APERA INSTRUMENTS DJS-0.1-F Conductivity-Temperature Electrode Instruction Manual
- June 3, 2024
- APERA INSTRUMENTS
Table of Contents
INSTRUMENTS DJS-0.1-F Conductivity-Temperature Electrode
Instruction Manual
DJS-0.1-F Conductivity-Temperature Electrode
Model| Cell
constant| Electrode size| Dimension size| Socket| Temp. Sensor| Electrode
body| Cable length| Range
---|---|---|---|---|---|---|---|---
DJS-0.1-F| 0.1| 7x18mm platinum| 012×155| BNC/RCA| 30K| Glass| 1m| (0-200)
pS/cm
Use and maintenance:
1.1 Connect the DJS-0.1-F conductivity electrode to your conductivity meter.
Set the cell constant to 0.1. Then calibrate the meter in the lower
calibration point (e.g. 84 µS/cm).
1.2 Keep the conductivity electrode clean, before and after the measurement,
rinse the electrode with deionized water/distilled water. In order to maintain
the highest measurement accuracy, we recommend rinsing the electrode several
times with your test sample solution.
1.3 A Flow cell (included) is recommended when testing ultra-low conductivity
solutions like ultrapure water. It helps reduce the influence from contact
with air and minimizes reading error. Use a rubber tube (7mm in diameter, not
included) to connect your water source on the low end of the flow cell. When
water flows through the glass cell, conductivity can be measured without
air.
1.4 How to clean a conductivity electrode:
(a) Use warm detergent water or alcohol to clean off organic contaminants on
the electrode.
(b) Use 10% lemon acid to clean calcium and magnesium sediments.
1.5 Conductivity electrode storage: Store dry when not in use.
Limited Warranty
We warrant this electrode to be free from defects in material and workmanship and agrees to repair or replace free of charge, at option of APERA INSTRUMENTS, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS for a period of six months. Warranty period is the time limit to provide free service for the products purchased by customers, not the service life of the tester or electrodes.
This limited warranty does not cover any damages due to:
- transportation;
- storage;
- improper use;
- failure to follow the product instructions or to perform any preventive maintenance;
- combination or use with any products, materials, processes, systems or other matter notprovided or authorized in writing by us;
- unauthorized repair;
- normal wear and tear; or
- external causes such as accidents, abuse, or other actions or events beyond our reasonable control.
APERA INSTRUMENTS (Europe) GmbH
Wilhelm-Muthmann-Str.18, 42329 Wuppertal, Germany
info@aperainst.de |
www.aperainst.de
Tel.: +49 202 51988998
Documents / Resources
|
APERA INSTRUMENTS DJS-0.1-F Conductivity-Temperature
Electrode
[pdf] Instruction Manual
DJS-0.1-F Conductivity-Temperature Electrode, DJS-0.1-F, Conductivity-
Temperature Electrode, Temperature Electrode, Electrode
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>