FKG RACE EVO Sequences Taper User Manual

June 12, 2024
FKG

FKG RACE EVO Sequences Taper

Description

Taper (1) and ISO diameter (2) identification

Depth marks ▲ (in millimetres)
Available on 21 / 25 / 31 mm instruments

Example on 25 / 31 mm instruments

Endo stop – ISO length / active part

Sequence Glide   path Shaping
RACE EVO 4% RE1
15/.04 RE2
25/.04 RE3

30/.04
RACE EVO 6%| RE1
15/.04| RE2
25/.04| RE3
25/.06

General information

Parameters for use :
Recommended speed range : 800-1000 rpm continuous rotation.

Torque : 1.5 Ncm.

Indications for use :
RACE® EVO endodontic instruments are intended for use in medical or hospital facilities, by qualified health professionals.

RACE® EVO instruments are designed for the glide path and shaping of root canals.

Contraindications :
Nickel-Titanium instruments must not be used on individuals with a known allergic sensitivity to these metals.

Adverse reactions : None known.

Notice and precautions :

  • Take several radiographs from different angles to determine the anatomy of the root canals (length, width and curvature).
  • The number of canals to be treated with a RACE® EVO instrument is limited to a maximum of 8 root canals.
  • Irrigate the canal thoroughly and frequently after the RACE® EVO instrument is extracted.
  • Regularly inspect the RACE® EVO instrument during use, and discard if it shows any signs of wear (for example : straightening).
  • If the RACE® EVO instrument is not progressing easily, remove it from the canal, clean it, inspect its cutting edges, then irrigate the canal and recapitulate with an ISO 010 hand file.
  • Always inspect the instrument(s) before use, and discard (it) them if there are any visible defect(s).
  • When an instrument reaches the end of its life, please dispose of it in accordance with the applicable laws and regulations.
  • Visit the FKG Dentaire SA website (www.fkg.ch) for all information relating to the applicable cleaning and sterilization methods.

Symbols

| Sterilized product (irradiation)
---|---
| Sterilizable in a steam sterilizer (autoclave) at the specified temperature
| To be used in continuous rotation at the recommended speed
| Warning : refer to the user manual
| Batch number
| Expiry date
| Do not use if package is damaged
| Nickel-Titanium instrument
| Reference number
| Manufacturer

Customer Support

FKG Dentaire Sàrl
Le Crêt-du-Locle 4, 2322 Le Crêt-du-Locle, Switzerland
Tel. +41 32 924 22 44, info@fkg.ch, www.fkg.ch

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals