tempmate M1 Multiple Use PDF Temperature Data Logger User Manual
- June 3, 2024
- tempmate
Table of Contents
- tempmate M1 Multiple Use PDF Temperature Data Logger
- Technical data
- Device operation instruction
- Initial use
- Key function description
- Battery management
- LCD display notice
- Appendix 1 – working status description
- Appendix 2 – other LCD display
- Appendix 3 – LCD page display
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
tempmate M1 Multiple Use PDF Temperature Data Logger
This data logger is mainly used to detect the temperature of food, pharmaceuticals, chemicals and other products during transportation or storage. The main features of this product: multiple use, automatically generated PDF report, high waterproof level, battery exchangeable.
Technical data
Technical Specifications
Temperature sensor | NTC internal and external optional |
---|---|
Measuring range | –30 °C to +70 °C |
Accuracy | ±0.5 °C (at –20 °C to + 40 °C) |
Resolution | 0.1 °C |
Data storage | 32,000 values |
Display | Multifunction LCD |
Start setting
| Manually by pressing a button or automatically on programmed start time
Recording time
| Freely programmable by customer/ up to 12 months
Interval| 10s. to 11h. 59m.
- Alarm settings Adjustable up to 5 alarm limits
- Alarm type Single alarm or cumulative
- Battery CR2032 / replaceable by customer
- Dimensions 79 mm x 33 mm x 14 mm (L x W x D)
- Weight 25 g
- Protection class IP67
- System Requirements PDF Reader
- Certification 12830, calibration certificate, CE, RoHS
- Software TempBase Lite 1.0 software / free download
- Interface to PC Integrated USB port
- Automatic PDF Reporting Yes
Device operation instruction
- Install tempbase.exe software (https://www.tempmate.com/de/download/), insert the tempmate.®-M1 logger to the computer via USB port, finish USB driver installation directly.
- Open tempbase.® data management software, after connection the logger with your computer, the data information will be automatically uploaded. Then you can click the “Logger Setting” button to enter the parameter configuration interface and configure the parameters according to specific application.
- After finishing configuration, click the “Save“ button to save the parameter setting, then it will open a window “Parameter Configuration completed”, click OK and close the interface.
Initial use
Configuration operation
Open tempbase.exe software, after connecting the tempmate.®-M1 logger with the
computer, the data information will be automatically uploaded. Then you can
click the “LoggerSetting” button to enter parameter configuration interface and
configure the parameters according to the specific application. After finishing
the configuration, click the “Save” button to save the parameter setting, then
it will open a window “Parameter Configuration completed”, click OK and close
the interface.
Logger start operation
The tempmate.®-M1 supports three start modes (manual start, start right now,
timing start), the specific start mode is defined by the parameter setting.
Manual start: Press the left key for 4 seconds to start the logger.
ATTENTION: The command done by the button press, will be accepted by the
device if the display has been activated by briefly pressing the left button
beforehand.
Start right now: Immediately start after the tempmate.®-M1 is
disconnected with the computer.
Timing start: tempmate.®-M1 starts when the set start time is reached
(Note: The set start time needs to be at least one minute).
- For one recording trip, the device can support a maximum of 10 marks.
- Under status of pause or sensor disconnected status (when external sensor is configured), the MARK operation is disabled.
Stop operation
M1 supports two stop modes(stop when reaches the max. record capacity, manual
stop), and the specific stop mode is determined by parameter setting.
Stop when reaches the max. record capacity: When record capacity reaches the
max. record capacity, the logger will stop automatically.
Manual stop: The device only stops when it is manually stopped except if the
battery is under 5%. If the recorded data reaches to its max. capacity, the
data will be overwritten (depends on the setting).
ATTENTION: The command done by the button press, will be accepted by the
device if the display has been activated by briefly pressing the left button
beforehand.
Note:
During the status of data overwriting (ring memory), MARK operation will not
be cleared. Saved marks still exist. The max. MARK events are still “10 times”
and every marked data will be saved without clearing during the transport
cycle.
Viewing operation
During tempmate.®-M1 is in recording or stopping status, insert the logger to
the computer, the data can be viewed by the tempbase.® software or the
generated PDF report in the USB device.
PDF reports are different if there is an alarm setting:
- If no alarm setting is programmed, there is no alarm information column and in data table, no alarm color marking, and at the left upper corner, it displays PDF in the black rectangle.
- If the alarm is set as upper/lower alarm, it has an alarm information column, and it has three lines of informations: upper alarm information, standard zone information, lower alarm information. The upper alarm recording data is displayed in red, and the lower alarm data is displayed in blue. In the left upper corner, if alarm occurs, the background of the rectangle is red and displays ALARM inside. If no alarm occurs, the background of the rectangle is green and displays OK inside.
- If the alarm is set as multiple zone alarm in the PDF alarm information column, it could have max. six lines: upper 3, upper 2,upper 1, standard zone; lower 1, lower 2 the upper alarm recording data is displayed in red, and the lower alarm data is displayed in blue. In the left upper corner, if alarm occurs, the background of rectangle is red and displays ALARM inside. If no alarm occurs, the background of rectangle is green and displays OK inside.
Note:
- Under all alarm modes, if data table zone for marked data is indicated in green. If the recorded points are invalid (USB connection (USB), pause data (PAUSE), sensor failure or sensor is not connected (NC)), then the record marking is gray. And in PDF curve zone, in case of USB data connection (USB), data pause (PAUSE), sensor failure (NC), all of their lines will be drawn as bold gray dotted lines.
- If the tempmate.®-M1 is connected to the computer during the recording period, it records no data during the connection time.
- During the tempmate.®-M1 is connected with the computer, the M1 is generating a PDF report depending on the configuration:
- If tempmate.®-M1 is stopped, it always generates a report when the M1 is plugged in the USB port
- If tempmate.®-M1 is not stopped, it only generates a PDF when it is enabled in the “Logger Setup”
Multiple start
The tempmate.®-M1 supports the function of continuous starting after last
logger stopping without the need to reconfigure the parameters.
Key function description
Left key: Start (restart) tempmate.®-M1, menu switch, pause
Right key: MARK, manual stop
Battery management
Battery level indication
Battery level indication | Battery capacity |
---|---|
40 % ~ 100 % | |
20 % ~ 40 % | |
5 % ~ 20 % | |
< 5 % |
Note:
When the battery capacity is lower or equal to 10%, please replace the battery
immediately. If the battery capacity is lower than 5%, the tempmate.®-M1 will
stop recording.
Battery replacement
Replacing steps:
Note:
It is recommended to check the battery status before restarting the logger to
ensure that the remaining battery life could finish the recording task. The
battery can be replaced before you configure the parameter. After replacement
of battery, the user needs to configure the parameter again.
When the logger is connected to computer under the status of recording or
pause status, it is prohibited to plug out the tempmate.®-M1 without battery
power supply.
LCD display notice
Alarm LCD display
When LCD display time is configured to 15 s, click left key to activate the
display. If over temperature incident occurs, it firstly displays alarm
interface for about 1 s, then skips to main interface automatically.
When display time is configured to “forever”, over temperature alarm occurs
permanently. Press left key to skip to main interface.
When display time is configured to “0”, there is no display available.
Appendix 1 – working status description
Device status | LCD display | Device status | LCD display |
---|
1 Start logger
| | |
5 MARK success
|
2 Start delay
• is flashing
| | |
6 MARK failure
|
3 Recording status
During recording status, in the middle of the first line, static display •
| | | 7 Device stop
In the middle of the first line, static display •
|
4 Pause
In the middle of the first line, blinking display •
| | |
8 USB connection
|
Appendix 2 – other LCD display
Device status | LCD display | Device status | LCD display |
---|
1 Erase data status
| | |
3 Alarm interface
Only exceed upper limit
|
2 PDF generation status
PDF file is under generation, PDF is in flash status
| | |
Only exceed lower limit
|
| | |
Both upper and lower limit occurs
|
Appendix 3 – LCD page display