FKG Dentaire D-Race Sequence Endodontic Instrument Instruction Manual

June 12, 2024
FKG Dentaire

Instructions for use
see www.fkg.ch for additional information
D-Race
Endodontic instruments
D-Race
Description – Beschreibung
FKG REF. 99.7AA.10.13A.XX – n°102 – 2021/12

D-Race Sequence Endodontic Instrument

DR1
Access
Size : ISO 30
Taper : 0.10 / 10%
Length : 15 / 8 mm
Tip  : active
Shank  : 12 mm
Speed  : 1000 rpm
Recomended torque  : 1.5 Ncm

DR2

Up to working length

Size  : ISO 25
Taper : 0.04 / 4%
Length  : 25 / 16 mm
Safety tip :  12 mm
Speed: 600 rpm
Torque : 1 Ncm

DR2 is for single use only
Depth marks▲ (in millimeters)
Example on instruments of 25 / 31 mmFKG Dentaire D-Race Sequence
Endodontic Instrument - instruments

Depth marks only on DR2
Golden rules

  • Let the instrument work.
  • Work 3-4 seconds in a row, get out, clean the blade and irrigate the canal.

D-Race 1 blister includes / contient / enthält : 2 x DR1 + DR2
« Desobturation » prior to retreatment, because access must be made, the entrance of the canal « desobturated » to reach inside the canal itself so that it could be properly retreated. Retreatment must be achieved because the first RCT failed or needs improvement.
D-Race will desobturate most canals, whatever the filling used – gutta-percha, carriers, paste, resin-based materials. D-Race removes a large part of the ancient filling material. With only 2 instruments time is saved and safety of treatment is increased.
Once desobturated, the canal will be retreated.

**General advice to take into consideration for removal of filling

material**

  1. Take 2 x-rays from different angles.
  2. A precise appreciation of canal anatomy is very important.
  3. Find entrance of canal and access the obturation material.
  4. Prepare, if needed, a cavity 1-2 mm with DR1 and drop solvent to soften obturation material. If necessary, use a heat plugger or ultrasonics.

Coronal third preparation

  1. Gently engage DR1 rotating at 1000 rpm (recom. torque 1.5 Ncm) into the obturation material.
    The active tip of DR1 facilitates the initial penetration.

Median and Apical thirds preparation

  1. Insert DR2 rotating at 600 rpm (1 Ncm).
  2. Don’t force the way, clean and check the blade regularly.
  3. Continue to advance apically as long as filling material is visible on the flute.
  4. If needed, drop solvent to help removal of obturation material.
  5. Establish final WL and finalise shaping with standard NiTi Race files.

Sterilization protocol for reusable medical devices
Non sterile products
First use
Carefully follow the rules herebeloCleaning and Sterilization
Sterile products
First use

  1. Open the alveolus
  2. Extract the instrument (use of gloves imperative)
  3. Use the instrument according to good dental practice
    Second use : Carefully follow the rules herebelow Second and following uses.

Cleaning

  1. Manual cleaning with/without Ultrasonic assistance
    Thorough cleaning allows efficient disinfection & sterilization.

  2. Rinse
    Under distilled/demineralized water at least 1 min.
    Then dry the instruments.

  3. Inspection – Check
    Sort out damaged or worn out instruments.

  4. Packing
    Place the instruments in a suitable support/container and pack the devices in sterilisation pouches ISO 11607-1.
    Sterilization

  5. According EN/ISO 17664 protocol Autoclave : 134°C/273°F, 2.2 bar during 18 min.

  6.  Storage
    Keep devices in sterilization wrap/pouch in a dry and clean environment.
    ** Always refer to manufacturer’s instructions for cycle & duration
    Second and following uses**

  7. Pre-disinfection
    Plunge instruments after use in detergent solution and eventually brush them manually.

  8. Rinse
    Under running water at least 1 min.
    Then dry the instruments.

  9. Follow steps 1 – 6 above
    **** Complete operating protocol and warnings on www.fkg.ch

General information

Pre-/Disinfection
DO NOT USE
solutions containing :
– phenol (corrosion)
– aldehyde (blood fixation)
– di-/triethanolamines (corrosion)
Caution : This product contains nickel and should not be used for individuals with known allergic sensitivity to this metal.
Sterilization with chemiclav or dry air devices have not been validated by FKG Dentaire.
Refer to manufacturer’s instructions for cycle & duration. Nevertheless our instruments withstand such methods.

A single patient use is recommended to avoid cross-contamination.

1639
FKG Dentaire Sàrl
Le Crêt-du-Locle 4, 2322 Le Crêt-du-Locle, Switzerland
Tel. +41 32 924 22 44,
[email protected],
www.fkg.ch

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

FKG Dentaire User Manuals

Related Manuals