**SONMOL SMPF-2S Electronic Peak Flow Meer Instruction Manual
**
Product Introduction
[Product Name] Peak Flow Meter
[Specification】 SMPF-2S
【Intended use] Measuring forced expiratory volume of 1 second (FEV1), peak
expiratory flow (PEF)
[The scope of application】 Apply to the test of forced vital capacity
Contraindications:
- Myocardial infarction, stroke, shock in the past 3 months
- Hemoptysis in the last 4 weeks
- Uncontrolled hypertension
- Severe hyperthyroidism
- Pneumothorax, huge lungs, and not ready for surgery
- Pregnant women
- Tympanic membrane perforation (measured after the affected ear canal is first blocked
- Severe cardiac insufficiency, severe arrhythmia, and unstable angina in the last 4 weeks
- Seizures require medication
- Aneurysm
- Heart rate 120 beats/min
Product Introduction
Warning:
- To ensure proper use of the product, please read the user manual thoroughly and store it safely.
- Dispose of used batteries in accordance with local regulations.
- For accurate testing, do not perform more than 5 consecutive tests as a single user.
- Do not use this product in environments containing enriched oxygen or flammable materials to prevent explosions.
- Avoid usage in areas with strong electromagnetic interference or excessive wind.
- Unauthorized disassembly or modification of the product is strictly prohibited.
Caution :
- Maintain a clean and free-of-vibration operating environment, with no corrosive or flamnable materials, and the temperature and humidity should not be too high or too low.
- When transitioning the product from a cold environment to a warm, humid one, allow it to acclimate before use.
- If the product fails to display data or exhibits other abnormalities during measurements, power it off and restart.
Product Structure and Components
The Peak Flow Meter consists of a Main Unit and a Mouthpiece.
Product Structure and Components
Screen Display
- FEVT Value
- PEF Value
- Current Month
- Current Date
- Current Year
- Current Time
- Low Battery Indicator
- Records
- Unit of PEF
- Unit of FEV1
- Risk Indicator
Operating Instructions
Warning: For personal use only
- The Peak Flow Meter (SMPF-25) is a handheld device designed for assessing lung function. It is intended for personal use, and users are only required to follow the instructions provided in the user manual. No specialized training is necessary. When the equipment is taken out of storage under extreme temperature conditions, it should be allowed to acclimate for at least half an hour before use
Measurement
Power on
Press the Home Button screen. the device is properly tumed on with 90-“showing
on the
Prepare
Sit straight/stand up, take a deep breath with your mouth, then cover the
Mouthpiece with your mouth completely
*Please don’t place your tongue or teeth against the Mouthpiece. Also
please do not cover the back of the Mouthpiece with your hand.
Operating Operating Instructions
Start
Blow as fast as you can, wait for 2 beeps from the device before you can read the result.
Note: Premeasurement is required in the following cases
- Cough
- Exhalation time is too short
- Exhale too slow
- Measured values deviate significantly from normal
Premeasurement
Press the Home Button”-” again, and repeat step 1.2 and step 1.3.
“It is recommended to perform at least three consecutive measurements
throughout the measurement, taking the maximum as the final result.
Power off
Press and then hold the Home Button for 2 to 3 seconds until the device is
turned off. Also the device will be automatically powered off after
approximately 1 minute of inactivity
Operating Instructions
Time setting
Press the Home Button”-“, the screen will be “fully bright fast, then it will
display “go.”
Press and hold the Home Button”-” and the Left Button” at the same time for
about 5 seconds, and you will see the year figure starts to flash at the top
of the screen. Adjust the value through the Left Button” and Right Button
Press the Home Button “-” to confirm, and then set the month, date and time in
turn.
Press and hold the Home Button “-” about 3 seconds to save the date and time
settings.
Setup your expected PEF value
The expected value affects the asthma control assessment results and needs to
be set in advance before the test. It is recommended to update the expected
value once a year due to changes in age, height, otc
Auto configuration
The device will be automatically configured based on your latest expected PEF
wallues set in the ePEF app once connected via Bluetooth 4.0.
Manual configuration
Make sure the device is powered off, press and hold the Home Button
ignore the “go” pagel for about 5 seconds. The display will show “”000″
indicating the parameter setting mode is properly activated, and the flashing
figure indicates that the value can be edited.
In the parameter setting mode, press the Right Button”” to increase the value
of the current digit, press the Left Button to decrease the value, press the
Home Button to switch to the next digit.
Press and hold the Home Button about 3 seconds to save the current value as
your expected PEF value.
Note:
- The maximum setting value is “999 L/MIN”, and the minimum is “60 L/MIN”
- No red, yellow and green Risk Indicators will be displayed on the device without properly setting the expected value.
How can I know my expected value?
-
Calculation formula provided by a specialized research institution, scan to calculate your expected value.
“The evaluation and measurement of FEVI values are more complicated than PEF values. It is recommended to consult a doctor or professional journals for detailed interpretation. -
Your physician recommend value
-
The optimal result you measured in good physical condition (in 2 weeks).
Equipment Maintenance
-
Cleaning
Main unit: Wipe with a soft dry cloth, do not put it in water. Cleaning once a week is recommended.
Mouthpiece: To clean the Mouthpiece intemal and external, a meidcal alcohol cotton of 75% concentration and larger than 30*60mm is recommended. To ensure the safety and hygiene during use, it must be cleaned after each use. -
About battery
When the icon “00” stay on continuously, it indicates a low battery, a 5V power adapter is required for charging.
“Note : “–” flashes sequentially, indicating that it is charging, and “–” stays on, indicating that it is fully charged.
-
Period of use
Peak Fow Meter life is three years. -
Maintenance
Before use, it is essential to perform a comprehensive inspection to ensure the equipment can operate normally. Please follow these steps.
Examine the equipment for any signs of mechanical damage.
Confirm that the display is functioning correctly, ensuring the equipment is in excellent working condition
Note : The equipment undergoes a thorough inspection at the factory before shipping, and during its intended period of use, no further calibrations are necessary.
-
Before use, please refer to the instructions the section (1) for cleaning.
When the low battery indicator lights up, please refer to the instructions the in section (2) for battery charge.
The maintenance of this product is limited to the qualified maintenance personnel designated by the manufacturer. The user canhut disassemble and repair. At the same time the manufacturer can provide the circuit diagram, components list, rectification rules, or necessary information to help qualified technical personnel for maintenance
Warning: Do not modify this equipment without authorization of the manufacturer. -
Storage
Please put the Peak Flow Meter in clean and dry place. Exposure to direct sun light or extreme high and low temperature, or violent impact may result in work failure of the Peak Flow Meter or even damage the device. -
Production date
Please refer to the production date label in detail
Definition
Glossary
PEF: Peak Expiratory Flow (PEF), also called Peak Expiratory Flow Rate
(PEFR), is a person’s maximum speed of expiration, as measured with a Peak
Flow Meter. It measures the airflow through the bronchi of the lungs and shows
the degree of obstruction in the airway.
FEVI: Forced Expiratory Volume (FEV1) calculates the amount of air that a person can force out of their lungs in 1 second. It is an indicator of the reversibility of airway obstruction and the primary indicator of impaired lung function,
PEF %: Measured PEF/Expected PEF 100%
FEV1%: Measured FEV1/Expected FEV1 100%
Expected value
The expected value (target value is usually calculated based on statistics. It
varies according to race, gender, age, and height According to the variation
between the actual tested value and the expected value, your PEF readings are
classified into three measurement zones, green, yellow, and red. The device
will show these colors with results accordingly.
Green: PEF6806, a peak fiow reading in the green zone indicates that the lung
function management is under good control,
Yellow: 50% PEF<80%, your peak flow reading indicates caution. It may
mean respiratory airways are narrowing, and additional medication may be
required.
Red: PEF-C50%, your peak flow reading indicates a medical emergency.
Severe amway narrowing may occur, and immediate action must be taken. This
would usually involve contacting a doctor or hospital
Note :
Recommendation from: National Institutes of Health (NIH)
FEV1 is generally used as an indicator to measure airway obstruction. Its
measurement, evaluation, and interpretation are complicated. It is recommended
to consult a doctor or consult professional publications,
Environmental protection instructions
From an environmental and resource standpoint, environmental disposal for
batteries should be in accordance with local regulations. The equipment in
lille end should be handled in accordance with local laws and regulations
Key symbols
- After the use of waste, please follow the regulations of local health or environmental protection agencies
- Applied part of type B
- Refer to instruction manual Warning, see the instructions for use
- Pollution control symbol of electronic, information products
- Indicating that this product is environmentally friendly for 10 years and is recyclable and should not be discarded
- Do not use if package is damaged
- Fragile, handle with care
- Keep dry
- Keep away from sunlight
Precautions and warnings
Adverse reactions during use are mostly mild, repeated deep breathing force,
hyperventilation may appear dizziness, hand and foot fingertips and facial
perioral numbness or acupuncture, slight hand tremor and other symptoms,
severe syncope may occur. At this point, the subject should be resting
quietly, and care should be taken to protect the subject from fall injuries.
Technical Specifications
Product name | Peak Flow Meter |
---|---|
Model | SMPF-2S |
Display | Segment LCD |
Test method | Pressure sensor |
Measurement range | Volume:0.5L –. 81 Flow rate: 60L/min –. 840 L/min |
Accuracy | Volume: ±3% or ±0.05L (whichever is greater)Flow rate: ±10% or |
±18L/min ( whichever is greater)
Repeatability| Volume: ±3% or ±0.05L of reading ( whichever is greater)Flow
rate: ±5% or ±10 L/min of reading (whichever is greater)
Airflow resistance| 0.25kPa/L/s (600 L/min)
Flow rate frequency response| ±12% or ±15L/min ( whichever is greater)
Working way| Continuous operation
Operating environment| ~10°C — +40°C; C80%RH
Operating atmospheric pressure| 700hPa-1060hPa
Transport, storage environment| -10°C — +55°C ; C95%RH
Transport, storage of atmospheric pressure| 500hPa-1060hPa
Software version| V2
Fault Analysis and Resolution
Troubles | Possible Reasons | Solutions |
---|---|---|
Display El | Initialization error | Turn off and restart |
Display H1 | The result is higher than 840L/min | Extremely high measurement |
The device cannot be powered on | Battery is low | Use 5V power adapter for |
charging
No data while blowing| Not enter the test mode| Press the Home Button again or
restart
Wrong inflatable posture| Refer to manual and blow properly
Sudden display disappearance| Shut down naturally while no operation for one
minute| Normal phenomenon
EMC Declaration
Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
Guidance and manufacturer’s declaration – electromagnetic emission
The SMPF-2S Peak Flow Meter is intended for use in the electromagnetic
environment specified below.The customer or the user of SMPF-2S Peak Flow
Meter should ensure that it is used in such an environment.
Emission test| Compliance| Electromagnetic environment – guidance
RF emissions CISPR 11| Group 1| The SMPF-2S Peak Flow Meter uses RF energy
only for its internal function. There for, its RF emissions are very low and
are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11| Class B| The SMPF-2S Peak Flow Meter is suitable for
use in all establish-ments other than domestic and those directly connected to
the public low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions IEC 61000-3-2| N/A
emissions IEC 61000-3-3 e fluctuations flicker| N/A
EMC Declaration
Guidance and manufacturer’s declaration — electromagnetic immunity
The SMPF-2S Peak Flow Meter is intended for use in the electromagnetic
environment specified below.The customer or the user of the SMPF-2S Peak Flow
Meter should ensure that it is used in such an environment.
Immunity test| IEC 60601test level| Compliance level| Electromagnetic
environment – guidance
Electrostatic discharge (ESD) I EC 61000-42| ± 8 kv contact± 2 kV, ± 4 kV, ± 8
kV, ± 15 kV air| ± 8 kv contact± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air| Floors
should be wood, concrete or ceramic tile. If floors are covered with synthetic
material, the relative humidity should be at least 30 %.
Electrostatic transient / burst I EC 61000-44| ± 2 kV for power supply lines±
1 kV for input/output lines| N/A| Mains power quality should be that of a
typical commercial or hospital environment.
Surge I EC 61000-45| ±1 kV differential mode± 2 kV common mode| N/A| Mains
power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input
linesIEC 61000-411| 0 % UT; 0,5 cycle g) At 0°, 45°, 90°, 135°, 180°, 225°,
270° and 315°0 % UT:1 cycle and 70% UT; 25/30 cycles Single phase: at 0°0 %
UT; 250/300 cycle| N/A| Mains power quality should be that of a typical
commercial or hospital environment. If the user of the SMPF-2S Peak Flow Meter
requires continued operation during power mains interruptions, it is
recommended that the SMPF-2S Peak Flow Meter be powered from an
uninterruptible power supply or a battery.
Power frequency (50/60 Hz) magnetic fieldIEC 61000-48| 30 A/m| 30 A/m| Power
frequency magnetic fields should be at levels characteristic of a typical
location in a typical commercial or hospital environment.
NOTE: UT is the a. c. mains voltage prior to application of the test level.
EMC Declaration
NOTE 1. At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
is affected by absorption and reflection from structures, objects and people.
a The ISM (Industrial, scientific and medicall bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz. 13,553 MHz to 13,567 MHz, 26,957 MHz to 27,283 MHz; and 40, 66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5, 4 MHz, 7 MHz to 7.3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14.2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
b Field strengths from fixed trarismitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should beconsidered. If the measured field strength in the location in which th SMPF-25 Peak Flow Meter is used exceeds the applicable RF compliance level above, the SMPF-25 Peak Flow Meter should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as réorlenting or relocating the SMPF-2S Peak Flow Meter.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3v/m.
Guidance and manufacturer’s declaration — electromagnetic immunity
The SMPF-2S Peak Flow Meter is intended for use in the electromagnetic
environment specified below.The customer or the user of the SMPF-2S Peak Flow
Meter should ensure that it is used in such an environment.
Immunity test| IEC 60601test level| Compliance level| Electromagnetic
environment – guidance
Conducted RF IEC 61000-4-6Radiated RFI EC 61000-4-3| 3 Vrms150 kHz to 80 MHz6V
in ISM and amateur radio bands between 0,15 MHz and 80 MHz10 V/m80 MHz to 2.7
GFi i385MHz-5785M Hz Test specifications for ENCLOSURE PORT IMMUNITY to RF
wireless communication equipment (Refer to table 9 of IEC60601-1-2:2014)|
N/A10 Vim80 MHz to 2.7 GHz385MHz-5785M Hz Test specifications for ENCLOSURE
PORT IMMUNITY to RF wireless communication equipment (Refer to table 9 of IEC
60601-1-2:2014)| Portable and mobile RF communications equipment should be
used no closer to any part of the SMPF-25 Peak Flow Meter, including cables,
than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.1.1.1 Recommended separation
distance3.5. •—
d=[— Vij //3 12 r
d=[ —,, ] /P 80 MHz to 800 MHzv2 dr[=j3 r“ F 800 MHz to 2.7 GHzE,
d=[ __FA —7]
E,where p is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in metres (m).bField strengths from fixed RF transmitters, as
determined by an electromagnetic
site survey,a should be less than the compliance level in each frequency
range.bInterference may occur in the vicinity of equipment marked with the
following symbol: i
Manufacturer Company: CHONGQING MOFFY INNOVATION TECHNOLOGY CO., LTD.
Address: No. 292 Jingdonglang Rd. Beibei Dist Chongqing, CN 400714
Email: service@sonmolnmed.com
Web: https://www.sonmoimed.com
Made in China