ACCU-SCOPE ACCU Fluor LED Fluorescence Illuminator User Manual

June 13, 2024
ACCU SCOPE

ACCU-SCOPE ACCU Fluor LED Fluorescence Illuminator User Manual

ACCU-SCOPE ACCU Fluor LED Fluorescence Illuminator User
Manual

www.accu-scope.com

Multi-wavelength LED epi-fluorescence illumination utilizes long lifespan LEDs as a light source, converting a traditional upright microscope (with infinity optical design) into an energy-saving and high- efficiency fluorescence microscope. Brightfield observation is preserved with either an open position in the fluorescence channel selector (for 1 and 2-channel illuminators), or it uses the emission characteristics of the UV illumination channel for brightfield observation. Our illuminators are available in 1, 2 or 3 channel configurations with a standard set of fluorescence channel specifications. Please contact your ACCU-SCOPE dealer if you have other requirements – special orders may be available.

SPECIFICATIONS

STANDARD FLUORESCENCE CHANNEL SPECIFICATION

ACCU-SCOPE ACCU Fluor LED Fluorescence Illuminator - STANDARD FLUORESCENCE
CHANNEL SPECIFICATION

Optional filters are available by special order.

EXAMPLES OF COMMON FLUOROPHORES (STAINS, DYES) BY EXCITATION COLOR

ACCU-SCOPE ACCU Fluor LED Fluorescence Illuminator - EXAMPLES OF COMMON
FLUOROPHORES \(STAINS, DYES\) BY EXCITATION COLOR

IN THE BOX

ACCU-SCOPE ACCU Fluor LED Fluorescence Illuminator -
Overview

COMPONENTS

ACCU-SCOPE ACCU Fluor LED Fluorescence Illuminator -
COMPONENTS

INSTALLATION INSTRUCTIONS

These installation instructions apply to all our LED fluorescence illuminators for upright microscopes, regardless of 1-, 2- or 3-channel configurations. Note that your illuminator may vary slightly in appearance than the one used to illustrate the following procedures.

1. Loosen the set screw that secures the head to the frame. Note that some microscopes have a thumb screw and others may use a hex screw.
2. Remove the two thumbscrews from the front of the fluorescence illuminator.

3. Attach UV shield to the front of the fluorescence illuminator using these two thumbscrews.

4. Remove viewing head from the microscope frame and place on the counter/bench top.

5. Place the LED fluorescence illuminator on the top of the microscope frame. Align the bottom dovetail of the illuminator to the dovetail receiver on the frame. If the dovetail does not fully enter the receiver, you may need to loosen the set screw some more.
6. When aligned, tighten the set screw on the microscope frame to secure the illuminator to the frame.

7. With the provided hex screwdriver, loosen the viewing head set screw. If the set screw is not already in place on the viewing head dovetail receiver, use the provided hex wrench to thread the provided set screw into the front of the hole for the viewing head dovetail receiver.

8. Align the viewing head dovetail to the dovetail receiver on the top of the illuminator. If the dovetail does not fully enter the receiver, you may need to loosen the set screw some more.
9. When aligned, tighten the set screw on the illuminator to secure the viewing head to the top of the illuminator.

ACCU-SCOPE ACCU Fluor LED Fluorescence Illuminator - When aligned, tighten
the set screw 2

10. Insert the round plug of the DC power adapter cable into the DC power port on the back of the illuminator. Insert the standard 2-prong plug of the DC power adapter into an AC outlet.

OPERATION INSTRUCTIONS

These operation instructions apply to all of our LED fluorescence illuminators for upright microscopes, regardless of 1-, 2- or 3-channel configurations. Note that your illuminator may vary slightly in appearance and features than the one used to illustrate the following procedures.

Good To Know

  • Main power button turns power on to the unit.

  • The LED display located on the front of the unit shows the current fluorescence channel selector position and LED power in % of maximum. If the LED display is not on, the main power to the unit is off.

  • LED intensity control knob has two functions.
    o Rotate the knob to increase or decrease the LED intensity for that fluorescence channel.
    o Press the knob to turn off the LED. Press again to turn the LED back on.

  • As long as the main power is on, the illuminator will “remember” the LED intensity that is set for each channel. This is ideal for repeatability and qualitative comparisons between samples.

  1. Place a positive control slide (i.e., demonstrated and known to have strong fluorescence staining) on the stage. Focus on the specimen. You may need to close the condenser aperture diaphragm to create enough contrast to see the specimen.
  2. Press the main power button on the left side of the illuminator to turn power on. The button will turn blue to indicate that main power is on.
  3. Turn off the transmitted light on the microscope.
  4. Slide the fluorescence channel selector in or out to the desired position.

NOTE: 1-channel illuminators have 2 positions – one for brightfield observation, and one for fluorescence. 2-channel illuminators have 3 positions – one for brightfield observation, one for one fluorescence channel, and another position for the second fluorescence channel. 3- channel illuminators also have 3 positions – the 3rd channel marked UV/O is used for both UV fluorescence observation of DAPI or similar UV fluorophores, and it is also used for brightfield observation. To use the UV/O position for brightfield, press the LED intensity control knob in to turn off the LED and confirm that the front LED display says “Off”.

  1. While viewing the sample, adjust LED intensity to the desired level by rotating the LED intensity control knob clockwise for more power, or counterclockwise to reduce intensity.

NOTE: The illuminator remembers the LED intensity by channel, until the main power is turned off.
NOTE: Press the intensity knob in to turn off the LED. Press the intensity knob again to turn the LED back on.

  1. Change fluorescence channels and/or samples as desired.

TROUBLESHOOTING

ACCU-SCOPE ACCU Fluor LED Fluorescence Illuminator -
TROUBLESHOOTING

ATTENTION

  1. LED brightness can be freely adjusted as needed and based on the sample being observed. As a general guideline, a setting of 80% maximum intensity is recommended to prolong the service life of the LED lamp.

  2. To reduce reflection and other extraneous light during fluorescence observation, block any fluorescence light from entering the condenser lens with non-reflective black paper or metal sheet.
    You may also close the condenser aperture diaphragm and lower the condenser.

  3. Prolonged illumination of the sample during fluorescence observation can result in fluorescence quenching or bleaching. Turn off the LED when not directly observing the sample. It is also recommended to establish optimal LED power settings on an area of the sample away from the region of interest.

  4. To use brightfield observation with a 3-channel fluorescence illuminator, switch the channel selection lever to the UV/O position and turn off the LED by pressing the LED intensity control knob, then turn on the transmitted light. Turn off the transmitted light and press the LED intensity control knob again to resume fluorescence observation in the UV channel.

  5. To avoid uneven illumination, ensure the microscope and LED fluorescence illuminator are secure and stable on an appropriate bench or work surface.

  6. Use on the provided 12V 2A power adapter that was included with the unit. The manufacture is not responsible for any damage to the illuminator if the incorrect power adapter is used.

LIMITED WARRANTY

This illuminator, the LED lamp and its electronic components are warranted to be free from defects in material and workmanship for a period of one year from the date of invoice to the original (end user purchaser. This warranty does not cover damage caused in-transit, misuse, neglect, abuse or damage resulting from improper servicing or modification by other then ACCU-SCOPE approved service personnel. This warranty does not cover any routine maintenance work or any other work, which is reasonably expected to be performed by the purchaser. Normal wear is excluded from this warranty. No responsibility is assumed for unsatisfactory operating performance due to environmental conditions such as humidity, dust, corrosive chemicals, deposition of oil or other foreign matter, spillage or other conditions beyond the control of ACCU- SCOPE INC. This warranty expressly excludes any liability by ACCU-SCOPE INC. for consequential loss or damage on any grounds, such as (but not limited to) the non-availability to the End User of the product(s) under warranty or the need to repair work processes. Should any defect in material, workmanship or electronic component occur under this warranty contact your ACCU-SCOPE distributor or ACCU-SCOPE at 631-864-1000. This warranty is limited to the continental United States of America. All items returned for warranty repair must be sent freight prepaid and insured to ACCU-SCOPE INC., 73 Mall Drive, Commack, NY 11725 – USA. All warranty repairs will be returned freight prepaid to any destination within the continental United States of America, for all foreign warranty repairs return freight charges are the responsibility of the individual/company who returned the merchandise for repair.