FKG ScoutRace and ISO 10 Endodontic Instruments Instruction Manual

June 12, 2024
FKG

FKG ScoutRace and ISO 10 Endodontic Instruments Instruction Manual
FKG ScoutRace and ISO 10 Endodontic Instruments

Description

Taper
Description

Depth marks (in millimetres)
Available on 21 / 25 / 31 mm instruments

Example on 25 / 31 mm instruments
Description

Safety Memo Disc (SMD)

Pull off petals to master metal fatigue:

1 x Simple case / cas simple / einfach Fall

Rubber stop – ISO length/active part

21/16 mm
25/16 mm
31/16 mm

Speed: 600-800 rpm
Torque: 1.5 Ncm

  • Long back and forth gentle strokes.
  • Light touch, let the instrument work.
  • Work 3-4 seconds in a row, get out, clean the blade and irrigate the canal.

Scout Race 1 blister includes

For canals with a severe apical

curvature or S-shape
Manual scouting with ISO 06-08 SSt files up to estimated working length (WL); confirm WL with an apex locator

  1. Introduce Scout Race (SR) 10/.02 in rotation and reach WL;
  2. Continue with SR 15/.02 up to WL;
  3. Finish glide path with SR 20/.02 up to WL;
  4. Final shaping with the main NiTi sequence (e.g. iRace or BioRace).

Case with an apical S-shape curvature

  1. Pre-op
    curvature or S-shape

  2. Difficulty
    curvature or S-shape

  3. Post-op : 35/.04
    curvature or S-shape

Race ISO 10 1 blister includes

For narrow and/or calcified canals

Manual scouting with ISO 06-08 SSt files up to estimated working length (WL); confirm WL with an apex locator

  1. Introduce Race 10/.02 in rotation and reach  WL;
  2. Continue with Race 10/.04 up to WL;
  3. Finish glide path with Race 10/.06 up to WL;
  4. Final shaping with the main NiTi sequence (e.g. iRace or BioRace)

Case with constricted and obliterated canals

  1. Pre-op
    calcified canals

  2. Difficulty
    calcified canals

  3. Post-op

Non sterile products First use
Carefully follow the rules herebelow
Cleaning and Sterilization

Sterile products First use

  1. Open the alveolus
  2. Extract the instrument (use of gloves imperative)
  3. Use the instrument according to good dental practice

Second use:
Carefully follow the rules herebelow
Second and following uses.

Cleaning

  1. Manual cleaning with/without
    Ultrasonic assistance Thorough cleaning allows efficient disinfection & sterilization.

  2. Rinse
    Under distilled/demineralized water at least 1 min. Then dry the instruments.

  3. Inspection – Check
    Sort out damaged or worn out instruments.

  4. Packing
    Place the instruments in a suitable support/container and pack the devices in sterilisation pouches ISO 11607-1.

  5. According EN/ISO 17664 protocol
    Autoclave : 134°C/273°F, 2.2 bar during 18 min.

  6. Storage
    Keep devices in sterilization wrap/pouch in a dry and clean environment.
    Always refer to manufacturer’s instructions for cycle & duration
    Second and following uses

  7. Pre-disinfection
    Plunge instruments after use in detergent solution and eventually brush them manually.

  8. Rinse
    Under running water at least 1 min. Then dry the instruments.

  9. Follow steps 1 – 6 above
    Complete operating protocol and warnings on www.fkg.ch

General information

Pre-/Disinfection
DO NOT USE solutions containing:

  • phenol (corrosion)
  • aldehyde (blood fixation)
  • di-/triethanolamines (corrosion)

Sterilization with chemiclav or dry air devices have not been validated by FKG Dentaire.
Refer to manufacturer’s instructions for cycle & duration. Nevertheless our instruments withstand such methods.

Caution:
This product contains nickel and should not be used for individuals with known allergic sensitivity to this metal.

200°C (392°F)

A single patient use is recommended to avoid cross-contamination.