Precision Medical PM5950 Oxygen Analyzer Instruction Manual
- June 9, 2024
- Precision Medical
Table of Contents
- Precision Medical PM5950 Oxygen Analyzer
- Receiving/Inspection
- Safety Signs and Warning\Caution Statements
- Specifications
- Component Description
- Oxygen Sensor Component Identification
- Operating Instructions
- Effects of Temperature
- Cleaning
- Maintenance
- Returns
- Troubleshooting
- Replacement Parts
- LIMITED WARRANTY AND LIMITATION OF LIABILITY
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Precision Medical PM5950 Oxygen Analyzer
Receiving/Inspection
- Remove the Precision Medical Oxygen Analyzer from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider.
Intended Use/Indications for Use
- Precision Medical, Inc. Oxygen Analyzer is intended as a tool for use by qualified personnel to spot-check or measure oxygen concentration of a delivered air/oxygen mixture.
Contraindication
- The Precision Medical, Inc. Oxygen Analyzer is not intended to actively monitor oxygen concentration or oxygen gas mixtures while being delivered to a patient. The Precision Medical, Inc. Oxygen Analyzer is not intended for use in a MRI environment.
Operator Profile
- The Oxygen Analyzer is to be used by trained healthcare professionals.
Read All Instructions Before Using
- This manual instructs a Professional how to operate the Oxygen Analyzer. This is provided for your safety and to prevent damage to the Oxygen Analyzer. If you do not understand this manual, DO NOT USE the Oxygen Analyzer and contact your Provider.
Safety Signs and Warning\Caution Statements
- DANGER
- Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
- WARNING: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION: Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
CAUTION
- Used without the safety alert symbol indicates a potentially hazardous situation which, if not avoided, may result in property damage.
Follow instructions for use
- General Mandatory Action Sign
- This device may contain electrical components that are hazardous to the environment. DO NOT dispose device into standard trash. Contact your local waste management for disposal of Electronic Equipment.
- Caution! U.S. Federal Law restricts this device to sale by or on the order of a physician.
Liquid ingress protection – Dripping water
- (vertically falling drops) shall have no harmful effect on the device when mounted in an upright position.
MR Unsafe
- An item marked MR unsafe is known to posed hazards in all MR environments.
WARNING
-
Read this User Manual before installing or operating the Oxygen Analyzer.
-
Only trained, qualified personnel should operate the Oxygen Analyzer.
-
Use this Oxygen Analyzer only for its intended use as described in this manual.
-
DO NOT use near any type of flame or flammable/ explosive substances, vapors or atmosphere.
-
DO NOT allow an excessive length of cable near anyone’s head or neck that could result in strangulation. Secure excess cable to bed rail or suitable object.
-
Medical Oxygen should meet the requirements of USP.
-
Always follow ANSI and CGA standards for Medical Gas Products, and Oxygen Handling.
-
The Oxygen Analyzer should only be serviced by a qualified service technician, or by Precision Medical, Inc.
-
DO NOT use Oxygen Analyzer with a cable that appears worn, cracked or has damaged insulation.
-
Never install the sensor in a location that will expose the sensor to exhaled breath or secretions, unless you intend to dispose of the sensor, flow diverter and tee adapter.
-
Improper use of this device can cause inaccurate oxygen readings which can lead to improper treatment, hypoxia or hyperoxia. Follow the procedures outlined in this user manual.
-
Not for use in an MRI environment. The Oxygen Analyzer contains magnetic, ferrous material that may affect the results of an MRI.
Device intended for use with dry gas only. -
Before use, all individuals who will be using the Oxygen Analyzer must become thoroughly familiar with the information contained in this User Manual. Strict adherence to the operating instructions is necessary for safe, effective product performance. This product will perform only as designed if installed and operated in accordance with these operating instructions.
-
Use only genuine Precision Medical Inc. accessories and replacement parts. Failure to do so may seriously impair the analyzer’s performance. Repair of this device must be performed by a qualified service technician.
-
Calibrate the Oxygen Analyzer before use, or weekly when in operation, or if environmental conditions change significantly. (i.e., Elevation, Temperature, Pressure, Humidity.
-
Use of the Oxygen Analyzer near devices that generate electrical fields may cause erratic readings.
-
Never autoclave, immerse or expose the Oxygen Analyzer (including sensor) to high temperatures (>70°C). Never expose the device to pressure, irradiation vacuum, steam, or chemicals.
-
This device does not contain automatic barometric pressure compensation.
-
Although the Sensor of this device has been tested with various anesthesia gases including nitrous oxide, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane and found to have acceptably low interference, the device in entirety (including electronics) is not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide. Only the threaded Sensor face, Flow Diverter, and Tee Adapter may be allowed to contact such a gas mixture.
-
NOT for use with inhalation agents. Operating the device in flammable or explosive atmospheres may result in fire or explosion.
-
No modification of this device is allowed.
CAUTION
DO NOT: use if dirt or contaminants are present on or around this Oxygen Analyzer or connecting devices. smoke in an area where oxygen is being administered. clean with aromatic hydrocarbons. steam autoclave. gas sterilize. immerse Oxygen Analyzer or Sensor in liquid. immerse the sensor in any cleaning solution, autoclave or expose the sensor to high temperatures.
- Store the Oxygen Analyzer in a clean, dry area when not in use.
- The Oxygen Sensor is a sealed device containing a mild acid electrolyte, lead (Pb), and lead acetate. Lead and lead acetate are hazardous waste and should be disposed of properly.
- Dropping Sensor can adversely affect its performance.
- Be sure to apply 100% oxygen, or ambient air concentration to the device during calibration or the device will not calibrate correctly.
Specifications
Base Device Specifications
- Dimensions (Analyzer without Cable and Sensor attached):
- Depth: 1.30” (3.32 cm)
- Width: 2.90” (7.30 cm)
- Height: 4.00” (10.23 cm)
- Cable Length: 10 ft. (3.05m) (fully extended)
- Weight: Device Weight: 0.35 lbs (5.60 oz / 0.15 kg)
- (includes: Analyzer, Sensor, Cable and batteries)
Operating Conditions:
- Temperature: 50ºF – 113ºF (10ºC – 45ºC)
- Altitude: Sea Level to 8000 feet
- Humidity: 0 – 95% non-condensing
Storage Conditions:
- Temperature: 5°F – 122°F (-15°C – 50°C)
- Humidity: 0 – 95% non-condensing
- Mode of Operation: Continuous
- Electrical Classification: Internally powered Medical Electrical equipment
- Diverter Fitting: fits industry standard, 15 mm Tee Adapter
- Measurement Range: 0.0 – 100% Oxygen
- Resolution: 0.1 %
- Total Accuracy: ± 3.0% Actual Oxygen Level over full operating temperature range
- Drift of Measurement: < +/-1% of full scale at constant temperature, pressure and humidity
- Response Time: 90% of final value in less than 12 seconds at 77°F (25ºC)
- Warm-up Time: Not required
- Low Battery Indication: Low battery icon displayed
- Patient Contact : Indirect contact via gas passing through sensor sampling site.
Sensor Specifications
- Sensor Type
- Galvanic Oxygen Sensor; Precision Medical PN 504877
- Expected Sensor Life
-
1,000,000 O2 % Hours
-
- Specifications are subject to change without notice.
Classifications
- Protection against
- electric shock: Internally powered equipment
- Protection against water: IPX1 (Drip Proof)
- Mode of Operation: Continuous
- Sterilization: Non-Sterile Device
- Flammable anesthetic mixture: Not suitable for use in presence of a flammable anesthetic mixture
Component Description
Oxygen Sensor Component Identification
ITEM NAME | DESCRIPTION |
---|
Coiled Cable
| The Coiled Cable allows the Sensor to be positioned up to 10 ft from the
side of the Oxygen Analyzer. There are Male Plugs at each end of the Coiled
Cable .
Locking Ring| Male Plugs have Locking Rings and must be engaged when in
use.
Oxygen Sensor| Galvanic Oxygen Sensor
Diverter Fitting| Fitting is used to connect to the Oxygen Source.
Tee Adapter
| The Tee Adapter is used to connect the Oxygen Sensor and Diverter Fitting to an oxygen pathway circuit.
Operating Instructions
Sensor Installation
CAUTION
-
Inspect the Oxygen Analyzer, Sensor and Cable for visual damage before use, DO NOT USE if damaged.
-
Inspect the Oxygen Sensor and Diverter Fitting for visual damage or electrolyte leakage before use. DO NOT USE if damaged.
-
Use ONLY an Oxygen Sensor specified by Precision Medical, Inc.
-
The Oxygen Sensor should not be used in the presence of flammable anesthetics such as Diethyl Ether or Cyclopropane.
-
DO NOT attempt to open or repair the Oxygen Sensor.
-
The Sensor electrolyte is corrosive and contains lead.
-
DO NOT let Sensor electrolyte come in contact with the skin. If it does, flush the affected area with water.
Check the Sensor regularly for leaks. -
If the Sensor is leaking, replace it with NEW Sensor. Leaking or used Sensors should be handled and disposed of in accordance with local regulations.
-
An SDS is available from Precision Medical, Inc.
-
If the Oxygen Sensor is used in breathing circuits, the Diverter must be attached to the Sensor and must be used with the Tee Adapter.
-
The Oxygen Sensor must be installed before the Oxygen Analyzer can be operated.
- Screw the Diverter to the bottom of the Oxygen Sensor, tighten until snug.
- If using the Tee Adapter, attach to the Diverter.
- Insert the one end of the Coiled Cable into the top of the Sensor, and secure by tightening the Locking Ring.
- Insert the other end of the Coiled Cable into the Sensor Cable Connection located on the top of the Oxygen Analyzer. Secure it in place by tightening the Locking Ring.
- Wait approximately 20 minutes for the NEW Sensor to stabilize to the environment.
- Calibrate the Oxygen Analyzer with the NEW Sensor.
CAUTION
- Calibrate the Oxygen Analyzer before each use, and when replacing the Oxygen Sensor or the batteries.
- To ensure accuracy, the Precision Medical Oxygen Analyzer should only be calibrated using 100% Oxygen.
- Using any other concentration will result in possible inaccurate readings.
- Air calibration is not recommended unless the Sensor can be exposed to a known source of clean air. Hospital room air is often enriched with excess oxygen.
- Calibrate the Oxygen Analyzer at a pressure and flow similar to your application.
- Before calibrating the Oxygen Analyzer, the oxygen concentration readout should be stable and not drifting more than 0.2%.
- DO NOT calibrate the Oxygen Analyzer in humidified gas.
Calibration
The new calibration is required when:
-
The measured O2 percentage in 100% O2 is below 97.0% O2. The measured O2 percentage in 100% O2 is above 100.0% O2.
-
The CAL reminder Icon is activated on the LCD.
If you are unsure about the displayed O2 percentage. -
A simple calibration may be made with the Sensor open to static ambient air. For optimum accuracy Precision Medical Inc. recommends that the Sensor be placed in a closed loop oxygen circuit where gas flow is moving across the sensor in a controlled manner. Calibrate with the same type of circuit and flow that you will use in taking your readings.
- Follow Sensor Installation instructions above.
- Attach an open-ended reservoir to the end of the Tee Adapter. Start flow of 100% oxygen at a pressure and flow similar to your application. Six to 10 inches of corrugated tubing works well as a reservoir.
- A calibration gas flow to the Oxygen Analyzer of two liters per minute or more is recommended to minimize the possibility of obtaining a “false” calibration value.
- Allow the oxygen to saturate the Sensor. Although a stable value is usually observed within 30 seconds, allow at least two minutes to ensure that the sensor is completely saturated with the calibration gas.
- If the Oxygen Analyzer is not already turned on, do so now by pressing the Analyzer “ON” button.
- Press and hold the Cal button on the Oxygen Analyzer for at least 3 seconds to activate calibration. The Analyzer will detect a stable Sensor signal. When obtained, the Analyzer will display the assumed calibration gas on the LCD.
Effects of Elevation/Barometric Pressure Changes
- This device does not automatically compensate for changes in barometric pressure. Calibration of the Analyzer shall be performed when elevation at which the device is being used changes more than 500 feet.
Effects of Temperature
To minimize temperature effects:
- In a breathing circuit, place the Oxygen Sensor upstream of the heater.
- Allow time for the Oxygen Sensor to stabilize to its new room temperature.
- Perform the calibration procedure at a temperature close to or similar to your clinical application
Effects of Humidity
- High Moisture levels will dilute the oxygen concentration, decreasing the concentration of oxygen being monitored by the Oxygen Sensor.
- High humidity can cause condensation to collect on the Oxygen Sensor, obstructing the passages and reducing the effectiveness of the Oxygen Sensor.
CAUTION
To reduce the effects of humidity on the Sensor:
- DO NOT USE the Oxygen Sensor in environments with greater than 95% humidity.
- Place the Oxygen Sensor upstream from the Humidifier in a breathing circuit.
Effects of Pressure
CAUTION
- The Oxygen Analyzer is not equipped with automatic barometric pressure compensation.
- The following recommendation is provided to reduce the chances of pressure causing false readings.
- Calibrate the Precision Medical Oxygen Analyzer using 100% oxygen or room air at the same pressure and flow as the gas to be analyzed.
Effects of Anesthetic Gases
Anesthetic Agent| Test Concentration| Oxygen Concentration
Error
---|---|---
Helium| 50%, Balance Oxygen| 0%
Nitrous Oxide| 80%, Balance Oxygen| 0%
Carbon Dioxide| 10%, Balance Oxygen| 0%
Halothane| 4%| <1.5% Oxygen
Enflurane| 5%| <1.5% Oxygen
Isoflurane| 5%| <1.5% Oxygen
Sevoflurane| 5%| <1.5% Oxygen
Desflurane| 15%| <1.5% Oxygen*
- Test mixture = 30% O2, balance 70% N2O except where noted.
- Errors may vary based on concentrations and exposure times.
CAUTION
- The Oxygen Sensor should not be used in the presence of flammable anesthetics such as Diethyl Ether or Cyclopropane.
Cleaning
CAUTION
- DO NOT steam autoclave.
- DO NOT immerse the Oxygen Analyzer into any liquid.
- DO NOT use any strong solvent or abrasive cleaners.
- DO NOT allow any liquid to enter the Oxygen Analyzer or the Oxygen Sensor; this will damage the Oxygen Analyzer or Oxygen Sensor and will void the Warranty.
- Disconnect all connections before cleaning.
- Clean exterior surfaces of the Oxygen Analyzer, Tee Adapter and Coiled Cable with a cloth dampened with mild detergent and water.
- Wipe dry with a clean cloth.
- Wipe Oxygen Sensor and Diverter using a lint free dry clean cloth.
Maintenance
Sensor Replacement
- Reference “SENSOR INSTALLATION”
CAUTION
- Sensor Replacement must be performed by Qualified Personnel.
- Replace batteries when “Low Battery” icon is displayed.
- Remove 4 screws located on the back and remove cover.
- Remove and replaced batteries with 2 new AA Alkaline batteries. Verify correct polarity.
- Secure back cover onto the Oxygen Analyzer with the 4 screws.
Returns
- Returned products require a Returned Goods Authorization (RGA) number, contact Precision Medical, Inc. All returns must be packaged in sealed containers to prevent damage.
- Precision Medical, Inc. will not be responsible for goods damaged in transit. Refer to Precision Medical, Inc. Return Policy available on the Internet, www.precisionmedical.com.
Disposal Instructions
- The Oxygen Analyzer may contain electrical components that are hazardous to the environment. DO NOT dispose device into standard waste stream.
- The Oxygen Analyzer contains internal batteries. Batteries contain materials which can contaminate the environment when improperly disposed of.
- The Oxygen Sensor contains lead. DO NOT dispose sensor into standard trash. Dispose in accordance with the local regulations.
- Contact your local waste management for disposal of Electronic Equipment.
Troubleshooting
- If the Oxygen Analyzer fails to function, consult the Troubleshooting Guide. If the problem cannot be solved by using Troubleshooting Guide, consult your Provider.
CAUTION
- DO NOT attempt to service the device while in use.
Problem | Probable Cause | Remedy |
---|---|---|
Low Battery Indicator is displayed | 1. Battery voltage too low | 1. Replace |
batteries
New Oxygen Sensor responds slowly or seems to drift| 1. Oxygen Sensor has NOT
temperature stabilized| 1. Wait approximately 20 minutes for Oxygen Sensor to
stabilize with the environment, and then recalibrate the Oxygen Analyzer
Oxygen Sensor does not react to changes in oxygen concentration| 1.
Condensation on the Oxygen Sensor
2. Nonfunctioning
Oxygen Sensor
| 1. Remove Condensation
2. Replace with New
Precision Medical Oxygen Sensor
No Display/ LCD screen will not power ON| 1. Dead batteries| 1. Replace
batteries
Displays “ – – – ”| 1. Defective Sensor
2. Bad Cable Connection
3. Invalid
Calibration
| 1. Replace with New Precision Medical Oxygen Sensor
2. Ensure Cable is secured to the Device and the Sensor. Replace Cable.
3. Recalibrate the device.
Replacement Parts
Description | Part # |
---|---|
User Manual | 508459 |
Precision Medical Oxygen Sensor with Diverter | 504877 |
Tee Adapter | 505126 |
Extendible Cable | 504937 |
Diverter | 505344 |
Batteries (2 required) | 505124 |
- Orders for replacement parts should include the part number, if available and the model and serial number of the instrument for which the parts are intended.
- 300 Held Drive
- Northampton PA 18067, USA
- www.precisionmedical.com
- T: (+001)610-262-6090
- F: (+001)610-262-6080
- Toll-Free: T:800-272-7285
- F:800-353-1240
- ISO 13485 Certified
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The PM5950 Acc O2 Analyzer is intended for use in the electromagnetic environment specified below. The user of the PM5950 Acc O2 Analyzer should make sure it is used in such an environment.
Immunity Test| IEC 60601 Test Level| Compliance Level|
Electromagnetic Environment – Guidance
---|---|---|---
Conducted RF IEC 61000-4-6| 3 Vrms
150 kHz
to 80 MHz
| 3 Vrms
3 V/m
| Portable and mobile RF communications equipment should be used no closer to
any part of the PM5950 Acc O2 Analyzer, including cables, than the recommended
separation distance calculated from the equation applicable to the frequency
of the transmitter.
Radiated RF IEC 61000-4-3| 3 Vrms
80 MHz
to 2.7GHz
| Recommended separation distance: d=1.2 √P 150 kHz to 80 MHz
d=1.2 √P 80 MHz to 800 MHz d=2.3 √P 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in
meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey a, should be less than the compliance level in
each frequency range b. Interference may occur in the vicinity of equipment marked with the following symbol:
Electrostatic Discharge (ESD)
IEC 61000-4-2
| ±8kV contact
±15kV air
| ±8kV contact
±15kV air
| Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be greater
than 5%.
Electrical Fast Transient/ burst
IEC 61000-4-4
| ± 2 kV for power supply lines
± 1 kV for input/ output lines
| ± 2 kV for power supply lines
± 1 kV for input/ output lines
| Mains power quality should be that of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
| ± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
| ± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
| Mains power quality should be that of a typical commercial or hospital
environment.
Voltage dips, short interruptions and voltage variations on power supply input
lines
IEC 61000-4-11
| <5% UT (>95% dip in UT ) for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT
) for 25 cycles
| <5% UT (>95% dip in UT ) for 0.5 cycle
40% UT (60% dip in UT) for 5 cycles
70% UT (30% dip in UT ) for 25 cycles
| Mains power quality should be that of a typical commercial or hospital
environment. If the user of the [ME EQUIPMENT or ME SYSTEM] requires continued
operation during power mains interruptions, it is recommended that the [ME
EQUIPMENT or ME SYSTEM] be powered from an uninterrupted power supply or a
battery.
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
| 30A/m| 30A/m| Power frequency magnetic fields should be at levels characteristic of a typical location in a typical hospital or home environment.
- At 80 MHz and 800 MHz, the higher frequency range applies.
- These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
- UT is the a.c. main voltage prior to application of the test level.
- a: Field strength from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcasts cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the concentrator is used exceeds the applicable RF compliance level above, the concentrator should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the PM5950 Acc O2 Analyzer.
- b: Over the frequency range 150 kHz to 80 MHz, the field strengths should be less than 3V/m.
- Recommended Separation Distances between Portable and Mobile RF Communications Equipment and This Device:
- This concentrator is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The user of the concentrator can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and this concentrator as recommended below, according to the maximum output power of the communications equipment.
Rated Maximum Power Output of Transmitter (W)| Recommended Separation
Distances for the device (meters)
---|---
150 kHz to 80 MHz d=1.2√P| 80 to 800MHz d=1.2√P| 800 MHz to 2.5 GHz d=2.3√P
0.01| 0.12| 0.12| 0.23
0.1| 0.38| 0.38| 0.73
1| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
- For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
- At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
- The guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
- Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
- The PM5950 Acc O2 Analyzer is intended for use in the electromagnetic environment specified below. The user of the concentrator should assure that it is used in such an environment.
Emissions Test| Compliance| Electromagnetic Environment –
Guidance
---|---|---
RF emissions CISPR 11| Group 1| The PM5950 Acc O2 Analyzer uses RF energy only
for its internal function. Therefore its RF emissions are very low and not
likely to cause any interference in nearby equipment.
RF emissions CISPR 11| Class B| The PM5950 Acc O2 Analyzer is suitable for use
in all establishments, including domestic establishments and those directly
connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic Emissions IEC 61000-3-2| Class A
Voltage fluctuations/flicker emissions IEC
61000-3-3
| Complies
LIMITED WARRANTY AND LIMITATION OF LIABILITY
- Precision Medical, Inc. warrants that the Oxygen Analyzer, (the Product), will be free of defects in workmanship and/or material for the following period:
- Two (2) years from shipment.
- Precision Medical, Inc. is NOT responsible for normal wear and tear, or any neglect or abuse of the product.
- The customer is responsible for the shipping costs of repairs back to Precision Medical, Inc.
- Precision Medical, Inc. will have in its sole and absolute discretion, the final determination if your product is covered under this limited warranty.
- Should any failure to conform to this warranty appear within the applicable period, Precision Medical, Inc. shall, upon written notification thereof (received by Precision
- Medical, Inc. within 30 days of the customer’s discovery of the alleged defect), along with return of the Product at the customer’s expense and substantiation that the goods have been stored, installed, maintained and operated in accordance with Precision Medical, Inc.’s instructions and standard industry practice, and that no modifications, substitutions, or alterations have been made to the goods, correct such defect by repair or replacement (at Precision Medical, Inc.’s option) at its own expense.
Precision Medical, Inc. warrants the 504877 Oxygen Sensor included with the PM5900 Oxygen Analyzer to be free from defects in material and workmanship for a period of sixteen (16) months, from date of shipment. Should any failure to conform to this warranty appear within the applicable period, Precision Medical, Inc. shall, upon written notification thereof (received by Precision Medical, Inc. within 30 days of the customer’s discovery of the alleged defect), along with return of the sensor at the customer’s expense and substantiation that the sensor has been stored, installed, maintained and operated in accordance with Precision Medical, Inc.’s instructions and standard industry practice, and that no modifications, substitutions, or alterations have been made to the sensor, correct such defect by repair or replacement (at Precision Medical, Inc.’s option) at its own expense. Should a sensor require repair or replacement due to said defects, the sensor is warranted only for the remainder of the original sensor warranty period. A sensor shall not be considered defective for failure to function beyond its normal estimated consumption capacity/rates, and this warranty does not cover normal wear due to consumption beyond the sensor’s estimated O2% hours.
ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.
- The representatives of Precision Medical, Inc. or any retailers are not authorized to make oral warranties about the merchandise described in this warranty, and any such statements shall not be relied upon and are not part of the contract for sale.
Thus, this writing is a final, complete and exclusive statement of the terms of the warranty for the products covered by the applicable contract.
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.
- Precision Medical, Inc. shall not under any circumstances be liable for special, incidental or consequential damages including but not limited to lost profits, lost sales, or injury to person or property. Correction of non-conformities as provided above shall constitute fulfillment of all liabilities of Precision Medical, Inc. whether based on contract, negligence, strict tort or otherwise. Precision Medical, Inc. reserves the right to discontinue manufacture of any product or change product materials, designs, or specifications without notice.
- Precision Medical, Inc. reserves the right to correct clerical or typographical errors without penalty.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>