LED Lamp Specifications
last up to 5 hours on a single charge depending on the brightness level used,
and is rated for 10,000 hours.
The battery should come fully charged from the factory. To recharge the
battery, follow the instructions below.
NOTE: The microscope can be used while recharging the battery with the
power cord, and it can still operate with the power cord when the battery can
no longer hold a charge and needs to be replaced.
If the LED lamp or battery needs to be replaced, contact your local authorized
ACCU-SCOPE dealer or call ACCU-SCOPE at
631-864-1000 for a dealer near you.
LED Lamp Assembly CAT# 120-3258-SLED; Battery Pack CAT# 02-1060 — available
only through authorized ACCU-SCOPE dealers.
LED Illumination
(For Corded or Cordless Operation)
Cordless Operation
To use the microscope for cordless operation (battery only), unplug the power
cord from the wall and from the base of the microscope. Turn the power switch
to the on “I” position. Corded Operation
(While Recharging the Battery)
You can recharge the battery while continuing to use the microscope. Turn the
main switch ① on the base of the microscope to the off “O” (middle) position,
connect the power cord to the base of the microscope and plug the cord into a
power supply outlet. Turn the main switch ① to the on “I” position. It takes
about 5-6 hours to fully recharge the battery.
Recharging the Battery
To recharge the battery without using the microscope, turn the main switch ①
on the base of the microscope to the off “O” (middle) position, connect the
power cord to the base of the microscope and plug the cord into a power supply
outlet. Turn the main switch ① to the “recharge” position “II”. It takes about
5-6 hours to fully recharge the battery.
Troubleshooting
PROBLEM | CAUSE | SOLUTION |
---|---|---|
LED lamp does not light | No power supply | Check the power cord or recharge the |
battery
The LED wire is disconnected| Reconnect the LED wire OR call your local
authorized ACCU-SCOPE dealer for service
The LED lamp is damaged| Change the LED lamp assembly — CAT# 120-3258-SLED
OR
call your local authorized ACCU-SCOPE dealer for service
The brightness of the LED lamp is not adequate| The rechargeable battery is
aging| Replace the battery pack — CAT# 02-1060
OR
call your local authorized ACCU-SCOPE dealer for service
The charging circuit is not working| Replace the charging circuit board
OR
call your local authorized ACCU-SCOPE dealer for service