KERN TKFP-V20-A KFP Platform Scales Instruction Manual
- June 13, 2024
- KERN
Table of Contents
- General hints
- Technical data
- Basic Information (General)
- Basic Safety Precautions
- Transport and storage
- Unpacking, Installation and Commissioning
- Operation limits
- Servicing, maintenance, disposal
- Instant help for troubleshooting
- Preload, Deadload and Overload settings
- Enclosure
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
**KERN TKFP-V20-A KFP Platform Scales
**
General hints
These installation instructions contain all data necessary for placing and
commissioning the following platforms:
KERN TKFP 3V20M-A
KERN TKFP 6V20M-A, TKFP 6V20LM-A
KERN TKFP 15V20M-A, TKFP 15V20LM-A
KERN TKFP 30V20M-A, TKFP 30V20LM-A
KERN TKFP 60V20M-A, TKFP 60V20LM-A
KERN TKFP 150V20M-A, TKFP 150V20LM-A
KERN TKFP 300V20M-A
KERN TKFP 600V20M-A
Technical data
Item no./ Type| Weighing range Max| Readability
d
| Verification value
e
| Minimum load Min| Cable length approx.| Transport Securing
---|---|---|---|---|---|---
kg| g| g| g| m
TKFP 3V20M-A| 3| 0,1| 1| 20| 2,5| yes
TKFP 6V20M-A| 6| 0,2| 1; 2| 20; 40| 2,5| yes
TKFP 6V20LM-A| 6| 0,2| 1; 2| 20; 40| 2,5| yes
TKFP 15V20M-A| 6; 15| 0,5| 2; 5| 40; 100| 2,5| yes
TKFP 15V20LM-A| 6; 15| 0,5| 2; 5| 40; 100| 2,5| yes
TKFP 30V20M-A| 15; 30| 1| 5; 10| 100; 200| 2,5| yes
TKFP 30V20LM-A| 15; 30| 1| 5; 10| 100; 200| 2,5| no
TKFP 60V20M-A| 30; 60| 2| 10; 20| 200; 400| 2,5| no
TKFP 60V20LM-A| 30; 60| 2| 10; 20| 200; 400| 2,5| no
TKFP 150V20M-A| 60; 150| 5| 20; 50| 400; 1000| 2,5| no
TKFP 150V20LM-A| 60; 150| 5| 20; 50| 400; 1000| 2,5| no
TKFP 300V20M-A| 150; 300| 10| 50; 100| 1000; 2000| 2,5| no
TKFP 600V20M-A| 600| 20| 200| 4000| 2,5| no
Technical data of the weighing cell
Sensitivity | 2 mV/V |
---|---|
Input resistance (all TKFP models*) | 409 Ω |
*excepted | TKFP 60V20M-A |
*excepted | TKFP 600V20M-A |
Output resistance (all TKFP models) | 350 Ω |
*excepted | TKFP 600V20M-A |
Power supply (all TKFP models*) | 5 – 12 V AC/DC |
*excepted | TKFP 15V20LM-A TKFP 30V20LM-A |
*excepted | TKFP 600V20M-A |
Accuracy class | C |
Basic Information (General)
Documentation
These installation instructions contain all data necessary for placing and
commissioning the platform KERN TKFP V40-A.
In combination with a display unit, described below as weighing system, for
operation configuration, please refer to the operating instructions of the
display unit.
Proper use
The platform purchased by you is intended to define the weighing value of weighed goods. It is intended to be used as a “non-automatic balance”, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing pan. As soon as a stable weighing value is reached, the weighing value can be read.
Improper Use
- Our balances are non-automatic balances, not provided for the use in dynamic weighing processes. However, the balances can also be used for dynamic weighing processes after verifying their individual operative range, and here especially the accuracy requirements of the application.
- Do not leave permanent load on the weighing pan. This can damage the measuring mechanism.
- Impacts and overloading exceeding the stated maximum load (max) of the weighing system, minus a possibly existing tare load, must be strictly avoided. The weighing system could be damaged.
- Never operate the weighing system in explosive environment. The serial version is not explosion protected.
- Changes to the weighing system’s design are not permitted. This may lead to incorrect weighing results, safety-related faults and destruction of the balance.
- The weighing system unit may only be operated in accordance with the described default settings. Other areas of use must be released by KERN in writing.
Warranty
Warranty claims shall be voided in case:
- Our conditions in the operation manual are ignored
- The appliance is used beyond the described intended use
- Structural changes of the device
- Mechanical damage and damage caused by media, liquids
- Natural wear and tear
- The appliance is improperly set up or incorrectly electrically connected
- Overload of the measuring system
Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the weighing system and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test.
Information is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of weighing system test substances and the test weights required for this. In KERN’s accredited DKD calibration laboratory test weights and weighing systems may be calibrated (return to the national standard) fast and at moderate cost.
Basic Safety Precautions
Pay attention to the instructions in the Operation Manual
Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances.
Personnel training
The appliance may only be operated and maintained by trained staff.
The installation and maintenance must only be carried out by a well acquainted
specialist with the workings of weighing balances.
Transport and storage
Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage.
Packaging / return transport
⇨ Keep all parts of the original packaging for a possibly required return.
⇨ Only use original packaging for returning.
⇨ Prior to dispatch disconnect all cables and remove loose/mobile parts.
⇨ Reattach possibly supplied transport securing devices.
⇨ Secure all parts against shifting and damage.
Unpacking, Installation and Commissioning
Installation Site, Location of Use
The platforms are designed in a way that reliable weighing results are
achieved in common conditions of use.
You will work accurately and fast if you select the right location for your
weighing system.
On the installation site observe the following:
the installation site observe the following:
- Place the weighing system on a firm, level surface. The floor must be able to carry safely the weight of the maximally loaded platform at the resting points. At the same time it should be sufficiently stable, that no vibrations may occur during weighing work. This is to be considered too when the platform is installed in conveyor systems and similar.
- In the installation site possibly no vibrations, e.g. by neighbouring machines should occur.
- Do not use the weighing system in an explosive environment.
- Avoid extreme heat as well as temperature fluctuation e.g. caused by installing next to a radiator or in the direct sunlight.
- Protect the weighing system against direct draughts e.g. due to open windows and doors.
- Use weighing system only in dry environment, protect it against high humidity, vapours and dust.
- The platform must have been stabilized to room temperature, before power supply is switched on.
- Do not expose the appliance to heavy moisture over a longer time period. Nonpermitted condensation (condensation of air humidity on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected appliance for ca. 2 hours at room temperature.
- Avoid jarring during weighing.
- Avoid static charge of weighed goods and weighing container.
- Keep away chemicals (such as liquids or gasses), which could attack and damage the balance inside or from outside.
- Keep IP protection of the device.
- If electromagnetic fields (e.g. by mobile phones or radio equipment), static charges (e.g. when weighing or countings plastic parts), as well as instable current supply occur, the displayed indications may differ considerably (wrong weighing results, damage to the balance). Change location or remove source of interference.
Unpacking, Scope of delivery
Remove the under construction and the weighing plate from the package, remove packaging material and place the devices at the planned workplace. Check if all parts of the scope of delivery are present and free of damage.
Scope of delivery:
- Under Construction (lacquered steel)
- Weighing pan (stainless steel)
- Transport securing (depending on model, see chap. 2)
- Installation Instructions
Placing
- Remove transport securing (see chap. 6.3.1).
- Put the weighing plate on the under construction.
Transport Securing
General hints for transport securing
-
Please consult the Technical Data whether your balance is equipped with a transport securing, chap. 2.
-
There are different types of transport securings in KERN balances. Please see the indications explained in this chapter.
-
Never remove other screws than these for the transport securing as otherwise the balance could be damaged.
-
The sealed screws (1) must not be unscrewed. The sealed screws can be recognized by the sealing varnish (different colours).
-
Do not remove the off-centre load attachments (2) as otherwise the balance could be damaged. The off-centre load attachments are located between the upper and the lower part of the platform. They consist of a screw and a nut.
Variant 1:
Note on transport securing|
---|---
Type of transport securing|
- 1 x hexagon screw on the upper side with nut
- 1 x threaded pin with inner hexagon on the lower side with nut
-
Remove the weighing plate.
-
Turn the scale over so that the bottom is facing you.
-
The position of the transport securings is marked by an arrow.
-
Remove the nuts of the transport securings on the lower side
-
Remove the threaded pin on the lower side.
-
Turn the balance over so that the top is facing you.
-
Remove the hexagon screw on the upper side
Variant 2:
Note on transport securing|
---|---
Type of transport securing|
- 4 x cylinder head screw with inner hexagon on the upper side
- Remove the weighing plate.
- Remove the transport securings.
Levelling
Accurate weighing results require a platform with perfect horizontal alignment. During initial installation and after each change of work area it is necessary to level the platform.
⇨ As the air bubble is located under the weighing plate, remove it.
⇨ Align the platform with the adjustment feet until the air bubble of the
water balance is in the prescribed circle
Connecting a display unit
Before any installation work, disconnect the appliance from the operating
voltage.
Put the connecting cable to the display unit in a manner that it is protected
against damage.
Connect the platform to the display unit according to the following table.
Output load cell | Connection of platform KERN KFP |
---|---|
EXC+ (5V) | see labelling of load cell |
EXC- (0)
SIG-
SIG+
SEN+
SEN-
Operation
Information about
- Mains connection (power is supplied via the connecting cable of the display unit)
- Initial Commissioning
- Connection of peripheral appliances
- Adjustment, linearization and verification (only the complete balance is verifiable, i.e. platform in conjunction with a
suitable display unit) and the correct operation you will find in the operating instructions included in the scope of delivery of the display unit.
Operation limits
The platform is designed so robust that a casual excess of the maximum
weighing load will not cause any damage.
The static carrying capacity, i.e. the maximum admitted load depends on the
type of load carrier (Position A – C). The maximum static carrying capacity
must not be exceeded.
| A = with centrical load
---|---
B = with side load
C = with one-side load
Avoid falling load, shock loads and impacts from the side!
Observe the following operation limits:
Item no./ Type | A | B | C |
---|---|---|---|
TKFP 3V20M-A | 4 kg | 3 kg | 1 kg |
TKFP 6V20M-A | 9 kg | 6 kg | 3 kg |
TKFP 6V20LM-A | 9 kg | 6 kg | 3 kg |
TKFP 15V20M-A | 22 kg | 15 kg | 7 kg |
TKFP 15V20LM-A | 22 kg | 15 kg | 7 kg |
TKFP 30V20M-A | 45 kg | 30 kg | 15 kg |
TKFP 30V20LM-A | 45 kg | 30 kg | 15 kg |
TKFP 60V20M-A | 90 kg | 60 kg | 30 kg |
TKFP 60V20LM-A | 90 kg | 60 kg | 30 kg |
TKFP 150V20M-A | 225 kg | 150 kg | 75 kg |
TKFP 150V20LM-A | 225 kg | 150 kg | 75 kg |
TKFP 300V20M-A | 450 kg | 300 kg | 150 kg |
TKFP 600V20M-A | 900 kg | 600 kg | 300 kg |
Servicing, maintenance, disposal
Before any maintenance, cleaning and repair work disconnect the appliance from the operating voltage.
Cleaning
⇨ Clean the platform with a soft cloth soaked with a mild cleaning agent.
⇨ Remove the weighing plate and wipe-off the dirt and the foreign bodies under
it.
Do not use hard or sharp objects for this.
⇨ Remove regularly corrosive substances.
⇨ Keep IP protection.
Servicing, maintenance
⇨ The appliance may only be opened by trained service technicians who are
authorized by KERN.
⇨ Ensure that the weighing system is regularly calibrated, see chap. 3.5
Testing instruments monitoring.
Disposal
⇨ Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.
Instant help for troubleshooting
In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The weighing process must then be restarted from the beginning
Help:
Fault | Possible cause |
---|---|
The displayed weight is permanently changing |
- Draught/air movement
- Floor vibrations • The platform has contact with foreign objects.
- Electromagnetic fields / static charging (choose different location/switch off interfering device if possible)
The weighing result is obviously wrong|
- No zero display with unloaded balance
- Adjustment is no longer correct.
- Great fluctuations in temperature. Platform on an uneven surface.
- Electromagnetic fields / static charging (choose different location/switch off interfering device if possible)
Should other error messages occur, switch balance off and then on again. If the error message remains inform manufacturer.
Preload, Deadload and Overload settings
Item no./ Type
| Headload ** (kg)
**= already applied preload
| ****
Centric overload protection (kg)
| ****
Off-centre overload protection (kg)
| ****
Minimum capacity of the load cell E min (kg)
| ****
Maximum capacity of the load cell E max (kg)
|
---|---|---|---|---|---|---
TKFP 3V20M-A| 2,2| –| 3,6| 0| 5|
TKFP 6V20M-A| 2,2| –| 7,2| 0| 10|
TKFP 6V20LM-A| 3,5| –| 7,2| 0| 10|
TKFP 15V20M-A| 3,5| –| 18| 0| 20|
TKFP 15V20LM-A| 2,7| –| 18| 0| 20|
TKFP 30V20M-A| 2,7| –| 36| 0| 50|
TKFP 30V20LM-A| 6,5| –| 36| 0| 50|
TKFP 60V20M-A| 2,7| –| 72| 0| 100|
TKFP 60V20LM-A| 6,5| –| 72| 0| 100|
TKFP 150V20M-A| 6,5| –| 180| 0| 200|
TKFP 150V20LM-A| 9| –| 180| 0| 200|
TKFP 300V20M-A| 9| –| 360| 0| 500|
TKFP 600V20M-A| 19| –| 660| 0| 750|
Item no./ Type| Dimensions of the platform (mm)| Load cell
type| Test certificteno.| Accur acy class| Ratio of min.
load cell verification interval| Maximum admitted number of interval
sizes| Allowable temperature range| Ratio of min. dead load output
return| Error percentage
Y| n LC /n max| T min| T max| Z|
PLC
| | ( °C)| ( °C)| |
TKFP 3V20M-A| 230 x 230 x 105| L6D 5KG| TC7868 rev. 1| C| 10000| 3000| -10|
40| 3000| 0,7
TKFP 6V20M-A| 230 x 230 x 105| L6D 10KG| TC7868 rev. 1| C| 10000| 3000| -10|
40| 3000| 0,7
TKFP 6V20LM-A| 300 x 240 x 105| L6D 10KG| TC7868 rev. 1| C| 10000| 3000| -10|
40| 3000| 0,7
TKFP 15V20M-A| 300 x 240 x 105| L6D 20KG| TC7868 rev. 1| C| 10000| 3000| -10|
40| 3000| 0,7
TKFP 15V20LM-A| 400 x 300 x 125| L6N 20KG| D09-06.11 rev. 1| C| 10000| 3000|
-10| 40| 3000| 0,7
TKFP 30V20M-A| 400 x 300 x 125| L6E 50KG| TC7838 rev. 1| C| 10000| 3000| -10|
40| 4200| 0,7
TKFP 30V20LM-A| 500 x 400 x 120| L6G 50KG| D09-03.22 rev. 2| C| 10000| 3000|
-10| 40| 4200| 0,7
TKFP 60V20M-A| 400 x 300 x 125| L6E 100KG| TC7838 rev. 1| C| 10000| 3000| -10|
40| 4200| 0,7
TKFP 60V20LM-A| 500 x 400 x 130| L6G 100KG| D09-03.22 rev. 2| C| 12000| 3000|
-10| 40| 3000| 0,7
TKFP 150V20M-A| 500 x 400 x 130| L6G 200KG| D09-03.22 rev. 2| C| 12000| 3000|
-10| 40| 3000| 0,7
TKFP 150V20LM-A| 650 x 500 x 135| L6G 200KG| D09-03.22 rev. 2| C| 12000| 3000|
-10| 40| 3000| 0,7
TKFP 300V20M-A| 650 x 500 x 135| L6G 200KG| D09-03.22 rev. 2| C| 12000| 3000|
-10| 40| 3000| 0,7
TKFP 600V20M-A| 800 x 600 x 190| PW12C 750KG| TC11749 rev. 0| C| 25000| 6000|
-10| 40| 8000| 0,7
Enclosure
Dimensions
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>