MEAN WELL RPS-120 Series 120W Reliable Green Medical Power Supply Owner’s Manual
- June 13, 2024
- MEAN WELL
Table of Contents
MEAN WELL RPS-120 Series 120W Reliable Green Medical Power Supply Owner’s
Manual
Features
- 4″x2″ compact size
- Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
- Suitable for BF application with appropriate system consideration
- 84W convention, 120W force air
- EMI Class B for both Class I (with FG) & Class II (no FG) configuration
- No load power consumption<0.3W
- Extremely low leakage current
- 12V/0.5A fan supply
- Protections: Short circuit/Overload/ Over voltage/Over temperature
- Operating altitude up to 4000 meters
- 3 years warranty
Applications
- Oral irrigator
- Hemodialysis machine
- Medical monitors
- Sleep apnea devices
- Pumps machine
Description
RPS-120 is a 120W highly reliable green PCB type medical power supply with a high power density on a 4″ by 2″ footprint. It accepts 80-264VAC input and offers various models with the output voltages between 12V and 48V. The working efficiency is up to 91% and the extremely low no load power consumption is down below 0.3W. RPS-120 is able to be used for both Class I (with FG) or Class II (no FG) system design. The extremely low leakage current is less than 150 A. In addition, it conforms to the international medical regulations (2*MOPP) and EMC BS EN/EN55011, perfectly fitting all kinds of BF rated “patient contact” medical system equipment
Model Encoding
SPECIFICATION
MODEL| RPS-120-12 □| RPS-120-15 □| RPS-120-24 □|
RPS-120-27 □| RPS-120-48 □
---|---|---|---|---|---
**OUTPUT| DC VOLTAGE| 12V| 15V| 24V|
27V| 48V
****CURRENT| 10CFM| 10A| 8A| 5A| 4.5A| 2.5A
Convection| 7.0A| 5.6A| 3.5A| 3.15A| 1.75A
RATED POWER| 10CFM| 120W| 120W| 120W| 121.5W| 120W
Convection| 84W| 84W| 84W| 85W| 84W
RIPPLE & NOISE (max.) Note.2| 100mVp-p| 120mVp-p| 150mVp-p|
150mVp-p| 150mVp-p
VOLTAGE ADJ. RANGE| 11.4~12.6V| 14.3~15.8V| 22.8~25.2V| 25.6 ~ 28.4V|
45.6 ~50.4V
VOLTAGE TOLERANCE Note.3| ±2.0%| ±2.0%| ±1.0%| ±1.0%| ±1.0%
LINE REGULATION| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%
LOAD REGULATION| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%
SETUP, RISE TIME| 500ms, 30ms/230VAC 500ms, 30ms/115VAC at full
load
HOLD UP TIME (Typ.)| 50ms/230VAC 10ms/115VAC at full load
INPUT| VOLTAGE RANGE Note.4| 80 ~ 264VAC 113 ~ 370VDC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY (Typ.)| 89%| 89%| 90%| 90%| 91%
AC CURRENT (Typ.)| 2.1A/115VAC 1.2A/230VAC
INRUSH CURRENT (Typ.)| COLD START 30A/115VAC 60A/230VAC
LEAKAGE CURRENT(max.) Note.5| Earth leakage current < 150μA/264VAC ,
Touch current < 80μA/264VAC
PROTECTION| OVERLOAD| 115~150% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is
removed
OVER VOLTAGE| 13.2 ~ 15.6V| 16.5 ~ 19.5V| 26.4 ~ 31.2V| 29.7 ~ 35V| 52.8
~ 62.4V
Protection type : Shut down o/p voltage, re-power on to recover
OVER TEMPERATURE| Protection type : Shut down o/p voltage, re-power on
to recover
FUNCTION| FAN SUPPLY| 12V@0.5A for driving a fan
; tolerance -15% ~ +10% at main output 40% rated current (10CFM)
ENVIRONMENT| WORKING TEMP.| -30 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 50℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes
OPERATING ALTITUDE Note.6| 4000 meters
SAFETY & EMC (Note 7)| SAFETY STANDARDS| IEC60601-1, TUV BS
EN/EN60601-1, EAC TP TC 004, UL ANSI / AAMI ES60601-1 (3.1
version),CAN/CSA-C22.2 No. 60601-1:14 – Edition 3 approved; Design refer
to BS EN/EN60335-1
ISOLATION RESISTANCE| Primary-Secondary: 2xMOPP, Primary-
Earth:1xMOPP, Secondary-Earth:1xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC
ISOLATION RESISTANCE| I/P-O/P, I/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011 (CISPR11)| Class B
Radiated emission| BS EN/EN55011 (CISPR11)| Class B
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–
EMC IMMUNITY| BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| ** BS EN/EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz
)Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 4, 4KV/Line-FG; 2KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| BS EN/EN61000-4-11| 100% dip 1 periods, 30% dip 25
periods,100% interruptions 250 periods
OTHERS| MTBF| 653.5Khrs min. MIL-HDBK-217F (25℃)
DIMENSION (LWH)| PCB:101.650.829mm or 4″ 2″ 1.141″ inch ;
Enclosed type:103.46240mm or 4.07″ 2.44″ 1.57″ inch
PACKING| PCB:0.15Kg; 72pcs/11.8Kg/0.84CUFT ; Enclosed type:0.24Kg;
60pcs/15.4Kg/1.06CUFT
NOTE| 1. All parameters NOT specially mentioned are measured at 230VAC
input, rated load and 25 of ambient temperature.2. Ripple & noise are measured
at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1μf
& 47μf parallel capacitor.3. Tolerance : includes set up tolerance, line
regulation and load regulation.4. Derating may be needed under low input
voltages. Please check the derating curve for more details.5. Touch current
was measured from primary input to DC output.6. The ambient temperature
derating of 3.5℃/1000m with finless models and of 5℃/1000m with fan models for
operating altitude higher than 2000m(6500ft).7. The power supply is
considered a component which will be installed into a final equipment. All the
EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate
with 1mm of thickness. The final equipment must be re-confirmed that it still
meets EMC directives. For guidance on how to perform these EMC tests, please
refer to “EMI testing of component power supplies.”(as available on
http://www.meanwell.com)※ Product Liability Disclaimer For detailed
information, please refer to
https://www.meanwell.com/serviceDisclaimer.aspx
Block Diagram
Derating Curve
Output Derating VS Input Voltage
Mechanical Specification
RPS-120 (PCB Type)
RPS-120-C (Enclosed Type)
AC Input Connector (CN1) : JST B3P-VH or equivalent
Pin No. | Assignment | Mating Housing | Terminal |
---|---|---|---|
1 | AC/N | JST VHRor equivalent | JST SVH-21T-P1.1or equivalent |
2 | No Pin | ||
3 | AC/L |
DC Output Connector (CN2) : JST B4P-VH or equivalent
Pin No. | Assignment | Mating Housing | Terminal |
---|---|---|---|
1,2 | +V | JST VHRor equivalent | JST SVH-21T-P1.1or equivalent |
3,4 | -V |
FAN Connector(CN101) : or equivalent JST S2B-PH-K-S
Pin No. | Assignment | Mating Housing | Terminal |
---|---|---|---|
1 | DC COM(FAN-) | JST PHR-2or equivalent | JST SPH-002T-P0.5Sor equivalent |
2 | +12V(FAN+) |
- HS1,HS2 cannot be shorted.
- HS1 must have safety isolation distance with system case.
Note
- The FAN supply is designed to serve as the source of the additive external fan for the cooling of the power supply, enabling the full load delivery and assuring the best life span of the product. Please do not use this FAN supply to drive other devices.
- The PCB type(Blank type)model delivers EMI Class B for both conducted emission and radiated emission for the power supply, when configured into either Class (with FG) or Class Ⅰ Ⅱ (no FG) system.
- The Enclosed type(-C type) model is not suitable for the configuration within a Class system but is Ⅱ (no FG) suggested to used within a Class (with FG) system.