PRECISION MEDICAL PM9000 VacTrap Vacuum Safety Trap Bottle User Manual
- June 9, 2024
- Precision Medical
Table of Contents
PRECISION MEDICAL PM9000 VacTrap Vacuum Safety Trap Bottle
RECEIVING / INSPECTION
Remove the Precision Medical, Inc. Vac Trap from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider.
INTENDED USE
To protect suction regulators and central vacuum systems from liquid contamination.
READ ALL INSTRUCTIONS BEFORE USING
This manual instructs a Professional to install and operate the Vac Trap. This is provided for your safety and to prevent damage to the Vac Trap. If you do not understand this manual, DO NOT USE the Vac Trap and contact your Provider.
SAFETY INFORMATION
WARNING Indicates a potentially hazardous situation which, if not avoided, may result in death or serious injury.
CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
CAUTION Used without the safety alert symbol, indicates a potentially hazardous situation which, if not avoided, could result in property damage.
WARNING Use this Vac Trap only for its “Intended Use” as described in this manual.
Only personnel instructed and trained in its use should operate this Vac Trap.
Be sure all connections are tight and leak free.
TYPICAL INSTALLATION DIAGRAM
CAUTION The Vac Trap must be in the vertical position to function.
PARTS LIST
Description | Part # |
---|---|
Float Assembly | 503234 |
Jar | 1731 |
Tubing Nipple | 9073 |
MAINTENANCE
Before each use, inspect the Vac Trap for any damage; DO NOT USE if damaged.
RETURNS
Returned products require a Returned Goods Authorization (RGA) number, contact
Precision Medical, Inc. All returns must be packaged in sealed containers to
prevent damage. Precision Medical, Inc. will not be responsible for goods
damaged in transit. Refer to Precision Medical, Inc. Return Policy is
available on the internet at
www.precisionmedical.com.
Manuals available on our website:
www.precisionmedical.com
DISPOSAL INSTRUCTIONS
Dispose of the Vac Trap in accordance with the local regulations.
Please Recycle
WARNING Dispose of the Vac Trap using Biohazard precautions.
TROUBLESHOOTING
Problem | Probable Cause | Remedy |
---|---|---|
Float Stuck |
- Float neck dirty
|
- Remove foreign substance
No Suction|
- Leaking connections
- Suction Source not turned “ON”
- Jar full
|
- Check connections & reconnect
- Tighten Jar
- Turn Suction Source “ON”
- Empty / clean Jar
CLEANING
External Cleaning Method:
- Unscrew the Jar.
- Remove the Float Assembly.
The vacuum trap can also be wiped down with quaternary cleaners, hypochlorite solutions, peroxide solutions or other hospital grade cleaners. Alcohol based cleaners should not be used.
Internal Cleaning Method:
The vacuum trap can be cleaned using an enzymatic detergent solution and
immersed during cleaning. It can be placed back into service using an
ultrasonic cleaner, washer/disinfector, and/or a hospital grade cleaner
providing high level disinfection, cleaners like Cidex OPA or the like.
Caution should be taken not to exceed 200 degrees F when cleaning the Vac
Trap, as temperatures higher than 200 degrees F will damage the device
Gas Sterilization:
Ethylene Oxide mixture can be used at temperatures up to 135 degrees F. Follow the procedures from the manufacturer of the sterilizer.
CAUTION Exceeding the recommended sterilization temperatures may shorten
the life of the Vac Trap.
DO NOT autoclave.
WARRANTY
Precision Medical, Inc. warrants that the Medical Vacuum Safety Trap will be
free of defects in workmanship and/or material for the following period: One
(1) year from date of shipment.
Should any failure to conform to this warranty appear within the applicable
period, Precision Medical, Inc. shall, upon written notification thereof and
substantiation that the goods have been stored, installed, maintained and
operated in accordance with Precision Medical, Inc.’s instructions and
standard industry practice, and that no modifications, substitutions, or
alterations have been made to the goods, correct such defect by suitable
repair or replacement at its own expense.
ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.
The representative of Precision Medical, Inc. or any retailers are not
authorized to make oral warranties about the merchandise described in this
contract, and any such statements shall not be relied upon and are not part of
the contract for sale. Thus, this writing is a final, complete and exclusive
statement of the terms of that contract.
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY OF MERCHANTABILITY,
FITNESS FOR A PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESS
OR IMPLIED.
Precision Medical, Inc. shall not under any circumstances be liable for
special,
incidental or consequential damages including but not limited to lost profits,
lost sales, or injury to person or property. Correction of non-conformities as
provided above shall constitute fulfillment of all liabilities of Precision
Medical, Inc. whether based on contract, negligence, strict tort or otherwise.
Precision Medical, Inc. reserves the right to discontinue manufacture of any
product or change product materials, designs, or specifications without
notice.
Precision Medical, Inc. reserves the right to correct clerical or
typographical errors without penalty
Customer Support
Precision Medical
300 Held Drive • Northampton, PA 18067 USA
T: 800-272-7285 • F:
800-352-1240
www.precisionmedical.com