FKG XP-endo Shaper Sequence Instruction Manual
- June 12, 2024
- FKG
Table of Contents
FKG XP-endo Shaper Sequence Instruction Manual
Golden rules
Speed : 800-1000 rpm
Torque : 1 Ncm
- Prior to using the XP-endo® Shaper, establish glide path to at least 15/.02.
- In multirooted teeth, begin with the largest canal.
- Never force the instrument.
- Irrigate copiously throughout the instrumentation protocol.
Description – Beschreibung
Taper
- and ISO diameter
- identification|
Depth marks (in millimetres)
Available on 21 / 25 / 31 mm instruments
Example on 25 / 31 mm instruments
Symbols
STERILE /R
Sterilized product, ready to use
Always check expiry date
Single use (max. 4 canals)
Do not use if package is damaged
Protocol for use
- Prior to using the XP-endo® Shaper (XP-S), establish glide path to at least 15/.02. In calcified/constricted canals or complex curvatures, a glide path to 10/.04 is recommended.
- In multirooted teeth, begin with the largest canal. The canal should always contain irrigant.
- Insert the tip of the XP-S into the canal until resistance (Fig. 1), retract (tip loose) and start the motor.
- Use long gentle strokes to progress down to working length (WL) (Fig. 2). If WL is not reached in 5 strokes, stop, irrigate, recapitulate and proceed again. Do not use pecking motion ! Never force the instrument and always keep it spinning and moving while in the canal.
- Once WL is reached, irrigate and work the instrument for 15 additional long gentle strokes to WL (Fig. 3).
- Irrigate the canal in order to eliminate suspended debris.
- Choose a gutta percha point that is tight and seats at the correct WL. If 30/.04 is desired and does not fit, work the XP-S for an additional 5-10 strokes. Once all the canals have been shaped, proceed with the cleaning of the canals. For an optimal result, the use of XP-endo® Finisher is recommended. Once the cleaning of the canals is completed, proceed to the next phase of treatment :
- Obturate with gutta percha and sealer. TotalFill® BC Points™ and TotalFill® BC Sealer™ are recommended.
General information
Indications for use :
These endodontic instruments are intended for use in medical or hospital
facilities, by qualified health professionals.
XP-endo® Shaper is used for endodontic treatments.
Contraindications :
This product contains nickel, and must not be used on individuals with a known
allergic sensitivity to this metal.
Adverse reactions : None known
Warnings and precautions :
- Single use instruments must not be retreated for subsequent use. That could compromise their integrity.
- Always inspect the instrument(s) before use, and discard (it) them if there are any visible defect(s).
- When an instrument reaches the end of its life, please dispose of it in accordance with the applicable laws and regulations.
FKG Dentaire Sàrl
Le Crêt-du-Locle 4, 2322 Le Crêt-du-Locle, Switzerland
Tel. +41 32 924 22 44, info@fkg.ch,
www.fkg.ch