APERA INSTRUMENTS MP511 pH-mV Benchtop Meter Instruction Manual
- June 3, 2024
- APERA INSTRUMENTS
Table of Contents
MP511 pH-mV Benchtop Meter
MP511 pH/mV Benchtop Meter Instruction Manual
APERA INSTRUMENTS (Europe) GmbH
www.aperainst.de
– 1 –
v4.1
Contents
1 Brief Introduction ………………………………………………………………………………………………- 3 2 Technical
Specifications……………………………………………………………………………………….- 4 –
2.1 Technical Parameters……………………………………………………………………………………………… – 4 2.2 Other
Technical Parameters …………………………………………………………………………………….. – 4 3 Complete
Kit……………………………………………………………………………………………………..- 5 4 The
Meter…………………………………………………………………………………………………………- 5 4.1 LCD Display
………………………………………………………………………………………………………….. – 5 4.2 Keypad
……………………………………………………………………………………………………………….. – 6 4.3 Store, recall and clear
readings ………………………………………………………………………………… – 7 4.4 Sockets:
………………………………………………………………………………………………………………. – 8 5 pH
Measurement……………………………………………………………………………………………….- 8 5.1
Preparation:…………………………………………………………………………………………………………. – 8 5.2 pH Electrode
Information ……………………………………………………………………………………….. – 9 5.3 Calibration:
………………………………………………………………………………………………………… – 10 5.4 Sample
Test………………………………………………………………………………………………………… – 11 5.5 Parameter Setting
……………………………………………………………………………………………….. – 12 5.6
Notes………………………………………………………………………………………………………………… – 14 5.7 Self-diagnostic
Information …………………………………………………………………………………… – 15 6 mV and ORP Measurement
……………………………………………………………………………….. – 16 6.1 Sample
Test………………………………………………………………………………………………………… – 16 6.2
Notes………………………………………………………………………………………………………………… – 16 6.3 Parameter Setting
……………………………………………………………………………………………….. – 17 7 RS232 Communication
……………………………………………………………………………………… – 17 7.1 System requirements
…………………………………………………………………………………………… – 17 7.2 Software
interface……………………………………………………………………………………………….. – 17 7.3 Load software
…………………………………………………………………………………………………….. – 18 7.4 Port Connection
………………………………………………………………………………………………….. – 18 7.5 Run Software
……………………………………………………………………………………………………… – 18 8 Recommended pH Electrodes for
Specific Applications …………………………………………… – 19 9 Warranty
……………………………………………………………………………………………………….. – 20 –
– 2 –
1 Brief Introduction
Thanks for purchasing APERA INSTRUMENTS MP511 pH/mV Benchtop Meter (referred
to “meter” as below). Before using this meter, please read this instruction
manual cafrefully in order to properly use and maintain it. Features This
meter has an outstanding combination of advanced electronic technology, sensor
technology and software design. It is suitable for laboratory use to measure
pH and ORP value of regular water solutions. With its built-in microprocessor
chip, elegant designs and user-friendliness, this meter has the following
features: Meets international GLP standards, with intelligent functions such
as automatic
calibration, automatic temperature compensation, data storage, auto. timing
measurement, RS232 output, clock display, parameter setting and self-
diagnosis. With digital processing technology, the response speed and accuracy
are greatly improved, appears when readings are stable. Automatic pH buffer
solution recognition. Buffer solution selectable: USA series, NIST series and
China series. The meter meets IP54 protection level. All of the sockets are
protected by the silica gel caps.
– 3 –
2 Technical Specifications
2.1 Technical Parameters
Range
Resolution
Accuracy
pH
Input Current
Input Impedance
Stability
(-2.00 19.99) pH 0.1/0.01 pH
±0.01pH±1 digit 2×10-12 A 1×1012
±0.01 pH±1 digit/3h
Temp. Compensation Range
(0 100)°C (Automatic or Manual)
mV (ORP)
Temp.
Range Resolution Accuracy
Range Resolution Accuracy
-1999mV 0 1999mV 1mV
±0.1% FS -10°C 110°C
0.1°C 5 60°C±0.5°C±1 digit Other±1°C
2.2 Other Technical Parameters Data storage
Storage content
Communication connector Power IP rated
Size and weight Quality and safety certification
600 groups serial number, date, time, measurement value,
temperature value, ATC or MTC state RS232
DC9V/0.5A IP 54 dustproof and splash-proof
160 × 190 × 70mm/880g ISO9001:2015 and CE
– 4 –
3 Complete Kit
3.1. MP511 pH/mV/Temp Meter 3.2. 602 flexible electrode holder 3.3. 201T-F
pH/ATC three-in-one combination electrode 3.4. pH standard buffer solution(pH
4.00, 7.00, 10.01/50ml) 3.5. 9V power adapter 3.6. RS232 data communication
cable 3.7. MP500 PC-Link software 3.8. Operation manual 3.9. Quick manual
4 The Meter
4.1 LCD Display
1 unit 1 pc 1 pc 1 set 1 pc 1 pc 1 pc 1 pc 1 pc
— Measurement mode icon — Measurement value — Time and date — Units of
measurement
– 5 –
— Temperature compensation icon: ATC — automatic temperature compensation; MTC
— manual temperature compensation
— Serial number and icon of data storage and recall M+ — icon for measurement
storage; RM — icon for reading recalls; Numbers on the left is serial number.
— emperature measurement and unit — RS232 communication icon When this icon
appears, the meter has been
connected to a computer or laptop. — Timing measuring icon — Electrode
calibration indication icon — Stability icon of readings
4.2 Keypad
Keypad operations: Short press– <1.5 seconds; Long press — >2seconds.
Keypad
ON OFF
Keypad operations and descriptions
Operations Short press
Descriptions
Turn on/off the power. Attention: Only under measuring mode can the meter be
turned off. If under calibration or parameter setting mode, the operation will
be invalid. You need to press
– 6 –
CAL MODE
UNIT
ENTER
M+/RM
Short press Short press Long Press
Short press
Short press
Short press Long press
Short press Long Press
The meter will go to calibration mode. Short pressing again starts calibration.
To select the parameters:
Enter into parameter setting mode P1. Once in this mode, short press to enter P2, P3,P4 and so on.
In pH measuring mode: press the button to change the
resolution repeatedly: 0.010.1pH
In mV measuring modepress the button to change
parameter mode:
In parameter setting mode: press the button to move the
number position.
In calibration mode: press the button to confirm 1-point or
2-point calibration and then return to measuring mode.
In parameter setting mode: press the button to confirm
parameter setting and then return to measuring mode.
In (RM) mode: press button to return to measuring mode.
In MTC mode: press the button to increase or decrease the temperature. The
temperature will be altered quickly with long press.
In parameter setting mode: press the button to changer numbers or ON/OFF
state.
In (RM) mode: press the button to alter the storage serial number. Long press
to alter quickly.
Short press to store measurements; long press to recall measurements.
4.3 Store, recall and clear readings 4.3.1 Store
(a) In measuring mode, once the measuring value is stable and the
appears, Short
press < M+/RM > key, LCD will display “M+” icon and storage serial number. These measurements will be stored in memory. The meter can store up to 600 groups of measurement values; 300 groups in each of the pH and mV measuring modes.
(b) In timing measuring mode,
icon will appear on the LCD screen. The meter will
continuously conduct measurements according to the set parameter (eg. Every 7 sec) and
stores the data simultaneously.
– 7 –
4.3.2 Recall measuring information: (a) In measurement mode, long press < M+/RM > key, the meter will recall the previous stored information. The storage serial number, RM icon and the complete measuring information will appear on the lower right corner of the LCD screen; measuring time and date will be displayed on the upper right corner. Then press < > or < > key and the measured information will be recalled accordingly. Long press < > or < > key to rapidly check the measuring information under other serial number.
(b) In the recalling mode, press < ENTER > key to return to the measuring mode.
4.3.3 Clear stored value In the recalling mode, long press < M+RM > key for 5s, for 2s. It means the internal storage has been cleared.
icon will appear on the LCD
4.4 Sockets:
REF — Reference electrode socket pH/mV — pH and ORP electrode socket (BNC
socket) TEMP — Temperature electrode socket (RCA socket) RS232 — RS232
communication connector socket DC9V — DC9V power socket, 2.5, inner “+” outer
“-”
5 pH Measurement
5.1 Preparation: 5.1.1 Install the electrode holder The electrode holder is
composed of a metal stand base, and a flexible electrode arm and holder.
Simply insert the arm and holder into the metal stand base. Adjust the
flexible arm and holder according to your needs.
– 8 –
5.1.2 Connect the power adapter Plug the power adapter tightly into the DC9V
socket. The power’s voltage should meet the requirements listed on the power
adapter. 5.1.3 Things needed in addition to what’s included in the box A clean
cup, distilled water (8-16oz), and tissue papers for rinsing and drying the
probe.
5.2 pH Electrode Information 5.2.1 pH Electrode The meter comes with the
201T-F 3-in-1 combination electrode with a built-in temperature sensor, which
enables the automatic temperature compensation. This electrode is only
suitable for regular water solutions’ pH testing. Please refer to Section 9
for ideal pH electrodes to use for other specific applications. The
electrode’s housing adopts polycarbonate materials, which is resistant to
shocks and corrosion. A 3M KCL storage bottle comes with the electrode, which
is for storing the electrode when not in use to keep the sensitivity of the
probe. 5.2.2 Technical Specifications of the 201T-F pH Electrode Measurement
Range: 0 – 14 pH, 0 – 80°C (32 176°F) Junction: Single Ceramic Reference
Electrode: Ag/AgCl Connector: BNC Size: ø12*160 mm Temperature unit: 30K
Thermistor
5.2.3 Electrode Connectors The pH electrode has two connectors: the BNC
connector connects the pH probe; the RCA connector connects the temperature
sensor. Plug these two connectors into `pH/mV” and “TEMP” sockets. Please note
not to pull the cables in case of poor contact. Please keep the connectors
clean and dry. Refer to section 4.7 regarding how to properly maintain the pH
electrode.
5.2.4 Use the Electrode Screw off the KCL storage bottle, and put it aside (do
not dump or spill the KCL solution). Prepare a cup of distilled water, and
rinse the electrode in it for a few seconds. Gently shake the probe to remove
excess water, and dry it with clean tissue paper (DO NOT rub or wipe the
probe, just use paper to dap off excess water). Gently stir the electrode for
a few seconds
– 9 –
after it’s dipped into the test solution and then let it stand still. Wait for
the stable measurement (a smiley face appears and stays) and then record the
readings. When test is finished, place the electrode in the storage bottle and
tighten the cap in order to keep the sensitivity of the pH sensor.
5.3 Calibration: ON
5.3.1. Switch in power, press < OFF > key to turn on.
5.3.2. Short press < MODE > key to switch to
mode.
5.3.3. Insert the pH electrode (remove the storage cap) into the meter’s socket.
5.3.4. Press < CAL > to enter calibration mode, LCD flashes the 1st point calibration.
indicating to make
5.3.5. Rinse pH electrode in distilled water, allow it to dry and submerge it in pH7.00 buffer
solution. Stir the solution briefly and allow it to stay in the buffer solution until reading is
stable and
icon appears and stays, then press < CAL >, LCD flashes 7.00 pH,
calibration finishes after several seconds and then
flashes, indicating the 1st point
calibration has been finished and the 2nd point calibration begins.
5.3.6. Take out the pH electrode, rinse it in distilled water, allow it to dry and submerge it in
pH4.00 buffer solution. Stir the solution briefly and allow it to stay in the buffer solution until
reading is stable and
icon appears and stays, then press < CAL >, LCD flashes
4.00pH, calibration finishes after several seconds and then
flashes, indicating the
2nd point calibration has been finished and the 3rd point calibration begins.
5.3.7. Take out the pH electrode, rinse it in distilled water, allow it to dry and submerge it in pH 10.01 buffer solution. Stir the solution briefly and allow it to stay in the buffer solution until reading is stable and icon appears and stays, then press < CAL >, LCD flashes 10.01pH,calibration finishes while
displaying a stable pH and temperature value.
icon appears on the screen,
indicating 3 points calibration has been finished. See picture (4-1).
5.3.8. Notes (a) The meter can perform 1 point, 2 points and 3 points of automatic calibration. When the 1st point calibration finished, press < ENTER
key to confirm and enter measuring mode.
– 10 –
icon for one-point calibration will appear on the lower left corner of the LCD. When the measuring accuracy is ±0.1pH, choose one kind of buffer solution according to the measuring range and then proceed with one-point calibration.
(b) When the 2nd calibration finishes, press < ENTER > key to confirm 2 points calibration
and enter into measuring mode. Icons or
for 2 points calibration will appear on
the lower left corner of the LCD. If measurement is within acid range, you can choose pH4.00
and pH7.00 for calibration. If the measurement is within alkalinity range, you can choose
pH7.00 and pH10.01 for calibration.
(c) How often to calibrate The frequency that you need to calibrate your meter
depends on the tested samples, condition of electrodes, and the requirement of
the accuracy. For High-Accuracy meaustements ( ±0.02pH), the meter should be
calibrated before test every time; For ordinary-accuracy measurements
(±0.1pH), once calibrated, the meter can be used for about a week or longer.
In the following cases, the meter must be re-calibrated: a) The electrode
hasn’t been used for a long time or a new electrode is connected. b) After
measuring strong acid (pH<2) or strong base (pH>12) solutions. c) After
measuring fluoride-containing solution and strong organic solution d) There is
a significant difference between the temperature of the test sample and the
temperature of the buffer solution that is used in the last calibration.
5.4 Sample Test Rinse pH electrode in distilled water, allow it to dry, and
submerge it in sample solution. Stir the solutionbriefly and allow it to stay
in the sample solution until the stable value and icon appears and stays on
LCD, get the reading which is pH value of sample solution. Please note that
the closer the temperature of the sample solution to the calibration
solution’s is, the more accurate readings will become.
5.4.1 Notes
-
This default 201T-F electrode will NOT give accurate and stable pH readings when testing purified water such as distilled or deionized water. This because those water do not have enough ions present for the electrode to function properly. To measure distilled or deionized water’s pH, users need to use a specialized pH electrode with the meter such as LabSenTM 803 Pure Water pH Electrode. Contact us info@aperainst.de for more details.
– 11 – -
When testing purified water like spring water or drinking water, it will take longer for the readings to get stabilized (typically 3-5 minutes) because there is very few ions left to be detected by the sensor in those purified water.
-
Purified water such as distilled water and deionized water are recommended for only rinsing the probe for the best result. Never store the electrode in distilled or deionized water, for it could cause permanent damage to the electrode.
-
The 201T-F pH Electrode is NOT suitable for testing high-temperature samples (>176°F). Testing high-temperature samples could cause permanent damage to the electrode. A specialized electrode such as LabSenTM 213 (up to 225°F) should be used in this situation.
-
The electrode housing is polycarbonate. When using cleaning solutions, take cautions on
carbon tetrachloride, trichlorethylene, tetrahydrofuran, acetone, etc., which will dissolve
the housing and invalidate the electrode.
5.5 Parameter Setting 5.5.1 pH measuring parameter setting
Chart (4-1)
Mode
Parameter Settings
Code
P1
pH buffer solution series selection
P2
Time setting for timing measuring
P3
Unit of temperature
P4
Date setting
Date
P5
Time setting
Time
P6
Restore factory setting
Parameters USA (Europe & U.S.A series)
NIS (NIST series) CH (China series)
0-99 min
°C – °F Month / Day / Year
Hour / Minute OFF – On
5.5.2. Select pH buffer solution standard (P1) with (a) Long press < MODE >
key to enter into P1 mode, see picture (4-2). (b) Press <> or <> key to choose
buffer solution series: (Europe & U.S.A series) — 1.68, 4.00, 7.00, 10.01 and
12.45 pH (NIST series) — 1.68, 4.01, 6.86, 9.18 and 12.45 pH (Chinese series)
— 1.68, 4.00, 6.86, 9.18 and 12.46 pH (c) Press < MODE > key to enter into
next parameter setting or press < ENTER > key to conform and return to
measuring mode.
– 12 –
5.5.3. Time setting for timing measurement (P2)
(a) Short press
(b) Press < UNIT > key, the ” ” will move to the right and flash. Press <> or
<> key to change when the number is flashing.(c) Press < MODE > key to enter into next parameter setting or press < ENTER > key to
conform and return to measuring mode.
(d) Factory setting is 0 second.
Notein the form of ”
“, on the left side of “” is minute, and its maximum setting
is 99; on the right side of “” is second, its maximum setting is 59. After set timing
measuring mode, icon will appear on the LCD screen.
5.5.4. Temperature unit °C/°F setting (P3)
(a) Short press
(b) Press <> or <> key to choose temperature unit: °C or °F.
(c) Press < MODE > key to enter into next parameter setting or press < ENTER > key to
conform and return to measuring mode.
5.5.5. Date setting (P4) (a) Short press
– 13 –
5.5.6. Time setting (P5) (a) Short press
key to enter into next parameter setting or press < ENTER > key to conform and return to measuring mode.
5.5.7. Restore to factory setting (P6) (a) Short presskey in mode P5 to enter into mode P6, see picture (4-7)
(b) Press <> key to choose “On”, indicating parameter setting has been restored
to the factory setting state. It will then return to measuring mode after 2s.
5.6 Notes 1) After immerging the pH combination electrode into the solution, please stir the solution briefly in order to remove air bubbles and obtain a faster response and stable measurement value. 2) If a temperature electrode is not present, users can press < > or < > key to perform manual temperature compensation. 3) There is a protection bottle surrounding the tip of the pH electrode. This contains the KCL soaking solution. The tip of the electrode should be immerged in the solution to keep the glass bulb junction’s activation. Unscrew the cap, pull out the electrode and rinse it in distilled water before measuring. Once done testing, put in the electrode in the soaking bottle and screw the cap tightly to prevent solution leakage. If the soaking solution is turbid or moldy, please replace the solution and clean the probe immediately. 4) Users should prevent the electrode from being soaked into distilled water, protein solution and acid fluoride solution and from contacting with organic solutions. 5) In order to improve the accuracy, the pH value of the buffer solution should be known and
– 14 –
reliable. Buffer solution should be changed after 10-15 times of calibrating. 6) Keep the meter clean and dry; especially the electrode and the socket of the electrode. Failure to do so may lead to an inaccurate measurement or invalidity. Clean stains with medical cotton and absolute alcohol and blow-dry afterwards. 7) The sensitive glass bulb in front of the combination electrode should not come in contact with hard surfaces (they might get broken). Scratches or cracks on the electrode will cause inaccurate readings. Before and after each measurement, rinse the electrode with distilled water and then throw off excess water on the electrode. Do not rub the glass bulb with a tissue for it will affect the stability of the electrode potential and increase the response time. The electrode should be thoroughly cleaned if there are residuals on the electrode. Use soap water and a soft brush if necessary. 8) The life span of pH electrode is about 1-2 years. This will be shortened if used in extreme condition or improperly maintained. The aged and invalid electrode should be replaced in time. 9) If the calibration or display of the meter occurs abnormal, please set P6 to be “On” to restore to factory setting and perform calibration again. 10) Please don’t disconnect the power plug until the meter is turned off. Disconnecting power when meter is still on will cause damage to the meter.
5.7 Self-diagnostic Information
In the process of using, some icons might appear. This is the self-diagnosis
for the meter, which will help to understand the problems of the meter or
electrode during usage:
5.7.1. The stationary”- 2.00 pH” or ” 19.99 pH ” icon — this means the measurements has exceeded the measuring range. If the meter is not properly connected with the electrode or the electrode is not immerged into the solution, such icons will also appear.
5.7.2.
— Electrode zero potential is exceeded (<-60mV or >60mV)
5.7.3.
— Electrode slope is exceeded (< 85% or >105%)
5.7.4. When
or
appears on the upper right corner, the meter will not work
normally, please check as following:
(a) Check the electrode bulb and see if there are any air bubbles. If any are present, shake
the electrode with your wrist unitl the bubbles have been removed. Be sure to have a firm
grip when doing so.
(b) Check the pH buffer solution and see if it goes bad or has bigger error.
(c) Restore the meter to factory setting mode, then recalibrate it. If the sensor doesn’t work
after the above checks, please replace the pH electrode.
– 15 –
6 mV and ORP Measurement
6.1 Sample Test
6.1.1. Press <
ON OFF
< UNIT > key to chooseto turn on the meter, press < MODE > to switch to
or ;
, then press
(a)
— This is the electrode potential measuring mode. mV value and the
temperature will be displayed at the same time;
(b)
— This is the ORP (redox) measurement mode. ORP measurement has no
temperature compensation, so there is no temperature appears in this mode;
Note: ORP is the abbreviation of “Oxidation-Reduction Potential”, representing the
Oxidation-Reduction Potential of the solution, ORP is the measuring parameter for the
oxidation reduction potential of the aqueous solution. Use mV as its unit.
6.1.2. Connect ORP electrode (sold separately) or ion electrode, immerge it into the sample
solution, stir briefly and then set it still. When smiling
icon appears and stays, the
reading is the ORP value or the potential value of the ion electrode.
6.2 Notes
6.2.1. ORP measurement does not require calibration. When the user is not sure
about ORP electrode quality or measuring value, use ORP standard solution to
test mV value and see whether ORP electrode or meter works properly. 6.2.2.
Clean and activate ORP electrode: After the electrode has been used over a
long period of time, the platinum surface will get polluted which causes
inaccurate measurement and slow response. Please refer to the following
methods to clean and activate ORP electrode: (a) For inorganic pollutant,
submerge the electrode in 0.1mol/L dilute hydrochloric acid for 30 minutes,
then wash it in distilled water, then submerge it in the soaking solution for
6 hours. (b) For organic or lipid pollutant, clean the platinum surface with
detergent, then wash it in distilled water, then submerge it in the soaking
solution for 6 hours. (c) For heavily polluted platinum surface on which there
is oxidation film, polish the platinum surface with toothpaste, then wash it
in distilled water, then submerge it in the soaking solution for 6 hours.
– 16 –
6.3 Parameter Setting
6.3.1. mV and ORP measurement parameter setting (chart (5-1))
Chart(5-1)
Mode
Contents
Code
P1
Time set for timing measuring
P2
Restore to factory setting
Parameters 0 to 99min OFF-On
6.3.2. Time set for timing measuring (P1): Please refer to 6.5.6. 6.3.3. Restore to factory setting (P2): Please refer to 6.5.7
7 RS232 Communication
7.1 System requirements
This meter uses “MP500 PC-Link” communication software for RS232
communication. This software requires the computer to meet such requirement:
Personal computer (Microsoft Excel 2000 or the version of higher rank) which
can operate Windows-based operation system, PCIBM compatible with XT and CD-
ROM driver, RS232 communication port. 7.2 Software interface Software
interface: refer to Picture (6-1).
— Meter Serial number
Picture (6-1) – 17 –
— Measurements and time curve — Stored value display area — Keys Clear — Press
the key to clear data Export — Press this key to export the stored value to
Microsoft Excel file Exit — Press this key, PC-Link program exits from the
computer interface Select CommID — Press the key to download the data from the
meter and upload it to PC — Parameters settingStandard of the buffer solution,
time setting etc. — Measurements display screen 7.3 Load software Put the
software disk into the computer and install the software as per the following
procedures: Open “PC-Link” file Double click “Setup” program Click “OK”Click
the icon (refer to Picture (6-2) ) Click “Continue” Click “Enter”.
Picture(6-2)
7.4 Port Connection Connect the meter and PC with RS232 cable and open the MP
500 PC Link Program, the PC will enter into program interface and then click”
Select CommID”. The default port of the PC is port#1, Icon will appear on the
lower left corner of the LCD screen. NOTE: If computer does not have a RS232
port for connecting, the use can connect to the computer via USB using a RS232
/ USB converter (Apera Instruments does not provide). 7.5 Run Software 7.5.1.
Upload the stored value
– 18 –
When the meter is connected with PC and icon appears on the LCD screen, the
data stored in the meter will be automatically uploaded to the PC. This
program will sort the pH and mV measurement and show the data in category.
7.5.2. Storage during operation When the program is running, press <M+/RM> key
to store data. All the measurement value will be uploaded to the PC through RS
232 and will not be stored in the meter. Storage data during operation will be
the same with which displays on the meter. If timing measuring mode is set,
the time curve will be showed on the interface as well as the measurements.
7.5.3. Data processing Press “Export” key to export the stored value to
Microsoft Excel file and then analyze or print the stored data.
8 Recommended pH Electrodes for Specific Applications
Application
Ideal Apera pH Electrodes to Use with MP511 Meter
Regular water solutions
201T-F, LabSen 213
Beverage, beer, or wine analysis
LabSen 213
Cosmetics Dairy products (milk, cream, yogurt, mayo, etc.) High-Temperature liquid
LabSen 851-1, (MP500 temp. probe required) LabSen 823
LabSen 213
Low-temperature liquid Meat
LabSen881 (MP500 temp. probe required) LabSen 763
Micro sample testing
LabSen 241-6, LabSen 241-3 (MP500 temp. probe required)
Purified Water (Low ion concentration LabSen 803, LabSen 813
samples)
Soil
LabSen 553
Solid or semi-solid samples (cheese, rice, fruit, etc.) Strong acid samples Strong alkalined samples Surface test (skin, paper, carpet, etc.) Titration TRIS buffer solutions
LabSen 753
LabSen 831 (MP500 temp. probe required) LabSen 841 (MP500 temp. probe
required) LabSen 371 (MP500 temp. probe required) LabSen 223 LabSen 213,
LabSen 223
Viscous liquid samples
LabSen 223, LabSen851-1
Wastewater or emulsion
LabSen 333
– 19 –
9 Warranty We warrant this instrument to be free from defects in material and
workmanship and agree to repair or replace free of charge, at option of APERA
INSTRUMENTS(Europe) GmbH, any malfunctioned or damaged product attributable to
responsibility of APERA INSTRUMENTS(Europe) GmbH for a period of TWO YEARS
(SIX MONTHS for the probe) from the delivery. This limited warranty does not
cover any damages due to: Transportation, storage, improper use, failure to
follow the product instructions or to perform any preventive maintenance,
modifications, combination or use with any products, materials, processes,
systems or other matter not provided or authorized in writing by us,
unauthorized repair, normal wear and tear, or external causes such as
accidents, abuse, or other actions or events beyond our reasonable control.
APERA INSTRUMENTS (Europe) GmbH Wilhelm-Muthmann-Straße 18 42329 Wuppertal,
Germany Email: info@aperainst.de Website: www.aperainst.de Tel. +49 202
51988998
– 20 –
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>