FKG XP-endo Finisher Endodontic Instruments Instruction Manual
- June 12, 2024
- FKG
Table of Contents
Instructions for use
XP-endo® Finisher
seewww.fkg.ch for additional information
Endodontic instruments
XP-endo Finisher Endodontic Instruments
XP-endo® Finisher
Description – Beschreibung
25/.00 Taper (1) and ISO diameter (2) identification
Durchmessers (2) Depth marks (in millimetres)
Available on 21 / 25 / 31 mm instruments
Example on 25 / 31 mm instruments
Symbols
Sterilized product, ready to use
Always check expiry date
Single use (max. 4 canals)
Do not use if package is damaged
Golden rules
Speed : 800-1000 rpm
Torque : 1 Ncm
- XP-endo® Finisher should be used only after canal preparation to at least ISO 25.
- In multirooted teeth, begin with the largest canal.
- Work along the entire length of the canal for approximately 1 minute.
XP-endo® Finisher
General information
Indications for use :
These endodontic instruments are intended for use in medical or hospital
facilities, by qualified health professionals. The XP-endo ® Finisher is used
as a final cleaning instrument in root canal treatments.
Contraindications :
This product contains nickel, and must not be used on individuals with a known
allergic sensitivity to this metal.
Adverse reactions : None known
Warnings and precautions :
- Single use instruments must not be retreated for subsequent use. That could compromise their integrity.
- Always inspect the instrument(s) before use, and discard (it) them if there are any visible defect(s).
- When an instrument reaches the end of its life, please dispose of it in accordance with the applicable laws and regulations.
Protocol for use
Speed : 800-1000 rpm
Torque : 1 Ncm
- XP-endo®Finisher (XP-F) must only be used only after preparing the canal to at least ISO 25.
- In multirooted teeth, begin with the largest canal. The canal should always contain irrigant.
- Fix the canal working length by using the plastic tube (a) to adjust the endo stop (b) (Fig. 1).
- Place the XP-F into the canal by lining up the tip facing the cusp, insert slightly into the canal (Fig. 2). Then start the motor and slowly thread the XP-F into the canal.
- Use the XP-F for 1 minute (approximately 60 strokes), using slow and gentle 7-8 mm longitudinal movements to contact the full length of the canal (Fig. 3). Make small multidirectional movements and avoid removing the file from the canal.
- Irrigate the canal to remove the suspended debris. A syringe tip 31G is recommended (or no bigger diameter than 30G). Once the cleaning of the canals is completed, proceed to the next phase of treatment :
- Obturate with gutta percha and sealer. TotalFill® BC Points™ and TotalFill ® BC Sealer™ are recommended.