Precision Medical PM50-3HG EasyComp Air Compressor Instruction Manual

June 9, 2024
Precision Medical

Precision Medical - logo Medical PM50-3HG EasyComp Air Compressor
Instruction ManualPrecision Medical PM50-3HG EasyComp Air
Compressor

RECEIVING/INSPECTION

Remove the Precision Medical, Inc. Easy Comp Compressor from packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider.

INTENDED USE

The basic function is to supply a portable compressed air source adequate to drive high volume nebulizers.

READ ALL INSTRUCTIONS BEFORE USING
This manual instructs the end user how to install, operate and safely use the PM50 Easy Comp Compressor. If you do not understand this manual, DO NOT USE the Easy  Comp Compressor and contact your Provider.

EXPLANATION OF ABBREVIATIONS

psi Pounds Per Square Inch
l/min Liters Per Minute

SAFETY INFORMATION – WARNINGS AND CAUTIONS

WARNING| Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
---|---
CAUTION| Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
CAUTION| Used without the safety alert symbol indicates a potentially hazardous situation which, if not avoided, may result in property damage.
| Type BF Equipment
| Class II with respect to protection from  electrical shock

DANGER
Caution must be exercised when using this Easy Comp Compressor where oxygen is being administered.
DO NOT use near any type of flame or flammable/explosive substances.
DO NOT allow the Compressor to exceed Operating or Storage
Temperature specifications.
TO REDUCE THE RISK OF ELECTROCUTION
Always unplug the Compressor immediately after using.
DO NOT use while bathing.
DO NOT place or store the Compressor where it can fall or be pulled into a tub or sink.
DO NOT place or drop into water or other liquid.
DO NOT reach for the Compressor that has fallen into water. Unplug immediately.
DO NOT attempt to repair the Easy Comp Compressor. Any attempt could result in electrical shock.

WARNING
To reduce risk of burns, electrocution, fire, or injury to persons:
Close supervision is necessary when this Easy Comp Compressor is used by, on, or near children or invalids.
Use this Easy Comp Compressor only for its “Intended Use” as described in this manual.
DO NOT use attachments not recommended by Precision Medical, Inc.
DO NOT use the Compressor with any equipment that requires “dry”
air, including ventilators, gas blenders, or flow generators, without a device to properly dehumidify the discharged air of the PM50.
DO NOT use if the Compressor has a damaged cord or plug.
DO NOT use if the Compressor has been dropped or damaged.
DO NOT drop or insert any object into any openings.
DO NOT cover air vents. Restriction of air flow will cause unit to overheat, causing the thermal reset to shut down the Compressor.

SPECIFICATIONS

Flow Output at 50 psig: ≥13 l/min
Maximum Rated Sustained Pressure: 60 psi
Duty Cycle: Continuous
Dimensions: Length: 10.25 in (26.0 cm)

Width: 5.25 in (13.3 cm)
Height: 9.5 in (24.1 cm)
Weight:| 9.7 lbs (4.4 kg)
Shipping Weight:| 11.2 lbs (5.1 kg)
Electrical:| 120 VAC, 60 Hz, 5 Amps
Operating Temperature Range:| 50°F to 104°F (10°C to 40°C)
Storage Temperature Range:| -4°F to 120°F (-20°C to 49°C)
Outlet:| 9/16 – 18 Male Threaded Fitting
Safety Systems:| High Temperature Motor Shutdown
Thermal Protector:| Class B, Thermal Limit Automatic Reset

This Easy Comp Compressor is not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide.
Specifications are subject to change without prior notice.

COMPONENT IDENTIFICATION

CAUTION
Missing or illegible labels must be replaced, contact Precision Medical, Inc.Precision Medical PM50-3HG EasyComp Air Compressor -
COMPONENT

OPERATING INSTRUCTIONS

CAUTION
Inspect the Easy Comp Compressor for visual damage, DO NOT USE if damaged.

  1. Become familiar with the Compressor (see Component Identification).

  2. Place Compressor near a properly grounded electrical receptacle on a table, floor or other horizontal, smooth, hard surface. Allow a minimum of 4 in. (10 cm) clearance on all sides to ensure adequate air flow for cooling of product.

  3. Press down on Compressor until Suction Feet are secure.

  4. Ensure the On/Off Switch is in the “OFF” position.

  5. Plug the Compressor into a receptacle.

  6. Attach appropriate device to the Outlet Fitting, securely, without overtightening.

  7. Press the On/Off Switch to the “ON” position.

  8. Adjust pressure.
    To INCREASE pressure – Turn Knob CLOCKWISE
    To DECREASE pressure – Turn Knob COUNTERCLOCKWISE

  9. Begin procedure.

  10. When procedure is complete, press the On/Off Switch to the “OFF” position.

CLEANING/MAINTENANCE

WARNING
Turn the Easy Comp Compressor off and disconnect from power source before cleaning.

To clean the filter (weekly):

  1. Unplug Compressor.
  2. Pry off filter end cap to check air filter for dust buildup.
  3. If dust buildup occurs, remove air filter and wash in warm, soapy water and rinse.
  4. Absorb excess water from filter with a dry towel.
  5. Return filter and end cap.

Cleaning for Exterior Components:
Use a clean cloth or sponge dampened with soap and water.
Wipe dry with a clean cloth or paper towel.

CAUTION
Excessive dust buildup on filter will reduce performance of Compressor. If this occurs, clean or replace with a new filter.
The air filter does not filter out saline. Treatments using nebulized saline will draw the saline mist back into the Compressor. Saline cause rapid deterioration of the compressor seals and will void the warranty.

RETURNS
Returned products require a Returned Goods Authorization (RGA) number, contact Precision Medical, Inc. All returns must be packaged in sealed containers to prevent  damage.
Precision Medical, Inc. will not be responsible for goods damaged in transit. Refer to Precision Medical, Inc. Return Policy available on the Internet, www.precisionmedical.com.
Manuals available on our website; www.precisionmedical.com
DISPOSAL INSTRUCTIONS
This device may contain electrical components that are hazardous to the environment. DO NOT dispose device into standard trash. Contact your local waste Management  for disposal of Electrical Equipment.
Please Recycle

REPLACEMENT PARTS

Description Part Number
Filter Assembly 1834
Filter (Case -10 ea.) 1841-10

TROUBLESHOOTING

If the PM50 Easy Comp Compressor fails to function, consult the Troubleshooting Guide below.
If the problem cannot be corrected, consult your Provider.

Problem Probable Cause Remedy
Compressor will not operate 1.On/Off switch in “OFF” position

2.Power cord not connected properly
3.Compressor overheated| 1.Push On/Off switch to “ON” position
2.Check power cord
3.Shut Compressor off, wait 15 minutes, Ensure vents are clear
Unable to adjust pressure| 1. Defective regulator| 1. Contact Provider
Overheating| 1. Poor Ventilation| 1. Move Compressor to a cooler, better ventilated area
Low pressure| 1.Loose connection at outlet fitting
2.Improper regulator adjustment| 1.Tighten connection
2.Readjust regulator

LIMITED WARRANTY AND LIMITATION OF LIABILITY

Precision Medical, Inc. warrants that the PM50 Easy Comp Compressor (the Product) will be free of defects in workmanship and/or material for the following period:
One (1) year from shipment
Should any failure to conform to this warranty appear within the applicable period, Precision Medical, Inc. shall, upon written notification thereof and substantiation that the  goods have been stored, installed, maintained and operated in accordance with Precision Medical, Inc.’s instructions and standard industry practice, and that no modifications,  substitutions, or alterations have been made to the goods, correct such defect by suitable repair or replacement at its own expense.
ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.
The representatives of Precision Medical, Inc. or any retailers are not authorized to make oral warranties about the merchandise described in this contract, and any such  statements shall not be relied upon and are not part of the contract for sale. Thus, this writing is a final, complete and exclusive statement of the terms of that contract.
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHER  WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.
Precision Medical, Inc. shall not under any circumstances be liable for special, incidental or consequential damages including but not limited to lost profits, lost sales, or injury to person or property.
Correction of non-conformities as provided above shall constitute fulfillment of all liabilities of Precision Medical, Inc. whether based on contract, negligence, strict tort or  otherwise. Precision Medical, Inc. reserves the right to discontinue manufacture of any product or change product materials, designs, or specifications without notice.
Precision Medical, Inc. reserves the right to correct clerical or typographical errors without penalty.

Precision Medical - logo300 Held Drive
Northampton PA 18067, USA
www.precisionmedical.com
T: (+001) 610-262-6090
F: (+001) 610-262-6080
ISO 13485 Certified
Tell us how we are doing!
www.precisionmedical.com

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Precision Medical User Manuals

Related Manuals