FKG RACE EVO Sequences Taper User Manual
- June 12, 2024
- FKG
Table of Contents
FKG RACE EVO Sequences Taper
Description
Taper (1) and ISO diameter (2) identification
Depth marks ▲ (in millimetres)
Available on 21 / 25 / 31 mm instruments
Example on 25 / 31 mm instruments
Endo stop – ISO length / active part
Sequence | Glide path | Shaping |
---|---|---|
RACE EVO 4% | RE1 | |
15/.04 | RE2 | |
25/.04 | RE3 |
30/.04
RACE EVO 6%| RE1
15/.04| RE2
25/.04| RE3
25/.06
General information
Parameters for use :
Recommended speed range : 800-1000 rpm continuous rotation.
Torque : 1.5 Ncm.
Indications for use :
RACE® EVO endodontic instruments are intended for use in medical or hospital
facilities, by qualified health professionals.
RACE® EVO instruments are designed for the glide path and shaping of root canals.
Contraindications :
Nickel-Titanium instruments must not be used on individuals with a known
allergic sensitivity to these metals.
Adverse reactions : None known.
Notice and precautions :
- Take several radiographs from different angles to determine the anatomy of the root canals (length, width and curvature).
- The number of canals to be treated with a RACE® EVO instrument is limited to a maximum of 8 root canals.
- Irrigate the canal thoroughly and frequently after the RACE® EVO instrument is extracted.
- Regularly inspect the RACE® EVO instrument during use, and discard if it shows any signs of wear (for example : straightening).
- If the RACE® EVO instrument is not progressing easily, remove it from the canal, clean it, inspect its cutting edges, then irrigate the canal and recapitulate with an ISO 010 hand file.
- Always inspect the instrument(s) before use, and discard (it) them if there are any visible defect(s).
- When an instrument reaches the end of its life, please dispose of it in accordance with the applicable laws and regulations.
- Visit the FKG Dentaire SA website (www.fkg.ch) for all information relating to the applicable cleaning and sterilization methods.
Symbols
| Sterilized product (irradiation)
---|---
| Sterilizable in a steam sterilizer (autoclave) at the specified temperature
| To be used in continuous rotation at the recommended speed
| Warning : refer to the user manual
| Batch number
| Expiry date
| Do not use if package is damaged
| Nickel-Titanium instrument
| Reference number
| Manufacturer
Customer Support
FKG Dentaire Sàrl
Le Crêt-du-Locle 4, 2322 Le Crêt-du-Locle, Switzerland
Tel. +41 32 924 22 44, info@fkg.ch,
www.fkg.ch