Ameda 17803 Platinum Breast Pump Instruction Manual
- June 3, 2024
- Ameda
Table of Contents
- ICON GLOSSARY
- PRODUCT DISPOSAL
- IMPORTANT SAFEGUARDS
- INTENDED USE
- AMEDA PLATINUM BREAST PUMP DIAGRAM
- AMEDA PLATINUM BREAST PUMP TROLLEY
- SET UP AND ASSEMBLY
- CONTROLS AND DISPLAYS
- CLEANING
- STORING AND TRANSPORTING
- TROUBLESHOOTING
- MAINTENANCE
- TECHNICAL INFORMATION
- GUIDANCE AND MANUFACTURER’S DECLARATION
- LIMITED WARRANTY
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
17803 Platinum Breast Pump
Instruction Manual
ICON GLOSSARY
| Double Insulated/Class II| | Waste Electrical and Electronic Equipment
---|---|---|---
| Type BF Applied Part| | Fragile
| Connected to Wall Outlet| | Protect from sunlight
| Washing Instruction| | Keep Dry
| % of Full Scale Suction 1% -100%; (approximately 30mmHg to 250mmHg/4 .0kPa
to 33.3kPa)| | Serial Number
| Cycles Per Minute (30-80 cpm)| | Lot Number
| Timer Display| | Catalogue Number
**| Consult Instructions For Use| | Temperature Limitation
| Fuse| | Humidity Limitation
| Caution Consult Documents| | Pressure Limitation For Operation/Transport/
Storage
IP21**| Solid Particle/Liquid Ingress Classification| | Manufacturer
| UL Certified E493585 MEDICAL – GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL
SHOCK, FIRE AND
MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH ANSI/AAMI ES 60601-1 (2005) + AMD1
(2012) and CAN/CSA-C22 .2 No . 60601-1 (2008) + (2014) and IEC 60601-1-6
(2010) + AMD1 (2013) and IEC 62366 (2007) + AMD1 (2014)| |
PRODUCT DISPOSAL
This product contains electrical and electronic equipment.
- The internal components of this product may contain hazardous materials. Dispose of this product according to local or regional waste administration systems and regulations.
- Do not dispose of as unsorted municipal waste .
- For further information regarding the environmental performance of this product please visit our minimum size website www.ameda.com.
IMPORTANT SAFEGUARDS
When using electrical products, especially if children are present, the following basic safety precautions should always be maintained.
READ ALL INSTRUCTIONS BEFORE USING
Danger: To reduce the risk of electrocution:
-
Always unplug electrical devices immediately after use.
-
Do not use while bathing, showering or swimming.
-
Do not place or store where product can fall or be pulled into bathtub, sink or pool.
-
Do not place or drop product into water or other liquid.
-
Do not reach for electrical product that has fallen into water . Unplug from wall outlet immediately .
Warning: To reduce the risk of burns, electrocution, fire, or injury to persons: -
To avoid the risk of electric shock, this equipment must only be connected to an electrical outlet with proper grounding . Use only the Ameda supplied 3-prong cord .
-
Never leave product unattended when plugged into electrical outlet.
-
Close supervision is necessary when product is used near children or invalids.
-
Use product only for intended use as described in this manual .
-
Do not use attachments or other milk collection kits other than those recommended by manufacturer.
-
Never operate if product has damaged cord or plug, is not working properly, or has been dropped, damaged, or becomes wet .
-
Keep cord and all attachments away from heated surfaces.
-
Never use while sleeping or drowsy.
-
Do not use outdoors with cord .
-
Do not operate where aerosol spray products are being used or oxygen is being administered.
-
Do not disassemble the Ameda Platinum Breast Pump.
-
Always place pump on a horizontal flat surface when using product.
-
Do not remove the breast flange from your breast while pumping. See Instructions for removing HygieniKit.
-
Pumping can induce labor, do not use while pregnant unless approved by your breastfeeding specialist or healthcare professional .
-
Strangulation Hazard: Keep tubing and cords out of reach of children.
-
Choking Hazard: Small parts, NOT for children under 3 years.
Contradictions for Use
There are no known contraindications for use with this product.
SAVE THESE INSTRUCTIONS
INTENDED USE
The Ameda Platinum Breast Pump is intended to express and collect milk from
the mother’s breast, to alleviate engorgement of the breast, maintain the
ability of lactation, and provide mother’s milk for future feedings when
separation of mother and baby occurs.
The Ameda Platinum Breast Pump is an electrically-powered vacuum device
intended to express milk from the breast of a nursing woman using the Ameda
HygieniKit Milk Collection System. The vacuum profile of the Ameda Platinum
Breast Pump is patterned after the sinusoidal waveform of the Ameda SMB Breast
Pump, and implemented using a motor-driven, piston pump mechanism that is
under microprocessor-based control. The electrical hardware/software
architecture enables the end-user to set cycle speed, labeled “speed”, and
vacuum level, labeled “suction”, from independent front panel controls, in
order to produce the necessary changes in vacuum output. Incorporating a
vacuum sensor and proportional valve inside a closed-loop control scheme
improves pump performance over previous pump designs in both single and double
pumping modes under varying ambient barometric conditions. Graphical displays
help the user to visualize pumping parameters and make repeatable adjustments.
The differentiating feature of an elapsed session timer display and its
control are intended ™ to facilitate a mother following a pumping protocol.
AMEDA PLATINUM BREAST PUMP DIAGRAM
1 . Suction Display (Maximum Shownin Percentage)| 10 . Milk Bottle/Breast
Flange Holders
---|---
2 . Speed Display (Maximum Cycles per minute shown)| 11 . Handle
3 . Speed Control| 12 . AC Power Inlet
4 . Power Button (Turn Pump ON/OFF)| 13 . Pump Feet
5 . AC Power Indicator| 14 . Trolley Mounting Hole
6 . Timer Display| 15 . Service Port Door
7 . Timer Reset Button| 16 . Pump Rear Foot Location
8 . Suction Control| 17 . Pump Front Foot Location
9 . Tubing Adapter Port| 18 . Pump Mounting Holes
AMEDA PLATINUM BREAST PUMP TROLLEY
Trolley Plate Top View (optional accessory)
Attaching the Ameda Platinum Breast Pump to Trolley (if applicable)
Trolley is available as an accessory item.
- Attach pump to trolley prior to plugging cord into wall outlet.
- Align front of the Ameda Platinum Breast Pump with wording, “Front of Ameda Platinum and Elite™ Pump” found on top of trolley mounting plate.
- Lower pump onto plate, inserting pump’s four feet into holes, one in each corner of trolley mounting plate.
- Locate mounting holes with wording, “Ameda Platinum Pump and Elite Pump Mounting” found on underside of trolley mounting plate.
- Use mounting screws for the Ameda Platinum Breast Pump supplied with trolley to secure pump to trolley base.
Trolley Operating Instructions
- Once pump is attached to trolley, ensure all wheels are unlocked to move the trolley-pump assembly. Wheels are unlocked by lifting the locking tab on each wheel.
- Move the trolley-pump assembly by gently pushing horizontally on the pump.
- When trolley-pump assembly is in the desired position, lock all wheels by pushing down on the locking tab on each wheel.
SET UP AND ASSEMBLY
Before Every Use
Always wash hands well with soap and water before handling the Ameda Platinum
Breast Pump and Ameda HygieniKit Milk Collection System.
IMPORTANT NOTE: ONLY USE THE AMEDA HYGIENIKIT MILK COLLECTION SYSTEM ON
AMEDA ELECTRIC BREAST PUMPS.
Prior to starting pump, assemble and connect the Ameda HygieniKit Milk
Collection System, per included instructions.
If the Ameda HygieniKit Milk Collection System is sterile, use as is.
If not sterile, see the Ameda HygieniKit Milk Collection System cleaning
instructions.
Ameda HygieniKit Milk Collection System
Ameda Platinum Breast Pump
- Place pump on a horizontal flat surface before use.
- Attach Ameda supplied grounded power cord to AC power inlet on rear of pump.
- Connect grounded plug to wall outlet.
- Ensure pump is positioned so that the power cord from the rear is accessible for removal.
- Insert tubing adapter into tubing adapter port.
Single and Dual Pumping
The Ameda HygieniKit Milk Collection System can be used to single or dual
pump.
To single pump: Remove one tube and close tubing adapter. To dual pump: Both tubes should be attached to tubing adapter.
CONTROLS AND DISPLAYS
Using “Up/Down” Buttons
Before starting pump, please note how the controls work.
- Press and release controls to increase/decrease settings by increments of 1.
- Press and hold controls to increase/decrease settings by increments of 10.
Operating the Pump
- When pump is plugged in, a light will appear next to the Power button.
- Press the Power button. The displays light up. Pump is now ON.
- Pump starts at lowest suction setting (1% or approximately 30mmHg/ 4 .0kPa) and maximum speed (80 cpm).
Setting Suction
- Pump starts at lowest suction setting (1% or approximately 30mmHg/4 .0kPa).
- Center nipple(s) in breast flange(s) and fill breast flange(s) with the breast(s) to create an air seal. Keep upper body and bottles/flanges upright while pumping.
- Press▲to increase suction until reaching maximum comfort level.
- If discomfort is felt in nipple(s) or breast(s), press▼to decrease suction. If discomfort is felt at all suction settings, turn pump off, insert a finger between breast(s) and breast flange(s) to break suction, remove breast flange(s) from breast(s). Discontinue pumping and contact a healthcare provider.
Setting Speed
- Pump starts at fastest speed (80 cpm).
- When milk flows (let-down), press▼ to decrease speed to help drain the breast faster .
- When milk flow slows, press▲ to increase speed to 60-80 cpm to help start another let-down .
- Repeat as milk flow changes to pump more milk.
NOTE : If “Check Kit” appears in suction window, see Troubleshooting section, on page 8.
Using Timer
- Timer starts when pump is on. Timer displays pumping duration in minutes : seconds.
- Timer can be reset by pressing Timer Reset button.
- Timer and pump automatically turn OFF after 60 minutes.
Turning Pump OFF
- When finished pumping, press and hold Power button to turn pump and timer functions OFF.
Removing Ameda HygieniKit Milk Collection System from Breast(s)
- After pump is OFF, if necessary, insert a finger between breast(s) and breast flanges(s) to break suction.
- Remove breast flanges(s) from breast(s).
- Place bottles with milk in bottle holders.
CLEANING
Cleaning the Ameda Platinum Breast Pump
For Personal Use
After each pumping, turn OFF the pump. Unplug cord from from wall outlet. Wipe
the Ameda Platinum Breast Pump with clean, damp cloth.
For Hospitals/Rental Stations
Turn OFF the pump. Unplug cord from wall outlet. Clean between users. Wearing
disposable gloves, wipe the Ameda Platinum Breast Pump with a hospital
approved antibacterial cleaner, such as CaviCide ,or PDI ®, CaviWipes®Sani-
Cloth®AF3.
STORING AND TRANSPORTING
Disconnect HygieniKit and power cord from pump before transporting or storing
the product.
When storing product, store in conditions as described in these instructions.
Handle with care when transporting or handling product.
TROUBLESHOOTING
Pump does not turn ON
- Ensure proper power cord connection to pump.
- Ensure grounded plug connection to electrical outlet.
- Ensure outlet has power .
Unable to adjust speed
- Ensure pump function is ON.
- Ensure “Check Kit” is not shown in suction display window.
Unable to adjust suction
- Ensure pump function is ON.
- Ensure “Check Kit” is not shown in suction display window.
Low or no suction
- Ensure “Check Kit” is not shown in suction display window.
- Check breast flange for adequate seal against breast.
- Check that the Ameda HygieniKit Milk Collection System is properly assembled.
- Check that the adapter cap is firmly snapped (attached) onto top of breast flange.
- Ensure silicone diaphragm is in place inside top of breast flange.
- Ensure tubing adapter is properly fitted into tubing adapter port on pump.
- Inspect valve for stretching or tears. If seen, replace valve.
- Ensure valve is pushed firmly into place in lower portion of breast flange.
- Ensure one side of tubing adapter is closed when single pumping .
“Check Kit” error message appears
- Ensure tubing adapter is properly fitted into tubing adapter port on pump.
- Ensure one side of tubing adapter is closed when single pumping .
- Check that adapter cap is firmly snapped onto top of breast flange.
- Ensure silicone diaphragm is in place inside top of breast flange.
- Check tubing and silicone diaphragm to ensure that parts are free of defects .
- Check tubing is firmly attached to adapter cap.
“Service Required” error message appears
- Call your local distributor. For listing of distributors worldwide, visit www.ameda.com.
The Ameda Platinum Breast Pump should be serviced only by Ameda, Inc. or authorized Service Centers approved by Ameda, Inc.
MAINTENANCE
It is recommended the Ameda Platinum Breast Pump be inspected at least once every year to include the following:
- Visually check pump housing for cracks, breakage, or damage and ensure Pump Feet are present.
- Visually check for power cord damage .
- Check for proper operation of controls and buttons.
- Check there are no missing screws .
- Check cover is present on Service Port Door.
- Check Tubing Adapter Port is free from obstruction.
Quick Functional Check
The following checks/tests should be performed prior to use by a new user:
- Visual checks for power cord damage .
- Check for abnormal noise .
- Check for proper operation of controls and buttons.
Service Port
- Do not open Service Port Door on bottom of the Pump.
- Do not connect any equipment to the service port when operating the pump.
- Service port is to be accessed only by authorized Service Centers approved by Ameda, Inc.
Replacing Detachable Power Cord:
- Power cord is detachable from power inlet .
- If power cord condition is in doubt, only replace with Ameda supplied 3-conductor cord equipped with protective ground conductor.
TECHNICAL INFORMATION
Protection Class II (Double Insulated), Type BF Applied Parts
Power Supply:
Operating Input Voltage 100 VAC to 240 VAC
Operating Frequency 50/60Hz
Input Power 80VA
Fuse Rating Quick Acting (F), 2.5A, 250V, 5 x 20m
Power Cord: Use only Ameda supplied power cord
For North America, use Ameda 24502059
Minimum Voltage Rating: 125V
Minimum Current Rating: 10A
Plug Type: IEC 60320 C13 (Hospital Grade)
Maximum Length: 10 feet (3.05m)
Cord Type: 3x18AWG, rated VW-1, 105C
Jacket Type: SJT or SJTW
Certification: UL Listed (ELBZ)
Detachable power cord provides mains isolation via appliance coupler.
Operation: Continuous
ISO 10079-1, Intermittent Suction, Medium Vacuum
Suction: Shown in percentage of full scale suction 1%-100%;
(approximately 30mmHg to 250mmHg/ 4.0kPa to 33.3kPa)
Speed: 30 to 80 cycles per minute (cpm)
Operational Conditions: Pump
Temperature +41°F (+5°C) to +90°F (+32°C) Relative Humidity 15% to 90% Atmospheric Pressure 0.69atm (70kPa) to 1.05atm (106kPa)
Transport and Storage Conditions: Pump
Temperature -20°F (-29°C) to +158°F (+70°C)
Relative Humidity 10% to 90%
Atmospheric Pressure 0.69atm (70kPa) to 1.05atm (106kPa)
Weight: Approximately 9.75 lbs. (approximately 4.42 kg)
Dimensions: Length 9.75 in. (24.7 cm) /Width 10.5 in.
(26.7 cm) / Height 8.0 in. (20.3 cm)
Standards:
UL Certified E493585 MEDICAL – GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL
SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH ANSI/AAMI ES
60601-1 (2005) + AMD1 (2012) and CAN/CSA-C22.2 No. 60601-1 (2008) + (2014) and
IEC 60601-1-6 (2010) + AMD1 (2013) and IEC 62366 (2007) + AMD1 (2014)
GUIDANCE AND MANUFACTURER’S DECLARATION
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC provided in
these Instructions for use. Portable and mobile RF communications equipment
can affect MEDICAL ELECTRICAL EQUIPMENT.
Warning: Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the Ameda Platinum Breast Pump, including cables
specified by the manufacturer. Otherwise, degradation of the performance of
this equipment could result.
Warning: Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of this equipment could result in
increased electromagnetic emissions or decreased electromagnetic immunity of
this equipment and result in improper operation.
Warning : Use of this equipment adjacent to or stacked with other
equipment should be avoided because it could result in improper operation. If
such use is necessary, this equipment and the other equipment should be
observed to verify that they are operating normally.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Ameda Platinum Breast Pump is intended for use in the electromagnetic
environment specified below. The customer or the user of the Ameda Platinum
Breast Pump should assure that it is used in such an environment.
Emissions test| Compliance| Electromagnetic environment – guidance
RF emissions CISPR 11| Group 1| The Ameda Platinum Breast Pump uses RF energy
only for its internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11| Class B| The Ameda Platinum Breast Pump is suitable for
use in all establishments, including domestic establishments and those
directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2| Class A
Voltage fluctuations/flicker emissions IEC 61000-3-3| Complies
Recommended separation distances between portable and mobile
RF communications equipment and the Ameda Platinum Breast Pump
The Ameda Platinum Breast Pump is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or
the user of the Ameda Platinum Breast Pump can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Ameda Platinum Breast Pump as
recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output power of
transmitter (W)| Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 1,2√V P| 80 MHz to 800 MHz
d = 1,2√P| 800 MHz to 2.7 GHz
d = 2,3√P
0,01| 0,12| 0,12| 0,23
0,1| 0,38| 0,38| 0,73
1| 1,2| 1,2| 2,3
10| 3,8| 3,8| 7,3
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.| NOTE 1 At 80 MHz and 800 MHz, the separation
distance for the higher frequency range applies.
NOTE 2These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Ameda Platinum Breast Pump is intended for use in the electromagnetic
environment specified below. The customer or the user of the Ameda Platinum
Breast Pump should assure that it is used in such an environment.
Immunity test| IEC 60601 test level| Compliance level|
Electromagnetic environment – guidance
Electrostatic discharge (ESD) IEC 61000-4-2| ± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air| ± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air| Floors should be wood, concrete or
ceramic tile. If floors are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical fast transient/burst IEC 61000-4-4| ± 2 kV for AC line; 100 kHz
repetition frequency| ± 2 kV for AC line; 100 kHz repetition frequency| Mains
power quality should be that of a typical commercial or hospital environment.
Surge IEC 61000- 4-5| ± 0,5 kV,
± 1kV line-to-line; ± 0,5 kV, ± 1kV,
± 2kV line-to-ground| t 0,5 kV,
± 1kV line-to-line; ± 0,5 kV, ± 1kV,
± 2kV line-to-ground| Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power
supply input lines IEC 61000-4-11| • 0% UT; 0,5 cycle At
0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
• 0% UT; 1 cycle and
70% UT; 25/30
cycles Single phase: at 0°
• 0% UT; 250/300 cycles| • 0% UT; 0,5 cycle At 0°, 45°, 90°, 135°, 180°, 225°,
270° and 315°
• 0% UT; 1 cycle and 70% UT; 25 cycles At 0°
• 0% UT; 250 cycles| Mains power quality should be that of a typical
commercial or hospital environment. If the user of the Ameda Platinum Breast
Pump requires continued operation during power mains interruptions, it is
recommended that the Ameda Platinum Breast Pump be powered from an
uninterruptible power supply.
Power frequency (50/60Hz)
magnetic field IEC
61000-4-8| 30 A/m
50 Hz or 60 Hz| 30 A/m
50 Hz and 60 Hz| Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or hospital
environment.
Conducted RF IEC
61000-4-6
Radiated RF IEC
61000-4-3| 3 V
0,15 MHz – 80 MHz 80% AM at 1 kHz
6V in ISM and amateur radio bands between 0,15 MHz and 80 MHz 80% AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz 80% AM at 1 kHz
9 – 28 V/m
At IEC 60601-1-2: 2014
Table 9 frequencies| 3 V
0,15 MHz – 80 MHz 80% AM at 1 kHz
6V in ISM and
amateur radio bands between 0,15 MHz and 80 MHz
80% AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz 80% AM at 1 kHz
9 – 28 V/m
At IEC 60601-1-2: 2014
Table 9 frequencies| Portable and mobile RF communications equipment should be
used no closer to any part of the Ameda Platinum Breast Pump. including
cables, than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance d = 1,24P
d = 1,2W) 80 MHz to 800 MHz
d = 2,34P 800 MHz to 2,7 GHz
Where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m). Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,a should be less than the
compliance level in each frequency range.b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE 1 UT is the a.c. mains voltage prior to application of the test
level.
NOTE 2 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 3 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
a Field strength from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the Ameda Platinum Breast
Pump is used exceeds the applicable RF compliance level above, the Ameda
Platinum Breast Pump should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary,
such as re-orienting or relocating the Ameda Platinum Breast Pump.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 10 V/m.
Caution : Changes or modifications to this device that are not expressly
approved by Ameda may void the user’s authority to use the device.
This device complies with part 15 of the FCC Rules. Operation is subject to
the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interference
that may cause undesired operation.
Note: This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses and
can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by
turning the equipment off
and on, the user is encouraged to try to correct the interference by one or
more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
LIMITED WARRANTY
The Ameda Platinum Electric Breast Pump (“Product”) is warranted to the
original user-whether institutional or individual (“First User”)-only.
From the date of First User’s purchase of this Product, Ameda, Inc. (“Ameda”)
warrants the Product to the First User against defects in material or
workmanship for three years on the pump mechanism. Ameda’s sole obligation
under this limited warranty shall be to repair or replace, at Ameda’s option,
any Product that is determined to be defective by Ameda and determined to be
covered by this express limited warranty. Repair or replacement under this
limited warranty is the sole and exclusive remedy of the First User. Proof of
purchase in the form of a receipted invoice or ill of sale evidencing that the
Product is within the warranty period must be presented to obtain warranty
service.
This limited warranty is extended by Ameda ONLY to the First User and is not
assignable or transferable. For warranty service contact Ameda at 1.866.99
AMEDA(26332).
EXCEPT TO THE EXTENT PROHIBITED BY APPLICABLE LAW, ANY IMPLIED WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE ON THE PRODUCT IS HEREBY
DISCLAIMED.
AMEDA SHALL NOT BE LIABLE FOR ANY INCIDENTAL, CONSEQUENTIAL, SPECIAL OR
PUNITIVE DAMAGES OF ANY KIND FOR BREACH OF THE EXPRESS LIMITED WARRANTY ON THE
PRODUCT OR ANY WARRANTY IMPLIED BY OPERATION OF LAW, OTHER THAN THE EXPRESS
LIMITED WARRANTY SET FORTH ABOVE, THERE ARE NO OTHER WARRANTIES THAT ACCOMPANY
THIS PRODUCT AND ANY ORAL, WRITTEN OR ANY OTHER EXPRESS REPRESENTATION OF ANY
KIND IS HEREBY DISCLAIMED.
Non-Applicability of Warranty
The above warranty shall not apply to consumable materials and attachment to
the Product. These include:
(i) the power cord; and
(ii) the Ameda HygieniKit Milk Collection System (sold separately).
In addition to any other limitations on the warranty discussed above, the
warranty shall expire and be of no force or effect upon the occurrence of any
of the following:
(i) the cause of damage to the Product is due to improper handling or use
(ii) the Product is not properly maintained by the First User, as determined
by Ameda
(iii) non-Ameda approved accessories or spare parts are used with the Product,
as determined by Ameda
(iv) the Product is repaired by a non-approved Ameda service provider.
If any of the above occurs, Ameda shall not be responsible for damage to the
Product, parts thereof or injuries arising therefrom, either directly or
indirectly. The warranty set forth above replaces all prior warranties with
respect to the purchased Product, whether in writing or otherwise.
Outside the United States: Call your local distributor or location where you
purchased the Product. For a listing of distributors in your country, please
visit www.ameda.com.
The Ameda Logo, Ameda, Ameda Mom Inspired. Hospital Trusted. Ameda Platinum,
Elite, SMB, Lact-e and HygieniKit are trademarks and/or registered trademarks
of Ameda, Inc. in the United States and other countries.
Any other trademarks, brand or images appearing herein are property of their
respective owners and are used herein with expressed or implied permission.
Distributed in USA by:
Ameda, Inc.
485 Half Day Road, Ste. 320
Buffalo Grove, IL 60089
1.866.99.AMEDA (26332)| Distributed in Canada by:
Mothers Choice Products Ltd.
2133-11871 Horseshoe Way
Richmond, BC V7A 5H5
1.800.604.6225
---|---
Distribution in other countries: For product information or feedback, call your local distributor or location where you purchased the product. For a listing of distributors in your country, please visit www.ameda.com.
Manufactured for: Ameda, Inc.
485 Half Day Road, Ste. 320
Buffalo Grove, IL 60089
USA
©2022 Ameda, Inc.
1200120-0222