APERA INSTRUMENTS LabSen 333 Plastic Premium pH Electrode User Manual
LabSen 333 Plastic Premium pH Electrode User Manual
LabSen electrochemical sensors are premium pH electrode with manufacturing
technology and key components imported from Switzerland. LabSen333 plastic
premium pH adopts
Polymer solid electrolyte, pore without diaphragm. It is suitable for
emulsion, suspension, effluent, samples containing protein and sulfide.
This probe has following features:
⚫ Anti-corrosion POM (Polyoxymethylene) shaft, have good mechanical
performance.
⚫ Blue gel inner solution, does not flow and will not cause bubble.
⚫ Intergrated ATC probe
1 Technical Data
Measuring Range | (0 ~ 14) pH | Electrolyte | Polymer |
---|---|---|---|
Temperature Range | (0 ~ 80) °C | Soaking Solution | 3M KCl |
Shaft Material | POM (Polyoxymethylene) | Temp. Sensor | NTC 30kΩ |
Membrane Type | S | Electrode Dimension | (Ø12×120) mm |
Membrane
Resistance
| <250MΩ| Connector| BNC/RCA
Reference| Long Life| Cable| Ø5×1m
Junction| Pore without diaphragm| |
2 Usage and Maintenance
2.1 When measuring, please unscrew the bottle cap, pull out the electrode and
rinse it with deionized water. After using, please put the electrode back into
the bottle and screw tight the cap.
2.2 Polymer reference electrode is a kind of high quality reference electrode
with the feature of nondiaphragm, hard to be contaminated, compression
resistance and resistance to organic solution. It is almost suitable for all
pH measurement solution such as solution containing fat, oils, sulfide and
protein, effluent, weak ion solution, emulsion and suspension.
2.3 The connector of the electrode should be kept clean and dry. If being
contaminated, please clean it with medical cotton and absolute alcohol and
blow dry to prevent the short circuit of the
electrode and slow reaction of electrode.
2.4 The electrode measuring tip should be soaked in the soaking bottle
containing certain amount of storage solution to keep the membrane hydrated
and junction unblocked. Clean the bottle and replace the storage solution if
the storage solution gets turbid and mildewed. The electrode should never be
sinked in pure water or buffer solution for long.
2.5 After long time exposure in air, Polymer electrolyte will cause shrinkage.
It will not happen in measurement status.
The electrode must be soaked in 3M KCl solution for storage. After long time
soaking, there will be some floccules dissolving out from 3M KCl solution,
this is normal phenomenon.
2.6 After 1-year of use, we recommend replacing the electrode to ensure the
best accuracy.
3 Warranty
We warrant this electrode free from defects in material and workmanship and
agrees to repair or replace free of charge, at option of APERA INSTRUMENTS any
malfunctioned or damaged product
attributable to responsibility of APERA INSTRUMENTS for a period of six
months. Warranty period is the time limit to provide free service for the
products purchased by customers, not the service life of the tester or
electrodes.
This limited warranty does not cover any damages due to: transportation;
storage; improper use; failure to follow the product instructions or to
perform any preventive maintenance; modifications; combination or use with any
products, materials, processes, systems or other matter not provided or
authorized in writing by us; unauthorized repair; normal wear and tear; or
external causes such as accidents, abuse, or other actions or events beyond
our reasonable control.
APERA INSTRUMENTS (Europe) GmbH
Address: Wilhelm-Muthmann-Str.18,
42329 Wuppertal, Germany
Phone: +49 202 51988998
Website: www.aperainst.de
Email: info@aperainst.de
Read More About This Manual & Download PDF:
References
- Apera Instruments® Meters and Sensors for pH, Conductivity, TDS, Salinity, ORP, DO, Turbidity, ISE
- Apera Instruments
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>