Beyes Maxso E800P Optic Electric Handpiece System Instruction Manual
- June 4, 2024
- Beyes
Table of Contents
- Beyes Maxso E800P Optic Electric Handpiece System
- Introduction
- Basic technical parameters
- Product performance structure and composition
- Main unit interface
- Function and operation
- Safety precautions
- Cleaning, disinfection and sterilization
- Troubleshooting
- Storage and transport
- Environment protection
- Symbol instruction
- EMC-Declaration of conformity
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Beyes Maxso E800P Optic Electric Handpiece System
Introduction
Brushless electric micromotor system for driving dental handpieces for dental
surgery at a constant torque and consistent speed at high load. Mainly used
for dental aesthetics restoration, crown preparation, and can be used for all
highspeed and low-speed handpiece applications.
Cautions for safe handling and operation Read these safety precautions first
before using this product. These instructions will show you how to operate
this product safely and prevent injury to yourself or others. They are
classified by the degree and/or severity of danger. All cautions regarding
safety should be observed.
Caution:
- To prevent electric shock: Do not pull or plug in the power cord with wet hands.
- To prevent electric shock: Prevent water from getting on to the control box.
- Keep away from explosives and flammable materials.
- Do not use this dental electric motor for patients who are anesthetized with nitrous oxide.
Warning:
- This dental electric motor may malfunction when used in an environment where electromagnetic interference occurs. This dental electric motor cannot be installed near devices that release magnetic waves. When using an ultrasonic vibrating device or an electrode knife in the vicinity, make sure the switch on the dental electric motor control panel is turned off.
- Maxso E800P requires special precautions for EMC and needs to be installed and put into use according to the EMC environment.
- Device with electromagnetic transmitting will affect the normal operation of Maxso E800P. Please do not run both devices at the same time.
- Do not use it in operating rooms that contain a mixture of potentially flammable gas.
- To avoid personal injury or damage to the dental electric motor, make sure that the motor handpiece (hereinafter referred to as the motor) has completely stopped when connecting or disconnecting the handpiece.
- A severe impact, such as a drop from high position, can result in damage to the dental electric motor.
- Do not try to disassemble the control panel or motor.
- After use, please immediately clean, lubricate and disinfect the dental handpiece (hereinafter referred to as the handpiece).
- Do not lubricate the motor. The lubricant can cause overheating and damage the motor.
- Do not use a solution with dissolving ability to clean the control panel.
- Do not remove the motor cable from the motor. 12. After each operation, turn off the power supply.
Intended use
Provide driving force for dental handpieces for the purpose of dental surgery.
Model Maxso E800P
Contraindications
Beyes Dental Canada Inc. will not be liable for any direct or indirect damages and losses under the following conditions:
- The device is used to the unmentioned usages or the usages outside the scope of applications.
- The operator did not use the device with the method in accordance with the procedures and requirements stipulated in the Instruction Manual.
- The wiring system of the room where the device is used does not meet the requirements of appropriate standard and other proper requirements.
- Assembling, operating, and repairing the equipment without the authorization of Beyes Dental Canada inc.
- The environmental conditions in which the equipment is located or stored do not meet the requirements mentioned in the section on technical requirements.
Safety requirements
Beyes Dental Canada Inc. will not be liable for any direct or indirect damages and losses under the following conditions:
- The device is used to the unmentioned usages or the usages outside the scope of application.
- The operator did not use the device with the method in accordance with the procedures and requirements stipulated in the Instruction Manual.
- The wiring system of the room where the device is used does not meet the requirements of appropriate standard and other proper requirements.
- Assembling, operating, and repairing the equipment without the authorization of the Beyes.
- The environmental conditions in which the equipment is located or stored do not meet the requirements mentioned in the section on technical requirements.
Basic technical parameters
Specification of main unit
- Model: Maxso E800P
- Size:165.5mm×129.7mm×77.6mm
- Power supply input: 100-240V~ 50/60Hz 2.5A
- Power supply output: DC 30V 3.2A
Motor specification
- Model: E800 MC
- Rotation speed: 2000-40000 rpm;
- Voltage input: DC 24V
- Size: 22×76.7mm
- Tail cord length: 1.8m
Use environment
- Ambient temperature: +5°C~ +40°C 2.3.2
- Relative humidity: 30% ~ 75% 2.3.3
- Atmospheric pressure: 70kPa ~ 106kPa
Device safety classification
- Type of protection against electric shock: Class II equipment
- Degree of protection against electric shock: B type applied part
- Degree of protection against harmful ingress of water: Ordinary equipment (IPX0). Not waterproof.
- Classified by operation mode: Intermittent operating device
- Degree of safety application in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide: Equipment cannot be used in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide.
Product performance structure and composition
Safety requirements during installation
Danger: Equipment is installed on the premise that the installation must
comply with appropriate standards and associated electric safety requirements.
Danger: Never install the unit in an explosive atmosphere and do not
operate in areas with flammable gases (anesthetic mixtures, oxygen, etc.).
Danger: The installation site should be protected from shocks and
splashes of water or other liquids.
Danger: Do not install the unit near or above the heat source. It must be
installed in a place where the surrounding air is sufficiently circulated.
There is enough space around it, especially for the exhaust fan and the back
position.
Warning: Do not place the parts directly under sunlight or ultraviolet
light.
Warning: This equipment is movable, but please handles with care.
Warning: Make sure the connection parts are dry before connecting the
wires to the unit. If necessary, blow to dry it with an air gun.
See the packing list for the machine configuration.
It mainly consists of main unit, motor, power adapter, etc.
Front View of the Main Unit
Rear view of the main unit
Schematic diagram of motor installation
Connecting/ Disconnecting the motor and the motor cord
- A. Align and insert the motor pin firmly into the pin holes of the motor cord connector and fasten the motor cord nut securely.
- B. To remove the motor cord from the motor, unscrew and detach the motor cord nut, and gently pull out the motor cord connector.
Connecting/Disconnecting the motor and handpiece
- A. To insert the E-Type handpiece into the motor, align the hand piece and motor, then turn until you hear a click (positioning pins are aligned)
- B. To remove the handpiece, simply pull the hand piece from the motor.
Installation steps
- First open the package to check whether the items of the equipment are complete according to the packing list, and place the main unit on a stable surface.
- Connect the four-hole handpiece tube to the control box through the fourhole handpiece connector and tighten it.
- Connect the power adapter to the DC adapter connector and connect the socket to the DC adapter with the power cord.
Main unit interface
Main interface
Setting interface
Manual mode interface
Interface of calibration mode
Enter the setting interface, click the calibration mode touch button, enter the air pressure calibration interface, click “Start”, and the interface will pop up as shown in Figure 9. Fully press down on the foot pedal until it displays 100%, and then release the pedal. The calibration screen should show it is successful.
Restore the factory setting
Are you sure you want to restore to factory setting?
Enter the setting interface, and click the restore factory settings touch button to enter the restore factory settings interface for confirmation. In confirming interface, click OK, the interface shown in Figure 10 will pop up. Select “OK” to restore to original factory settings parameters or select “Cancel” to quit restoring factory settings.
Language selecting interface
Enter the setting interface, click the language selection touch button so that the interface shown in Figure 11 will pop up, and select the desired language. When you click “OK”, the corresponding language selected will save. When you select “Cancel”, the original language setting would be maintained.
Function and operation
-
Install the product correctly according to the product installation steps, and the operator should face the screen.
-
Turn on the power switch on the main unit, the screen will display and enter the main control interface (Figure 1).
-
The electric motor operation is controlled by the foot pedal of the dental chair.
-
Make sure that the pedal control calibration is performed before using the unit for the first time.
-
Icon description Icon
Icon| Name| Function
---|---|---
| Mode| Select the preset fixed speed. (P1/P2/P3)
| Speed ratio| Select the speed of contraangle
(16:1/1:1/1:5)
| Speed adjustment| Increase speed
| Speed adjustment| Decrease speed
| Store| Store the set parameters
| Forward/Reverse
rotation
| Control the forward and reverse rotation of motor
| Speed| Display the set operating speed
| Setting| Enter the setting interface
DEMO
| Manual mode| Enter manual mode adjusting interface
| Calibration| Enter calibration interface
| Restore factory setting| Restore the system to factory setting.
| Language selection| Enter language setting interface
| Exit| Exit the submenu setting mode
| Start| Start to activate the motor
| Stop| Stop the motor
| LED| Turn on/off the motor LED
| Speed adjustment| Set the speed using the Speed
Adjustment Key (+/-).
Basic function adjustment on main unit controlling interface
-
Select speed ratio by through touching key “1”. The color of the key will change when touching the key.
-
Control the ON and OFF of LED through key “2”.
-
Adjust the speed of the motor by using key “3”. Increase or decrease speed with key “+” and “-“.
-
Set forward rotation or reverse rotation by using key “4”. 5.6.5 Set the working speed of corresponding mode by touching key “5”.
-
Store the set speed ratio, speed, mode, forward rotation/reverse rotation, and the ON/OFF of LED by touching key “6”.
-
Enter setting interface by touching key “7”.
Basic function adjustment in setting interface
-
In setting interface, touch key “8” Manual Mode key to enter demo mode interface. Under the demo mode state, the motor rotation can be directly controlled without foot pedal. Touch Speed Ratio key to select corresponding speed ratio. Touch Speed key to adjust the speed. Touch Start key, Stop key, Forward/Reverse Rotation key, and LED key to control the output of motor. Touch the Exit key to exit the demo mode interface.
-
In setting interface, touch key “9” Calibration key, touch Start key, and the interface shown in Figure 8, 9 will pop up. Fully step on the foot pedal until the screen displays 100%, and release the foot pedal to finish the calibration. 5.7.3 In setting interface, touch key “10” Restore Factory Setting key, click OK to enter interface shown in Figure 10, and decide whether to restore the factory setting by clicking “OK” or “Cancel” key.
-
In setting interface, touch key “11” to enter interface shown in Figure 7 for language selection.
-
In the setting interface, touch key “12” to exit submenu setting mode.
Safety precautions
Cautions:
- For repairs and purchase of spare parts, please contact our authorized dealers.
- The accuracy of the speed monitoring depends on the high precision performance of the handpiece installed on the micro motor. To ensure the accuracy of the display speed, please use Beyes electric handpieces with the motor.
- Read this operating manual before use and fully understand the functions of each part.
- Check the operating status of the dental electric motor before use to confirm that there is no abnormality.
- Test the dental electric motor before use to ensure accurate operation. 6.6 If the dental electric motor is permanently malfunctioning (excessive vibration, noise and heat generation, etc.), please immediately close it and return it to the authorized dealer. 6.7 Clean the control panel with a damp cloth and turn off the power before cleaning.
Cleaning, disinfection and sterilization
-
Cleaning
Cleaning the motor surface with a water-absorbent cloth to remove any liquid residue from the motor surface and wipe it back and forth 5 times. -
Disinfection
Use a chlorine-free disinfectant to wipe the motor. It is recommended to use Cavicide disinfectant or 75% alcohol, but do not soak the motor in the disinfectant. -
Drying
Drying should be conducted after cleaning and disinfection. It is recommended to use compressed air to dry. -
Sterilization
Before sterilization, first remove the motor, install a sterile plug and a sterile aluminum sleeve, place the motor in a high pressure steam sterilization bag, and seal it. And then sterilize it under the temperature and pressure of 134°C (273 ° F) and 2.0 bar ~ 2.3 bar (0.20MPa ~ 0.23MPa) ) for not less than 4 minutes. Drying is needed after sterilization, The sterilization of the dental contra-angle handpiece is detailed in the instruction manual of the contra-angle handpiece.
The installation is complete
Warning:
- Install disinfection plug and disinfection aluminum sleeve at the motor interface before sterilization. The installation method is shown above.
- Before sterilization, the motor cord should be removed.
- Do not lubricate the inside of the motor.
- The motor can be sterilized repeatedly for at least 250 cycles of sterilization.
- Sterilize the motor only. Do not autoclave (or any high temperature sterilization) control box, AC adaptor, or motor cord.
Troubleshooting
Fault | Cause | Solution |
---|---|---|
Error 01 | Step on the foot pedal before turning on the device, so that the | |
voltage is too high. | Release the foot pedal and reboot. | |
Error 02 | The input voltage is too high or too low. | Check whether the power |
adapter is right.
Error 03| Motor abnormality| Check whether the motor is well connected or
replace the motor.
If the problem still cannot be solved, please contact our local distributor or our company.
Note: The user must use the original accessories. Please contact our local dealer or the company for purchase. It is forbidden to use related accessories of other brands, so as to avoid damage to the electric motor or other dangers.
Storage and transport
- The device should be handled carefully and lightly. Be sure that it is far from the vibration, and installed or kept in a cool, dry, and ventilated place.
- Do not store the machine together with articles that is poisonous, combustible, caustic, or explosive.
- This machine should be stored in a room where the relative humidity is10%~93%, atmospheric pressure is 70kPa~106kPa, and the temperature is -20°C~+55°C.
- Excessive impact and shake should be prevented during transport. Lay it carefully and lightly.
- Do not put it together with dangerous goods during transport. 9.6 Avoid being exposed to sun, rain, and snow during transport.
After-sales service
Beyes Dental Canada Inc. is responsible for repairs under the warranty period if the device cannot work normally due to a quality defect. The device must be registered with Warranty Card from the date of invoice to the original user. Please refer to the Warranty Card for details of warranty.
Environment protection
Part | Toxic or harmful substances or elements |
---|---|
(Pb) | (Hg) |
Main unit | ○ |
Motor handpiece | ○ |
Power adapter | ○ |
Dental contra-angle | ○ |
Mechanical elements, including bolts, nuts, washers, etc. | ○ |
- Indicates that the content of the toxic substance in all homogeneous materials of the part is below the limit requirement stipulated in SJ/T-11363- 2006 Limit Requirements for Toxic and Hazardous Substances in Electronic Information Products.
×: indicates that the content of the toxic substance in at least one of the homogeneous materials of the part exceeds the limit requirement specified in SJ/T-11363-2006. (This product meets EU RoHS environmental protection requirements; there is currently no mature technology in the world to replace or reduce the content of lead in electronic ceramics, optical glass, steel and copper alloy.) According to the Administrative Measures on the Restriction
of the Use of Hazardous Substances in Electric and Electronic Products and the Regulations on the Management of the Recycling of Waste Electric and Electronic Products and related standards, please observe the safety and precautions of the products, and after use, please recycle or dispose this product after according to the methods in local laws and regulations.
Symbol instruction
- Date of manufacture
- Type BF applied part
- Used indoor only 93 %
- Humidity limitation 10 %
- 106kPa Atmospheric pressure for storage 70kPa
- IPXO Ordinary equipment
- Manufacturer
- Class II equipment
- Follow Instructions for Use 55 max
- Temperature limitation -20 min
- Can be autoclaved
- Appliance compliance WEEE directive
EMC-Declaration of conformity
The device has been tested and homologated in accordance with EN 60601-1-2 for EMC. This does not guarantee in any way that this device will not be effected by electromagnetic interference Avoid using the device in high electromagnetic environment.
Serial number | Cable name | cable length (m) | Cable type |
---|---|---|---|
1 | Power cord (input) | 1.2m | Unshielded parallel line |
2 | Power cord (output) | 1.2m | Unshielded parallel line |
3 | Handle tail | 1.8m | Unshielded parallel line |
Beyes Dental Canada Inc. 23-595 Middlefield Road Toronto, Ontario, M1V 3S2 Canada Tel: 1-855-603-1888 Fax: 1-855-720-1228 Email: info@beyes.ca Website: www.beyes.ca
Lotus NL B.V. Koningin Julianaplein 10, Ie Verd, 2595AA, The Hague,
Netherlands Tel: +31645171879 (English)
+31626669008 (Dutch) Email: info@beyes.ca Website:
www.beyes.ca
Printed in Canada ENI044
Rev.1/13.08.21