MACHEREY-NAGEL Alkalinity TA Visocolor ECO Test Kit Instructions

June 4, 2024
MACHEREY-NAGEL

Alkalinity TA Visocolor ECO Test Kit
Instruction Manual

Test kit for the photometric determination of total alkalinity (acid capacity)

Method:
Alkalinity (acid capacity / acid binding capacity) describes the capability of water to neutralize acids with a pH range of up to 4.3.
The capability results from the sum of all bases (hydroxide ions) and buffers (carbonates, phosphates, etc.).
The photometric determination of alkalinity uses bromophenol blue as an indicator.

Measurement range

0.4–17.5 °e
5–250 mg/L CaCO3

Contents:

sufficient for 100 tests
23 mL TA- 1
2 x 50 NANOFIX TA-2
1 syringe 5 mL
1 syringe 1 mL
2 tips for syringe 1 mL
1 instruction leaflet

Hazard warning:
Information regarding safety can be found on the box label and in the safety data sheet. You can download the SDS from www.mn-net.com/SDS.

Procedure

Required equipment: Test tubes 16 mm OD (REF 91680)

  1. Rinse the test tube several times with the test sample and fill it with a 5 mL sample (syringe 5 mL).
  2. Place the test tube in the photometer and measure ZERO.
  3. Add 0.2 mL TA-1 (syringe 1 mL + syringe tip)
  4. Add 1 NANOFIX TA-2, close, and shake vigorously.
  5. Clean the outside of the test tube and measure after 2 min.
  6. Rinse and close the tubes after use. The method can be applied also to the analysis of seawater.
    To achieve more accurate results, the usage of pipettes is recommended.

Measurement:
see VISOCOLOR® ECO test instructions for compact photometers PF-3 / PF-12 Plus

Conversion table

°d °e °f mg/L CaCO3 mmol/L H+ gpg
1 1.3 1.8 18 0.36 1
2 2.5 3.6 36 0.72 2
3 3.8 5.4 54 1.08 3
4 5 7.1 71 1.42 4
5 6.3 8.9 89 1.78 5
6 7.5 10.7 107 2.14 6
7 8.8 12.5 125 2.5 7
8 10 14.3 143 2.86 8
9 11.3 16.1 161 3.22 9
10 12.5 17.8 178 3.56 10

Sample disposal:
Information regarding disposal can be found in the safety data sheet.
You can download the SDS from www.mn-net.com/SDS.
Storage:
Store the test kit in a cool (< 25 °C) and dry place.

MACHEREY-NAGEL GmbH & Co. KG · Neumann-Neander-Str. 6–8 · 52355 Düren · Germany
Tel.: +49 24 21 969-0
Fax: +49 24 21 969-199
[email protected]
www.mn-net.com