MEAN WELL MPS-65 Series 65W Single Output Medical Type Instruction Manual

June 4, 2024
MEAN WELL

MPS-65 Series 65W Single Output Medical Type

65W Single Output Medical Type
MPS-65 seriesMEAN WELL MPS 65 Series 65W Single Output Medical
Type

Features :

  • Universal AC input / Full range
  • Low leakage current <250pA
  • Protections: Short circuit / Overload / Over voltage
  • Cooling by free air convection
  • Medical safety approved (2 x MOPP between primary to secondary)
  • 100% full load burn-in test
  • Fixed switching frequency at 45KHz
  • 3 years warranty

MEAN WELL MPS 65 Series 65W Single Output Medical Type -
icon

SPECIFICATION

MODEL| MPS-65-3.3| MPS-65-5| MPS-65-7.5| MPS-65-12| MPS-65-13.5| MPS-65-15| MPS-65-24| MPS-65-27| MPS-65-48
---|---|---|---|---|---|---|---|---|---
OUTPUT| DC VOLTAGE| 3.3V| 5V| 7.5V| 12V| 13. W| 15V| 24V| 27V| 48V
RATED CURRENT| 12A| 12A| 13A| 5.2A| 4.7A| 4.2A| 2.7A| 24A| 1.35A
CURRENT RANGE| 0^ 15.2A| 0 – 13.8A| 0^9.6A| 0 – 6A| 0 – 5.1A| 0 – 4.8A| 0^3A| 0 ^ 2.7A| 0^ 1.5A
RATED POWER| 39.6W| 60W| 60W| 62.4W| 63.45W| 63W| 641W| 64.8W| 64.8W
OUTPUT POWER (max.)| 72W(.3.3V5OW:•5V:69Whviti 18CFM min. Faced air contraction
RIPPLE 6 NOISE (max.) Note.2| 80mVp.p| 100mVp-p| 100mVp-p| 100mVp-p| 100mVp-p| 100mVp-p| 100mVp-p| 100mVp-p| 100mVp-p
VOLTAGE ADJ. RANGE| 2.97 – 3.63V| 4.5^5.5V| 6.75^825V| 10.8^132V| 122 – 14.85‘| 13,5 – 165V| 21.6-26.4V| 243-29.7V| 43.2^52.8V
VOLTAGE TOLERANCE Note.3| 23.%| ±3.0%| 23.%| ±2.0%| ±2.0%| ±2.0%| ±2.0%| ±2.0%| ±2.0%
LINE REGULATION| 21.%| ±1.0%| ±1.0%| 21.%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%
LOAD REGULATION| ±3.0%| 23.%| 23.%| ±2.0%| ±2.0%| ±2.0%| ±2.0%| ±2.0%| ±2.0%
SETUP. RISE TIME| 800ms. 30ms230VAC 800ms. 30ms/115VAC al Woad
HOLD UP TIME (Typ.)| 50ms1230VAC 16ms/115VAC at full load
INPUT| VOLTAGE RANGE| 93 – 264VAC 127 – 370VDC
FREQUENCY RANGE| 47 -63Hz
EFFICIENCY(Typ.)| 66% 174% 176% 177% 178% 179% 180% 180% 180%
AC CURRENT (Typ)| 1.6A1115VAC 0.9A/230VAC
INRUSH CURRENT (Typ.)| COLD START 174/115VAC 40N230VAC
LEAKAGE CURRENT Note.?| Each leakage current < 25011A/264VAC . Touch current < 600/264VAC
PROTECTION| OVERLOAD| 73 – 105W (3.3V:51 – 75W)(5V:70 – 105W) rated output power
Protection type: Hkcup mode. recovers automatically after the fault condition is removed
OVERVOLTAGE| 3.8 – 4.46V I 5.75 – 6.75V 1 8.63- 10.1V I 13.8 – 16.2V I 15.5 – 18.2V11725 – 20.25VI 27.6 – 32.4VI 31 – 36.45V I 55.2 -64.8V
Protection type: Hiccup mode. recovers automatically after the fault condition is removed
EIWIRONVENT| WORKING TEMP.| -10^ +60t (Refer b ‘Derating Curve’)
WORKING HUMIDITY| 20 – 93% RH non-condensing
STORAGE TEMP.. HUMIDITY| -20^ +85t, 10 – 95% FtH
TEMP. COEFFICIENT| ±0.04%!C (0 – 501: )
VIBRATION| 10 – 500Hz. 2G 10min.71cycle. period Ice 60min. each along X. Y. Z axes
SAFETY 8
EMC (Note 4)| SAFETY STANDARDS| ANSI/AAMI E560601.1. TUV EN69501.1. IEC60601-1 approved
ISOLATION LEVEL| Primary-Secondary: 2xMOPP. Primary.Earth:14MOPP
WITHSTAND VOLTAGE| UP.O’P:4KVAC lP.FG:2KVAC OP.FG:0.5KVAC
 ISOLATION RESISTANCE| VP-0/P. 1/P-FG:10041 Ohms / 500VDC I 25t170% RH
EMC EMISSION| Compliance to EN 55011 (CISPR1I) Class B. 8461000-3-2.-3
EMC IMMUNITY| Compliance to EN6100041-211551.11. EN60601-1-2, medical level. criteria
OTHERS| MTBF| 359.7Khrs min. MIL-HDBK-217F (251.:)
DIMENSION| 127’76’42mm (L’WH)
PACKING| 0.23Kg: 54pcs/14.6Kg/128CUFT
NOTE| l. Al parameters NOT specially mentioned are measured at 230VAC input rated load end 251:41 ambient temperature.
a Ripple 8 noise is measured at 20MHz of banchwth by using an It twisted parties terminated with a Glut & 47uf parallel capacitor. 1 Tolerance: colas set up a tolerance. line regulation and load regulation.
4. The power supply is considered a component that will be instated into the final equipment Al the EMC tests are been executed by mounting the unit on a 360mml6Ornm metal plate with 1mm of thickness. The real equipment must be re- confirmed that it still meets EMC’s duties. For guidance on how to do these EMC tests. refer 4 -EMI testing or component sc./Vies.” (as avaiablo on http:/Mwer ineanwettcom) perform please pont
S. Mounting holes MI and 142 should be grounded for EMI purposes
6 Heat Sink HSI.HS2 can not be shorted
7 Touch °went vas measured from primary input to DC output
,.• Product Liability Disclaimer: For detailed intonation. Please refer to haps Jiwew meant/ell caWserviceDIsclauner aspx

Mechanical Specification

MEAN WELL MPS 65 Series 65W Single Output Medical Type -
Mechanical

Block Diagram

MEAN WELL MPS 65 Series 65W Single Output Medical Type -
Diagram

Derating Curve

MEAN WELL MPS 65 Series 65W Single Output Medical Type - Derating
Curve

Static Characteristics

MEAN WELL MPS 65 Series 65W Single Output Medical Type - Derating Curve
1

https://www.meanwell.com/Upload/PDF/Medical_EN.pdf

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

MEAN WELL User Manuals

Related Manuals