tempmate S1 Single Use Temperature Logger
Product Overview
This document describes specific validation procedures for template A-5l data loggers, for the purpose to ensure proper operation and provide a reliable temperature and time record during transportation and storage.
Technical Specifications
- Recording Options: Single-Use
- Temperature Range: -300C to 70 •c
- Temperature Accuracy: ±0.5 (-200C/+400C); ±I.O (other range)
- Temperature Resolution: 0.1 •c
- Data Storage Capacity: 16.000 readiness
- Shelf Life / Battery: 2 years / CR2450 button cell
- Recording Interval: 10 minutes (standard, others on request)
- Recording Duration: up to 1 0 days (standard, others on request)
- Startup Mode: Button or software
- Stop Mode: Button, software, or stop when full
- Protection Class: IP67 / NEMA 6
- Dimensions: 83 mm x 47 mm x 7 mm (L x w x H)
- Weight: 14.6g
Operation Instructions
- Keep the button pressed for a minimum of S seconds to start recording. Then the RECORD LED flashes 10 times, which indicates a successful starting. If the STATUS LED is lit for 3 seconds, please do not use the logger (Indicating low battery)! Tear off the serial number label for your reference.
- Press the button once at any time to check the current status of the logger. Double-click the button to set a mark in the reports. See details of these optional features in the tables below.
- Keep the button pressed for a minimum of S seconds to stop recording. Then the STATUS LED flashes 10 times, which indicates a successful stopping.
- After stopping the record, t ear Off the transparent plastic cover and connect the logger to a USB port of the computer. The PDF and CSV reports will be generated automatically and can be found in the device’s .,t empmate· folder.
Operation Indication
Start/ Stop
- Status Action LED confirmation
- Start Logger 5 sec pushing button RECORD LED flashing 10 times
- Set Mark double-click button STATUS CED + RECORD LED flashing 5 times
- Stop Logger 5-sec pushlng button STATUS LED flashing 10 times
Status Request by pushing the Button 1 time
Status Action LED confirmation
- Not Started to push button 1-time STATUS LED + RECORD LED flashing 1 time
- Recording – 0K push button 1-time RECORD LED flashing I time
- Recording – Alarm push button 1-time STATUS LED flashing I time
- Stopped – 0K push button 1-time RECORD LED flashing 2 times
- Stopped – Alarm push button 1-time STATUS LED flashing 2 times