KERN FKB-BA-e-2111 High-Resolution Bench Scale Instruction Manual
- June 3, 2024
- KERN
Table of Contents
FKB-BA-e-2111 High-Resolution Bench Scale
KERN & Sohn GmbH
Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com
Phone: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern- sohn.com
Operating manual Table scales
KERN FKB
Version 1.1 2021-07 GB
FKB-BA-e-2111
KERN FKB
GB
Version 1.1 2021-07
Operating manual
Table scales
Content
1 Technical data……………………………………………………………………………………………..5 2 Declaration of
conformity ………………………………………………………………………………8 3 Appliance overview
………………………………………………………………………………………9
3.1 Components ………………………………………………………………………………………….9 3.2 Operating elements
………………………………………………………………………………10
3.2.1 Keyboard overview …………………………………………………………………………10 3.2.2 Numeric
entry…………………………………………………………………………………11 3.2.3 Display overview
…………………………………………………………………………….11 4 Basic Information (General)
…………………………………………………………………………12 4.1 Proper
use…………………………………………………………………………………………..12 4.2 Improper
Use……………………………………………………………………………………….12 4.3
Warranty……………………………………………………………………………………………..12 4.4 Monitoring of Test
Resources…………………………………………………………………13 5 Basic Safety Precautions
…………………………………………………………………………….13 5.1 Pay attention to the instructions in the
Operation Manual …………………………..13 5.2 Personnel
training…………………………………………………………………………………13 6 Transport and
storage…………………………………………………………………………………13 6.1 Testing upon acceptance
………………………………………………………………………13 6.2 Packaging / return transport
…………………………………………………………………..13 7 Unpacking, Setup and Commissioning
………………………………………………………….14 7.1 Installation Site, Location of Use
…………………………………………………………….14 7.2 Unpacking and checking
……………………………………………………………………….14 7.3 Assembling, Installation and
Levelling……………………………………………………..15 7.4 Mains
connection………………………………………………………………………………….15 7.5 Battery operation (optional)
……………………………………………………………………15 7.6 Rechargeable battery operation (optional)
……………………………………………….16 7.6.1 Recharge battery ……………………………………………………………………………16
7.7 Connection of peripheral devices ……………………………………………………………17
2
FKB-BA-e-2111
7.8 Initial Commissioning…………………………………………………………………………….17 7.9 Adjustment
………………………………………………………………………………………….17
7.9.1 External adjustment
FKB-BA-e-2111
3
14.4.3 Automatic data output < auto > ………………………………………………….39 14.4.4 Continuous data output < cont > ………………………………………………..39 14.5 Data format ………………………………………………………………………………………40 15 Servicing, maintenance, disposal ………………………………………………………………41 15.1 Cleaning …………………………………………………………………………………………..41 15.2 Servicing, maintenance ………………………………………………………………………41 15.3 Disposal …………………………………………………………………………………………..41 16 Instant help …………………………………………………………………………………………….42
4
FKB-BA-e-2111
1 Technical data
KERN
Item no./ Type
Readability (d)
Weighing range (max)
Taring range (subtractive)
Reproducibility
Linearity
Stabilization time (typical)
Smallest part weight for piece counting – under lab conditions*
Smallest part weight for piece counting – under normal conditions**
Adjustment points
Recommended adjustment weight (not supplied)
Warm-up time
Weighing Units
Humidity of air
Allowable ambient temperature
Input voltage Appliance
Input voltage Mains adapter
Batteries (option)
FKB 6K0.02
TFKB 6K-5-A 0.02 g 6,000 g 6,000 g 0.02 g ± 0.06 g 3 sec.
20 mg
200 mg
1 / 3 / 5 / 6 kg
5 kg; 1 kg (F1)
FKB 8K0.1
TFKB 8K-4-A 0.1 g
8,000 g 8,000 g
0.1 g ± 0.3 g 2 sec.
FKB 8K0.05
TFKB 8K-5-A 0.05 g 8,000 g 8,000 g 0.05g ± 0.15g 3 sec.
100 mg
50 mg
1 g
500 m
2 / 5 / 7 / 8 kg
2/4/5/7/8 kg
5 kg; 2 kg; 1 kg 5 kg; 2 kg; 1 kg
(F1)
(F1)
2 hrs. kg, g, gn, dwt, ozt, lb, oz max. 80% rel. (non-condensing)
-10 °C … + 40 °C
9 V, 1 A 110V 240V AC; 50Hz/60Hz
6 x 1.5V AA
FKB 15K0.5
TFKB 15K-4-A 0.5 g
15,000 g 15,000 g
0.5 g ± 0.15 g 3 sec.
1 g
10 g
5/10/15 kg
15 kg (F2)
Storage battery operation (optional)
Operating period 90 h (background illumination OFF) Operating period 40 h
(background illumination ON)
Loading time approx. 10 hrs.
Auto-Off (battery, rechargeable battery)
3 min
Auto off (net))
Selectable 30s, 1, 2, 5, 30, 60 min
Dimensions housing
350 x 390 x 120 (W x D x H) [mm]
Weighing pan stainless steel mm
340 x 240
Net weight (kg)
7
7
7
6
Interfaces Underfloor weighing device
· RS-232 (DB9 female), as per series · USB-appliance connection (USB B),
Factory option · Ethernet, Factory option · WLAN, Factory option
yes (hook supplied)
FKB-BA-e-2111
5
KERN
Item no./ Type Readability (d) Weighing range (max) Taring range (subtractive)
Reproducibility Linearity Stabilization time (typical) Smallest part weight
for piece counting – under lab conditions* Smallest part weight for piece
counting – under normal conditions** Adjustment points Recommended adjustment
weight (not supplied) Warm-up time Weighing Units Humidity of air Allowable
ambient temperature Input voltage Appliance Input voltage Mains adapter
Batteries (option)
Storage battery operation (optional)
Auto-Off (battery, rechargeable battery) Auto off (net)) Dimensions caisse (l
x L x h) [mm] Weighing pan stainless steel mm Net weight (kg)
Interfaces
Underfloor weighing device
FKB 16K0.1
TFKB 16K-4-A 0.1 g
16,000 g 16,000 g
0.1 g ± 0.3 g 3 sec.
FKB 16K0.05
TFKB 16K-5-A 0.05 g 8,000 g 8,000 g 0.05g ± 0.15 g 3 sec.
FKB 30K1
TFKB 30K-3-A 1 g
30,000 g 30,000 g
1 g ± 2 g 2 sec.
FKB 36K0.1
TFKB 36K-4-A 0.0001 kg 36 kg 36 kg 0.0001 kg ± 0.0003 kg 3 sec.
100 mg
50 mg
2 g
100 mg
1 g
500 mg
20 g
1 g
5/10/15/16 kg 5/10/15/16 kg
10/20/30 kg 10/15/30/36 kg
10 kg; 5 kg; 1 kg (F1)
10 kg; 5 kg; 1 kg (F1)
30 kg (F2)
20 kg + 10 kg (E2)
4 hrs.
2 hrs.
2 hrs.
kg, g, gn, dwt, ozt, lb, oz
max. 80% rel. (non-condensing)
2 hrs.
-10 °C … + 40 °C
9 V, 1 A 110V 240V AC; 50Hz/60Hz
6 x 1.5V AA
Operating period 90 h (background illumination OFF) Operating period 40 h
(background illumination ON)
Loading time approx. 10 hrs.
3 min
Selectable 30s, 1, 2, 5, 30, 60 min
350 x 390 x 120
340 x 240
7
7
6
7
· RS-232 (DB9 female), as per series · USB-appliance connection (USB B), Factory option · Ethernet, Factory option · WLAN, Factory option
yes (hook supplied)
6
FKB-BA-e-2111
KERN
Item no./ Type Readability (d) Weighing range (max) Taring range (subtractive)
Reproducibility Linearity Stabilization time (typical) Smallest part weight
for piece counting – under lab conditions* Smallest part weight for piece
counting – under normal conditions** Adjustment points Recommended adjusting
weight F1 (not supplied) Warm-up time Weighing Units Humidity of air Allowable
ambient temperature Input voltage Appliance Input voltage Mains adapter
Batteries (option)
Storage battery operation (optional)
Auto-Off (battery, rechargeable battery) Auto off (net)) Dimensions caisse (l
x L x h) [mm] Weighing pan stainless steel mm Net weight (kg)
Interfaces
Underfloor weighing device
FKB 36K0.2
TFKB 36K-4B-A 0.0002 kg 36 kg 36 kg 0.0002 kg ± 0.0006 kg
200 mg
FKB 65K1
TFKB 65K-3-A 0.001 kg 65 kg 65 kg 0.001 kg ± 0.003 kg 3 sec.
2 g
FKB 65K0.2
TFKB 65K-4-A 0.0002 kg 65 kg 65 kg 0.0002 kg ± 0.0006 kg
200 mg
2 g
20 g
2 g
10 / 20 / 30 / 36 kg
20/40/60 kg
15 / 30 / 50 / 60 kg
20 kg; 10 kg (F1)
60 kg (F2)
50 kg; 10 kg (E2)
2 hrs. kg, g, gn, dwt, ozt, lb, oz max. 80% rel. (non-condensing)
-10 °C … + 40 °C
9 V, 1 A
100 V – 240V AC 50/60Hz 0.3A
6 x 1.5V AA
Operating period 90 h (background illumination OFF) Operating period 40 h
(background illumination ON)
Loading time approx. 10 hrs.
3 min
Selectable 30s, 1, 2, 5, 30, 60 min
350 x 390 x 120
340 x 240
7
6
7
· RS-232 (DB9 female), as per series · USB-appliance connection (USB B), Factory option · Ethernet, Factory option · WLAN, Factory option
yes (hook supplied)
FKB-BA-e-2111
7
- Smallest component weight for part counting – under lab conditions: There are ideal ambient conditions for high-resolution counting The parts to be counted have no variation
** Smallest component part for part counting under normal conditions: There are unsteady ambient conditions (draft, vibrations) The parts to be counted are subject to variation
2 Declaration of conformity
The current EC/EU Conformity declaration can be found online in:
www.kern-sohn.com/ce
For verified weighing scales (= weighing scales assessed for conformity) a declaration of conformity is included in the scope of delivery.
8
FKB-BA-e-2111
3 Appliance overview
3.1 Components
Pos.
1 2 3 4 5 6 7 8 9 10
Designation
Weighing pan Display Keyboard Levelling screw Mains adapter connection Bubble
level Anti-theft protection device connection USB-interface (Factory option)
RS 232 interface Ethernet (Factory option)
FKB-BA-e-2111
9
3.2 Operating elements
3.2.1 Keyboard overview Button Name
ON/OFF button
TARE-key
Function in Operating mode
Function in Menu
Switch on/off (press button long time)
Switch on/off the display background illumination (press button long time)
Menu level back
Exit menu / back to weighing mode.
Taring
PRE-TARE (press button long time)
Invoke application menu (press button long time)
Navigation key
Select menu item
ZERO key
Zeroing
F-key
Navigation key Select menu item
Switch over between
Navigation key
-key
weight display and
Activate menu item
piece quantity display Confirm selection
PRINT button
Calculate weighing data via interface
10
FKB-BA-e-2111
3.2.2 Numeric entry Button Designation Navigation key
Navigation key
Function
Select cipher Confirm entry. Press button repeatedly for every digit. Wait
until the numeric input window extinguishes.
Reduce flashing cipher (0 9)
Navigation key Increase flashing cipher (0 9)
3.2.3 Display overview
1
2
3
5
4 6
Position
1 2 3
4
5
Display
0<
NET
HI OK LO
Units display / Pcs/ %
6
–
G
–
Description Stability display Zero indicator Minus display Net weight value
display
Tolerance marks for check weighing
selectable g, kg, lb, gn, dwt, oz,ozt or
Application icon [Pcs] for piece counting or [%] for determination of
percentage Rechargeable battery charge indicator Optional reference piece
number enabled Weighing data can be found in the sum
memory
FKB-BA-e-2111
11
4 Basic Information (General)
4.1 Proper use The balance you purchased is intended to determine the weighing
value of material to be weighed. It is intended to be used as a “non-automatic
balance”, i.e. the material to be weighed is manually and carefully placed in
the centre of the weighing pan. As soon as a stable weighing value is reached,
the weighing value can be read.
4.2 Improper Use Do not use balance for dynamic add-on weighing procedures, if
small amounts of goods to be weighed are removed or added. The “stability
compensation” installed in the balance may result in displaying an incorrect
measuring value! (Example: Slowly draining fluids from a container on the
balance.) Do not leave permanent load on the weighing pan. This may damage the
measuring system. Impacts and overloading exceeding the stated maximum load
(max) of the balance, minus a possibly existing tare load, must be strictly
avoided. Balance may be damage by this. Never operate balance in explosive
environment. The serial version is not explosion protected. The structure of
the balance may not be modified. This may lead to incorrect weighing results,
safety-related faults and destruction of the balance. The balance may only be
used according to the described conditions. Other areas of use must be
released by KERN in writing.
4.3 Warranty Warranty claims shall be voided in case
· Our conditions in the operation manual are ignored
· The appliance is used outside the described uses
· The appliance is modified or opened
· Mechanical damage and damage by media, liquids, natural wear and tear
· The appliance is improperly set up or incorrectly electrically connected
· The measuring system is overloaded
12
FKB-BA-e-2111
4.4 Monitoring of Test Resources In the framework of quality assurance the
measuring-related properties of the balance and, if applicable, the testing
weight, must be checked regularly. The responsible user must define a suitable
interval as well as type and scope of this test. Information is available on
KERN’s home page (www.kern-sohn.com) with regard to the monitoring of balance
test substances and the test weights required for this. In KERN’s accredited
DKD calibration laboratory test weights and balances may be calibrated (return
to the national standard) fast and at moderate cost.
5 Basic Safety Precautions
5.1 Pay attention to the instructions in the Operation Manual
Carefully read this operation manual before setup and
commissioning, even if you are already familiar with KERN balances.
All language versions contain a non-binding translation.
The original German is binding.
5.2 Personnel training The appliance may only be operated and maintained by
trained personnel.
6 Transport and storage
6.1 Testing upon acceptance When receiving the appliance, please check
packaging immediately, and the appliance itself when unpacking for possible
visible damage.
6.2 Packaging / return transport
Keep all parts of the original packaging for a possibly required return. Only
use original packaging for returning. Prior to dispatch disconnect all cables
and remove loose/mobile parts. Reattach possibly supplied transport securing
devices. Secure all parts such as wind screen, weighing pan, power unit etc.
against shifting and damage.
FKB-BA-e-2111
13
7 Unpacking, Setup and Commissioning
7.1 Installation Site, Location of Use The balances are designed in a way that
reliable weighing results are achieved in common conditions of use. You will
work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the installation site:
· Place the balance on a firm, level surface. · Avoid extreme heat as well as
temperature fluctuation caused by installing next
to a radiator or in the direct sunlight. · Protect the balance against direct
draughts due to open windows and doors. · Avoid jarring during weighing; ·
Protect the balance against high humidity, vapours and dust. · Do not expose
the device to extreme dampness for longer periods of time. Non-
permitted condensation (condensation of air humidity on the appliance) may
occur if a cold appliance is taken to a considerably warmer environment. In
this case, acclimatize the disconnected appliance for ca. 2 hours at room
temperature. · Avoid static charge of goods to be weighed or weighing
container. If electro-magnetic fields or static charge occur, or if the power
supply is unstable major deviations on the display (incorrect weighing
results) are possible. In that case, the location must be changed.
7.2 Unpacking and checking Remove device and accessories carefully from
packaging, remove packaging material and install the device at the planned
work place. Check if that there has been no damage and that all packing items
are present.
Scope of delivery / serial accessories:
· Balance, see chap. 3.1 · Mains adapter · Operating manual · Protective cover
· Flush-mounted hook
14
FKB-BA-e-2111
7.3 Assembling, Installation and Levelling
Remove the four transport securing devices above the supports of the weighing
pan
Install weighing pan and wind shield if necessary. Ensure that the balance is
installed in a level position. Level balance with foot screws until the air
bubble of the water balance is in the
prescribed circle.
Check levelling regularly
7.4 Mains connection
Select a country-specific power plug and insert it in the mains adapter.
Check, whether the voltage acceptance on the scales is set correctly. Do not
connect the scales to the power mains unless the information on the scales
(sticker) matches the local mains voltage. Only use KERN original mains
adapter. Using other makes requires consent by KERN.
Important: Before starting your weighing balance, check the mains cable for
damage. Ensure that the power unit does not come into contact with liquids.
Ensure access to mains plug at all times.
7.5 Battery operation (optional) When the batteries are exhausted, in the
display will appear < instab>.
Rotate the balance carefully in a way that the bottom of the balance is freely
accessible.
Open the battery compartment and exchange the batteries.
Ensure correct polarisation.
Close again the lid.
· To save the battery, in menu (see chap. 13.3.1.) the automatic switch-off
function < Autoff> can be activated.
· If the balance is not used for a longer time, take out the battery and store
it separately. Leaking battery liquid could damage the balance.
FKB-BA-e-2111
15
7.6 Rechargeable battery operation (optional)
ATTENTION
The rechargeable battery and the battery match with each
other. Only use the delivered mains adapter.
Do not use the balance during the loading process. The rechargeable can only
be replaced by the same or by a
type recommended by the manufacturer.
The rechargeable battery is not protected against all environmental
influences. If the rechargeable battery is exposed to certain environmental
influences, it may set on fire or explode. Persons may be injured or material
damage may occur.
Protect the rechargeable battery against fire and heat. Do not bring the
rechargeable battery in contact with fluids, chemical
substances or salt.
Do not expose the rechargeable battery to high pressure or
microwaves.
Under no circumstances the rechargeable batteries and the charging
unit may be modified or manipulated.
Do not use a defective, damaged or deformed rechargeable battery.
Do not connect or short-circuit the electrical contacts of the
rechargeable battery with metallic objects.
Liquid may squirt out from a damaged rechargeable battery.
If the liquid gets into contact with the skin or the eyes, the skin and the
eyes may be irritated.
Ensure the correct polarity when inserting or changing the rechargeable
battery (see instructions in the battery compartment)
The rechargeable battery operation is overridden when the mains
adapter is connected. For weighing in mains operation > 48 hrs., the
rechargeable batteries must be removed! (Danger of overheating).
If the rechargeable battery starts to smell, being hot, changing
the colour or being deformed, it must be immediately unplugged from mains
supply and from the balance if possible.
7.6.1 Recharge battery The rechargeable battery package is charged via the
delivered power cable.
Before the first use, the rechargeable battery package should be charged by
connecting it to the mains power cable for at least 15 hours.
To save the rechargeable battery, in menu (see chap.0.) the automatic switch-
off function < Autoff> can be activated.
If the capacity of the rechargeable battery is exhausted, in the display
16
FKB-BA-e-2111
7.7 Connection of peripheral devices
Before connecting or disconnecting of additional devices (printer, PC) to the
data interface, always disconnect the balance from the power supply.
With your balance, only use accessories and peripheral devices by KERN, as
they are ideally tuned to your balance.
7.8 Initial Commissioning
In order to obtain exact results with the electronic balances, your balance
must have reached the operating temperature (see warming up time chap. 1).
During this warming up time the balance must be connected to the power supply
(mains, accumulator or battery).
The accuracy of the balance depends on the local acceleration of gravity.
Strictly observe hints in chapter Adjustment.
7.9 Adjustment
As the acceleration value due to gravity is not the same at every location on
earth, each display unit with connected weighing pan must be coordinated – in
compliance with the underlying physical weighing principle – to the existing
acceleration due to gravity at its place of location (only if the weighing
system has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature.
To receive accurate measuring values it is also recommended to adjust the
display unit periodically in weighing operation.
· Arrange the required adjustment weight, see chap. 1. Carry out adjustment as
near as possible to the highest load of the balance (recommended adjustment
weight see chap. 1). Weights of different nominal values or tolerance classes
may be used for adjustment but are not optimal for technical measuring. The
accuracy of the adjustment weight must correspond approximately to or, if
possible, be better than, the readability [d] of the balance. Info about test
weights can be found on the Internet at: http://www.kern-sohn.com
· Observe stable environmental conditions. A warming up time (see chapter 1)
is required for stabilization.
· Ensure that there are no objects on the weighing pan.
7.9.1 External adjustment
Observe stable environmental conditions. A warm up time (see chapter 1) is
required for stabilization.
Ensure that there are no objects on the weighing pan. To invoke the setup menu
press the TARE and ON/OFF button at the same time
and keep them pressed until the first menu item < Cal> will be displayed.
Navigation key Press button,
FKB-BA-e-2111
17
Navigation key Press button, the first selectable adjustment weight will be
displayed.
Use the navigation keys to select the desired adjustment weight, see chap. 1
,,Adjustment points” or ,,Recommended adjustment weight”
Prepare the required adjustment weight.
Acknowledge selection by button. < Zero >, < Pt ld > followed by the
weight value of the adjustment weight to be placed will be displayed.
Place the adjustment weight and confirm with button,
18
FKB-BA-e-2111
8 Basic Operation
8.1 Turn on/off Start-up:
Press the ON/OFF button.
The display lights up and the balance carries out a selftest. Wait until the
weight display appears, then the balance is ready for weighing.
Switch off:
Keep ON/OFF button pressed until the display disappears
8.2 Simple weighing
Check zero display [>0<] and set to zero with the help of the ZEROkey, as
required.
Place goods to be weighed on balance. Wait until the stability display appears
( ). Read weighing result.
Overload warning Overloading exceeding the stated maximum load (max) of the
device, minus a possibly existing tare load, must be strictly avoided. This
could damage the instrument. Exceeding the maximum load is indicated by
. Unload balance or reduce preload.
8.3 Weighing with tare
8.3.1 Taring The dead weight of any weighing container may be tared away by
pressing a button, so that the following weighing procedures show the net
weight of the goods to be weighed.
Put the weighing container on the weighing pan Wait until the stability
display appears ( ), then press TARE key.
The weight of the container is now internally saved. The zero display and the
indicator ,,NET” will appear. ,,NET” informs that all shown weight values are
net values.
Weigh the material. Wait until the stability display appears ( ). Read net
weight.
FKB-BA-e-2111
19
· When the balance is unloaded the saved taring value is displayed with
negative sign.
· To delete the stored tare value, remove load from weighing pan and press the
TARE button.
· The taring process can be repeated any number of times, e.g. when adding
several components for a mixture (adding). The limit is reached when the
taring range capacity is full.
· Numeric input of the tare weight (PRE-TARE), see chap. 10.1.2.2
8.4 Underfloor weighing Objects unsuitable for placing on the weighing scale
due to size or shape may be weighed with the help of the flush-mounted
platform. Proceed as follows:
Switch off the balance. Open closing cover (1) at the balance bottom. Place
weighing balance over an opening. Completely screw-in the hook Hook-on the
material to be weighed and carry out weighing.
CAUTION
· Always ensure that all suspended objects are stable enough to hold the
desired goods to be weighed safely (danger of breaking).
· Never suspend loads that exceed the stated maximum load (max) (danger of
breaking)
Always ensure that there are no persons, animals or objects that might be
damaged underneath the load.
NOTICE
After completing the underfloor weighing the opening on the bottom of the
balance must always be closed (dust protection).
20
FKB-BA-e-2111
9 Operating concept
From factory the balance is delivered with various applications (normal
weighing, check weighing, counting). After the first start-up the balance is
in the
In the application menu (see chap.13.2.) however, you can define, selecting an
application, in which mode the balance after switching-on has to continue
working. Either as per standard in weighing mode or e.g. as check balance or
counting balance
Note: The number of the installed applications depends on the model.
Selecting an application:
Press the TARE button and keep it pressed until the first menu item will be
displayed
Use the TARE button to select the menu setting < mode > and acknowledge with
button. The current setting will be displayed.
Use the TARE button to select the desired mode, selectable
weig h Weighing mode
count
Counting mode
check Check mode
Confirm with button.
According to the selected application in the application menu just appear the
application-specific settings, so that you reach the target quickly without
detour.
· Information about the application-specific settings you will find in the
description of the respective application.
· All basic settings and parameters, which influence the whole operation of
the balance, are resumed in the Setup Menu (see chap.13.3.). These settings
remain valid for all applications.
Change application:
Press and hold TARE button until the first setup menu item is displayed
Use the TARE button to select the menu setting < mode > and acknowledge with
button. The current setting will be displayed.
Use the TARE button to select the desired mode and acknowledge
with button.
FKB-BA-e-2111
21
10 Application
How to carry out a simple weighing and taring, please refer to chap. 8.2 or
8.3. Further specific settings you will find in the following chapters.
Shouldn’t the application
10.1 Application-specific settings + Navigation in the application menu see chap. 13.1
10.1.1 Overview
Press the TARE button and keep it pressed until the first menu item <
pretare > will be displayed
Level 1 Ptare PRE-TARE
hold Unit Units
Mode Applications see chap. 9.
Level 2
actuAl
manuAl CLEAR
–
g kg lb gn dwt ozt oz % Free factor
multiplication factor
Pcs weigh count check
Level 3 Description / Chapter Take over the placed weight as PRE-TARE value,
see chap. 10.1.2.1 Numeric input of the tare weight, see chap. 10.1.2.2.
Delete PRE-TARE value Start-Hold function This function defines in which
weighing unit the result will be displayed. In verified models not all the
weighing units are available, see chap. 1.
Application
22
FKB-BA-e-2111
10.1.2 Description of individual functions
10.1.2.1 Take over the placed weight as PRE-TARE value, < Ptare actuel >
Deposit weighing container Invoke menu setting < Ptare > and confirm by
button. To take over the placed weight as a PRE-TARE value, use the navigation
keys
to select < actuel >
Confirm with button. < wait > is displayed. The weight of the weighing
container is stored as tare weight. Remove the weighing container, the
indicator (NET) and the tare weight with
minus sign will appear.
Place the filled weighing container. Wait until the stability display appears
( ). Read net weight.
The entered tare weight remains invalid until a new tare weight is input. To
delete press the TARE key or confirm the menu setting < clear > using the
button.
FKB-BA-e-2111
23
10.1.2.2 Enter the known tare weight numerically < Ptare manuel >
Invoke menu setting < Ptare > and confirm by button. Use the navigation keys
to select the setting < manuel > and confirm with
button.
Enter the known tare weight, numeric input see chap. 3.2.2 The input weight is
automatically saved as tare weight, the indicator (NET) and the
tare weight with minus sign will appear.
Place the filled weighing container. Wait until the stability display appears
( ). Read net weight.
The entered tare weight remains invalid until a new tare weight is input. To
delete enter the zero value or confirm the menu setting
10.1.2.3 Data-Hold function < hold >
Invoke menu setting < hold > Place goods to be weighed on balance. Confirm
with button. The first stable weighing value is kept for 15 s, symbolised by
[HOLD] in the upper
edge of the display.
24
FKB-BA-e-2111
11 Application
Shouldn’t the application
11.1 Application-specific settings + Navigation in menu see chap. 13.1
11.1.1 Overview
Press the TARE button and keep it pressed until the first menu item < ref >
will
be displayed
Level 1 Ref Reference quantity
Ptare
Level 2
5 10 20 50 free input
actuAl
Level 3
Description / Chapter
Reference piece number 5 Reference piece number 10 Reference piece number 20 Reference piece number 50 Optional, numeric input, see chap. 3.2.2. Input unit weight
Take over the placed weight as PRE-TARE value, see chap. 10.1.2.1
manuAl CLEAR
Numerical input of the tare weight, see chap. 10.1.2.2.
Delete PRE-TARE value
FKB-BA-e-2111
25
11.2 Using the application
11.2.1 Parts counting
Before the balance can count parts, it must know the average piece weight
(i.e. reference). Proceed by putting on a certain number of the parts to be
counted. The balance determines the total weight and divides it by the number
of parts, the so-called reference quantity. Counting is then carried out on
the basis of the calculated average piece weight.
· The higher the reference quantity the higher the counting exactness.
· Especially high reference must be selected for small parts or parts with
considerably different sizes.
· Smallest counting weight see table ,,Technical data”
Procedure:
1. Calculate reference
Reference piece quantity 5, 10, 20 or 50:
If necessary, put on and tare the weighing container. Put on the desired
quantity of reference pieces. Press the TARE button and keep it pressed until
the first menu item < ref > will be
displayed
Confirm with button. Use the navigation keys to select the reference piece
quantity (5, 10, 20, 50)
according to the placed reference and confirm with the button.
The balance will calculate the average item weight and then displays the
quantity of
pieces.
Remove reference weight. The balance is now in parts counting mode counting
all
units on the weighing pan.
Reference piece quantity user-defined:
If necessary, put on and tare the weighing container. Put on the desired
quantity of reference pieces. Press the TARE button and keep it pressed until
the first menu item < ref > will be
displayed
Confirm with button. Use the navigation keys to select the reference piece
quantity < free > and
confirm with button.
The numeric input window appears. Enter and confirm the quantity of the placed
reference parts, numeric input see
chap. 3.2.2
26
FKB-BA-e-2111
The balance will calculate the average item weight and then displays the
quantity of
pieces.
Remove reference weight. The balance is now in parts counting mode counting
all
units on the weighing pan.
Numerical input of the reference piece count::
In Iweighing mode, press and hold the TARE key until the first menu item < ref
is displayed.
Confirm with the key.. Select unit and confirm with key Select the input of the piece weight with the navigation keys and confirm
with -key.
Enter the desired reference sample quantity and confirm with the TARE key. The reference sample quantity is now set
2. Parts counting
Put the empty container on the weighing pan and press the TARE button. The
container is tared, the zero display will appear.
Fill the counting quantity. The piece quantity is shown in the display.
With the – key you can switch between number of items and weight display
FKB-BA-e-2111
27
12 Application
Shouldn’t the application
12.1 Application-specific settings + Navigation in menu see chap. 13.1
12.1.1 Overview
Press the TARE button and keep it pressed until the first menu item < limits
will be displayed
Level 1 limitS
Level 2
limlow
Level 3
Description / Chapter
Lower limit value, numeric input see chap. 3.2.2
Ptare PRE-TARE
limupp actuel manuel CLEAR
Upper limit value, numeric input see chap. 3.2.2 Take over the placed weight as PRE-TARE value, see chap. 10.1.2.1 Numeric input of the tare weight, see chap. 10.1.2.2. Delete PRE-TARE value
12.2 Using the application
With the application
When limit values are exceeded below or above, an optical and acoustic signal
will be displayed (if enabled in menu) Optical signal:
The tolerance marks provide the following information:
Target quantity exceeds defined tolerance
Target quantity within defined tolerance
Target quantity below defined tolerance
28
FKB-BA-e-2111
Audio signal: The acoustic signal depends on the menu setting < setup beeper
, see chap. 13.3.1.
Procedure: 1. Define upper and lower limit value:
Make sure that the balance is in Checkweighing mode. If necessary, switch over
with the Select weighing unit for checkweighing mode
Press the TARE button and keep it pressed until the first menu item < limit >
will be displayed.
Confirm with button. < limlow > will be displayed. Press button to confirm,
the numeric input window for entering the lower limit
value will appear. Enter the lower limit value for the target piece quantity
(numeric input see chap. 3.2.2) and confirm with button. The balance returns
to the < limlow > menu.
Use the navigation keys to select the setting < limupp > and confirm with
button.
The numeric input window for entering the upper limit value < will appear.
Enter the
upper limit value for the target piece quantity (numeric input see chap.
3.2.2) and confirm with button. < limupp > will appear.
Press repeatedly button to exit menu.
Finished the setting works, the weighing balance is ready now for check
weighing.
2. Start tolerance check:
Place the weighed material and check by means of the tolerance marks /
acoustic
signal if the weighed material is within the defined tolerance.
Load below specified tolerance
Load within specified tolerance
Load exceeds specified tolerance
FKB-BA-e-2111
29
13 Menu
13.1 Navigation in the menu Call up menu:
Application menu
Setup menu
Press the TARE button and keep it pressed until the first menu item will be displayed
Press the TARE and ON/OFF button at the same time and keep it pressed until the first menu item will be displayed
Select and adjust parameter:
Scrolling on one level
Use the navigation buttons to select the individual menu blocks one by one.
Use the navigation key to scroll down.
Use the navigation key to scroll up.
Activate menu item / Confirm selection
Press navigation key
Menu level back / back to weighing mode
Press navigation key
13.2 Application menu The application menu allows you a fast and targeted
access to the respectively selected application (see chap. 9.).
· An overview of the application-specific settings you will find in the
description of the respective application.
30
FKB-BA-e-2111
13.3 Setup menu In the setup menu you have the possibility to adapt the
behaviour of the balance to your requirements (e.g. environmental conditions,
especial weighing processes). These settings are global and do not depend on
the selected application.
13.3.1 Overview < setup >
Level 1 Cal Adjustment
com Communication
Level 2
calext caleud graadj grause
Rs232
Usb-d
WLAN
Level 3 Level 4 / Description
Description
For external adjustment, see chap. 0
External adjustment, user-defined, see chap. 7.9.2
Constant of gravity adjustment place numeric input see chap. 3.2.2
Constant of gravity installation place, numeric input see chap. 3.2.2
baud
data parity stop handsh protoc on
300 600 1200 2400 4800 9600 14400 19200 38400 57600 115200 128000 256000
7dbits 8dbits None Odd
even 1bit 2bits none kcp
off
FKB-BA-e-2111
31
Print Data output
BEEPER Audio signal
intfce sum prmode
keys Check
Rs232 USB-d on off manual auto
cont format Layout
off on ok
low
high
RS 232 interface
USB device interface
Switch on / off add-up mode, see chap. 14.4.1
On, off
Data output by pressing the PRINT button (see chapter 14.4.2)
On, off
Automatic data output with stable and positive weighing value see chap.14.4.3.
Another output only after zero display and stabilisation, depending on the
settings < zRange >, selectable (off, 1, 2, 3,4,5)
On, off
Continuous data output depending on the set time interval < speed >, see chap.
14.4.4
short long
Standard measuring protocol Detailed measuring protocol
Not documented
Switch on / off acoustic signal by pressing button
off Slow Std Fast Cont. off Slow Std Fast Cont. off Slow Std Fast Cont.
Acoustic signal off Slow Standard Fast Continuous Acoustic signal off Slow Standard Fast Continuous Acoustic signal off Slow Standard Fast Continuous
32
FKB-BA-e-2111
Autoff Automatic
switch-off function
mode
off Auto
Automatic switch-off function switched off
The balance is automatically switchedoff according to the time without load
change or without operation defined in menu item < Time >
Only0
buttons Function button allocation
Blight Display background illumination
Time
30 s 1 min
2 min
5 min
30 min 60 min
Not documented
mode
always
timer
Automatic switch-off only with zero display After the set time without load
change or operation the balance will switch off automatically
Background lighting of display is switched on permanently The background
illumination is automatically switched-off according to the time without load
change or without operation defined in menu item < Time >
No bl
Display background illumination always switched off
tarerg Taring range
Time
100%
10%
5 s 10 s 30 s 1 min 2 min 5 min 30 min
The background illumination of the display will switch off automatically after the set time without load change or operation
Defining the maximum taring range, you can choose 10100%. To enter the numerical value
FKB-BA-e-2111
33
ztrack Maintaining zero
reset
on
Automatic maintaining zero [< 3 d]
off
If the amount of the weighed material is reduced or increased significantly,
the scale’s “stabilizing and compensating” mechanism can result in displaying
erroneous weighing results! (e.g.: slow outflow of the liquid from the
container placed on the scale, evaporating processes).
When dosing with small weight fluctuations, it is recommended to switch this
function off.
Reset balance settings to factory settings
34
FKB-BA-e-2111
14 Interfaces Via the interfaces weighing data may be exchanged with connected
peripheral devices. Issue may be made to a printer, PC or control displays. In
the same way, control commands and data inputs may occur via the connected
devices (such as PC, keyboard, barcode reader).
The available interfaces can be used in a parallel manner.
14.1 Interface cable (RS232)
Connection
KERN Standard setting
Sub-D bushing 9 pol. (bushing = to balance)
Pin 1: VB Pin 2: TXD (RS232) Pin 3: RXD (RS232) Pin 4: VCC Pin 5: Signal
ground (RS232) Pin 6: Low Signal (signal light “IN4”) Pin 7: Hi Signal (signal
light “IN2”) Pin 8: OK Signal (signal light “IN1″) Pin 9: No assignment
· 8 Data bit
· 1 Stop bit
· No parity
FKB-BA-e-2111
35
14.2 Connect printer
Turn off scale and printer. Use a suitable cable to connect the weighing
balance to the interface of the
printer. Faultless operation requires an adequate KERN interface cable
(optional).
Turn on scale and printer.
Communication parameters (baud rate, bits and parity) of balance and printer
must match; see menu item < com Rs232. >. (chap. 13.3.1)
Printout examples KERN YKB-01N
N:
S S
2.998 kg Net weight (stable weighing value)
T:
0.3000 kg Tare weight
G
3.2999 kg Gross weight
N:
S D
2.998 kg Net weight (instable weighing value)
T:
0.3000 kg Tare weight
G
3.2999 kg Gross weight
N: T: G: PCS: UW: REF:
S D
3.2998 kg Net weight (instable weighing value)
0.0000 kg Tare weight
3.2998 kg Gross weight
33 pcs Quantity placed on balance
0.1000 kg Average piece weight
10 pcs Reference quantity
14.3 KCP-interface commands
A detailed description you will find in the ,,KERN Communications Protocol”
manual, available in the download area on our KERN homepage.
36
FKB-BA-e-2111
14.4 Issue functions
14.4.1 Add-up mode < sum > With this function the individual weighing values
are added into the summation memory by pressing a button and edited, when an
optional printer is connected. Activate function:
In Setup menu invoke the menu setting < Print sum> and confirm with
button .
Use the navigation keys to select the setting < on > and confirm with button.
To exit the menu press the navigation key repeatedly
Condition: Menu setting prmode manual on >
Add-up weighed goods:
If required, place empty container on scale and tare. Place first good to be
weighed on balance. Wait until stability display ( )
appears and then press the PRINT button. The display changes to < sum1 >,
followed by the current weighing value. The weighing value is stored and
edited by the printer. The symbol pops up. Remove the weighed good.
Place second good to be weighed on balance. Wait until stability display ( )
appears and then press the PRINT button. The display changes to < sum2 >,
followed by the current weighing value. The weighing value is stored and
edited by the printer. Remove the weighed good.
Add-up more weighed goods as described above. You can repeat this process
until the capacity of the scales is exhausted.
Display and edit sum ,,Total”:
Press the PRINT key long time. The number of weighings and the total weight
are
edited. The sum memory is deleted; the symbol [..] extinguishes.
FKB-BA-e-2111
37
Sample log (KERN YKB-01N): Menu setting prmode format short
First weighing
Second weighing
Third weighing
Number of weighings/ Total sum
14.4.2 Data output after pressing the PRINT button < manual > Activate
function:
In Setup menu invoke the menu setting
button .
For a manual data output select the menu setting < manual > with the
navigation
keys and confirm with the button.
Use the navigation keys to select the setting < on > and confirm with button.
To exit the menu press the navigation key repeatedly.
Place goods to be weighed on balance:
If required, place empty container on scale and tare. Place goods to be
weighed on balance. The weighing value is edited by pressing the
PRINT button.
38
FKB-BA-e-2111
14.4.3 Automatic data output < auto >
Data output happens automatically without having to press the PRINT-key as
soon as the corresponding output condition has been met, dependent on the
setting in the menu.
Enable function and set the output condition:
In Setup menu invoke the menu setting < Print PrMode > and
confirm with button .
For an automatic data output select the menu setting < auto > with the
navigation keys and confirm with the button.
Use the navigation keys to select the setting < on > and confirm with button.
Place goods to be weighed on balance:
If required, place empty container on scale and tare. Place weighed goods and
wait until the stability display (
The weighing value is issued automatically.
) appears.
14.4.4 Continuous data output < cont >
Enable function and set the output interval:
In Setup menu invoke the menu setting < Print PrMode > and confirm
with button.
For a continuous data output select the menu setting < cont > with the
navigation keys and confirm with the button.
Use the navigation keys to select the setting < on > and confirm with button.
Place goods to be weighed on balance.
If required, place empty container on scale and tare. Place goods to be
weighed on balance. The weighing values are issued according to the defined
interval.
FKB-BA-e-2111
39
Sample log (KERN YKB-01N):
14.5 Data format
In Setup menu invoke the menu setting < Print PrMode > and confirm
with button .
Use the navigation keys to select the menu setting < format > and confirm
with button.
Use the navigation buttons to select the desired setting.
Options:
Sample log (KERN YKB-01N): format short
format long
40
FKB-BA-e-2111
15 Servicing, maintenance, disposal
Before any maintenance, cleaning and repair work disconnect the appliance from
the operating voltage.
15.1 Cleaning Please do not use aggressive cleaning agents (solvents or
similar agents), but a cloth dampened with mild soap suds. Ensure that no
liquid penetrates into the device. Polish with a dry soft cloth. Loose residue
sample/powder can be removed carefully with a brush or manual vacuum cleaner.
Spilled weighing goods must be removed immediately.
15.2 Servicing, maintenance
The appliance may only be opened by trained service technicians who are
authorized by KERN.
Before opening, disconnect from power supply.
15.3 Disposal Disposal of packaging and appliance must be carried out by
operator according to valid national or regional law of the location where the
appliance is used.
FKB-BA-e-2111
41
16 Instant help
In case of an error in the program process, briefly turn off the balance and
disconnect from power supply. The weighing process must then be restarted from
the beginning.
Fault
Possible cause
The displayed weight does not glow.
· The balance is not switched on.
· The mains supply connection has been interrupted (mains cable not plugged
in/faulty).
· Power supply interrupted.
The displayed weight is permanently changing
· Draught/air movement
· Table/floor vibrations
· Weighing pan has contact with other objects. · Electromagnetic fields /
static charging (choose
different location/switch off interfering device if possible)
The weighing result is obviously incorrect
· The display of the balance is not at zero
· Adjustment is no longer correct.
· The balance is on an uneven surface.
· Great fluctuations in temperature.
· Warm-up time was ignored.
· Electromagnetic fields / static charging (choose different location/switch
off interfering device if possible)
42