FKG ScoutRace and ISO 10 Endodontic Instruments Instruction Manual
Description
Taper
Depth marks (in millimetres)
Available on 21 / 25 / 31 mm instruments
Example on 25 / 31 mm instruments
Safety Memo Disc (SMD)
Pull off petals to master metal fatigue:
1 x Simple case / cas simple / einfach Fall
Rubber stop – ISO length/active part
21/16 mm
25/16 mm
31/16 mm
Speed: 600-800 rpm
Torque: 1.5 Ncm
- Long back and forth gentle strokes.
- Light touch, let the instrument work.
- Work 3-4 seconds in a row, get out, clean the blade and irrigate the canal.
Scout Race 1 blister includes
For canals with a severe apical
curvature or S-shape
Manual scouting with ISO 06-08 SSt files up to estimated working length (WL);
confirm WL with an apex locator
- Introduce Scout Race (SR) 10/.02 in rotation and reach WL;
- Continue with SR 15/.02 up to WL;
- Finish glide path with SR 20/.02 up to WL;
- Final shaping with the main NiTi sequence (e.g. iRace or BioRace).
Case with an apical S-shape curvature
-
Pre-op
-
Difficulty
-
Post-op : 35/.04
Race ISO 10 1 blister includes
For narrow and/or calcified canals
Manual scouting with ISO 06-08 SSt files up to estimated working length (WL); confirm WL with an apex locator
- Introduce Race 10/.02 in rotation and reach WL;
- Continue with Race 10/.04 up to WL;
- Finish glide path with Race 10/.06 up to WL;
- Final shaping with the main NiTi sequence (e.g. iRace or BioRace)
Case with constricted and obliterated canals
-
Pre-op
-
Difficulty
-
Post-op
Non sterile products First use
Carefully follow the rules herebelow
Cleaning and Sterilization
Sterile products First use
- Open the alveolus
- Extract the instrument (use of gloves imperative)
- Use the instrument according to good dental practice
Second use:
Carefully follow the rules herebelow
Second and following uses.
Cleaning
-
Manual cleaning with/without
Ultrasonic assistance Thorough cleaning allows efficient disinfection & sterilization. -
Rinse
Under distilled/demineralized water at least 1 min. Then dry the instruments. -
Inspection – Check
Sort out damaged or worn out instruments. -
Packing
Place the instruments in a suitable support/container and pack the devices in sterilisation pouches ISO 11607-1. -
According EN/ISO 17664 protocol
Autoclave : 134°C/273°F, 2.2 bar during 18 min. -
Storage
Keep devices in sterilization wrap/pouch in a dry and clean environment.
Always refer to manufacturer’s instructions for cycle & duration
Second and following uses -
Pre-disinfection
Plunge instruments after use in detergent solution and eventually brush them manually. -
Rinse
Under running water at least 1 min. Then dry the instruments. -
Follow steps 1 – 6 above
Complete operating protocol and warnings on www.fkg.ch
General information
Pre-/Disinfection
DO NOT USE solutions containing:
- phenol (corrosion)
- aldehyde (blood fixation)
- di-/triethanolamines (corrosion)
Sterilization with chemiclav or dry air devices have not been validated
by FKG Dentaire.
Refer to manufacturer’s instructions for cycle & duration. Nevertheless our
instruments withstand such methods.
Caution:
This product contains nickel and should not be used for individuals with known
allergic sensitivity to this metal.
200°C (392°F)
A single patient use is recommended to avoid cross-contamination.