Whitecroft lighting Mirage 3 MRI Recessed Downlighter Healthcare MRI Rooms Installation Guide

October 29, 2023
Whitecroft lighting

Mirage 3 – MRI
Installation Instructions

PK-NPD134-MRI-IL
Revision 4 – 10/01/2023

INSTALLATION DETAILS

  1. Open packing and locate the disposable gloves, this is to stop grease from skin contact getting onto the reflector surface. These must be worn throughout the installation.

  2. Inspect the contents of the box for any damage in transit.

  3. Prepare aperture using appropriate tool, circular fittings are based around standard size hole saws. See Dimensions on last page for cut sizes.

  4. Ceiling thickness is show under Product and Driver Dimensions. See fig 1.Whitecroft lighting Mirage 3 MRI Recessed Downlighter Healthcare MRI Rooms - Fig

  5. Check void dimensions. Ensure 100mm clearance from the down lighter is given in all directions to walls/insulation and MRI Link Connection box. See Fig.2. Insulation  cannot directly cover the back of the down lighter.Whitecroft lighting Mirage 3 MRI Recessed Downlighter Healthcare MRI Rooms - Fig1

  6. Untighten the 3 cable glands fig.4 and remove the brass screw fig.5. Open the box to show the PCB. Cables from external driver into room should be routed through the  IN gland and connected to terminal block beneath. To connect to the next fitting, connect cable to the terminal block through the OUT gland and to the IN gland on  the next fitting.Whitecroft lighting Mirage 3 MRI Recessed Downlighter Healthcare MRI Rooms - Fig2

  7. Wire the products as shown below (typical room schematic).
    Note that the driver is installed outside the MRI room with Filter bridging into the room (supplied by others) then connecting to the MRI link connection boxes that  loop in and loop out to the next fitting. Maximum number of fittings that can be run by driver and maximum circuit length listed below.

  8. Maximum Number of fittings that can be run by driver:***
    Driver| Maximum Low Output| Maximum High Output| Minimum
    Low Output| Minimum* High Output
    ---|---|---|---|---
    M5D100W| 6| 3| 2| 1
    M5D185W| 10| 6| 3| 2
    M5D320W| 18| 10| 4| 3

Recommended maximum circuit lengths:
30m using 100W driver with 1.5mm² cable
20m using 185W driver with 1.5mm² cable
12m using 320W driver with 1.5mm² cable
18m using 320W driver with 2.5mm² cable
Note: These figures are based on default voltage output from the driver. This can be increased on installation and can increase circuit length. Please contact Technical  support  for more information.

  1. Install the product in the ceiling by placing the MRI Link box in the ceiling first then installing the fitting. To install the fitting place the tops of the plastic wings it the cut  aperture then push up until the product is fully secure. See fig.6.
  2. To remove the fitting for maintenance, first ensure clean gloves are used to protect fitting from dirt and grease. Then place a flat bladed screw driver carefully behind the  bezel and ceiling to open up a gap. Use fingers to grip the product and pull down. You can also use your free hand to ease the spring through the aperture to stop damage to the ceiling.

SPECIFICATION

Lamp type:
Output:
Colour:
Materials
Supply
Operating temp
Power
Connection
Ingress:
Protection| LED
Refer to website for lumen outputs
3000k/4000K RA90
Acrylic Diffuser Polycarbonate Reflector (Semi Specular) Graphite  Composite Heatsink
MRI 36V DC
0˚c to 25˚c
Refer to website for wattages
Terminal block/Driver 0.75mm² to 1.5mm², 10-11mm stripped
IP44
---|---

PRODUCT DIMENSIONS

Version| Ø(mm)| W(mm)| H(mm)| L(mm)| Ceiling
Depth(mm)| Cut Aperture
CAر1mm| KG
---|---|---|---|---|---|---|---
C165 MRI Fitting| 173| 173| 106| 173|  1-50| 165| 1
MRI Link Box|      n/a| 55| 59| 107|      n/a|      n/a| 0.09
MRI Driver 100W|      n/a| 68| 38.8| 220|      n/a|     n/a| 1.12
MRI Driver 185W|      n/a| 68| 38.8| 228|      n/a|     n/a| 1.55
MRI Driver 320W|      n/a| 90| 43.8| 252|      n/a|      n/a| 1.88

Whitecroft lighting Mirage 3 MRI Recessed Downlighter Healthcare MRI Rooms -
Fig5

M5D100W (100w Driver)
Whitecroft lighting Mirage 3 MRI Recessed Downlighter Healthcare MRI Rooms -
Fig6

M5D185W (185w Driver)

Whitecroft lighting Mirage 3 MRI Recessed Downlighter Healthcare MRI Rooms -
Fig7

M5D320W (320w Driver)

Whitecroft lighting Mirage 3 MRI Recessed Downlighter Healthcare MRI Rooms -
Fi8

GENERAL INFORMATION

  1. A qualified electrician, in accordance with IEE wiring regulations should carry out connection to mains wiring.
  2. Observe ESD precautions during installation.
  3. Ensure that the rated voltage and frequency requirements are compatible with the available mains supply.
  4. Cleaning of reflectors and lenses should be carried out using clean, soft and lint free cloths. Use of cleaning fluids can mark reflector surfaces.
  5. Do not carry out high voltage insulation test, i.e. 500/1000v this may damage internal components.
  6. For secondary suspension points please refer to fig.7. The holes are design to accommodate steel cable (supplied by others).
  7. Allow sufficient time for product to cool before performing maintenance. Heatsink may be hot.
  8. The light source contained in this luminaire shall only be replaced by the manufacturer or a designated service agent or a suitably qualified person.

Whitecroft Lighting Limited
Burlington Street, Ashton-Under-Lyne, Lancashire OL7 0AX
Telephone +44 (0)161 330 6811Facsimile: +44 (0)161 331 5855
www.whitecroftlighting.com

TECHNICAL SUPPORT
Telephone: 0161 331 5700
E-mail: technical@whitecroftlight.com
http://www.whitecroftlighting.com/