tempmate M2 TH Use USB Temperature Data Logger User Manual

June 13, 2024
tempmate

tempmate logo User MANUAL

M2 TH Use USB Temperature Data Logger

tempmate M2 TH Use USB Temperature Data Logger

Introduction

The tempmate.®-M2 is designed to be mounted on a shipment or stationary and measure relevant parameters such as temperature and optionally relative humidity. The device records data and stores it on an internal memory.

Intended Use

The tempmate.®-M2 is designed to be mounted on shipments or stationary and to record relevant parameters as mentioned in the data sheet. Any use or operation r equiring specificrequirementsandstandardsnot explicitly mentioned in the data sheet must be validated and tested at the customer‘s own responsibility.

tempmate.®-M2 Model

tempmate M2 TH Use USB Temperature Data Logger - M2
Model

Multi-Use| |
---|---|---
Temperature| |
Rel. Humidity| |
LCD| |

Device Description

tempmate M2 TH Use USB Temperature Data Logger - Device
Description

Display

tempmate M2 TH Use USB Temperature Data Logger -
Display

Operation and Usage

*STEP 1 Configuration optional**
This step is only necessary if you want to adapt the pre-installed configuration to your application.

  • Download the free tempbase 2 software – https://www.tempmate.com/de/download/
  • Install the tempbase 2 software on your PC.
  • Remove the cap and connect the not started logger to your PC.
  • Open the tempbase 2 software and select the „Logger Setup“ button“ (1).
  • Make the desired settings and save them via the „Save Parameter“ button (2).
  • Remove the logger from your PC and replace the cap securely.

tempmate M2 TH Use USB Temperature Data Logger - Operation and
Usage

STEP 2 Start Logger

  • Press and hold the green start button for 5 seconds.
  • A successful start is indicated by the green LED on your device flashing 10 times.
  • Note: If another or no flashing signal appears, do not use the logger and contact support.

STEP 3 Set Mark

  • Briefly press the green start button twice in succession to set a mark.
  • A successfully set mark is indicated by the word „MARK“ and the number of marks set so far in your display.
  • Note: Up to 10 marks can be set per operation.

STEP 4 Stop Logger

  • Press and hold the red stop button for 5 seconds.
  • A successful stop is indicated by the red LED on your device flashing 10 times.

Alternative stop modes
Automatic Stop (default setting)

  • The device will stop automatically when the maximum number of measured values in thedata memory is reached and no manual stop is performed beforehand.
  • This stop mode works in addition to the manual stop.

Software Stop (optional)

  • This setting can be made in the tempbase 2 software. (see STEP 1)
  • The stop is triggered automatically by connecting the logger to the PC and opening  the  software.
  • A manual stop is not possible in this configuration.

STEP 5 Manual Readout of Data

  • Remove the cap and connect the logger to your PC.
  • A successful connection is indicated by both LEDs flashing. The abbreviations CSV and PDF appear one after the other in the display.
  • The logger automatically opens as an external drive on your PC. This process may take a few minutes depending on the amount of data.
  • Open the drive and copy the PDF and CSV report stored on it for your filing.
  • Note: A report is automatically generated as PDF and/or CSV when the device is stopped. The device can still be read out during a running measurement and an intermediate report can be downloaded.
  • Note: Already generated reports are automatically overwritten and deleted when the device ted. is restared.

Readout with the tempbase 2 software (optional)

  • Remove the cap and connect the logger to your PC.
  • Open the tempbase 2 software and select the „Export/Import“ button (3).tempmate M2 TH Use USB Temperature Data Logger - icon 1
  • Select the desired file format (PDF/XLS/IME) for export and the file location and confirm the download.

tempmate M2 TH Use USB Temperature Data Logger - Operation and Usage
2

External Sensors

  • Remove the cap and connect the not started logger to your PC.
  • Open the tempbase 2 software and select the „Logger Setup“ button.
  • In the „Sensor Type“ area, select the sensor type you want to work with.
  • Confirm your configuration by clicking on „Save Parameter“ and remove the device from your PC.
  • To record with an external sensor, use a screwdriver to loosen the screw on the bottom of the device and remove the standard cap.
  • Replace it with the external sensor of your choice and screw it again.

Replace Battery

  • Open the cover on the back of the device by turning it counterclockwise.
  • Remove the old battery and dispose of it according to national regulations.
  • Insert the new battery and replace the cover, closing it clockwise.
  • Remove the cap and connect the logger to your PC.
  • Open the tempbase 2 software to synchronize date & time again. This process is automatically trigge  ed when the logger is connected to PC and software.
  • Caution: Back up your data and download your last report before removing the battery  from the instrument.

Important Notes

  • Configuration cannot be changed during recording.
  • We recommend recalibration after 1 year.
  • Always dispose of batteries according to your country‘s regulations.
  • Do not place the device in corrosive liquids and do not expose it to direct heat.

tempmate M2 TH Use USB Temperature Data Logger - Main
Technical

Main Technical Specifications tempmate.®-M2

Temperature Sensor| HQ Digital Temperature sensor (internal and external optional)
---|---
Temperature Range| -30°C to +70°C (-40°C to +90°C with ext. T Sensor) (-80°C to +200°C with ext. PT100 Sensor)
Temperature Accuracy| ±0.3°C (at -20°C to + 40°C, other 0.5°C)
Temperature Resolution| 0.1°C
Humidity Sensor| n/a
Humidity Range| n/a
Humidity Accuracy| n/a
Humidity Resolution| n/a
Data Storage| 60,000 values
Display| Big Multifunction LCD
Start Setting| Manually by pressing button, by software or timed
Recording Time| Up to 6 months
Interval| 10sec. up to 11 h 59min. (default 10 min.)
Alarm Settings| Up to 6 points customizable
Alarm Type| Single alarm or cumulative
Battery| CR2450 / replaceable by customer
Dimensions| 100 x 53 x 12 mm
Weight| 54g
Protection Class| IP65
Connection Interface| USB 2.0, A-Type
Conformity| EN 12830, CE, RoHS
Software| PDF or CSV reader or tempbase 2 software / free download
Interface to PC| Integrated USB port
Reprogrammable| Yes, with internal HTML tool* or optional tempbase 2 Software
Automatic Reporting| PDF & CSV

tempmate M2 TH Use USB Temperature Data Logger - Main Technical
1

Main Technical Specifications tempmate.®-M2

Temperature Sensor| HQ Digital Temperature sensor (internal and external optional)
---|---
Temperature Range| -30°C to +70°C (-40°C to +90°C with ext. T Sensor) (-80°C to +200°C with ext. PT100 Sensor)
Temperature Accuracy| ±0.3°C (at -20°C to + 40°C, other 0.5°C)
Temperature Resolution| 0.1°C
Humidity Sensor| HQ Digital Temperature/rel. Humidity sensor (internal and extern
Humidity Range| 0%rH to 100%rH
Humidity Accuracy| ±3%rH (20 to 80%rH), 5% others (at 25°C)
Humidity Resolution| 0.1%rH
Data Storage| 60,000 values
Display| Big Multifunction LCD
Start Setting| Manually by pressing button, by software or timed
Recording Time| Up to 6 months
Interval| 1 Osec. up to 11 h 59min. (default 10 min.)
Alarm Settings| up to 6 points temperature and 2 points humidity customizable
Alarm Type| Single alarm or cumulative
Battery| CR2450 / replaceable by customer
Dimensions| 100 x 53 x 12 mm
Weight| 54g
Protection Class| IP65
Connection Interface| USB 2.0, A-Type
Conformity| EN 12830, CE, RoHS
Software| PDF or CSV reader or tempbase 2 software / free download
Interface to PC| integrated USB port
Reprogrammable| Yes, with internal HTML tool* or optional tempbase 2 Software
Automatic Reporting| PDF & CSV

Main Technical Specificationstempmate.®-M2 Accessory

tempmate.®-M2 External T-Sensor

Sensor| HQ Digital Temperature Sensor
Temperature Range| –40°C to +90°C
Temperature Accuracy| 0.3°C (at –20 ° C to + 40 ° C, other 0.5°C)
Temperature Resolution| 0.1°C
Sensor Tip| Stainless Steel (30 x 5 mm)
Sensor Connetion| M2-USB Connection
Cable length| 1.2 m
Cable Diameter| 3 mm
Cable Material| PVC

tempmate.®-M2 External High/Low T-Sensor

Temperature Sensor PT100 Sensor
Temperature Range -80°C to +200°C
Temperature Accuracy ±1°C
Temperature Resolution 0,1°C
Sensor Tip Stainless Steel (30 x 5 mm)
Sensor Connetion M2-USB Connection
Cable Diameter 3 mm
Cable length 1.2 m
Cable Material PTFE

tempmate.®-M2 External T/rH-Sensor

Sensor HQ Digital Temperature/rel. Humidity Sensor
Temperature Range –40°C to +90°C
Temperature Accuracy 0.3°C (at –20 ° C to + 40 ° C, other 0.5°C)
Temperature Resolution 0,1°C
Humidity Range 0 – 100 %rH
Humidity Accuracy  ±3%rH (10% to 70%), 5% others (at +25°C)
Humidity Resolution 0.1 %rH
Sensor Tip Stainless Steel (30 x 5 mm)
Sensor Connetion M2-USB Connection
Cable length 1.2 m
Cable Diameter 3 mm
Cable Material PVC

Contact

tempmate M2 TH Use USB Temperature Data Logger -
Contact

Do you have any questions? Please contact us – our experienced team will be happy to support you.
[email protected]
+49 7131 6354 0
empower your supply chain.
V1.0-12/2021-DE · Technical Changes and Errors excepted

tempmate logotempmate GmbH
Wannenäckerstr. 41
74078 Heilbronn, Germany
Tel. +49-7131-6354-0
[email protected]
www.tempmate.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

tempmate User Manuals

Related Manuals