Beyes Maxso E600B Built-in Electric Micromotor System Instructions

June 4, 2024
Beyes

Maxso® E600B
Electric Micromotor System (Built-in)
Instruction for UseBeyes Maxso E600B Built in Electric Micromotor
System -

Introduction

Maxson E600B is designed to be built into the dentist delivery system and operates as a seamlessly integrated electric motor system. The integrated motor has an intuitive touchpad interface and is engineered to reduce complexity and provide increased efficiency for optimal treatment results.
For your safety, before operating this product for the first time, please read through this manual thoroughly.

1.1 Package includes

  • instruction for use –this manual
  • user control panel
  • control board
  • cable
  • o-ring
  • moun,ting rack
  • screws, 2-pin terminal and 3-way air connector
  • motor with seal blocker and cap
  • jumper connector

1.2 Symbol
The following sign may appear in this manual, before or in attached files. Some signs stand for this product’s standard and conformity.

| Conformity to Medical Device Directive 93/42/EEC.
---|---
| Date of Manufacturing
| Manufacturer
| Serial number
| Type B
| Do not trash
****| Consult instruction for use
| Classification: Class II, Type B

Safety and warning

This product should be used for dental operations and to be operated only by dentists or dental-related professional

Warning

  • Install this product by an authorized third-party technician
  • Follow the instruction for installation
  • Clean and maintain this product following this manual
  • Do not drop or crash this product.
  • Do not spray oil into this motor.
  • Do not use chemical disinfection on motor
  • Do not trash this product. Check local available rubbish classified collection for disposal

Safety

  • Do not operate this product in an environment mixed with air, oxygen, nitric oxide, or a flammable anesthetic.
  • Do stop the motor before detaching the motor or cable

Installation requirement

Warning

  • This module should be installed inside the dental delivery unit by an authorized third-party technician
  • This module requires AC24V/DC36V power, water, and air from the delivery unit

3.1 Connection inside the delivery unit
a. Board to air supply, control panel cable, motor cable, and power cable

Beyes Maxso E600B Built in Electric Micromotor System - inside delivery
unit

b. Base, air, and water tubing connection

Beyes Maxso E600B Built in Electric Micromotor System - tubing
connection

3.2 Connection outside delivery unit
a. Control panel

b.Motor

Beyes Maxso E600B Built in Electric Micromotor System -
Motor

c. Control panel mounting

  • Screw panel on the holder
  • Install the plastic cap on the left or right side of the delivery unit
  • Adjust the angle and tighten the holder and cap

Beyes Maxso E600B Built in Electric Micromotor System - holder and
cap

Operation

Beyes Maxso E600B Built in Electric Micromotor System -
Operation

4.1 Start Motor

  • Select the gear ratio 16:1, 1:1, or 1:5 to match the handpiece
Gear ratio Speed range
16:01 100-2,500 RPM
1:01 2,000-40,000 RPM
1:05 10,000-200,000 RPM
  • Press REV to select motor direction – green for clockwise, yellow for counter-clockwise
  • Press +/- to adjust the speed – the higher number indicates a higher speed
  • Step on the foot paddle to start the motor
  • During motor running, the light is on and the speed can be adjusted

4.2 Stop Motor
To stop the motor, release the foot paddle

Warning

  • Do not press REV while the motor is running.
  • To reverse direction, first, stop the motor.
  • Shut down power before detaching the motor

4.3 Program Preset

  • Select the gear ratio
  • Set the speed
  • Press and hold P1 or P2 key 3 seconds or more until a beep
  • Press P1 or P2 to restore the setting when needed

Note: motor direction is not stored

Maintenance

5.1 Control Panel
Wipe control panel with ethyl alcohol
5.2 Motor
Wipe the surface with ethyl alcohol

Caution

  • Do not use strong oxidant water or chemical disinfectants to clean motor
  • Do not dip or wash motor
  • Inspect o-ring – replace if worn or broken

5.3 Sterilization

  • Wipe the surface with ethyl alcohol
  • Put seal blocker and seal cap on the motor

Beyes Maxso E600B Built in Electric Micromotor System -
Sterilization

  • Put the motor in the autoclave pouch and seal
  • Autoclave 20 min @121°C or 15min @ 132°C

Caution

  • Must close the seal blocker and seal cap for autoclave
  • Must clean the motor surface before autoclave disinfection

Technical specification

Control system: Vector control system
Motor type: DC brushless motor
Speed: 2,000-40,000rpm
Torque: 3.4Ncm
Power input: AC 24V/DC36
Motor size: φ22  x L71mm
Motor weight: 67g
Noise: < 60dB
Light source: LED (white light)
Air supply: ht) >1.5L/min (0.2MPa)
Water supply: >65mL/min (0.2MPa)
Control box size: L130 x W55 x H21mm
Total weight: 632.00g
Environment temperature: -40°C~+70°C
Humidity: 10%~100%
Air pressure: 177~216KPa (1.8~3.0 kgf/cm2)

Beyes limited warranty statement

7.1 Scope of Warranty
BEYES Dental Canada Inc. (‘BEYES’) warrants to the original retail purchaser that it will be at BEYES’s option to repair or replace components of the dental products manufactured by BEYES (except for components not warranted under ‘Exclusions’) that are defective in material or workmanship under normal use and service. BEYES’ obligation under this limited warranty is limited to the repair or replacement of the applicable components. This limited warranty shall only apply to defects that are reported to BEYES within the applicable warranty period and which, upon examination by Beyes, prove to be defective. This warranty extends only to the first retail purchaser of a product and is not transferable or assignable. Replacement components or products may be used and/or refurbished components or products, provided they are of like quality and specifications as new components or products.

7.2 Applicable Warranty Period
The applicable warranty period, measured from the date of invoice to the original user, shall be warranted for a period of 25 months.
7.3 Exclusions
This limited warranty does not cover and BEYES shall not be liable for the following;

  1. Defects, damage, or other conditions caused, in whole or in part, by misuse, abuse, negligence, alteration, accident, freight damage, negligent storage, tampering or failure to seek and obtain repair or replacement in a timely manner;
  2. Products that are not installed, used, and properly cleaned and maintained as required or recommended in the BEYES ‘Installation’ and/or ‘Installation/User’s Manual for the applicable product, including the specified structural and operational environment conditions and electrical power requirements;
  3. Products considered to be of a consumable or sterile nature;
  4. Accessories or parts not manufactured by BEYES;
  5. Charges by anyone for adjustments, repairs, replacement parts, installation or other work performed upon or in connection with such products which are not expressly authorized in writing in advance by BEYES;
  6. Costs and expenses of routine maintenance and cleaning;
  7. Representations and warranties made by any person or entity other than BEYES;
  8. Matching of color, grain, or texture except to commercially acceptable standards;
  9. Changes in color caused by natural or artificial light;
  10. Custom manufactured products;
  11. Alterations or modifications to the product by any person or entity other than BEYES;
  12. Products that would otherwise be covered under Sections 1 and 2 of this limited warranty, but are acquired: (i) from a person or entity that is not BEYES or one of its authorized dealers; or (ii) from a BEYES dealer that is not authorized to sell the product at issue in the geographic territory where the purchaser is located or is not authorized to sell the product at issue within the medical, animal health or dental market, as the case may be, in which purchaser intends to use the product.

7.4 Exclusive Remedy; Consequential Damages Disclaimer
BEYES’ OBLIGATION UNDER THIS LIMITED WARRANTY IS THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS. BEYES SHALL NOT BE LIABLE FOR AND HEREBY DISCLAIMS ANY DIRECT, SPECIAL, INDIRECT, INCIDENTAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES OR DELAYS, INCLUDING, BUT NOT LIMITED TO, DAMAGES FOR LOSS OF PROFITS OR INCOME, LOSS OF USE, DOWNTIME, COVER AND EMPLOYEE OR INDEPENDENT CONTRACTOR WAGES, PAYMENTS AND BENEFITS.

7.5 Warranty Disclaimer
THIS LIMITED WARRANTY IS BEYES’ ONLY WARRANTY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED. BEYES MAKES NO IMPLIED WARRANTIES OF ANY KIND INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS.

Federal law restricts this device to sale by or on the order of a dentist, physician, or any other practitioner licensed by the law of the states in which he or she practices to use or order the use of this device. Beyes Dental Canada Inc. is not responsible for any typographical errors. Subjects to technical changes without notice. *All brands are holders of their respective trademarks.

Beyes Dental Canada Inc.
23-595 Middlefield Road
Toronto, Ontario, P1V 3S2
Canada
Tel: 1-855-603-1888
Fax: 1-855-720-1228
Email: info@beyes.ca
Web: www.beyes.ca
Lotus NL B.V.
Koningin Julianaplein 10, Ie Verd,
2595AA, The Hague,
Netherlands
Tel: 3-164-517-1879 (English)
3-162-666-9008 (Dutch)
Email: info@beyes.ca
Web: www.beyes.ca

Printed in Canada
File #: ENI025
Rev.2/01.03.21

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>