IMPORTANT SAFEGUARDS
Do not use Comfort Glide Air Blower in the presence of flammable anesthetics.
Precaution – Comfort Glide Repositioning Sheet:
- Caregivers must verify that all caster brakes on the bed have been engaged prior to patient movement.
- For safety, always use at least two people during patient movement.
- Never leave patient unattended on an inflated device.
- Use this product only for its intended use as described in this manual. Do not use attachments not recommended by the manufacturer.
- The Comfort Glide Repositioning sheet is designed to be used with 1 Air Blower.
Precaution – Comfort Glide Air Blower – To reduce the risk of burns,
electrocution, fire or injury to persons:
- Use this product only for its intended use as described in this manual. Do not use attachments not recommended by the manufacturer.
- Never operate this product if it has a damaged cord or plug, if it is not working properly, if it is has been dropped or damaged, or ropped into water.
- Keep the cord away from heated surfaces.
- Never drop or insert any object into any openings.
- Route the power cord in a manner to ensure it is free from hazards.
- Disconnect from power supply circuit before opening.
- Keep Comfort Glide Air Blower away from sources of liquid and open flames.
- The Air Blower needs to be used with a stable power supply or in conjunction with a UPS
THE PURPOSE OF THIS MANUAL
This operation manual is mainly focused on the set-up, cleaning and routine maintenance of the Comfort Glide Air Blower. We recommend keeping this manual handy to answer questions related to the system.
PRODUCT DESCRIPTION – INTENDED USE
The Comfort Glide Air Blower is a component of the Comfort Glide air system
designed to help clinicians safely reposition and move patients.
The system has been designed to reduce risks and make patient repositioning
and transfers easier and safer.
GENERAL SAFETY
It is important to read the information in this user manual before you use
your Comfort Glide Air Blower.
Please follow the guidelines below for your added safety and maintaining
system performance.
Avoid exposing the Air Blower to flammable anesthetics.
DEVICE FEATURE
Part Identification:
INFLATING COMFORT GLIDE AIR SHEET
a. Patient should be in a supine position with the bed at a 0 degree incline
b. Place Comfort Glide Air Sheet underneath the patient.
c. Place the Comfort Glide Air Blower directly on flat ground or use the
carabiner to hang the unit on the bed frame.
d. Plug the power cord into the nearest standard wall outlet
e. Connect the air hose to one side of the Comfort Glide Air Sheet air intake
opening and securely fasten into place. Choose the air input on the side that
will permit free movement of the hose.
CAUTION: It is important to routinely inspect power cable to ensure it is not obstructed or causing a tripping hazard. Check the power cable is not under strain or damaged.
f. If using for lateral transfer, ensure that transfer surfaces are touching
and lock all wheels.
g. Turn on the green power switch and the air blower will inflate the Comfort
Glide Air sheet. When the Comfort Glide Air sheet is fully inflated, the sheet
is ready to use and the patient can be
repositioned.
h. Ensure patient is centered on the surface prior to deflation.
i. Turn off Air Blower and raise the bed or stretcher rails.
CAUTION: Never leave the patient unattended with the Air Blower on.
MAINTENANCE & TROUBLESHOOTING
No daily maintenance is required. It is intended this equipment should only be serviced by qualified and authorized technical personnel. Please contact your Medline representative if the below troubleshooting solutions do not work
Fault/Problem Description | Possible Root Cause | Solution |
---|---|---|
The Air Blower does not turn on. | The Air Blower is not receiving power. |
Check power cord plug is securely connected and fits snugly into the outlet.
Secure plug connection and turn on Air Blower power switch.
Blown fuses in pump and plug.| Check for blown fuses in pump and plug. Replace
fuse if blown.
Power outlet is faulty.| Try another power outlet.
Defective Air Blower Unit.| Replace with new device. Contact your Medline
representative.
The Comfort Glide Repositioning Sheet cannot be inflated.| Hose not properly
connected.| Verify that the hose is properly con- nected to the Air Blower.
The hose has 2 prongs that snap into place to connect it to the Air Blower
unit.
Air Flow is not sufficient.| 1. Check hose connection points and re-
connect if needed.
2. Check hose for tears or kinking.
3. Check for punctures or rips in the Comfort Glide Air Sheet.
4. Check the air filter for re- strictions. Clean/replace as necessary.
WARNING: Disassembly or attempts to make repairs may void
Manufacturer’s Warranty. Please consult the Statement of Warranty in this
Manual for more information.
If the problem is not resolved, please contact your Medline representative
CLEANING & DISINFECTION PROTOCOL
-
Please follow your facility’s guidelines for cleaning and disinfecting the Comfort Glide Air Blower. Ensure that the power unit is off.
-
Unplug the power cord from the wall outlet.
-
Examine the surface of the Air Blower for visible dirt and other substances.
-
Refer to your Bio-Medical Engineering or Infection Prevention
Department for your facility’s protocol for cleaning and disinfecting. -
Perform the following steps to clean the power unit and hose fittings:
– Wipe all controls, chassis and hose fittings with a damp cloth and a mild detergent.
– Air dry all treated surfaces.
GENERAL DISASSEMBLY OF PARTS
Before disassembling any parts, ensure that the power unit is off and unplug the power cord from the wall outlet.
-
Changing the air hose:
a. Use a flathead screwdriver to disconnect the prongs on the hose from the blower unit.
b. Push the hose out to replace with a new one.
-
2. Changing filter:
a. Unscrew 3 screws on the bottom cover.
b. Open the filter cover and identify the filter.
c. Replace the filter and assemble the parts in reverse order
WARNING: Switch off the electrical supply to the pump and disconnect the power
cable from the electrical outlet before cleaning and inspection.
All equipment should be inspected. Any item that is visibly dirty should be
properly cleaned or removed. It is recommended that the system is cleaned
regularly and after each patient use.
SPECIFICATION
Control unit Specification
Model Name | Comfort Glide Air Blower |
---|---|
Size in CM (L x W x H) | 32.5 X 18 X 24.8 |
Weight | 5 kgs |
Air Hose | 37mm dia x 1.5m length |
Min/Max Pressure | 20 ~ 30 mmHg ± 5mmHg |
Flow Rate | ≥ 2500 L/min |
Rated Voltage | AC110 – 120V /60Hz |
Max Current | 10A |
Fuse rating | F12AH 250V |
Protection Type | Class II Type B [](https://manuals.plus/wp- |
content/uploads/2022/12/MEDLINE-MSC60090-Comfort-GLIDE-Air-Blower-User-
Manual_8.jpg)
Environment (Temperature)| Operation: 15 oC to 35 oC (59 oF to 95 oF)
Storage: 5 oC to 60 oC (41 oF to 140 oF)
Environment (Humidity)| Operation: 30% to 75% non-condensing
Storage: 30% to 90% non-condensing
Filter Air| 5 microns
Operation Atmospheric Pressure Range| 700 hPa to 1060 hPa
Operation altitude| -1017 feet to 9,843 feet
Test Standard| IEC60601-1, IEC60601-1-2
WARNING: Medical electrical equipment needs special precautions regarding EMC
and needs to be installed according to the EMC information provided. Careful
consideration of this information is
essential when stacking or collocating equipment and when routing cables and
accessories.
WARNING: RF mobile communications equipment can effect medical electrical
equipment.
ELECTROMAGNETIC COMPATIBILITY
Electric devices may interact due to electro-magnetic radiation. We recommend a safety distance of at least 1 meter, especially for sensitive equipment. Upon request, we can provide a table with more detailed electromagnetic information.
Recommended separation distances between
portable and mobile RF communications equipment and the Comfort Glide Air Blower
The Comfort Glide Air Blower is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or
the user of the Comfort Glide Air Blower can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Comfort Glide Air Blower
as rec- ommended below, according to the maximum output power of the
communications equipment.
Rated maximum output power of transmitter
W
| Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d = 1,2
| 80 MHz to 800 MHz
d = 1,2
| 800 MHz to 2,7 GHz
d = 2,3
0,01| 0,12| 0,12| 0,23
0,1| 0,38| 0,38| 0,73
1| 1,2| 1,2| 2,3
10| 3,8| 3,8| 7,3
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where p is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies. NOTE 2: These guidelines may not apply in all
situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Manufacturer’s declaration – electromagnetic emissions
The Comfort Glide Air Blower is intended for use in the electromagnetic environment (for profes- sional healthcare) specified below.
The customer or the user of the Comfort Glide Air Blower should assure that it is used in such an environment.
Emissions test| Compliance| Electromagnetic environment-
guidance (for professional healthcare environment)
RF emissions CISPR 11| Group 1| The Comfort Glide Air Blower uses RF
energy only for its internal function.
Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11| Class A| The Comfort Glide Air Blower is suitable
for use in all establishments other than domestic establishments and those
directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2| N/A
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
| N/A
Manufacturer’s declaration – electromagnetic immunity
The Comfort Glide Air Blower is intended for use in the electromagnetic environment (for professional healthcare) specified below.
The customer or the user of the Comfort Glide Air Blower should assure that it is used in such an environment.
Immunity test| IEC 60601
test level
| Compliance level| Electromagnetic environment
-guidance (for professional healthcare environment)
Electrostatic discharge (ESD) IEC 61000-4-2| Contact:±8 kV Air±2 kV,±4 kV,
±8 kV,±15 kV
| Contact:±8 kV Air±2 kV,±4 kV,
±8 kV,±15 kV
| Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30%
Electrical fast transient/burst IEC 61000-4-4| ± 2 kV for power supply lines
± 1 kV for input/ output lines
| ± 2kV for power supply lines Not applicable| Mains power quality should be
that of a typical professional healthcare environment.
Surge
IEC 61000-4-5
| ± 0.5kV, ± 1kV
line(s) to line(s)
± 0.5kV, ± 1kV,
± 2kV line(s) to earth
| ± 0.5kV, ± 1kV
line(s) to line(s) Not applicable
| Mains power quality should be that of a typical professional healthcare
environment.
Voltage Dips, short interruptions and voltage variations on power supply input
lines
IEC 61000-4-11
| Voltage dips:
0 % UT; 0,5 cycle
0 % UT; 1 cycle
70 % UT; 25/30 cycles
Voltage interruptions: 0 % UT;
250/300 cycle
| Voltage dips:
0 % UT; 0,5 cycle
0 % UT; 1 cycle
70 % UT; 30 cycles
Voltage interruptions: 0 % UT; 300 cycle
| Mains power quality should be that of a typical professional healthcare
environment. If the user of the Comfort Glide Air Blower requires
continued operation during power mains interruptions, it is recommended that
the Comfort Glide Air Blower be powered from an uninterruptible power
supply or a battery.
Power frequency (50, 60 Hz)
magnetic field IEC 61000-4-8
| 30 A/m
50 Hz or 60 Hz
| 30 A/m
60 Hz
| The Comfort Glide Air Blower power frequency magnetic fields should be
at levels characteristic of a typical location in a typical profes- sional
healthcare environment.
---|---|---|---
NOTE: UT is the a.c. mains voltage prior to application of the test
level.
Manufacturer’s declaration-electromagnetic immunity Test specifications
for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
The Comfort Glide Air Blower is intended for use in the electromagnetic environment (for professional healthcare) specified below.
The customer or the user of the Comfort Glide Air Blower should assure that it is used in such an environment.
Test frequency (MHz)
|
Band a) (MHz)
|
Service a)
|
Modulation b)
|
Maximum power (W)
|
Distance
| IMMUNITY TEST LEVEL
(V/m)
| Compliance LEVEL
(V/m)
(for professional healthcare)
385
|
380-390
|
TETRA 400
| Pulse modulation b) 18 Hz|
1 , 8
|
0 , 3
|
27
|
27
450
|
430-470
| GMRS 460,
FRS 460
| FM c)±5 kHz deviation 1 kHz sine|
2
|
0 , 3
|
28
|
28
710|
704-787
|
LTE Band 13,
17
|
Pulse modulation b) 217 Hz
|
0 , 2
|
0 , 3
|
9
|
9
745
780
810|
800-960
| GSM 800/900, TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
|
Pulse modulation b) 18 Hz
|
2
|
0 , 3
|
28
|
28
870
930
1720|
1700-
1900
| GSM 1800;
CDMA 1900;
GSM 1900; DECT;
LTE Band 1, 3,
4, 25; UMTS
|
Pulse modulation b) 217 Hz
|
2
|
0 , 3
|
28
|
28
1845
1970
2450
|
2400-
2570
| Bluetooth, WLAN,
802.11 b/g/n, RFID 2450, LTE Band 7
|
Pulse modulation b) 217 Hz
|
2
|
0 , 3
|
28
|
28
5240|
5100-
5800
|
WLAN 802.11
a/n
|
Pulse modulation b) 217 Hz
|
0 , 2
|
0 , 3
|
9
|
9
5500
5785
NOTE : If necessary to achieve the IMMUNITY TEST LEVEL, the distance
between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be
reduced to 1 m. The 1 m test distance is permitted by
IEC 61000-4-3.
a. For some services, only the uplink frequencies are included.
b. The carrier shall be modulated using a 50 % duty cycle square wave signal.
c. As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it would be worst case.
Manufacturer’s declaration – electromagnetic immunity
The Comfort Glide Air Blower is intended for use in the electromagnetic
environment (for professional healthcare) specified below. The customer or the
user of the Comfort Glide Air Blower should assure that it is used in such
an environment.
Immunity test| IEC 60601 test level| Compliance
level| Electromagnetic environment-guidance (for professional
healthcare environment)
Conducted RF
IEC 61000-
4-6
| 3 Vrms:
0,15 MHz – 80 MHz
6 Vrms:
in ISM bands between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
| 3 Vrms:
0,15 MHz – 80 MHz
6 Vrms:
in ISM bands between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
| Portable and mobile RF communications equipment should be used no closer to
any part of the Comfort Glide Air Blower including cables, than the
recommended separation distance calculated from the equation applicable to the
frequency of the transmitter.
Radiated RF
IEC 61000-
4-3
| 3 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
| 3 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
| Recommended separation distance:
d = 1,2
d = 1,2 80 MHz to 800 MHz
d = 2,3 800 MHz to 2,7 GHz
| | | Where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer and d is the recommended
separation distance in metres (m).
| | | Interference may occur in the vicinity of equipment marked with the
following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
WASTE DISPOSAL
This Product has been supplied from an environmentally aware manufacturer that
complies with the Waste Electrical and Electronic Equipment Directive (WEEE).
This product may contain substances that could be harmful to the environment
if disposed of in places (landfills) that are not appropriate according local
regulations. Please be environmentally responsible and recycle this product
through your recycling facility at its end of life.
STORAGE AND CARE
Comfort Glide Air Blower:
- Check the power cord and plug for abrasions or excessive
- Plug in the unit and verify air flows from the units hose connection ports
- Place in box for storage
It is recommended the following guidelines are used whenever this sys- tem is
being stored or transported another location:
Temperature limitations: 5 oC to 60 oC (41 oF to 140 oF) Relative Humidity:
30% ~90%
EXPECTED SERVICE LIFE
The Comfort Glide Air Blower has an expected service life of two years. Replace the air filter every 6 months. Do NOT use unapproved accessories or attempt to modify, disassemble or otherwise misuse the device.
WARRANTY
- The manufacturer warrants this equipment to be free from defects in material and workmanship for 2 years from the date of
- At Manufacturer’s discretion, we agree to service, repair or replace any equipment or part found to be defective at no
- This warranty excludes equipment damaged through shipping, tampering, improper maintenance, carelessness, accidents, negligence, misuse, or which has been altered, repaired or dismantled other than with the manufacture’s written authorization and by its approved procedures and by properly qualified
- In no event shall the manufacturer be liable for any direct, indirect or consequential damage or loss resulting from the use of
Medline Industries, LP, Three Lakes Drive, Northfield, IL 60093 USA
Medline United States
1-800-MEDLINE (633-5463)
www.medline.com |
info@medline.com
Read More About This Manual & Download PDF:
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>