Operation Manual

  • Equipment Classification
  •  Class II equipment:
  •  Type B shock protection:
  •  Recommended sterilization: See 7. Sterilization
  •  Do not use in the presence of flammable liquids or gases.

Cautions for safe handling and operation
Read these safety precautions first before using this product. These instructions will show you how to operate this product safely and prevent injury to yourself or others. They are classified by the degree and/or severity of danger. All caution regarding safety should be observed.

Classification Degree and severity of danger
   Warning Provides an instruction where personal injury may occur.
   Caution Provides an instruction where minor to medium injury may

occur.
   Notice| Provides an instruction that should be observed for safety reasons.

Warning

  •  To Prevent Electric Shock: Do not plug in the power cord with wet hands.
  •  To Prevent Electric Shock: Prevent water from getting onto the control box.
  •  To Prevent Electric Shock: Plug in and remove the power by holding on to the base of the plug and not the cord.
  •  To Prevent Electric Shock: Use an electrical outlet that is grounded.
  •  If you feel any abnormality such as excessive vibration, heat, or noise, stop using the unit immediately.

Caution

  •  When operating this system, always consider the safety of the patient.
  •  The product is designed for clinical dental use by qualified personnel
  •  Every connection must be firm. If it is not firmly connected, it may cause an air or water leak, or the LED might not turn on.
  •  Do not drop the product or cause it to be impacted in any way. This could cause a malfunction and damage.
  •  This system is to be used with ISO 9168, type B or type C hose connectors.
  •  This system is meant to function in temperatures of 0-40°C (32-104°F), humidity at 10-85% RH, atmospheric pressure 700-1060hPa and no moisture or condensation in the Control Unit.
  • Operating outside of these parameters can cause a malfunction.
  •  When installing the Control Box or Motor, do not bend or twist the tubing or the cord. No excessive force is need to do the installation.
  •  Do not use contaminated air (by dust or moisture). If the air contains water or dust , it might cause a malfunction or overheat.

Unit

  •  Grasp cord by the plug to remove from the outlet. Do not pull or yank the cord itself. It may cause the wire to malfunction.
  •  Care should be taken not to place the AC cord near a gas burner. Never attempt to repair a burned motor cord. Always replace it with a new cord.
  •  Prior to use, always check for vibration, noise and overheating. If any abnormalities are detected, stop using immediately and contact your dealer.
  •  Be careful not to spill water onto the unit as it may result in a fire or an electric shock due to a short circuit.

Motor, Handpiece

  •  Do not use this product under strong stress for extended periods of time. It can cause overheating.
  •  Do not connect or disconnect the cord until the drive motor has completely stopped.
  •  Do not connect/disconnect the hand piece during operation.
  •  Verify that the Speed Control Switch is adj
  • sted within the allowable speed before use.
  • Connect only 4-hole, 5-hole or 6-pin tubing.
  •  Air Requirements: Dry, free from contamination and oil. Use a compressor with a dry air system/Install an air filter if necessary/ Blow out the lines before installation.
  •  Do not autoclave (or any other high-temperature sterilization) Control Box, AC Adaptor, or Motor Code.
  •  The user shall be responsible for operation and maintenance and inspection.

Specifications

Model Maxso E600P
Speed 2,000 ~ 40,000 rpm
Torque 3.4 Ncm
Rated Input AC 24V / DC36
Noise Level <60dB
Dimensions W127 x D149 x H54 (mm)

Component Names

Beyes-Maxso-E600P-Electric-Micromo

  1.  Control Box
  2.  Digital Speed Display
  3.  Gear Ratio Lamp
  4.  Gear Ration Select Key
  5.  FWD/REV Select Key
  6.  Program 1 Button
  7.  Program 2 Button
  8.  Speed Adjusters
  9.  Motor
  10.  Motor Cord
  11.  Power Switch
  12. AC Adapter Connector
  13.  Motor Cord Connector
  14.  Turbine Hose Connector
  15.  AC Cord
  16.  AC Adaptor

Control Box Operation

Caution
Make sure there is no residual air or water in the turbine hose when connecting it to the control box. Hand tightens the nut properly without excessive force when connecting the turbine hose and the motor cord connector.

3-1 Connect the handpiece
tubing from your delivery unit. Fit the turbine hose plug from the delivery unit to the turbine hose connector and tighten it completely.

Caution

  •  Do not use any other hose than the ones provided with the unit
  •  Do not use contaminated air
  •  Blow out the lines before installation

3-2 Connecting the Motor.
Connect the motor by aligning the motor cord plug and insert firmly.

3-3 Connecting the AC adapter
Insert the AC adapter plug into the AC adapter connector before plugging it into the wall outlet.

Caution
Never use any AC adapter other than the one provided with this product.

3-4 Connecting the Power Cord
Insert the power cord into the inlet of the AC adapter.

Handling the Motor

4-1 Connecting/Disconnecting the motor and motor cord.
Align and insert the motor pin firmly into the pin holes of the motor cord connector and fasten the motor cord nut securely. To remove the motor cord from the motor, unscrew and detach the motor cord nut, and gently pull out the motor cord connector.

4-2 Connecting/ Disconnecting the motor and handpiece
To insert the E-Type handpiece into the motor, align the handpiece and motor, then turn until you hear a click (positioning pins are aligned) To remove the handpiece, simply pull the handpiece from the motor.

Caution

  •  Care should be taken when using any handpiece attachment which requires lubrication (E-Type micro motor).
  •  Following lubrication, stand the handpiece attachment vertically until the excess solution drains. Wipe the handpiece then attach it to the motor.
  •  Do not connect or disconnect the handpiece until the motor has completely stopped.

Operation

5-1 Insert the power cord plug into the AC outlet and turn the power switch on. When the power switch is on, the lamps and indicator will come on.
5-2 Before using this system, make sure to operate the Foot Air Calibration function.
5-3 Push the Gear Ratio Select Key to indicate the speed of the attachment you will use.
Check the Table Below for Speeds

Hand Piece Gear Ratio Indicator Speed (min-1(rpm))
16:1 1 – 25 100 – 2,500
1:1 2 – 40 2,000 – 40,000
1:5 1 – 20 10,000 – 200,000

5-4 Set the maximum speed using the Rotation Adjustment Key (+/-).
5-5 Select FWD/REV rotation using the FWD/REV Select key.
5-6 The micro motor operation is controlled by the air switch/foot pedal of the delivery unit.
5-7 Program Function.

You can memorize the setting:

  1.  set the values for gear ratio, speed
  2.  Press P1 or P2 key for more than 3 seconds until it beeps, the setting is completed.

Sterilization

Sterilize the motor only.
For sterilization, we recommend the autoclave sterilization method. Sterilization is required the first time you use the unit as well as after each patient as noted below.

Autoclaving

  1.  Turn off the power.
  2.  Detach the motor from the motor cord.
  3.  Clean the surface of the motor with a brush (Do not use a metal brush), and wipe it with cotton and disinfectant alcohol.
  4.  Screw the Motor Cap to the Motor. Put the Autoclave Plug onto the motor insert.
  5.  Insert into autoclave ouch. Seal the pouch.
  6.  Autoclavable up to a maximum of 135°C.
  7.  Keep the Motor in the autoclavable pouch until you are ready to use it to keep it clean.

Caution

  •  Do not autoclave (or any other high-temperature sterilization) the Control Box, AC Adaptor, or Motor Cord.
  •  Do not lubricate the motor.
  •  Do not wipe or immerse the system in acidic water or acidic solutions.
  •  Do not sterilize with dirt on the surface, it may cause rust.

Maintenance

If the O-Rings become worn, the motor might not connect well to the handpiece, or water may leak. Replace the O-Rings.

Changing the O-Ring
Remove the O-Ring at the motor insert with a pointed tool, and mount the new O-rings into the groove.

Caution
There are 4 O-Rings on the Motor Insert. The blue one is thinner than the other 3. Make sure to install them in the proper spot.

Caution
If the O-Rings become worn, it may cause:

  •  Air/Water leak
  •  Difficulty attaching/disconnecting the handpiece
  •  Excess Vibration

Error Codes

Error Code Issue Cause Solution
E0 Overcurrent error (In Circuit) Detected an overcurrent in the circuit

Contact your dealer
E1| Overcurrent error. (In Software)| Detected an overcurrent in the software| Contact your dealer
E2| Retry Error| Motor was overloaded and stopped and can not restart.| Contact your dealer
E3| ITRIP Error| Faulty motor or internal circuit| Contact your Dealer
E5| Overvoltage Input

Error

| Overvoltage to the control box| Contact your dealer
E6| LED voltage error| Overvoltage in the internal cir- cuit.(more than 3.7V)| Contact your dealer
E8| Overheating error| High temperature safety system

activated due to extended use .

| Wait for a while and let it cool down before using again.
E9| Motor start error| Motor didn’t reach the preset speed in the  prescribed amount of time.

Break down of the motor cord wire, or internal circuit malfunc- tion

| Contact your dealer
---|---|---|---
EA| LED voltage error| Voltage for the LED isn’t high enough| Contact your dealer
Eb| Air offset error| Lower than preset ‘Lower limit of air pressure”| Displayed by foot/air calibration function. Not a Malfunction
Ec| Air full scale error| Higher than preset “Upper limit of

air pressure”

Air pressure higher than 0.4MPa

| Displayed by foot/air calibration function. Not a malfunction
EE| EEPROM Error| Malfunction of the settings

memory

| Contact your dealer

Troubleshooting

When having an issue with the unit, check this guide before contacting your dealer. These are the most common issues which may occur.

Issue Cause Solution
Pilot Lamp does not come on Power switch is off

-AC Adaptor is not connected

properly

-Internal fuse is blown

| Turn on the switch

-Check the connection

-Contact your dealer

Motor doesn’t not run| Tubing, Motor Cord, AC  adapt- er is not connected properly

-Air pressure is not sufficient

| Check the connection

-Check the air pressure from the delivery unit

-refer to Error code 9

The rotation speed is too slow| -Air pressure from delivery unit

is too low

-Air pressure on the delivery unit is lower than the “Lower limit of air pressure”.

| Check the air pressure on the delivery system.

-Do “Foot/Air calibration

Beeps when turning on the

switch

| You are stepping on the foot pedal when turning on the power switch| Do not step on the foot pedal

while turning on the switch

LED does not light up| Reached the end of life expec- tancy| Contact your dealer
---|---|---
Motor heats up abnormally during rotation| Coolant air is not coming out of

the delivery unit

| Check the air pressure of the delivery system
Water Leakage| Turbine, Motor cord is not

connected correctly

| Check the connection
When turned on, the setting values are different from what they were set to.| Turned off the motor while it

was rotating

| Do not turn off the unit until

after the motor stops spinning

Beyes Limited Warranty Statement

Scope of Warranty
BEYES Dental Canada Inc. (‘BEYES’) warrants to the original retail purchaser that it will be at BEYES’s option to repair or replace components of the dental products manufactured by BEYES (except for components not warranted under ‘Exclusions’) that are defective in material or workmanship under normal use and service. BEYES’ obligation under this limited warranty is limited to the repair or replacement of the applicable components. This limited warranty shall only apply to defects that are reported to BEYES within the applicable warranty period and which, upon examination by Beyes, prove to be defective. This warranty extends only to the first retail purchaser of a product and is not transferable or assignable. Replacement components or products may be used and/or refurbished components or products, provided they are of like quality and specifications as new components or products.

Applicable Warranty Period
The applicable warranty period, measured from the date of invoice to the original user, shall be warranted for a period of 25 months.

Exclusions
This limited warranty does not cover and BEYES shall not be liable for the following;

  1.  Defects, damage or other conditions caused, in whole or in part, by misuse, abuse, negligence, alteration, accident, freight damage, negligent storage, tampering or failure to seek and obtain repair or replacement in a timely manner;
  2.  Products which are not installed, used, and properly cleaned and maintained as required or recommended in the BEYES ‘Installation’ and/or ‘Installation/User’s Manual’ for the applicable product, including the specified structural and operational environment conditions and electrical power requirements;
  3.  Products considered to be of a consumable or sterile nature;
  4. Accessories or parts not manufactured by BEYES;
  5. Charges by anyone for adjustments, repairs, replacement parts, installation or other work performed upon or in connection with such products which are not expressly authorized in writing in advance by BEYES;
  6.  Costs and expenses of routine maintenance and cleaning;
  7.  Representations and warranties made by any person or entity other than BEYES;
  8.  Matching of color, grain or texture except to commercially acceptable standards;
  9. Changes in color caused by natural or artificial light;
  10.  Custom manufactured products;
  11. Alterations or modifications to the product by any person or entity other than BEYES;
  12. Products that would otherwise be covered under Sections 1 and 2 of this limited warranty, but are acquired: (i) from a person or entity that is not BEYES or one of its authorized dealers; or (ii) from a BEYES dealer that is not authorized to sell the product at issue in the geographic territory where the purchaser is located, or is not authorized to sell the product at issue within the medical, animal health or dental market, as the case may be, in which purchaser intends to use the product.

Exclusive Remedy; Consequential Damages Disclaimer

BEYES’ OBLIGATION UNDER THIS LIMITED WARRANTY IS THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS. BEYES SHALL NOT BE LIABLE FOR AND HEREBY DISCLAIMS ANY DIRECT, SPECIAL, INDIRECT, INCIDENTAL, EXEMPLARY OR CONSEQUENTIAL DAMAGES OR DELAYS, INCLUDING, BUT NOT LIMITED TO, DAMAGES FOR LOSS OF PROFITS OR INCOME, LOSS OF USE, DOWNTIME, COVER AND EMPLOYEE OR INDEPENDENT CONTRACTOR WAGES, PAYMENTS AND BENEFITS.

Warranty Disclaimer
THIS LIMITED WARRANTY IS BEYES’ ONLY WARRANTY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED. BEYES MAKES NO IMPLIED WARRANTIES OF ANY KIND INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS.Federal law restricts this device to sale by or on the order of a dentist, physician, or any other practitioner licensed by the law of the states in which he or she practices to use or order the use of this device. Beyes Dental Canada Inc. is not responsible for any typographical errors. Subjects to technical changes without notice. *All brands are holders of their respective trademarks.