sauermann CLASS 50 Portables Thermo-Anemometer User Guide

June 12, 2024
sauermann

sauermann CLASS 50 Portables Thermo-Anemometer User Guide
sauermann CLASS 50 Portables Thermo-Anemometer

These portable instruments are used to measure the following parameters: temperature, relative humidity, CO, pressure, air velocity and light.

Operating temperature, protection of the instruments and information

about storage

Conditions of use (°C/%RH/m): from 0 to +50 °C. In non-condensing conditions. From 0 to 2000 m. Protection: IP54 

Storage temperature: HD 50, LX 50: from -20 to +70 °C. CT 50, CO 50, HM 50, LV 50, MP 50, MP 51, MP 55, VT 50: from -20 to +80 °C. If the instrument is stored outside the operating temperature (for example in a van, a warehouse…), please wait for 10 minutes in its operating temperature before start using it.

Keys

Select

On/Off/Hold

Take a measurement

Take a measurement

VT 50: before using the instrument with the hot wire probe, please lower the protection tube (1) of the sensing element.
VT 50: always use the hot wire probe with the red point in front of flow.

Switch on the instrument by pressing briefly on “On/Off/Hold”.

The instrument displays its name then the measurement.

CO 50: the instrument displays its name then a 5 second countdown. This count corresponds to the preheating time of the sensor. At the end of the countdown, the measured value of CO is displayed.

MP 50, MP 51: the instrument displays its name then an autozero is performed. At the end of the autozero, the measured value displays.

Press briefly on “Select” to switch measurement parameter display.

MP 50, MP 51, MP 55: press briefly and successively on “Select” to change the unit.

  • Available units for MP 50: Pa, mmH2 0, inWg, daPa.
  • Available units for MP 51: mmH2 0, kPa, in Wg, mbar, mmHg, daPa.
  • Available units for MP 55: mbar, mmHg, hPa.

HM 50: press briefly on “Select” to select a type of material.
Press the spikes in the material to be measured. When a measurement is over- range, above the maximum range, the value flashes. Below the minimum range, the value is no more measurable and the device indicates “—“, it means that for:

  • Wood type 1: the measurement is < than 10%
  • Wood type 2: the measurement is < than 9%
  • Plaster, brick and concrete: the measurement is < than 0.1%

Carry out a manual autozero (MP 50, MP 51)

During measurement:
Remove pressure tubes. Carry out a long press on “Select”.
The instrument makes an autozero then returns to the measurement display.

Hold the measurement

During measurement, press briefly on “On/Off/Hold” to hold measurement: “Hold” appears on the screen and measurements are fixed.

CO 50, HD 50, HM 50, LX 50, MP 55, VT 50: press briefly on “Select” to move from one parameter to the other. Press again briefly on “On/Off/Hold” to exit the Hold function.

Change the unit

The instrument is on.

CO 50, CT 50, HD 50, LX 50:
Carry out a long press on “Select” to change the unit.
Press briefly on “Select” to select the required unit of measurement.
Carry out a long press on “Select” to confirm the unit and return to the normal measuring mode.

VT 50:

Change the air velocity unit:
Press briefly on “Select” up to display the air velocity.
Carry out a long press on “Select” to change the air velocity unit.
Press briefly on “Select” to select the required unit of measurement.
Carry out a long press on “Select” to confirm the unit and return to the normal measuring mode.

Change the temperature unit:
Press briefly on “Select” up to display the temperature.
Carry out a long press on “Select” to change the temperature unit.
Press briefly on “Select” to select the required unit of measurement.
Carry out a long press on “Select” to confirm the unit and return to the normal measuring mode.

Select the velocity type: contact or optical (CT 50)

The instrument is on.
Carry out a long press on “Select”: “Cont” or “Optic” flashes.
Press briefly on “Select” to select contact velocity (“Cont”) or optical velocity (“Optic”).
Carry out a long press on “Select” to confirm.

Perform a self-test (HM 50)

Turn the instrument on. Remove the cap and turn the cap.Press “Select” to select the type of material to verify. Press the 2 pins of the instrument into the contact plugs of the cap. Read the value indicated by the instrument:

  • If the indicated value matches with the given value in the table below, the instrument runs properly.
  • If the indicated value does not match with the given value in the table below, clean the tips and carry out a new self-test. If the value is still wrong, change the pins. If it does not work, please contact the after sales service.
Resistance Material
Wood type 1 Wood type 2
1 MΩ 30.8% ±1%

Materials

Wood Plaster Brick and concrete
Moisture content State of the wood Moisture content
Moisture content State of the brick/concrete
< 15% Dry < 2%
From 15 to 30% Moist From 2 to 4%
> 30% High moisture > 4%

Batteries replacement

The instrument is OFF.
Remove the front part at the back of the instrument. Replace used batteries by new batteries AAA LR03 1.5 V. Replace the front.

**** To ensure safety, replace batteries with the same type.
Not doing so might lead to explosion. Dispose of used batteries according to instructio

Auto shut-off

The instrument switches off if it is not used during 20 minutes.

Download the data sheet

Sauermann Industrie
ZA Bernard Moulinet 24700 Montpon France
T. +33 (0)5 53 80 85 00
services@sauermanngroup.com

Sauermann NA
140 Fell Court, Ste. 302 Hauppauge, New York 11788
T. (+1) 631-234-7600
F. (+1) 631-234-7605
info.usa@sauermanngroup.com

Sauermann GmbH
Leibnizstraße 6
D – 74211 Leingarten
T. +49 (0)7131/399990
F. +49 (0)7131/399992
info.germany@sauermanngroup.com

Sauermann UK
Units 7-9, Trident Business Park Amy Johnson Way Blackpool – FY4 2RP
T. +44 (0) 870 950 6378
F. +44 (0) 870 950 6379
info.uk@sauermanngroup.com

Sauermann Italia srl S.U
Via Golini 61/10 40024 Castel S.Pietro Terme (BO)
T. (+39)-051-6951033
F. (+39)-051-942254
info.italy@sauermanngroup.com

Sauermann Ibérica
C/Albert Einstein 33. Planta 3. P. I. Santa Margarida II08223 Terrassa (Spain)
T. +34 931 016 975
info.spain@sauermanngroup.com

Sauermann Australia
1/36 Campbell Avenue, Cromer ,2099, NSW, Sydney
T. (+612) 8880 4631
sauermannpumps.com.au

Sauermann China
No. 1, Changxu Road, Juyuan New Park, Jiading District, Shanghai
T. (+86) 6100 1877
F. (+86) 6100 1870
info.china@sauermanngroup.com

sauermann Logo

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>