KERN PEJ Precision Balance User Manual

June 9, 2024
KERN

KERN PEJ Precision Balance

PEJ Precision Balance

Features

  • KERN PEJ: Automatic internal adjustment, guarantees high degree of accuracy and makes the balance independent of its location of use

  • KERN PES: Adjusting program CAL for quick setting of the balance accuracy using an external test weight

  • Metal housing: robust and sturdy

  • Weighing with tolerance range (check weighing):
    Input of two upper and two lower limit values through four arrow keys. An audible and visual signal assists with the portioing, dispensing or grading

  • 1 Draught shield standard for models with [d] = 0,001 g, weighing space W×D×H 170×150×100 mm

  • Underfloor weighing: load support with hook on the underside of the balance, standard for models with [d] = 0,001 g, hook not included with the delivery

  • A, B Protective working cover included with delivery

Technical data

  • Fluorescent display, bright with high contrast, digit height 14 mm

  • Dimensions weighing surface, Stainless Steel
    A W×D 140×120 mm
    B W×D 200×200 mm, see larger picture
    C W×D 250×220

  • Overall dimensions W×D×H, without draught shield
    A 220×330×90 mm
    B 220×339×80 mm
    C 260×330×113 mm
    Net weight approx. A 4,0 kg, B 4,4 kg, C 10 kg

  • Permissible ambient temperature 10 °C/30 °C
    Technical data-Logo

  • Net weight approx. A 4,0 kg, B 4,4 kg, C 10 kg

  • Permissible ambient temperature 10 °C/30 °C

Accessories

  • Protective working cover, scope of delivery: 5 items, KERN PES-A04S05
  • KERN PES: Internal rechargeable battery pack, operating time up to 32 h, without backlight, charging time approx. 15 h, KERN PES-A01
  • Loop for underfloor weighing, for models with weighing plate size B, KERN PES-A03
  • Relay output to connect relays, signal lamps, valves etc., 5 outputs for weighing in 3 tolerance ranges, must be ordered at purchase, KERN PES-A02
  • RS-232/Ethernet adapter for connection to an IP-based Ethernet network, KERN YKI-01
  • Minimum weight of sample, smallest weight to be weighed, depending on the required process accuracy, only in combination with a DAkkS calibration certificate, KERN 969-103
  • Equipment qualification: compliant qualification concept which includes the following validation services: Installation Qualification (IQ), Operating Qualification (OQ), Further details see 208
  • Further details, plenty of further accessories and suitable printers see Accessories

Function icon

Model

KERN

| Weighing capacity [Max]

g

| Readability [d] g| Verification value [e]

g

| Minimal load

[Min] g| Linearity

g

| Weighing plate| | Option
---|---|---|---|---|---|---|---|---
Verification| DAkkS Calibr. Certificate

KERN

| |

KERN

|
PES 620-3M| 620| 0,001| –| –| ± 0,003| A| | –| | 963-103|
PES 2200-2M| 2200| 0,01| –| –| ± 0,02| B| | –| | 963-127|
PES 4200-2M| 4200| 0,01| –| –| ± 0,02| B| | –| | 963-127|
PES 6200-2M| 6200| 0,01| –| –| ± 0,03| B| | –| | 963-104|
PES 15000-1M| 15000| 0,1| –| –| ± 0,2| B| | –| | 963-128|
PES 31000-1M| 31000| 0,1| –| –| ± 0,4| C| | –| | 963-128|
Note: For applications that require verification, please order verification at the same time, initial verification at a later date is not possible.
Verification at the factory, we need to know the full address of the location of use.
PEJ 620-3M| 620| 0,001| 0,01| 0,1| ± 0,003| A| | 965-201| | 963-103|
PEJ 2200-2M| 2200| 0,01| 0,1| 0,5| ± 0,02| B| | 965-216| | 963-127|
PEJ 4200-2M| 4200| 0,01| 0,1| 0,5| ± 0,02| B| | 965-216| | 963-127|

  • ONLY WHILE STOCKS LAST !

Pictograms

Internal adjusting:
Quick setting up of the balance‘s accuracy with internal adjusting weight (moto driven )

Adjusting program CAL:
For quick setting up of the balance‘s accuracy.
External adjusting weight required

Easy Touch:
Suitable for the connection, data transmission and control through PC or tablet.

Memory:
Balance memory capacity, e.g. for article data, weighing data, tare weights, PLU etc.

Alibi memory:
Secure, electronic archiving of weighing results, complying with the 2014/31/EU standard.

Data interface RS-232:
To connect the balance to a printer, PC or network
RS-485 data interface:
To connect the balance to a printer, PC or other peripherals. Suitable for data transfer over large distances. Network in bus topology is possible

USB data interface:
To connect the balance to a printer, PC or other peripherals

*Bluetooth data interface:**
To transfer data from the balance to a printer, PC or other peripherals

WiFi data interface:
To transfer data from the balance to a printer, PC or other peripherals

Control outputs
(optocoupler, digital I/O ):
To connect relays, signal lamps, valves, etc.

Analogue interface:
to connect a suitable peripheral device for analogue processing of the measurements

Interface for second balance:
For direct connection of a second balance

Network interface:
For connecting the scale to an Ethernet network
KERN Communication Protocol (KCP):
It is a standardized interface command set for KERN balances and other instruments, which allows retrieving and controlling all relevant parameters and functions of the device. KERN devices featuring KCP are thus easily integrated with computers, industrial controllers and other digital systems
GLP/ISO log:
The balance displays serial number, user ID, weight, date and time, regardless of a printer connection
GLP/ISO log:
With weight, date and time. Only with KERN
printers.
Piece counting:
Reference quantities selectable. Display can be switched from piece to weight
Recipe level A:
The weights of the recipe ingredients can be added together and the total weight of the recipe can be printed out
Recipe level B:
Internal memory for complete recipes with name and target value of the recipe ingredients. User guidance through display
Totalizing level A:
The weights of similar items can be added together and the total can be printed out
Percentage determination:
Determining the deviation in % from the target value (100 % )
Weighing units:
Can be switched to e.g. nonmetric units. See balance model. Please refer to KERN‘s website for more details
Weighing with tolerance range:
(Check weighing) Upper and lower limiting can be programmed individually, e.g. for sorting and dosing. The process is supported by an audible or visual signal, see the relevant model
Hold function:
(Animal weighing program ) When the weighing conditions are un stable, a stable weight is calculated as an average value
**Protection against dust and water splashes IPEX:
The type of protection is shown in the pictogram.
Suspended weighing:
Load support with hook on the underside of the balance
Battery operation:
Ready for battery operation. The battery type is specified for each device
Rechargeable battery pack:
Rechargeable set
Universal plug-in power supply:
with universal input and optional input socket adapters for A ) EU, CH, GB; B ) EU, CH, GB, USA; C ) EU, CH, GB, USA, AUS
Plug-in power supply:
230V/50Hz in standard version for EU, CH.
On request GB, USA or AUS version available
Integrated power supply unit:
Integrated in balance. 230V/50Hz standard EU.
More standards e.g. GB, USA or AUS on request
Weighing principle: Strain gauges:
Electrical resistor on an elastic deforming body
Weighing principle: Tuning fork:
A resonating body is electromagnetically excited, causing it to oscillate
Weighing principle: Electromagnetic force compensation:
Coil inside a permanent magnet. For the most accurate weighings
Weighing principle: Single cell technology:
Advanced version of the force compensation principle with the highest level of precision
Verification possible:
The time required for verification is specified in the pictogram
DAkkS calibration possible (DKD ):
The time required for DAkkS calibration is shown in days in the pictogram
Factory calibration (ISO ):
The time required for Factory calibration is shown in days in the pictogram
Package shipment:
The time required for internal shipping preparations is shown in days in the pictogram
Pallet shipment:**
The time required for internal shipping preparations is shown in days in the pictogram

KERN – Precision is our business

 Your KERN specialist dealer:

To ensure the high precision of your balance KERN offers you the the appropriate test weight in the international OIML error limit classes E1-M3 from 1 mg – 2500 kg. In combination with a DAkkS calibration certificate the best pre-requisite for proper balance calibration. The KERN DAkkS calibration laboratory today is one of the most modern and bestequipped DAkkS calibration laboratories for balances, test weights and force-measurement in Europe. Thanks to the high level of automation, we can carry out DAkkS calibration of balances, test weights and force-measuring devices 24 hours a day, 7 days a week.

Range of services:
  • DAkkS calibration of balances with a maximum load of up to 50 t
  • DAkkS calibration of weights in the range of 1 mg – 2500 kg
  • Volume determination and measuring of magnetic susceptibility (magnetic characteristics ) for test weights
  • Database supported management of checking equipment and reminder service · Calibration of force-measuring devices
  • DAkkS calibration certificates in the following languages DE, EN, FR, IT, ES, NL, PL
  • Conformity evaluation and reverification of balances and test weights

         KERN & SOHN GmbH · Ziegelei 1 · 72336 Balingen ·
Germany · Tel. +49 7433 9933 - 0
www.kern-sohn.com · info@kern- sohn.comOrder

Read User Manual Online (PDF format)