all programmed temperatures are readjusted using the base
calibration stored in the (USB) T‑Set.
After finishing a calibration, the acceptance of the new values must be
confirmed with ****.
Pressing the button in the calibration menu will restore the device
calibration to factory default. With the heating block NANOCOLOR ® VARIO 4,
the block to be tested (left or right) can be selected in the two menus and
with the keys .
Press the key to start the selected program. The determined temperature
deviations are displayed in different colors depending on the deviation.
Deviation | Color |
---|---|
< 1 °C | Green |
1 < x < 2 °C | Yellow |
> 2 °C | Red |
A report is stored in the device for both temperature testing and calibration. This report can be output using the software supplied (see Data Export).
Data export
The data generated in the course of quality control can be output to a PC using the “Ka‑ lib‑Software” program supplied. Proceed as follows:
- Connect the heating block to a computer using the USB‑A/B cable provided (see scope of delivery) and check the connection (device manager).
- Start the “Kalbi‑Software” from the supplied DVD. The connected heating block is automatically searched and the data is read in the background.
- Click on “Testing” in the software to generate a certificate of temperature verification. Click on “Calibration” to generate a certificate of heating block adjustment.
- Then disconnect the device from the computer.
Update
NOTICE
We recommend to make a data backup before the update (see chapter “Data
export”).
MACHEREY‑NAGEL offers updates for the heating blocks on its website. Download
the folder with the update files from the MACHEREY‑NAGEL website. Unzip the
folder and follow the instructions in the update guide.
Maintenance and cleaning of the device
17.1 Cleaning of the device
NOTICE
**For all cleaning work, the unit must be switched off and disconnected
from the power supply.
WARNING
**Fire hazard: Do not use flammable cleaning agents or organic solvents to
clean the unit.
WARNING
**Danger of burns: Do not clean the device while it is still hot.
WARNING
** Risk of chemical burns: If a cuvette breaks, make sure that the liquid
does not come into contact with the skin. If necessary, use an exhaust system
to remove chemical vapors.
WARNING
**** Dispose of chemicals, waste and used cuvettes according to national,
regional and local regulations.
WARNING
Danger of boiling delay: Always keep the bores of the heating block dry.
NOTICE
Do not clean the device with acetone or similar products.
Clean the unit regularly to ensure continuous and correct operation.
Turn off the unit by disconnecting it from the power source.
Wipe the outside of the device with a damp cloth or a mild detergent and then
dry the device with a soft cloth. Remove splashes on the unit immediately. If
necessary, clean or dry the touch screen with a soft, lint‑free cotton cloth.
Light dirt in the holes can be removed with a soft brush.
If the inside of the device is contaminated by a cuvette leak, contact the
manufacturer or local distributor and arrange for the following measures to be
taken:
a) Switch off the device and disconnect it from the power source.
b) Wait until the temperature of the heating block has dropped to room
temperature.
c) Use a pipette to remove any remaining liquid in the holes and dispose it of
according to the applicable regulations.
d) Remove glass splinters with tweezers and remove any remaining liquid on the
device. Avoid contact of the liquid with the skin.
17.2 Maintenance
NOTICE
Malfunction or damage of the device due to incorrect maintenance. Maintenance
of the device may only be carried out by qualified personnel.
WARNING
Contaminated bores can damage the cuvettes used. Check the cleanliness of the
holes at regular intervals and as part of maintenance. Contaminated wells must
not be used for sample digestion. Clean the dirty wells.
Establish maintenance intervals according to the regulations that apply to
you. Regardless, MACHEREY‑NAGEL recommends at least one annual maintenance of
the unit. The following points should be part of a maintenance of the unit:
- Check the power supply
- Check the function of the display and touch screen
- Check the cleanliness of the unit and the bores
- Check the individual temperatures (see chapter Quality Control)
- Check the heating time (against stopwatch)
- Check time and date
- Check the automatic switch‑off after the time has expired
If you have any questions about maintenance of the unit, contact MACHEREY‑NAGEL or your local representative.
Errors, sources and solutions
Depending on the operating status, different messages can be displayed. The source of the error can be either an operating error or a malfunction of the device.
Error | Source | Solution |
---|---|---|
Current temperature is too high to start the selected program | The actual | |
temperature of the heating block is warmer than the selected temperature |
Wait until the heating block has cooled down and start the program again. The
current temperature of the block can
be displayed by pressing .
Digestion aborted| The temperature has deviated from the setpoint by more than
4°C during digestion. The digestion was automatically interrupted.| Do not
insert cuvettes into the holes of the heating block during digestion.
The entered temperature/time is already in the memory.| The newly entered
temperature/time already exists.| Use the temperature/time already programmed.
The entered temperature is outside the possible range.| The entered
temperature does not correspond to the possible temperature limits.| Enter a
temperature in the range 40–160 °C.
Critical hardware error.| There’s a hardware problem.| There’s a hardware
problem.
The unit does not start correctly after being switched on.| The fuse is
defective.
The power supply is defective.| Check the power supply to the unit and check
the fuses on the back of the unit.
Only one T‑set can be used. If you want to use the (USB) T‑set, disconnect
the (USB) T‑set first.| Two T‑sets are connected for temperature testing.|
Disconnect the connection to the T‑set that you do not want to use.
The device connected to the USB port is not supported| The used device is not
recognized because it is unknown or defective.| Connect the device again.
Check the function of the unit on another instance (e.g. PC).
The last test data will be lost if you leave the test now. Press to
continue.| The data of the current test (see chapter Quality Control) has not
yet been saved.| Wait until the test is finished before leaving the menu.
The last calibration data will be lost if you leave the calibration now.
Press to continue.| The data of the current calibration (see chapter
Quality Control) have not yet been saved.| Wait until the calibration is
finished before leaving the menu.
The T‑set must be connected to finish the calibration/test. Press to stop
the process.| The (USB) T‑Set was removed during calibration/test.| Reconnect
the (USB) T‑Set to the device to finish the calibration.
The T‑Set must be connected to start the calibration/test.| No (USB) T‑Set is
connected.| Connect a (USB) T‑Set to the heating block and repeat the
procedure.
The difference between the heating block and T‑set temperature is outside the
permissible range. Check whether the T‑set is connected correctly.| The
difference between the heating block and T‑set temperature is outside the
permissible range. Check whether the T‑set is connected correctly.| Check that
the (USB) T‑set is properly plugged in.
Check that the temperature sensor is inserted correctly.
Check that the temperature sensor is inserted on the correct side (VARIO 4
only).
Overtemperature sensor was activated (185°C). Take the
heating block out of operation.| The temperature sensor is defective and the
block is overheated| Contact your distributor or MACHEREY‑NAGEL to request
repair of your unit.
Service
19.1 Spare parts, accessories and consumables
NOTICE
Only use accessories and spare parts approved by the manufacturer. If parts
not approved by the manufacturer are used, the operator is responsible for the
conformity of the device and the warranty will be voided.
WARNING
Risk of injury: Injury to persons and damage or malfunction of the device
or equipment when using parts not approved by the manufacturer. The
replacement parts in this section are approved by the manufacturer.
Description | REF |
---|
NANOCOLOR® USB T‑Set for electronic temperature control and calibration for
heating blocks NANOCOLOR ® VARIO 4 / VARIO C2 / VARIO C2 M / VARIO HC / VARIO
Mini| 919921
NANOCOLOR® T‑Set for electronic temperature control and calibration for
heating blocks NANOCOLOR ® VARIO 4 / VARIO C2 / VARIO C2 M / VARIO HC / VARIO
3 / VARIO compact| 919917
T‑Set Adapter 16 mm| 919924
Protective cover for NANOCOLOR® VARIO 4 / VARIO C2 / VARIO C2 M / VARIO HC,
transparent| 919310
Protective cover with bores for TOC tests for NANOCOLOR® VARIO 4 / VARIO C2 /
VARIO C2 M / VARIO HC, transparent| 919309
Safety cover sheet for NANOCOLOR® VARIO 4 / VARIO C2 / VARIO HC / VARIO 3 /
VARIO compact| 916598
Reducing adaptors 16 → 13 mm for NANOCOLOR® heating blocks (8 pieces)| 916910
Reducing adaptors 22 → 16 mm for NANOCOLOR® heating blocks (2 pieces)| 919916
Tube for sample decomposition 22 mm Ø, NS 19/26 with glass stopper for
NANOCOLOR® VARIO C2 M (2 pieces)| 91666
Empty reaction tubes with 16 mm outer diameter (20 pieces)| 91680
Empty reaction tubes with 22 mm outer diameter for NANOCOLOR® VARIO C2 M (2
pieces)| 91622
USB‑serial‑adaptor for NANOCOLOR® VARIO 4 / VARIO C2 / VARIO C2 M and
NANOCOLOR ® T‑Set| 919926
USB‑cable A/B for NANOCOLOR® VARIO 4 / VARIO C2 / VARIO C2 M| 919687
Transport case for NANOCOLOR® heating blocks VARIO 4 / VARIO C2 / VARIO C2 M|
91938
NOTICE
**** Disposal via public disposal systems is not permitted. Contact your
local MACHEREY‑NAGEL representative.
Disposal according to EU Directive 2012/19/EU. In accordance with the EU Directive 2012/19/EU MACHEREY‑NAGEL takes back the old device and disposes it free of charge.
19.3 Warranty, liability and complaints
The warranty for this device is 24 months from date of purchase. The original
invoice serves as proof and must be presented when making a claim. In case of
improper handling and / or maintenance of the device, the warranty expires. It
does not cover defects that are due to a power supply other than the external
power supply included in the delivery.
The warranty is limited to the repair of defective parts or ‑at
MACHEREY‑NAGEL’s discretion ‑ the delivery of a faultless replacement unit.
The warranty period of 24 months shall not be affected by any warranty claim.
There is no right of withdrawal. Further claims are excluded.
These include in particular all claims for damages resulting from
consequential or indirect damages. In addition, our general terms and
conditions of sale and delivery apply in the currently valid version as
printed on all price lists.
19.4 Contact
MACHEREY‑NAGEL GmbH & Co. KG
If you still have questions or need technical assistance after reading the
operating instruction, contact:
MACHEREY‑NAGEL GmbH & Co. KG
Valenciennes Str. 11 · 52355 Duren · Germany
Phone: +49 2421 969‑0
E‑Mail: info@mn-net.com
www.mn‑net.com
19.5 Version history
Manual Heating block NANOCOLOR VARIO C2 / VARIO 4 EN, V1.00 / 03.2008, March
2008
Manual Heating block NANOCOLOR VARIO C2 / VARIO 4 EN, V2.00 / 08.2021, August
2021
Manual heating block NANOCOLOR®
VARIO C2 / VARIO 4 EN, 08.2021