POWER PROBE PPDMM Digital Multimeter User Manual

June 3, 2024
POWER PROBE

POWER-PROB-logo

POWER PROBE PPDMM Digital Multimeter

GENERAL SPECIFICATIONS

  • Operating Altitude: 2000m
  • Relative Humidity: 75% max operating
  • Operating Temperature: 0°C-40°C/32°F-104°F (<80% RH)
  • Storage Temperature: -10°C-60°C/14°F-140°F (<70% RH)
  • Accuracy Temperature: -18°C-28°C/64°F-82°F (<80% RH)
  • Temperature Coefficient: 0.1x(specified accuracy)/°C (<18°C or >28°C)
  • Sampling Frequency: approx. 3 times/sec.
  • Fuse Protection: µNmA input: F600mN600V 10A input: F10N600V
  • DC/AC Voltage: 600V
  • DC/AC Current: 10A
  • Resistance: SOMO
  • Capacitance: 60mF
  • Frequency: 10kHz
  • Diodes: 2.7V
  • Continuity: <500
  • Temperature: -2o·c-1ooo·c1-4°F-1832°F
  • LCD Display: 3 ¾ digit display (6000 counts)
  • Product Supply: 3×1.SV AAA batteries
  • Product Size: 160mmx74mmx49mm / 6.3″x2.9″x1 .9″
  • Product Weight: 482g / 1.06Ib
  • Safety Rating: CAT IV 600V; pollution degree: II
  • Safety Standards: IEC61010-1
  • Pollution Degree: 2
  • Accuracy: ±(of reading+# of least significant digits)

WARNINGS
To avoid electric shock and injury or damage to the meter, observe the following safety methods:

  • Check the meter before use to make sure there was no damage during transit.
  • Check that the insulation on the test leads is not damaged and/or wires are not exposed.
  • If any faults or abnormalities are observed, the meter should not be used and should be checked out prior to use.
  • Never exceed the protection limit values indicated in specifications for each range of measurement.
  • Always be careful when working voltages above 60V DC or 30V AC rrns, keep fingers behind the probe barrier while measuring.
  • Make sure the rotary switch is in the correct position before measurement.
  • Never use the meter in an environment with explosive gas, vapor or dust.
  • Always keep fingers behind probe barriers when making measurements

WARNINGS

  • When connecting test leads to a circuit, connect the black test lead first, then the red lead. Disconnect in the opposite order.
  • Turn off power and discharge all capacitors first before measuring resistance, continuity or diodes.
  • Failure to follow safety guidelines could compromise the safety features of this meter.
  • Do not use the meter without the battery cover in place.
  • Replace the batteries as soon as the low battery symbol “” to avoid false reading that could lead to electric shock and injury.
  • Remove test leads from all circuits before opening the battery cover.

SAFETY SYMBOLS

  Im       tant safety information
  Ground
  DC(direct current)
  AC (alternating current)
  DC/AC Voltage or Current
  Double insulation protection

| ****

Fuse

 | ****

Compliance with EU regulations

FEATURE DETAILS

  1. LCD Display

  2. Hold/Backlight Button
    Press to hold the current reading on the display. Press the button again to release the hold Hold “ to turn on the backlight. Hold the button again to manually turn off the backlight.

  3. Max/Min Button
    In all modes (except continuity, diode, capacitance), press Max/Min” and the display will show the maximum value recorded since the button was pressed. Press the button again and the display will show the minimum value recorded. Pressing the button a third time will show the difference between the max and min value. Hold “Max/Min” to return the display to normal readings.

  4. Function Switch Button
    Press “FUNC” to switch between functions or between AC/DC current.

  5. Frequency/Duty Cycle Button (Hz/%)
    In AC voltage/current modes, press “Hz%” and the display will show the frequency measurement. Press the button again to switch to duty cycle. Press the button a third time to return to normal display.

  6. Range Button
    In voltage, current and resistance modes, the default range is auto. To enter manual range, press “RAN”. Each press of the button increases the range, and returns to the lowest range when pressed in the highest range. Hold “RAN” to return to auto range. (Only auto range is available in capacitance mode)

  7. Relative Measurement Button
    In all modes (except resistance, continuity, diode), press “REL” and the display will show the relative value, i.e. the difference between |the stored value when the button was pressed and the currently measured value. (REL = stored value – currently measured value) Press the button again to return the display to normal. (In REL mode, auto range is disabled)

  8. Rotary Switch

  9. Input Jack (all measurements; current below 600mA)

  10. A Jack (current measurements between 600mA-10A only)

  11. Common Jack (all measurements)

DISPLAY FUNCTION INSTRUCTIONS

  1. Direct Current
  2. Alternating Current
  3. Negative DC Value
  4. Auto Power Off
  5. Data Hold
  6. Auto Range Active
  7. Maximum Display Minimum Display
  8. Relative Display Diode Test
  9. Continuity Test
  10. Low Battery
  11. Temperature in Celsius
  12. Resistance
  13. Frequency
  14. Capacitance
  15. DCIAC Current
  16. DC/AC Voltage
  17. Main Display

ROTARY SWITCH FUNCTION INSTRUCTIONS

DC Voltage: <600V

POWER-PROBE-PPDMM-Digital-Multimeter-fig5

AC Voltage: <600V

POWER-PROBE-PPDMM-Digital-Multimeter-fig6

DC Current (large): <10A

POWER-PROBE-PPDMM-Digital-Multimeter-fig7

AC Current (large): <10A

POWER-PROBE-PPDMM-Digital-Multimeter-fig8

DC Current (middle): <600mA

POWER-PROBE-PPDMM-Digital-Multimeter-fig9

AC Current (middle): <600mA

POWER-PROBE-PPDMM-Digital-Multimeter-fig10

DC Current (smallM): <6000uA

POWER-PROBE-PPDMM-Digital-Multimeter-fig11

AC Current (small): <6000uA

POWER-PROBE-PPDMM-Digital-Multimeter-fig12

Resistance: <60MQ

POWER-PROBE-PPDMM-Digital-Multimeter-fig13

Continuity: <500

POWER-PROBE-PPDMM-Digital-Multimeter-fig14

Diode Test: <1VPOWER-PROBE-PPDMM-Digital-Multimeter-
fig15

Capacitance: <60mFPOWER-PROBE-PPDMM-Digital-Multimeter-
fig16

Temperature: -20°C-1000°C/-4°F-1832°FPOWER-PROBE-PPDMM-Digital-
Multimeter-fig17

ELECTRICAL SPECIFICATIONS

DC Voltage Measurement

Range| Resolution| Accuracy
600mV| 0.1mV| ****


% of reading +2 digits)

6V|
60V| 0.01V
600mV| 0.1V

  • Input impedance: 1OMO
  • Max. input voltage: 600V rms

Range

V

| Resolution

0.1mV

| Accuracy



.0% of reading +5 digits)

---|---|---
6V|
60V| 0.01V
600mV| 0.1V

  • Input impedance: 1OMQ
  • Max. input voltage: 600V rms
  • Frequency response: 40-400Hz, calibrated to rms of sine wave (average response)

Range| Resolution| Accuracy
600µA| 0.1µA|

of reading +5 digits)

6000µA|
60mA| 0.01µA
600mA| 0.1µA
10A| 10mA| ±(2.0% of reading +10 digits)

  • Overload protection: PA/mA input: Fuse (F600mA/600V) 10A input Fuse (F10A/600V)
  • Max. input current: UA/mA input: 600mA rms 10A input: 10A rms

AC Voltage Measuremen t

e| Resolution| Accuracy
 | 0.1µA|

of reading +5 digits)

6000µA|
60mA| 0.01µA
600mA| 0.1µA
10A| 10mA| ±(2.5% of reading +10 digits)

  • Overload protection: HA/mA input: Fuse (F600mA/60OV) 10A input Fuse (F10A/600V)
  • Frequency response: 40-400Hz calibrated to rms of sine wave (average response)
  • Max. input current: |HA/mA input: 600mA rms 10A input: 10A rms

e Measuremen t

Rang| Resolution| Accuracy
6000| 0.10|

±(0.8% of reading +5 digits)

6kO|
60kO| 0.01kO
600kO| 0.1kO
6MQ| 0.001MO
60MQ| 0.01MO| ±(2.0% of reading +5 digits)

  • Max. input voltage: 600V rms

POWER-PROBE-PPDMM-Digital-Multimeter-fig18

  • Max. input voltage: 600V rms

POWER-PROBE-PPDMM-Digital-Multimeter-fig19

  • Max. input voltage: 600V rms

Range| Resolution

0.001nF

| Accuracy

±(4.0% of reading +10 digits)

---|---|---
10nF|  | ****



±(3.0% of reading +10 digits)

100nF| 0.1nF
1µF| 1nF
10µF| 10nF
100µF| 100nF
1mF| 1µF
10mF| 10µF| ****

±(4.0% of reading +10 digits)

60mF| 10µF

  • Max. input voltage: 600V rms
equency (cont.) (AC voltage)
  Resolution Accuracy
99.99H 0.01Hz

of reading +5 digits)

999.9Hz|
9.999kHz| 0.001kHz

10kHz| 0.01kHz| Reference only

  • Signal input range: 20.2V AC rms (voltage input will increase as frequency increases)
  • Input impedance: 10MQ
  • Max. input voltage: 600V rms

ency (cont.) (AC voltage )

Ran| Resolution| Accuracy
99.99Hz| 0.01Hz|

reading +5 digits)

999.9Hz|

1kHz| 0.001kHz| Reference only

  • Signal input range: uA: 2 60uA rms mA: 2 6mA RMS A: 20.6A rms (current input will increase as frequency increases)
  • Max. input voltage: 10A rms
Range I Resolution I Accuracy
-20°C- 1000°C I I  

±(2. 0% of reading +3 digits)

-4 ° F- 1832 °| I| I

  • Overload protection: Fuse (F600mA600V)

CANADA & USA

Power Probe Group, Inc. cs.na@mgl-intl.com 2810 Coliseum Centre Drive, Ste. 100. Charlotte, North Carolina 28217 USA Tel: +1 833 533-5899

700020287 SEP 2021 V1

2021 MGL International Group Limited. All rights reserved. Specifications are subject to change without notification.

Read User Manual Online (PDF format)