APERA INSTRUMENTS LabSen 241-6 Semi-Micro pH Electrode User Manual
- June 3, 2024
- APERA INSTRUMENTS
Table of Contents
APERA INSTRUMENTS LabSen 241-6 Semi-Micro pH Electrode
LabSen electrochemical sensors are premium pH electrodes with manufacturing technology and key components imported from Switzerland. been 241-6 semi-micro pH electrode is suitable for micro samples (≥0.2ml) in small containers, such as tubes and vials. This probe has the following features:
- Impact-resist membrane, there is no danger of electrode breakage during normal use.
- Blue gel inner solution does not flow and will not cause a bubble.
- Long life reference system, has better stability and service life.
Technical Data
Measuring Range | (0 ~ 14) pH | Electrolyte | 3M KCl |
---|---|---|---|
Temperature Range | (0 ~ 100) °C | Soaking Solution | 3M KCl |
Membrane Type | S | Membrane Resistance | <350MΩ |
Shaft Material | Lead-free Glass | Electrode Dimension | Electrode (Ø12×150) mm |
Measuring Tip (Ø6×100) mm
Reference| Long Life| Connector| BNC
Junction| Ceramic| Cable| Ø3×1m
Usage and Maintenance
-
Connect BNC plug to the input on the pH meter.
-
When measuring, please unscrew the bottle cap, pull out the electrode and rinse it with deionized water. After using, please put the electrode back into the bottle and screw tight the cap.
Before pulling out or putting back the electrode, fully loosen the blue cap on the storage sleeve so that the electrode can move in and out smoothly. -
Prior to measurement, remove the rubber plug to maintain the pressure of the reference solution, and keep a consistent flow rate of reference solution and stable potentials of the junction.
-
After a period of usage, the reference solution will run low. Whenever the level falls to 1/2 height of the electrode, add 3M KCL solution to the refilling hole by using a syringe or pipette.
-
The connector of the electrode should keep clean and dry. If being contaminated, please clean it with medical cotton and absolute alcohol and blow dry to prevent a short circuit of the electrode and slow reaction of the electrode.
-
The electrode’s measuring tip should be soaked in the soaking bottle containing a certain amount of storage solution to keep the membrane hydrated and the junction unblocked. Clean the bottle and replace the storage solution if the storage solution gets turbid and mildewed. The electrode should never be sunk in pure water or buffer solution for long.
-
After 1-year of use, we recommend replacing the electrode to ensure the best accuracy.
Warranty
We warrant this electrode free from defects in material and workmanship and agrees to repair or replace free of charge, at option of APERA INSTRUMENTS any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS for a period of six months. Warranty period is the time limit to provide free service for the products purchased by customers, not the service life of the tester or electrodes. This limited warranty does not cover any damages due to: transportation; storage; improper use; failure to follow the product instructions or to perform any preventive maintenance; modifications; combination or use with any products, materials, processes, systems or other matter not provided or authorized in writing by us; unauthorized repair; normal wear and tear; or external causes such as accidents, abuse, or other actions or events beyond our reasonable control.
APERA INSTRUMENTS (Europe) GmbH
Address: Wilhelm-Muthmann-Str.18,
42329 Wuppertal, Germany
Phone: +49 202 51988998
Website: www.aperainst.de
Email: info@aperainst.de
References
- Apera Instruments® Meters and Sensors for pH, Conductivity, TDS, Salinity, ORP, DO, Turbidity, ISE
- Apera Instruments
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>