FKG RACE EVO Endodontic System Instruction Manual
- June 12, 2024
- FKG
Table of Contents
EVO Endodontic System
Instruction Manual
see www.fkg.ch for additional information
Description
Taper (1) and ISO diameter (2) identification conicity
Depth marks (in millimetres)
Available on 21 / 25 / 31 mm instruments
Example on 25 / 31 mm instruments
Endo
stop – ISO length/active part
| 19/8 mm| | 25/16 mm
---|---|---|---
| 21/16 mm| | 31/16 mm
General information
Parameters for use:
Recommended speed range: 800-1000 rpm continuous rotation.
Torque: 1.5 Ncm.
Indications for use :
RACE® EVO endodontic instruments are intended for use in medical or hospital
facilities, by qualified health professionals.
RACE® EVO instruments are designed for accessing, glide path and shaping of
root canals.
RACE® EVO instruments are to be used with gentle strokes of 2-3 mm, applying
very light apical pressure and allowing the instrument to progress passively
along the canal. After 3 strokes, remove and clean the instrument, then
irrigate ; recapitulate with an ISO 010 hand file. If necessary, a brushing
motion may be applied to remove coronal interferences and/or ensure uniform
shaping of irregular canals.
Contraindications:
Nickel-Titanium instruments must not be used on individuals with a known
allergic sensitivity to these metals.
Adverse reactions: None known.
Notice and precautions:
- Take several radiographs from different angles to determine the anatomy of the root canals (length, width and curvature).
- The number of canals to be treated with a RACE® EVO instrument is limited to a maximum of 8 root canals.
- Irrigate the canal thoroughly and frequently after the RACE® EVO instrument is extracted.
- Regularly inspect the RACE® EVO instrument during use, and discard if it shows any signs of wear (for example : straightening).
- If the RACE® EVO instrument is not progressing easily, remove it from the canal, clean it, inspect its cutting edges, then irrigate the canal and recapitulate with an ISO 010 hand file.
- Always inspect the instrument(s) before use, and discard (it) them if there are any visible defect(s).
- When an instrument reaches the end of its life, please dispose of it in accordance with the applicable laws and regulations.
- Visit the FKG Dentaire SA website (www.fkg.ch) for all information relating to the applicable cleaning and sterilization methods.
Symbols
| Sterilized product (irradiation)| | Nickel-Titanium instrument
---|---|---|---
| Expiry date| | Warning : refer to the user manual
| Sterilizable in a steam sterilizer (autoclave) at the specified temperature|
| Reference number
| Do not use if package is damaged| | Batch number
| To be used in continuous rotation at the recommended speed| | Manufacturer
FKG Dentaire Sàrl
Le Cret-du-Locle 4, 2322
Le Cret-du-Locle
Switzerland
Tel. +41 32 924 22 44,
info@fkg.ch
www.fkg.ch