MACHEREY-NAGEL 920182 Visocolor HE Phosphate Colorimetric Test Kit Instruction Manual
- June 4, 2024
- MACHEREY-NAGEL
Table of Contents
Phosphate
High sensitivity test kit for the determination in the range of
0.05–1.0 mg/L P
Method:
Phosphomolybdenum blue Contents of test kit (*refill pack): sufficient for 300 tests
100 mL P-1
100 mL P-2
2 round glass tubes with screw caps
1 comparator block
1 color comparison disc Phosphate
Hazard warning:
P-1 contains sulfuric acid 15–30 %, P-2 contains sodium disulfite 10–25 %. H314 Causes severe skin burns and eye damage. P260sh, P280sh, P303+361+353, P305+351+338, P310 Do not breathe dust/vapors. Wear protective gloves/eye protection. IF ON SKIN (or hair): Take off immediately all contaminated clothing. Rinse skin with water [or shower]. IF IN EYES: Rinse cautiously with water for several minutes. Remove contact lenses, if present, and easy to do. Continue rinsing. Immediately call a POISON CENTER/doctor.
Procedure:
- Insert color comparison disc (see illustration).
- Open both round glass tubes, rinse several times with the water sample and fill up to the mark with the sample.
- Add 6 drops of P-1 to the right glass tube, close, and mix.
- Add 6 drops of P-2 to the right glass tube, close, and mix. Wait 10 min.
- Reading: Turn the color disc until both colors match by transmitted light from above. Read test results from the mark on the front side of the comparator (see illustration). Intermediate values can be estimated.
- After use clean both round glass tubes thoroughly and close.
mg/L P | mg/L PO₄³– | mg/L P₂O₅ | mmol/m3 |
---|
0.05
0.10
0.15
0.20
0.3
0.4
0.6
0.8
1.0| 0.15
0.31
0.46
0.61
0.9
1.2
1.8
2.5
3.1| 0.11
0.23
0.34
0.46
0.7
0.9
1.4
1.8
2.3| 1.6
3.2
4.8
6.5
10
13
19
26
32
The method can also be applied for the analysis of seawater.
Disposing of the samples:
The used analysis specimens can be flushed down the drain with tap water and channeled off to the local sewage treatment works.
Interferences:
Oxidizing agents in larger amounts inhibit the formation of the blue color complex. They must be destroyed prior to the determination. The following ions will not interfere:
≤ 1 mg/L Si(IV)
≤ 2 mg/L H2S
≤ 10 mg/L of heavy metals
The temperature of the water sample should be between 18 and 30 °C. Outside
this range, the rate of the reaction decreases by finding less phosphate than
actually present.
MACHEREY-NAGEL GmbH & Co. KG · Neumann-Neander-Str. 6–8 · 52355 Düren ·
Germany
Tel.: +49 24 21 969–0 · Fax: +49 24 21 969–199 · info@mn-net.com · www.mn-net.com
PD17576 / A063329 / 920082 / 1080.3