Ameda Penguin Deluxe Nutritional Warmers Instructions
- June 14, 2024
- Ameda
Table of Contents
- WELCOME TO THE AMEDA FAMILY
- WARNINGS AND CAUTIONS
- INTENDED USE
- PRODUCT DESCRIPTION Penguin Nutritional Warmer
- SET-UP INSTRUCTIONS
- INSPECTING BEFORE OPERATION
- OPERATING INSTRUCTIONS
- MAINTENANCE
- SERVICING AND SPARE PARTS
- WARRANTY
- TECHNICAL DATA, STORAGE AND OPERATING CONDITIONS
- AMEDA ACCESSORIES
- EMC INFORMATION
- Documents / Resources
Ameda Penguin
Deluxe Nutritional Warmers
Instructions for Use
WELCOME TO THE AMEDA FAMILY
We are excited to help you meet your nutritional warming needs. With the Ameda
Deluxe Single Well or Four-Well Penguin® Nutritional Warmers, you can safely
warm feedings and protect vital nutrients.
The Ameda Deluxe Nutritional Warmers work with the patented design Ameda
Therma-Liner® which isolates the feeding container from the water while
providing an encompassing kangaroolike warming environment using water to
safely transfer heat.
For any questions regarding your Ameda Deluxe Penguin Nutritional Warmer,
CustomerCare™ is available during business hours 8am-4:30pm CST on weekends at
1.877.99.AMEDA (26332).
Have your serial number for warmers and lot number for Therma-Liners when
calling. Please check www.ameda.com for up-to-date
hours of operation.
Icon Glossary
| Warning| | Keep Dry| | Serial Number
---|---|---|---|---|---
| Bail| | Manufacturer| | Catalog Number
| Important| | Temperature Limit| | Unit of Measure
| Hot Surface| | Humidity Limitation| | CUL Listed
| Grounded| | Pressure Limitation For Operation/Transport/Storage| |
Recyclable
(Therma-Liners)
WARNINGS AND CAUTIONS
IMPORTANT SAFEGUARDS
When using electrical products, the following basic safety precautions should
always be followed.
READ ALL INSTRUCTIONS BEFORE USING THIS PRODUCT
DANGER To reduce the risk of electrocution:
- DO NOT place or store where heater can fall or be pulled into a sink.
- DO NOT place or drop heater into water or other liquid.
- DO NOT reach for electrical product that has fallen into water. Immediately unplug from electrical outlet.
The WARNING symbol identifies important safety instructions that if not followed can lead to injury or damage to the heater. The warning symbols when used with the following words stand for:
WARNING Can lead to serious injury or death.
CAUTION Can lead to minor injury.
IMPORTANT Can lead to material damage.
WARNING To avoid fire, electrocution, serious burns or injury.
- ENSURE the power cord is easily accessible and removable, in the event of an emergency requiring immediate and total power disconnection.
- DO NOT use the equipment in a manner not specified.
- DO NOT use AC power source that applies more than 132 VAC between the supply conductors or between supply conductor and ground. A protective ground connection by way of the grounding conductor in the power cord is required for safe operation.
- DO NOT use parts or accessories other than those that are recommended by the manufacturer.
- If the equipment appears damaged or operates abnormally, protection may be impaired.
DO NOT attempt to continue operation. Have the equipment serviced.
WARNING To avoid fire, electrocution, serious burns or injury.
- ALWAYS operate heater on a flat, horizontal surface.
- Close supervision is necessary if product is used near children or infants.
- Use product for its intended use ONLY, as described in this manual.
- Prior to use, always inspect power cord for damage or exposed wires. If damage and/or exposed wires are found, DO NOT use and call Ameda CustomerCare at 1.877.99.AMEDA (26332).
- NEVER operate electric heater
- if it has a damaged cord or plug
- if it is not working properly
- if it has been dropped, damaged, or becomes wet without being properly and fully assembled
- ALWAYS keep cord away from heated surface.
- ALWAYS ensure the power cord is disconnected before cleaning or servicing.
- ALWAYS ensure the voltage of the product is compatible with the power source.
- ALWAYS ensure the power cord is inserted in a socket-outlet with a protective earth contact for grounding. Intentional AC interruption is prohibited.
- ALWAYS ensure servicing is performed by qualified service personnel only using the Manufacturer’s Service Manual. There are no user-serviceable parts.
CAUTION Can lead to minor injury.
- DO NOT use Therma-Liner
- for more than one shift, not to exceed 12 hours
- on more than one patient, baby, or child
IMPORTANT Can lead to material damage.
- NEVER put electrical product in water or a sterilizer, as you can cause permanent damage to the nutritional heater.
- DO NOT use abrasive cleaners/detergents when cleaning the nutritional warming unit or accessories. CaviWipe™ or SaniWipe™ are recommended for surface cleaning.
- DO NOT use sharp objects on the device buttons or wells.
- DO NOT use Therma-Liner for more than one shift or 12 hours.
- DO NOT overfill or under-fill Therma-Liner.
- If servicing is required, ensure servicing is performed by qualified service personnel.
SAVE THESIS INSTRUCTIONS
INTENDED USE
The Ameda Penguin® Nutritional Warmer is intended to warm foods at a range that includes body temperature.
PRODUCT DESCRIPTION Penguin Nutritional Warmer
Your Penguin Nutritional Warmer is a feeding warming device and is customized with settings and warming profiles according to your unique facility needs. The nutritional warmer operates with gentle vibration which helps preserve and suspend feeding nutrients and promote even warming.
3.1 BUTTON DESCRIPTION
Navigation Buttons
top| | Scroll Up| | Scroll Up|
---|---|---|---|---|---
Middle| Select| Go Back
Bottom| Scroll Down| Scroll Down
Beak Buttons
SET-UP INSTRUCTIONS
4.1 PREPARING THE WARMER
- PLACE heater onto a stable, flat, horizontal surface.
- PLUG the power cord into the power source.
- CONNECT the power cord into the back of the heater device.
4.2 CUSTOMIZING ROOM/BED/POD INFORMATION
- FLIP the rear power switch into the “ON” position.
- PRESS and HOLD the bottom navigation button.
- FOLLOW the on-screen instructions to ENTER up to 5 spaces for Room, up to 5 spaces for Bed and up to 5 spaces for Pod: •
+abc/123
• =return to previous entry
• -abc/123
• Enter button=advance to next entry
• On/Off button=edits complete, return to Standby mode
4.3 MODES
STANDBY MODE
The Penguin Nutritional Warmer will be in Standby mode and cycle through the
following screens whenever powered on and not warming:
![Ameda Penguin Deluxe Nutritional Warmers - STANDBY MODE](https://manuals.plus/wp-content/uploads/2023/11/Ameda-Penguin-Deluxe- Nutritional-Warmers-STANDBY-MODE.jpg)
WARMING MODES
| Fridge to Feed
Feed Temp range of 90-100°F/32-38°C
Average Time 5-12 minutes depending on volume
---|---
| Frozen to Feed
Feed Temp range of 90-100°F/32-38°C
Average Time 10-18 minutes depending on volume
| Thaw
After Thaw Temp range of 38-45°F/3-7°C
Time 15-25 minutes depending on volume
READY TO FEED MODE – HOLD TIME (Fridge to Feed and Frozen to Feed only) The feeding is ready when the screen background turns green with the following words displayed:
REMAINING HOLD TIME
Displays a countdown of the remaining hold time. The warmer emits a ready tone
at the beginning of the hold time and periodically throughout. The warmer will
keep the feeding in a safe temperature range for 30 minutes (until turned off,
returned to Standby mode, or extends to Overtime mode).
THAW COMPLETE MODE
When the Thaw feature is used, the feeding should be removed immediately upon
completion to prevent over-warming. The warmer emits a ready tone. The heater
immediately shuts off, and milk must be refrigerated. The screen background
will turn green and the phrase shown will appear on the screen.
WARNING Milk must be refrigerated immediately.
OVERTIME MODE Fridge to Feed and Frozen to Feed only
After 30 minutes in Ready to Feed HOLD TIME, the warmer will enter Overtime.
Warming and vibration will turn off. The screen background will turn red and
the word “Overtime” will appear in the middle of the screen.
The Overtime screen will remain on until the unit is turned off or is returned
to Standby by PUSHING the On/Off button or Enter button.
Note: As the manufacturer, we recommend not using the feeding after the
30 minute hold time, but that feeding should be handled according to the
hospital policy.
DISPLAYED ERRORS
If the Penguin Nutritional Warmer encounters an error, the error will be
displayed on a red screen with a warning tone. If this should occur,
discontinue use of the warmer and call Ameda CustomerCare at 1.877.99.AMEDA
(26332).
4.4 THERMA-LINER PREPARATION
-
At the beginning of a shift, take out one Therma-Liner per user.
-
Using a permanent marker, enter the following information onto the Therma-Liner:
a. Patient Information b. Date/Time c. Information -
Open the water compartment (blue zipper) and fill to water line using room temperature water.
-
Pinch one end of the blue zipper between your thumb and forefinger and slide down the length of the zipper to close the water compartment.
-
Repeat to ensure a proper seal. This compartment will remain sealed for the duration of the 12 hour shift.
Note: Use care when removing and re-inserting milk container to prevent damage to interior of water compartment.
INSPECTING BEFORE OPERATION
Inspect the product for damage. Do not use if damaged.
Call Ameda CustomerCare at 1.877.99.AMEDA (26332).
OPERATING INSTRUCTIONS
-
Flip the rear power switch into the ‘ON’ position.
-
Press On/Off Button to begin.
-
Scroll to the warming mode of interest:
a. Room Temperature/Fridge to Feed
b. Frozen to Feed
c. Thaw -
Advance by pressing Enter or Select (green check mark)*.
- Note that if a wrong entry was made, you may press the Go Back selection button to go back a screen.
6.1 WARM REFRIGERATED OR FROZEN FEEDINGS
- Scroll to choose the container type and size of interest.
- Press Enter.
- Only when ready to begin warming the feeding, place the feeding container into the inner compartment of the Therma-Liner and gently push it to the bottom. Use care when removing and re-inserting feeding containers to prevent damage to the interior water compartment.
- Place the prepared Therma-Liner in the well and gently push it to the bottom.
- Press Enter*.
- The device will begin warming, vibrating, and alternating screens between the profile selected and elapsed time.
- When the feeding has been warmed to a range that includes body temperature, the device will enter Ready to Feed mode.
- Please remove the feeding from the Therma-Liner and set the Therma-Liner aside until next use during the shift.
- Press On/Off to return the Penguin to Standby mode.
- Note that if a wrong entry was made, you may press the Cancel selection button to go back a screen.
6.2 THAW FROZEN MILK TO COOL LIQUID STATE
Caution: the Penguin Nutritional Warmer is not recommended for thawing less
than 15 mL of breast milk in storage bags!
- Scroll to choose the container type and size of interest.
- Press Enter.
- Only when ready to begin warming the feeding, place the feeding container into the inner compartment of the Therma-Liner and gently push it to the bottom. Use care when removing and re-inserting feeding containers to prevent damage to the interior water compartment.
- Place the prepared Therma-Liner in the well and gently push it to the bottom.
- Press Enter*.
- The device will begin warming, vibrating, and alternating screens between the profile selected and elapsed time.
- When the feeding has been warmed to a liquid cool state, the device will enter Thaw Complete mode.
- Please immediately remove the feeding** from the Therma-Liner and set the Therma-Liner aside until next use during the shift.
- Press On/Off to return the heater to Standby mode.
*Note that if a wrong entry was made, you may press the Cancel selection button to go back a screen.
**If there are still some frozen particles in the feeding, let it sit until
the remainder thaws, or rotate the container gently to thaw the rest of the
way. It is better to have a few remaining ice crystals than to overshoot the
refrigerated range. Any portion that is meant for immediate feeding may be
placed in a bottle or syringe for further warming. Feeding for the future
should be placed immediately in the refrigerator for up to 24 hours. After 24
hours, it should be discarded.
CLEANING
- Disconnect Power Cord from the wall receptacle.
- Prepare reagent wipes (eg CaviWipe™ or SaniWipe™) containing Isopropanol, Ethylene Glycol Monobutyl Ether (2-Butoxyethanol), Diisobutylphenoxyethoxyethyldimethyl benzylammonium chloride.
- Using the prepared reagent wipe, clean and remove all surface debris from the device, including the power cord.
- Repeat as deemed necessary using a newly prepared reagent wipe each time.
- After the last cleaning wipe is used, allow device to air dry for a minimum of 10 minutes or dry with a soft cloth before returning the device to service.
MAINTENANCE
The Penguin Nutritional Warmer requires no preventative maintenance.
SERVICING AND SPARE PARTS
All heater service and repair can be accomplished by contacting Ameda. CustomerCare will issue a Return Material Authorization (RMA) or provide your facility’s biomedical department with parts needed to repair the heater and return it to operation.
WARRANTY
THE AMEDA WARRANTY IS NONTRANSFERABLE. THE AMEDA PENGUIN NUTRITIONAL WARMER IS
WARRANTED TO THE ORIGINAL OWNER ONLY.
From the date of the purchase of this product, Ameda, Inc. (“Ameda”) warrants
the Product against defects in material or workmanship for a period of two
years. Ameda’s sole obligation under this limited warranty shall be to repair
or replace, at Ameda’s option, any Product that is determined to be defective
by Ameda and determined to be covered by this express limited warranty. Repair
or replacement under this limited warranty is the sole and exclusive remedy of
the First User. ”
PROOF OF PURCHASE IN THE FORM OF A RECEIPTED INVOICE OR BILL OF SALE
EVIDENCING THAT THE PRODUCT IS WITHIN THE WARRANTY PERIOD MUST BE PRESENTED TO
OBTAIN WARRANTY SERVICE. This limited warranty is extended by Ameda ONLY
to the First User and is not assignable or transferable . For warranty service
contact Ameda at 1.877.99.AMEDA (26332).
EXCEPT TO THE EXTENT PROHIBITED BY APPLICABLE LAW, ANY IMPLIED WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE ON THE PRODUCT IS HEREBY
DISCLAIMER. INCIDENTAL, CONSEQUENTIAL, SPECIAL OR PUNITIVE DAMAGES OF ANY KIND
FOR BREACH OF THE EXPRESS LIMITED WARRANTY ON THE PRODUCT OR ANY WARRANTY
IMPLIED BY OPERATION OF LAW. OTHER THAN THE EXPRESS LIMITED WARRANTY SET FORTH
ABOVE, THERE ARE NO OTHER WARRANTIES THAT ACCOMPANY THIS PRODUCT AND ANY ORAL,
WRITTEN OR OTHER REPRESENTATION OF ANY KIND IS HEREBY DISCLAIMED.
The user/owner of this equipment shall have sole responsibility and liability
for any injury to person or damage to property resulting from: 1) Operation
not in accordance with supplied operating instructions, 2) Maintenance not in
accordance with authorized maintenance service instructions, 3) Service by
anyone other than manufacturer authorized qualified service personnel, 4)
Modification of the equipment or accessories and 5) Use of damaged or
unauthorized components and accessories.
TECHNICAL DATA, STORAGE AND OPERATING CONDITIONS
Penguin Single Well Nutritional Warmer
Power Rating: 120VAC +/- 10%, 60 Hz, 1.25A
Relative Humidity: 80% up to 31°C to 50% to 40°C / 80% up to 87.8°F to 50% up
to 104%F
Altitude: 2000 meters
Operating Temperature: 5 to 40°C/41 to 104°F
Storage Temperature: -40 to 75°C/-40 to 167°F
Line Fuse: 2 each, type “T” 1.25A, 250VAC
Mass: 3.96 kg
Size: 21.8 cm HX 19.4 cm WX 21.9 cm D
Pollution degree 2
Installation category H
Indoor use only
Penguin Four Well Nutritional Warmer
Power Rating: 120VAC+/- 10%, 60 Hz, 4A
Relative Humidity: 80% up to 31°C to 50% to 40°C / 80% up to 87.8°F to 50% up
to 104% F
Altitude: 2000 meters
Operating Temperature: 5 to 40°C/41 to 104°F
Storage Temperature: -40 to 75°C/-40 to 167°F
Line Fuse: 2 each, type “T” 5.0A, 250VAC
Mass : 9.06 kg
Size: 24.8 cm HX 54.0 cm WX 21.9 cm D
Pollution degree 2
Installation category H
Indoor use only
AMEDA ACCESSORIES
You may also order additional warmer accessories: Therma-Liner replacements or Perche Shelf through your sales specialist or Ameda CustomerCare at 1.877.99.AMEDA (26332).
EMC INFORMATION
THIS ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be installed and put into service according to the EMC provided in these Instructions for use. Portable and mobile RF communications equipment can affect this ELECTRICAL EQUIPMENT. Warning: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Penguin Nutritional Warmer, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result. Warning: Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation. Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
Guidance and manufacturer’s declaration – electromagnetic emissions
The Ameda Penguin Warmer is intended for use in the electromagnetic
environment specified below.
The customer or the user of the Ameda Penguin Warmer should ensures that it is
used in such an environment.
Test emissions| Compliance| Electromagnetic environment –
guidance
RF emissions CISPR 11| Group 1| The Ameda Penguin Warmer uses RF energy only
for its internal function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11| Class B| The Ameda Penguin Warmer is suitable for use
in all establishments, including domestic establishments and those directly
connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2| Class A
Voltage fluctuations/flicker emissions
IEC 61000-3-3| Compline
Guidance and manufacturer’s declaration – electromagnetic immunity
The Ameda Penguin Warmer is intended for use in the electromagnetic
environment specified below.
The customer or the user of the Ameda Penguin Warmer should ensures that it is
used in such an environment.
Immunity test| IEC 60601 test level| Compliance level|
Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2| ± 8 kV contact
±2kV, ±4kV, ±8kV, ± 15 kV air| ± 8 kV contact
±2kV, ±4kV, ±8kV, ± 15 kV air| Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4| ± 2 kV for AC line; 100 kHz repetition frequency| ± 2 kV for AC
line; 100 kHz repetition frequency| Mains power quality should be that of a
typical commercial or hospital environment.
Surge IEC 61000-4-5| ±0.5kV, ±1kV line-to-line; ±0.5kV, ±1kV, ±2kV line-to-
ground| ±0.5kV, ±1kV line-to-line; ±0.5kV, ±1kV, ±2kV line-to-ground| Mains
power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input
lines IEC 61000-4-11| • 0% UT; 0.5 cycle At 0°, 45°, 90°, 135°, 180°, 225°,
270° and 315°
• 0% UT; 1 cycle and 70% UT; 25/30 cycles Single phase: at 0°
• 0% UT; 250/300 cycles| • 0% UT; 0.5 cycle At 0°, 45°, 90°, 135°, 180°, 225°,
270° and 315°
• 0% UT; 1 cycle and 70% UT; 25 cycles At 0°
• 0% UT; 250 cycles| Mains power quality should be that of a typical
commercial or hospital environment. If the user of the Ameda Penguin Warmer
requires continued operation during power mains interruptions, it is
recommended that the Ameda Penguin Warmer be powered from an uninterruptible
power supply.
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8| 30 A/m
50Hz or 60Hz| 30 A/m
50Hz| Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
NOTE UT is the ac mains voltage priority to application of the test
level.
Guidance and manufacturer’s declaration – electromagnetic immunity
The Ameda Penguin Warmer is intended for use in the electromagnetic
environment specified below.
The customer or the user of the Ameda Penguin Warmer should ensures that it is
used in such an environment.
Immunity test| IEC 60601 test level| Compliance level|
Electromagnetic environment – guidance
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3| 3V
0.15MHz – 80MHz
80% AM at 1 kHz
6V in ISM and amateur radio bands between 0.15 MHz and 80 MHz 80% AM at 1 kHz
10V/m
80 MHz – 2.7 GHz
80% AM at 1 kHz
9 – 28 V/m
At IEC 60601-1-2:2014
Table 9 frequencies
| 3V
0.15MHz – 80MHz
80% AM at 1 kHz
6V in ISM and amateur radio bands between 0.15 MHz and 80 MHz 80% AM at 1 kHz
10V/m
80 MHz – 2.7 GHz
80% AM at 1 kHz
9 – 28 V/m
At IEC 60601-1-2:2014
Table 9 frequencies
| Portable and mobile RF communications equipment should be used no closer to
any part of the Ameda Penguin Warmer, including cables, than the recommended
separation distance calculated from the equation applicable to the frequency
of the transmitter.
Recommended separation distance d = 1.2√P
d = 1.2√P 80 MHz to 800 MHz
d = 2.3√P 800 MHz to 2.7 GHz
Where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, a should be less than the compliance level in
each frequency range. b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
a Field strength from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the Ameda Penguin Warmer is
used exceeds the applicable RF compliance level above, the Ameda Penguin
Warmer should be observed to verify normal operation. If abnormal performance
is observed, additional measures may be necessary, such as re-orienting or
relocating the Ameda Penguin Warmer.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be
less than 10 V/m.
Recommended separation distances between portable and mobile RF communications equipment and the Ameda Penguin Warmer
The Ameda Penguin Warmer is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the user of
the Ameda Penguin Warmer can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Ameda Penguin Warmer as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter (W)| Separation distance
according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 1.2√P| 80MHz to 800MHz
d = 1.2√P| 800 MHz to 2.7 GHz
d = 2.3√P
0.01| 0.12| 0.12| 0.23
0.1| 0.38| 0.38| 0.73
1| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Distributed in USA by:
Distribution in other countries: For product information or feedback, call
your local distributor or location where you purchased the product.
For a listing of distributors in your country, please visit
www.ameda.com
The Ameda Logo, Ameda, Ameda Mom Inspired.
Hospital Trusted, Flexishield, Finesse™, Cool’N Carry, Store‘N Pour,
HygieniKit and/or registered trademarks of Ameda, Inc. in the United States
and other countries.
Ameda, Inc.
485 Half Day Road. Suite 320 Buffalo Grove, IL 60089
1-866-992-6332(AMEDA)
Mothers Choice Products Ltd.
2133-11871 Horseshoe Way Richmond, BC V7A5H5
1.800.604.6225
Manufactured for:
Ameda, Inc.
485 Half Day Road, Suite 320 Buffalo Grove, IL 60089 USA
**1200066-1222**
Documents / Resources
|
Ameda Penguin Deluxe Nutritional
Warmers
[pdf] Instructions
Penguin Deluxe Nutritional Warmers, Penguin, Deluxe Nutritional Warmers,
Nutritional Warmers, Warmers
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>