Precision Medical PM15 Series EasyAir Compressor Instruction Manual
- June 9, 2024
- Precision Medical
Table of Contents
- Precision Medical PM15 Series EasyAir Compressor
- Safety Information
- Specifications
- Component Description
- Grounding Instructions
- Operating Instructions
- Cleaning/Maintenance
- Replacement Parts
- Troubleshooting
- Limited Warranty and Limitation of Liability
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Precision Medical PM15 Series EasyAir Compressor
Receiving/Inspection
Remove the Precision Medical, Inc. Easy Air Compressor (PM15-F, PM15EF, PM15-P
or PM15EP) from the packaging and inspect for damage. If there is any damage,
DO NOT USE and contact your Provider.
Intended Use
The basic function is to supply a portable compressed air source adequate to
drive the high-volume nebulizers.
Read All Instructions Before Using
This manual instructs a Professional to install and operate the Easy Air
Compressor (PM15-F, PM15EF, PM15-P or PM15EP). This is provided for your
safety and to prevent damage to the Easy Air Compressor. If you do not
understand this manual, DO NOT USE the Easy Air Compressor and contact your
Provider.
Explanation of Abbreviation
- psi Pounds Per Square Inch
- l/min Liters Per Minute
Safety Information
Warnings and Cautions
- DANGER Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
- WARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
- CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
- CAUTION Used without the safety alert symbol indicates a potentially hazardous situation which, if not avoided, may result in property damage.
- CONSULT ACCOMPANYING DOCUMENTS
- Type B Equipment
- Protective Earth Ground
- Power On
- Power Off
DANGER
Caution must be exercised when using this Compressor where oxygen is being
administered.
- DO NOT use near any type of flame or flammable/explosive substances.
- DO NOT allow the Compressor to exceed Operating or Storage Temperature specifications.
TO REDUCE THE RISK OF ELECTROCUTION
Always unplug the Compressor immediately after using.
- DO NOT use while bathing.
- DO NOT place or store the Compressor where it can fall or be pulled into a tub or sink.
- DO NOT place or drop into water or other liquid.
- DO NOT reach for the Compressor that has fallen into water. Unplug immediately.
- DO NOT attempt to repair the Easy Air Compressor. Any attempt could result in electrical shock.
WARNING
To reduce risk of burns, electrocution, fire, or injury to persons:
- Close supervision is necessary when this Compressor is used by, on, or near children or invalids. Use this Easy Air Compressor only for its “Intended Use” as described in this manual.
- DO NOT use attachments not recommended by Precision Medical, Inc.
- DO NOT use the Compressor with any equipment that requires “dry” air, including ventilators, gas blenders, or flow generators, without a device to properly dehumidify the discharged air of the PM15.
- DO NOT use if the Compressor has a damaged cord or plug.
- DO NOT use if the Compressor has been dropped or damaged.
- DO NOT drop or insert any object into any openings.
- DO NOT cover air vents. Restriction of air flow will cause unit to overheat, causing the thermal reset to shut down the Compressor.
Specifications
-
Dimensions:
- Length: 16 in (40.6 cm)
- Width: 7.25 in (18.4 cm)
- Height: 10.25 in (26.0 cm)
-
Weight: Weight: 19.5 lbs (8.8 kg)
-
Shipping Weight: 23 lbs (10.4 kg)
-
Electrical:
- PM15-F & PM15-P: 120 VAC ±10%, 60 Hz, 5 Amps
- PM15E: 230 VAC ±10%, 50 Hz, 3.2 Amps
- PM15EFGB: 220/230 VAC, 50 Hz, 2.8 Amps
-
Outlet: Grounded 9/16 – 18 Male Threaded Fitting
Models Equipped with a Pressure Gauge- Maximum Rated Sustained Pressure: 60 psi
- PM15 Output: 50 psi………………………..Minimum 13 l/min
- PM15E/PM15EFGB Output: 50 psi….Minimum 9 l/min
-
Operating Range:
- Models Equipped with a Flow Gauge
- PM15-F & PM15EF Output: (± 1 l/min)
- l/min – as set and read on the gauge
-
Operating Temperature Range: 50°F to 104°F (10°C to 40°C)
-
Storage Temperature Range: -4°F to 104°F (-20°C to 40°C)
-
Storage/Operating Humidity: Maximum 95% Noncondensing
-
Safety Systems: High-Temperature Motor Shutdown
-
Thermal Protector: Class B, Thermal Limit Automatic Reset
- Class I – with respect to protection from electrical shock
-
Equipment Classification:
- Type B – degree of protection against electrical shock
- IPX0 – degree of protection against ingress of liquids
- Mode of Operation – continuous
-
The Compressor meets the EMC requirements of EN 55011: Group 1 Class B.
-
The Compressor meets the requirements of CGA V-5: section 5.4.
This Compressor is not suitable for use in the presence of a flammable anesthetic mixture with air, with oxygen or with nitrous oxide.
Specifications are subject to change without prior notice.
Component Description
CAUTION
Missing or illegible labels must be replaced, contact Precision Medical, Inc
.
Grounding Instructions
The Easy Air Compressor must be plugged into an electrical outlet that is properly grounded. In the event of an electrical short circuit, grounding reduces the risk of electrical shock. This product is equipped with a grounding circuit.
DANGER
Improper use of the grounding plug can result in a risk of electrical shock.
WARNING
Ground reliability can only be achieved when the power cord is connected to an
equivalent receptacle marked “Hospital Grade.”
Operating Instructions
CAUTION
Inspect the Easy Air Compressor for visible damage, DO NOT USE if damaged.
- Become familiar with the Compressor (see Component Identification).
- Position the Compressor near a properly grounded electrical receptacle. Allow a minimum of 4 in (10 cm) clearance on all sides to ensure adequate air flow for cooling of the Compressor.
- Ensure the On/Off (I/O) Switch is in the “OFF” (O) position.
- Ensure the Power Cord is properly connected.
- Attach appropriate device to the Outlet Fitting securely, without overtightening.
- Pull Regulator Knob “OUT” to unlock regulator, then turn Knob fully counterclockwise. (Failure to do so may cause the device not to start.)
- Press the On/Off (I/O) Switch to the “ON” (I) position.
- Adjust pressure/flow.
- To INCREASE pressure/flow – Turn Knob CLOCKWISE
- To DECREASE pressure/flow – Turn Knob COUNTERCLOCKWISE
- Push Regulator Knob “IN” to lock pressure/flow setting.
- Compressor is now ready to use.
- When the Compressor is not in use, press the On/Off (I/O) Switch to the “OFF” (O) position.
Cleaning/Maintenance
WARNING
Turn the Easy Air Compressor off and disconnect from any external power source
before cleaning. DO NOT attempt to repair. There are no serviceable parts
inside.
To clean the filter (weekly):
- Unplug Compressor.
- Check air inlet filters for dust buildup.
- If dust buildup occurs, remove air inlet filters and wash in warm, soapy water and rinse.
- Absorb excess water from filter with a dry towel.
- Return the air inlet filters.
Cleaning for Exterior Components:
Use a clean cloth or sponge dampened with soap and water. Wipe dry with a
clean cloth or paper towel.
CAUTION
Excessive dust buildup on filter will reduce the performance of the
Compressor. If this occurs, clean or replace it with a new filter. The air
filter does not filter out saline. Treatments using nebulized saline will draw
the saline mist back into the Compressor. Saline causes rapid deterioration of
the compressor seals and will void the warranty.
Maintenance
- The EasyAir Compressor should be serviced if the device does not meet flow specifications.
- Return the Easy Air Compressor to your Provider for Compressor Maintenance (including rebuilding pump).
- Contact Precision Medical, Inc. for Preventative Maintenance Services.
- Compressor Rebuild Kit – Part #502060C
Returns
Returned products require a Return Goods Authorization (RGA) number, contact
Precision Medical, Inc. All returns must be packaged in sealed containers to
prevent damage. Precision Medical, Inc. will not be responsible for goods
damaged in transit. Refer to Precision Medical Inc. Return Policy available on
the Internet, www.precisionmedical.com.
Manuals available on our website; www.precisionmedical.com.
NOTE:
When returning this Easy Air Compressor, disconnect the power cord from the
back of the device.
Disposal Instructions
This device and its packaging contain no hazardous materials. No special
precautions need to be taken when disposing the device and/or its packaging.
Replacement Parts
Part #/Description
- 503237 Large Filter
- 502391 Small Filter
Troubleshooting
If the Easy Air Compressor fails to function, consult the Troubleshooting Guide below. If problem cannot be solved, consult your Provider.
Problem | Probable Cause | Remedy |
---|---|---|
On/Off Switch in “OFF” position | Push On/Off Switch to “ON” position | |
Power Cord not connected |
properly
| Check Power Cord
connections
Shut Compressor Off, Wait 15 minutes, Ensure vents are
clear.
Replace Fuse
| Compressor overheated
Compressor does not operate| Blown Fuse
Power Interruption
| Turn off Compressor. When Power restored, repeat Start
Up Procedure.
|
Trapped back pressure
| BEFORE STARTING COMPRESSOR. Pull
“OUT” regulator Knob to unlock, then turn Knob fully counterclockwise.
Unable to adjust pressure/flow|
Locked Regulator Knob
|
Pull Regulator Knob out
Overheating
|
Poor Ventilation
| Allow a minimum of 4 in (10 cm) clearance on all sides to ensure adequate air flow for
cooling of the Compressor.
Clean Air Filters
Dirty Air Filters
Low pressure/flow
| Loose connection at outlet fitting| Tighten connection
Improper regulator adjustment| Re-adjust regulator
Limited Warranty and Limitation of Liability
Precision Medical, Inc. warrants that EASY AIR 15 Compressor (the Product) will be free of defects in workmanship and/or material for the following period:
- One (1) year from shipment
Should any failure to conform to this warranty appear within the applicable period, Precision Medical, Inc. shall, upon written notification thereof and substantiation that the goods have been stored, installed, maintained and operated in accordance with Precision Medical, Inc.’s instructions and standard industry practice, and that no modifications, substitutions, or alterations have been made to the goods, correct such defect by suitable repair or replacement at its own expense.
ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.
The representatives of Precision Medical, Inc. or any retailers are not
authorized to make oral warranties about the merchandise described in this
contract, and any such statements shall not be relied upon and are not part of
the contract for sale. Thus, this writing is a final, complete and exclusive
statement of the terms of that contract.
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.
Precision Medical, Inc. shall not under any circumstances be liable for special, incidental or consequential damages including but not limited to lost profits, lost sales, or injury to person or property. Correction of non- conformities as provided above shall constitute fulfillment of all liabilities of Precision Medical, Inc. whether based on contract, negligence, strict tort or otherwise. Precision Medical, Inc. reserves the right to discontinue the manufacture of any product or change product materials, designs, or specifications without notice. Precision Medical, Inc. reserves the right to correct clerical or typographical errors without penalty.
Precision Medical, Inc.
- 300 Held Drive Northampton, PA 18067 USA.
- Toll Free Tel: 800-272-7285
- Toll-Free Fax: 800-352-1240
- Tel: (+001) 610-262-6090
- Fax: (+001) 610-262-6080
ISO 13485 Certified
Tell us how we are doing!
Visit us at www.precisionmedical.com.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>