Laserliner 082.035.1, 082.035.2 ThermoMaster User Manual

June 8, 2024
Laserliner

Innovation in Tools thermometer
User Manual
Laserliner 082 035A ThermoMaster Thermometer

082.035.1, 082.035.2 ThermoMaster

ThermoSensor Air / 082.035.1Laserliner 082 035A ThermoMaster
Thermometer - ThermoSensor ThermoSensor Tip / 082.035.2

Laserliner 082 035A ThermoMaster Thermometer - ThermoSensor
Tip

**!**  Read the operating instructions and the enclosed brochure „Guarantee and additional notices“ completely. Follow the instructions they contain. Safely keep these documents for future reference.

Function/Application
The digital thermometer is used to measure temperature and temperature differences with the aid of exchangeable type K thermocouples/sensors. The digital thermometer is predominantly used for temperature measurement in laboratories and in industrial applications. With the aid of the MAX function, it is possible to determine limit temperature infringements in extended series of measurements. The 1/4“ tripod connection provides a flexible locking facility for stationary setups.

  • Only hold the test prods by the handles. Do not touch the test prods, they can become very hot during the measuring procedure.
  • Wherever possible the test prods should not be operated with an external voltage supply; if this is unavoidable pay particular attention to the maximum voltage with respect to earth potential and the maximum voltage between the two inputs T1 and T2. To avoid damaging the device, the external voltage supply must not be higher than 24 V AC or 60 V DC. Please note that the test prods are electrically conductive and may irreparably damage the test piece when measuring in connection with an external voltage supply.
  • Use under high voltages or under high electromagnetic alternating fields can have an influence on the measuring accuracy.
  • The device must only be used in accordance with its intended purpose and within the scope of the specifi cations.
  • If you are working with voltages higher than 24 V AC/60 V DC, exercise extreme caution. Touching the electrical conductors at such voltages poses a risk of life-threatening electric shocks.
  • If the device comes into contact with moisture or another conductive residue, work must not be carried out under voltage. When using the device outdoors, make sure that the weather conditions are appropriate and/or that suitable protection measures are taken.
  • The measuring tools and accessories are not toys. Keep out of reach of children.

Symbols
Warning about hazardous electrical voltage: Unprotected live components inside the device housing are capable of posing a risk of electric shock.
Danger area warning
 Protection class II: The test device has reinforced or double insulation.
 Earth potential
**!**  Important notes. Must be observed.

General information

  • If the thermometer is subject to large fluctuations in the ambient temperature, after the temperature has stabilized, wait 20 minutes before performing a measurement.
  • Always ensure adequate thermal coupling at the measuring point in order to avoid measuring errors caused by temperature loss.
  • Please note that all thermometers with a contact sensor infl uence the measurement and their thermal capacity can reduce the actual temperature. More thermal energy should therefore be applied to the thermocouple than it can dissipate.
  • OL is shown in the display when no measuring sensor is connected.
  • The thermometer shows OL when the measured temperature is outside the measuring range.
  • Only use the correct types of thermocouples plugged in at the K-type connection; the wrong type of thermocouple can cause considerable measuring errors.
  • In addition to being subject to aging, a thermocouple is also greatly dependent on the operating conditions and should therefore be checked regularly.
  •  High pressure or mechanical deformation can change the grid structure and therefore have an  infl uence on the thermoelectric voltage output.

Laserliner 082 035A ThermoMaster Thermometer - fig

2 Connection of K-type thermocouplesLaserliner 082 035A ThermoMaster
Thermometer - thermocouples ! Pay attention to the polarity information on the thermocouple as well as on the device connection.

Laserliner 082 035A ThermoMaster Thermometer -
thermocou

Laserliner 082 035A ThermoMaster Thermometer - LC
display

1 LC display, 3.5 segments, max. 1999
2 ON/OFF
3 Degrees Celsius
4 Hold current measured value
5 Degrees Fahrenheit
6 Temperature K-type input T2
7 Differential temperature K-type input T1-T2| 8 Temperature K-type input T1
9 Measured value resolution
10 MAX value
11 Display illumination
12 1/4“ tripod connection
13 Battery compartment
14 K-type input T2
15 K-type input T1
---|---

a Units:
Degrees Celsius/Fahrenheit
b Measured value
c Negative measured value
d MAX value
e Hold current measured value
f Temperature K-type input T1
g Temperature K-type input T2
h Differential temperature Ktype input T1-T2

3 ON/OFFLaserliner 082 035A ThermoMaster Thermometer -
display This display appears when no corresponding thermocouple is connected or the measuring range has been exceeded.
4 Temperature measurements (T1, T2, T1-T2)
The temperature T1 or T2 is measured and displayed by pressing buttons „T1“ or „T2“. A thermocouple (type K) must be plugged in at connection T1 or T2. The differential temperature can be determined by pressing the button „T1-T2“ when thermocouples are plugged in at both connections.

The maximum temperature in an extended series of measurements is determined by pressing the „MAX“ button. The HOLD function shows the last measurement or measured value in the display.
6 Measured value resolution (0.1 °C/1.0 °C)
This function changes the measurement resolution is changed between 0.1 °C and 1.0 °C. The resolution of 0.1 °C is particularly useful for measurements ≤100 °C where it is necessary to determine the temperature exactly. A  resolution of 1.0 °C is recommended for measuring higher temperatures.
**!**  The device should be checked regularly to ensure it is functioning correctly. Calibration intervals of 1 year are recommended.

Technical data Subject to technical alterations 03.11.
Measuring range, measuring system -50 °C … 1300 °C

Accuracy, measuring system
-50 °C … 0 °C
0 °C … 1000 °C
1000 °C … 1300 °C| ±2 °C
±(0.5 % of displayed value +1 °C)
±(0.8 % of displayed value +1 °C)
Resolution| 0.1 °C/1 °C selectable
Unit of measurement| °C/°F selectable
Measuring range, thermocouple
ThermoProbe K
ThermoSensor Air
ThermoSensor Tip| -50 °C … 800 °C
-50 °C … 800 °C
-50 °C … 800 °C
Accuracy, thermocouple
ThermoProbe K
ThermoSensor Air
ThermoSensor Tip| ±2.5 K (-50 °C … 200 °C)
±2.5 K (-50 °C … 333 °C)
±0.75 % (333 °C … 800 °C)
±2.5 K (-50 °C … 333 °C)
±0.75 % (333 °C … 1200 °C)
Sensor (EN 60584-2)| Replaceable sensor, type K
Display| LC display, 3.5 segments, (max. 1999)
Power supply| 1x 9 V alkaline battery (NEDA 1604, IEC 6F22)
Operating temperature| 0 °C … 50 °C
Storage temperature| -20 °C … 60 °C, 10 % … 70 % rH (non-condensating)
Dimensions| 162 mm x 76 mm x 38.5 mm
Weight (without battery)| 0.21 kg

EU directives and disposal
 This device complies with all necessary standards for the free movement of goods within the EU.
This product is an electric device and must be collected separately for disposal according to the European Directive on waste electrical and electronic equipment.
Further safety and supplementary notices at: www.laserliner.com/infoLaserliner 082 035A
ThermoMaster Thermometer - ThermoMaster

SERVICE
Umarex GmbH & Co KG
– Laserliner –
Möhnestraße 149, 59755 Arnsberg, Germany
Tel.: +49 2932 638-300, Fax: +49 2932 638-333
[email protected]
082.035A / 8.082.96.38.1 / Rev.0311

Umarex GmbH & Co KG
Donnerfeld 2
59757 Arnsberg, Germany
Tel.: +49 2932 638-300, Fax: -333
www.laserliner.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Laserliner User Manuals

Related Manuals