Warnings and safety instructions
- Read the intended use before using.
- The unit must only be connected to a limited power supply of 15 W.
- Avoid direct sunlight.
Intended use
POS901 is a transmitter designed to transmit positions to KNOP 901 series
products. These products can be set to different operating modes when a
position is received.
The POS901/POS901L requires the advanced settings of the product to be set
during installation and commissioning. Please check the Knoptool manual and
the POS901/POS901L installation manual.
POS901/POS901L requires no subsequent operation or adjustment.
POS901 has a built-in antenna and POS901L is used for a loop antenna, e.g.
outdoors under a pavement.
Indicator
The product’s LED can show the following:
- Lights green when POS901/POS901L is running under normal conditions.
- Flashes green when POS901/POS901L is being calibrated.
- Lights yellow when POS901/POS901L has reduced power, e.g. too hot.
- Flashes red when POS901/POS901L is faulty – contact your retailer or Knop Elektronik.
- Flashes rapidly red when POS901/POS901L is connected to KnopTool.
Note that the indicator can be turned off in advanced settings (see the KnopTool manual), but not in case of error.
Spare parts and accessories
Spare parts and accessories can be ordered from Knop Elektronik A/S at
www.knop.dk.
Cleaning
The product can be cleaned with a moist cloth or a wet wipe.
Technical data
Frequency HF: | 869.2125 MHz |
---|---|
Range: | Up to 1500m of unobstructed view for an RX901B. |
Frequency LF/Loop: | 18kHz +/- 1kHz |
Range: | Up to 240 cm depending on wall conditions. |
External power supply: | 10-24VDC (supplied with max. 15W) |
Power consumption, typical: | 150 mA |
Surrounding environment: | Indoor use. < 90% humidity without condensation. |
Ambient temperature: | 0-40° C |
Cabinet type: | White RAL9010 ABS |
Cabinet dimensions: | W: 27mm, H: 237mm, D: 21mm. |
Density: | IP20 |
Connections: | External floating inputs, outputs, and power supply. |
External rod or loop.
Weight:| POS901: 150g
POS901L: 80g
The right to make changes is reserved.
(¹) Measured outdoors with an unobstructed view between transmitter and
receiver. In buildings, the range is reduced.
Explanation of symbols and approvals
| This product complies with:
Directive 2017/745/EU: MDR
Directive 1907/2006/EU: REACH
Directive 2011/65/EU:RoHS
ISO 14971:2019: Risk Management for Medical devices
EN 301 498-1 V2.2.3: Electro Magnetic Compatibility
EN 301 489-3 V3.1.1: Electro Magnetic Compatibility
EN 50130-4:2011 + 2014: Immunity alarm systems
EN 300 220-1 V3.1.1: Short Range Devices
EN 300 220-2 V3.1.1: Short Range Devices
EN 300 220-3 V2.1.1: Short Range Devices
EN 60601-1-2:2014 + 2015: Electromagnetic compatibility (Medical)
EN 62368-1:2020: Electrical safety
EN 50581:2012: Hazardous substances
---|---
| The product must not be used if the packaging is damaged.
| Medical Device
Klass 1
| Manufacturer
KNOP Elektronik A/S, Fabriksvej 20, DK-7600 Struer, Denmark
| Read the manual(s) before installation and commissioning.
| Interference may occur in an environment with equipment marked with this
symbol.
| Can be protected against liquids before installation.
| 0 °C to +40 °C, temperature limit for transport/storage and use.
| The product must not be disposed of with normal household waste.
| Single registration number
DK-MF-000025631
| Unique Device Identifier
05744002850441
05744002853442
05744002855446
| Product reference
POS901, Position Rod
POS901L, Position Loop
POS901S, Position Slave
| Serial number
Placed on the product
| Ingress Protection Code
IP20