WOHLER GS 220 Gas Sniffer User Manual
- June 9, 2024
- WOHLER
Table of Contents
GS 220 Gas Sniffer
User Manual
General Information
Operation Manual Information
This operation manual allows you to work safely with the Wohler GS 220 Gas
Sniffer. Please keep this manual for your information.
The Wohler GS 220 should be employed by professionals for its intended use
only.
Liability is void for any damages caused by not following this manual.
Notes
WARNING!
Not following this warning can cause injury or death.
ATTENTION!
Not following this note can cause permanent damage to the device.
NOTE!
Useful information
Intended Use
The Wohler GS 220 Gas Sniffer is a gas leak detector for the tightness control and for the leakage testing of gas tubes. The meter shows the concentration of methane and propane on a bar graph display from 0 ppm to 1200 ppm. It is suitable for the leakage detection of gas lines carrying combustible gases. The meter is not suitable for the measurement of the exact gas concentration. The meter can be used indoors only. Do not use the meter for any other use than set out in this manual.
Information on disposal
Electronic equipment does not belong into domestic waste, but must be disposed in accordance with the applicable statutory provisions. You may hand in any defective batteries taken out of the unit to our company as well as to recycling places of public disposal systems or to selling points of new batteries or storage batteries.
Direction
Wohler Technik GmbH
Wohler-Platz 1
33181 Bad Wünnenberg
Tel.: +49 2953 73-100
Fax: +49 2953 73-96100
E-Mail: info@woehler.com
Technical Data
Description | Date |
---|---|
Sensor | Semiconductor |
Range Methane | 40 … 640 ppm |
Propane | 0 … 640 ppm |
Indicators | Acoustic: intermittent beep, the frequency will increase according |
to an increasing gas concentration
Visual: 5 level LED
Sensitivity| 35 ppm
Response time| approx. 60 Seconded
Operation Temp.| -5 to 45 °C
Storage Temp.| -20 to 45 °C
Auto-Off-Function| 10 minutes from Power On
Power supply| 4 AA alkaline batteries
Power consumption| ca. 72 mA
Battery life| 14 hours (continuous working), depending on the type of battery
Size| 175 x 70 x 38 mm
Probe length| 440 mm
Important notes
WARNING!
The Wohler GS 220 may never be used in an environment where there is the risk
of explosion. If the environment is suspected to conglomerate potentially
explosive gases, in no case you must use the Wohler GS 220. Batteries may not
be taken from the device and/or be changed.
ATTENTION!
The Wohler GS 220 indicates the approximate concentration of gas in a 5-level
LED.
However, the indication may be rather imprecise, according to the sensor, age,
ambient conditions etc. The Wohler GS 220 is a gas leak detector, and as such
it detects leaks, but it does not measure the exact concentration of gas.
High-density exposure to corrosive materials such as H2S, Sox, Cl2, Hall, etc.
may cause corrosion or breakage of the lead wires or heater material.
Sensor drift may occur when the sensor is contaminated by alkaline metals,
especially salt water spray.
Sensor drift may occur due to soak or splash the sensor with water.
If water freezes on the sensing surface, the sensor would crack and alter
characteristics.
This meter requires to operate under around 21% ambient oxygen environment in
order to function properly. This meter cannot work well in zero or low oxygen
atmosphere.
Light condensation under indoor usage should not be a problem for sensor.
However, if water condenses on the sensor‘s surface for a while, the sensor
characteristics may still drift.
Sensor performance may be also affected if exposed to a high density gas for a
long time, regardless of the powering condition.
When meter is off for a long time, the sensor may show a drift in resistance
according to the storage environment. So, the meter should be stored in a
sealed bag with clean air.
Controls
Meter
- Flexible Probe
- Sensor cap
- Indicator LED
- ON/OFF key
- Ear phone jack (3,5 mm)
- Mains connection
- Mute-Slide Switch
Display
- Methane and Propane gas concentrations
- Sensor ready LED
- Bat Low LED
Function Check
We recommend to carry out a Function Check of the gas detector before each use. To do this, proceed as follows.
- Switch on the gas detector.
- Check the external condition of the device, the function of all operating elements, the display and the charge status of the batteries.
- Hold the sensor in fresh air. The display must now read “0”.
- Check the sensitivity of the gas detector in a ventilated room as follows:
- Let gas flow out at a gas lighter (do not ignite!).
- Bring the probe of the gas detector close to the escaping gas.
- The gas detector must now display a clear signal.
Power ON/OFF
- Press ON/OFF key more than 0.2 second to turn on the meter. While the meter is on and electrical power is enough, the BAT. OK LED will stay on.
- To ensure the accuracy, users must perform the warm-up period in the clean air.
- The meter will be automatically switched off in 10 minutes. Or you could press ON/OFF key again to turn off.
While the meter is powered on but the „ready“LED is off, the sensor is under warm-up process. The warm-up time is approximately 60 seconds. While „ready“- LED is on, the meter is ready to measure. Even you turn off the meter for a short time and turn on again, it is always necessary to wait for 60 seconds warm-up time..
NOTE!
If the Wohler GS 220 has not been used for a long time, we recommend to repeat
the warm up/calibration phase several times to achieve a higher sensitivity
and accuracy. To do so, switch on the meter 2 or 3 times and off again, then
the „ready“- LED is on.
- The meter beeps every one second to assure the meter is in normal working status. The beeps frequency will speed up while the detected gas concentration becomes higher.
- While user use earphone, the meter will automatically mute but user could hear the beeper through earphone.
- While not using earphone, user could mute the beeper by switching the mute to ON position.
Leakage Checking
- To check the pipe leakage proceed as follows:
- Approaching the sensor to the pipe, slide slowly the sensor along the pipe; repeat the procedure from the other side of the pipe.
When the sensor is near to a leak, the LED will light and beep frequency increased.
After the Leakage Checking
- It takes at least 2 min. to wait for sensor recovering after removing the sensor out of leak point.
Battery Replacement
- While the Low battery LED is on, it is suggested to always have fresh batteries to ensure the accuracy. Open the battery compartment from the meter rear side and replace 4 new AA alkaline batteries.
Troubleshooting
The meter does not turn on.
Ready LED is not on after 60 seconds warm up.
- Make sure the time of pressing „ON/OFF“ key is more than 200 ms.
- Check the batteries are in place, good contact and correct polarity.
- Replace a new battery and try again.
- Check the sensor is in good contact. If LED is still off, please return the meter to the dealer for repairing.
Warranty and Service
Warranty
Each Wohler GS 220 Gas Sniffer will be tested in all functions and will leave
our factory only after extensive quality control testing.
If used properly, the warranty period for the Wohler GS 220 Gas Sniffer will
be twelve month from the date of sale.
Batteries are not covered by this warranty. Service by non authorized
personnel or making modifications to the analyzer voids any warranty.
Service
Wohler has built our reputation on excellence in customer service. Therefore, of course, we are readily available to assist you after the warranty period ends.
- Send us the device and we will repair it and return it to you with our package service.
- Immediate help is provided by our technical staff over the telephone.
Declaration of conformity
The product
Name: Wohler GS 300 Leakage Tester complies with the key safety
requirements set down in the guidelines of the Council for the Harmonization
of the Legal Requirements of the Member States in relation to the
electromagnetic compatibility (2014/30/EU).
The following standards were availed of to evaluate the product in respect of
the electromagnetic compatibility:
EN 61326-1 (2013)
Bad Wünnenberg, 18/01/2023 Johannes Lottering, Geschäftsführer/Managing
Director
Points of sale and service Germany
USA
Wohler USA Inc.
208 S Main Street
Middleton, MA 01949
Tel.: +1 978 750 9876
www.wohlerusa.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>