PRECISION MEDICAL 195405 EasyPulse5 Oxygen Conserving Regulator User Manual
- June 9, 2024
- Precision Medical
Table of Contents
PRECISION MEDICAL 195405 EasyPulse5 Oxygen Conserving Regulator
PARTS
How To Use
- Inspect Conserver and Cylinder Valve for damage or contaminants. Do NOT use if any are found. “Crack” cylinder valve by opening the valve slightly and closing it.
- Ensure Conserver is in the OFF position and O-ring Seal is in place. Align the Nipple with a recessed hole in the cylinder. Hand tightens the Knob. Do NOT use tools.
- When using a Precision Medical Carry Bag, slide the cylinder into the bag with the gauge facing the mesh. Attach the cannula to the Conserver’s Outlet Connection.
- Slowly open the cylinder valve. Turn Dial to the prescribed setting. Zip the bag closed.
- Place cannula in nostrils and breathe through nose to begin Oxygen fl ow. If using a bag adjust the straps for maximum comfort.
- When finished using:
Turn the cylinder valve to the OFF position. Turn Conserver Dial to the “2 LPM Continuous” position. Wait for Oxygen to stop fl owing from Conserver. Remove Conserver from Cylinder Valve. Store Conserver and Carry Bag in a safe place.
Troubleshooting
If the problem cannot be corrected, consult your Dealer/Home Care equipment provider.
PROBLEM | PROBABLE CAUSE | REMEDY |
---|---|---|
A. No flow | 1. Cylinder valve closed | 1. Turn on the cylinder |
2. Conserver in “OFF” | 2. Set to the prescribed position setting | |
3. Cylinder empty | 3. Replace the cylinder | |
4. Dial not at setting | 4. Set the dial so the indicator points to a |
setting
| 5. Conserver not sensing breath| 5. Check the position of the cannula in
the nose Do not breathe through the mouth
B. Leaking at Cylinder Connection| 1. Missing or defective O-ring Seal| 1.
Replace the O-ring Seal