FKG RACE EVO Endodontic System Instruction Manual

June 12, 2024
FKG

EVO Endodontic System
Instruction Manual

see www.fkg.ch for additional information

Description

FKG RACE EVO Endodontic System

Taper (1) and ISO diameter (2) identification conicity

FKG RACE EVO Endodontic System - Tapered
identification

Depth marks (in millimetres)
Available on 21 / 25 / 31 mm instruments
Example on 25 / 31 mm instruments
FKG RACE EVO Endodontic System - Depth marks Endo stop – ISO length/active part

| 19/8 mm| | 25/16 mm
---|---|---|---
| 21/16 mm| | 31/16 mm

General information

Parameters for use:
Recommended speed range: 800-1000  rpm continuous rotation.
Torque: 1.5 Ncm.
Indications for use :
RACE® EVO endodontic instruments are intended for use in medical or hospital facilities, by qualified health professionals.
RACE® EVO instruments are designed for accessing, glide path and shaping of root canals.
RACE® EVO instruments are to be used with gentle strokes of 2-3  mm, applying very light apical pressure and allowing the instrument to progress passively along the canal. After 3 strokes, remove and clean the instrument, then irrigate ; recapitulate with an ISO  010 hand file. If necessary, a brushing motion may be applied to remove coronal interferences and/or ensure uniform shaping of irregular canals.
Contraindications:
Nickel-Titanium instruments must not be used on individuals with a known allergic sensitivity to these metals.
Adverse reactions: None known.

Notice and precautions:

  • Take several radiographs from different angles to determine the anatomy of the root canals (length, width and curvature).
  • The number of canals to be treated with a RACE® EVO instrument is limited to a maximum of 8 root canals.
  • Irrigate the canal thoroughly and frequently after the RACE® EVO instrument is extracted.
  • Regularly inspect the RACE® EVO instrument during use, and discard if it shows any signs of wear (for example : straightening).
  • If the RACE® EVO instrument is not progressing easily, remove it from the canal, clean it, inspect its cutting edges, then irrigate the canal and recapitulate with an ISO 010 hand file.
  • Always inspect the instrument(s) before use, and discard (it) them if there are any visible defect(s).
  • When an instrument reaches the end of its life, please dispose of it in accordance with the applicable laws and regulations.
  • Visit the FKG Dentaire SA website (www.fkg.ch) for all information relating to the applicable cleaning and sterilization methods.

Symbols

| Sterilized product (irradiation)| | Nickel-Titanium instrument
---|---|---|---
| Expiry date| | Warning : refer to the user manual
| Sterilizable in a steam sterilizer (autoclave) at the specified temperature| | Reference number
| Do not use if package is damaged| | Batch number
| To be used in continuous rotation at the recommended speed| | Manufacturer

FKG Dentaire Sàrl
Le Cret-du-Locle 4, 2322
Le Cret-du-Locle
Switzerland
Tel. +41 32 924 22 44,
[email protected]
www.fkg.ch

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

FKG User Manuals

Related Manuals