KNOP LS901 Light Unit User Manual

June 3, 2024
KNOP

KNOP logo LS901 Light Unit V3.1
User Manual
KNOP LS901 Light Unit

Warning and safety instructions:

  • The battery in this product complies with the relevant IEC safety standards for batteries.
  • Please read the safety instructions on the power supply before use.
  • The power supply unit must meet the relevant IEC safety standards.
  • The battery must not be replaced, and the product must be sent for repair.
  • The product must not be disassembled. Risk of electric shock!
  • Read the intended use before putting into service.

Intended use

LS901 is a light source that can help minimize fall injuries. For example, it can switch on the light when a person leaves the bed or when he or she presses the wristband transmitter.
LS901 has a built-in backup battery, so it will work even in the event of a power failure.
The product is designed to work together with other products in the KNOP 901 series.
It is possible to program up to five different transmitters to activate the light and how long the light should stay on.

Commissioning

LS901 only needs to be connected to a power outlet. There are no additional connection options.
It switches on automatically when the power is turned on.
We recommend that the product is updated with the latest software version to ensure that all functions work as described.

The LED

The LED shows the status of the product.

Green: Everything is OK.
Flashing green: Encoding active.
Red: Battery operation.

Programming

To program the device, simultaneously press until the LED flashes green.
On the left side, there is a dim light where there is already a coded transmitter. The selected memory location is displayed with flashing. If an already encoded memory  location is selected, it is overwritten.
On the right side, the time the light is on for the current memory location is displayed.
Choose between 30 seconds (lower part), 2, 5, 10, and 30 minutes (upper part).

  • First, select the memory you want to use by pressing .
  • Then activate the transmitter to be encoded.
  • Then select with button for how long the light should stay on.

If more codings are desired, this can be done afterward by repeating the procedure.
Programming is automatically terminated if the buttons are not pressed again after one minute.
Alternatively, both buttons can be held down for 3 seconds.

Deletion

It is possible to restart the product by pressing and holding until the LED goes out – about 2030 seconds.
If the product is unplugged, it switches off afterward.

Turn off

To switch off the product, unplug it from the wall socket. Then press and hold until the LED flashes red. You can switch the product on again by plugging it into the main socket.

Maintenance

Update of LS901
It is possible to update the software of LS901. For further description please refer to KNOPtool.

Check the coverage area

One person activates the product at short intervals, while another person systematically walks around the area and marks on a sketch the building and the area where there is coverage. The map is placed in the office so that anyone who needs to use the system can see where there is coverage. Outside the coverage area, alarms cannot be received.
Contact your retailer for more information.

Cleaning

The product can be cleaned with a moist cloth or a wet wipe.

Spare parts and accessories

Spare parts and accessories can be ordered from Knop Elektronik A/S at www.knop.dk.

Technical data

Frequency HF: 869.2125 MHz
Range: Up to 1500 meters between LS901 and TX901 (¹)
Power supply: 230VAC
Usage: <1W (st.by) max 5W
Plug: 2-pole EURO
Battery backup: Approx. 15 hours without light, about 3 hours with light.
Charging time: Approx. 1-day
Surrounding environment: Indoor use. ≤ 90% non-condensing
Ambient temperature: 0°C to +40°C
Cabinet type: Color RAL9001 ABS
Cabinet dimensions: W: mm, H: mm, D: mm.
Density: IP40
Weight: 135g

The right to make changes is reserved.
All rights reserved.
© KNOP Elektronik A/S
(¹) Measured outdoors with unobstructed view between transmitter and receiver. In buildings the range is reduced.

Explanation of symbols and approvals

| This product complies with:
Directive 2017/745/EU: MDR
Directive 1907/2006/EU: REACH
Directive 2011/65/EU: RoHS
ISO 14971:2019: Risk Management for Medical devices
EN 301 498-1 V2.2.3: Electro Magnetic Compatibility
EN 301 489-3 V3.1.1: Electro Magnetic Compatibility
EN 50130-4:2011 + 2014: Immunity alarm systems
EN 300 220-1 V3.1.1: Short Range Devices
EN 300 220-2 V3.1.1: Short Range Devices
EN 300 220-3 V2.1.1: Short Range Devices
EN 60601-1-2:2014 + 2015: Electromagnetic compatibility (Medical)
EN 62368-1:2020: Electrical safety
EN 50581:2012: Hazardous substances
---|---
| The product must not be used if the packaging is damaged
| Medical Device
Class 1
| Manufacturer
KNOP Elektronik A/S, Fabriksvej 20, DK-7600 Struer, Denmark
| Read the manual(s) before installation and commissioning.
| Interference may occur in an environment with equipment marked with this symbol.
| Must be protected against liquids.
| 0 °C to +40 °C, temperature limit for transport/storage and use.
| The product must not be disposed of with normal household waste.
| Single Registration Number
DK-MF-000025631
| Unique Device Identifier
05744002850427
| Product reference
LS901 Light Guide
| Serial number
Placed on the product
| Ingress Protection Code
IP40

© KNOP Elektronik A/S
Fabriksvej 20
DK-7600 Struer
knop@knop.dk
knop.dk
+45 97840444

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals