Eickemeyer EICKWARM H₂O Warming Unit For Small Animals User Guide

September 17, 2024
Eickemeyer

EICKWARM H₂O Warming Unit For Small Animals

Specifications

  • Product Name: EICKWARM HO
  • Product Type: Warming Unit for Small Animals
  • Art. No.: 648020

Product Usage Instructions

Device Overview

The EICKWARM HO warming unit features front and back components:

  • Front: Light strip, Power button (On / Off), Timer mode
  • Back: Water inlet, Temperature display, Rubber Stopper, Hose Connection, Mains cable

Special Settings

Setting the Timer: Follow the instructions in the manual to set the timer as needed.
Setting the Childproof Lock: Instructions
for setting the childproof lock are included in the manual.

Spare Parts
For replacement parts, the following are available:

  • Art. No. E64802001 – Warming Mat, small, L 55 x W 55 cm
  • Art. No. E64802002 – Warming Mat, large, L 55 x W 100 cm

Frequently Asked Questions (FAQ)
Q: Can this warming unit be used for animals other than small  animals?
A: The EICKWARM HO is designed specifically for small animals and should only be used as intended.

EICKWARM H₂O
WARMING UNIT FOR SMALL ANIMALS

DEVICE

FrontEickemeye 0EICKWARM-H₂O -Warming-Unit-For-Small-Animals-
\(2\)

Back Eickemeye 0EICKWARM-H₂O -Warming-Unit-For-Small-Animals-
\(3\)

 PREPARATION FOR USE

  1. Step 1: Place the device on a fl at surface
    • Ensure that there is a distance more than 5 cm between the device and the heating mat.
    • The device should be at the same height or slightly higher than the heating mat.
    • Do not place the device on an uneven surface (automatic switch-off may be triggered).
  2. Step 2: Connect the shut-off valves and the hose to the device
    • Connect the two shut-off valves to the device.
    • Connect the hoses of the heating mat to the shut-off valves.
    • Open the shut-off valves.
    • To change the mat, close the shut-off valves.
  3. Step 3: Connect the device to the power supply and press the power button
    • The „beep“ signals that the power is switched on.
    • During the initial setup, the error code „E1“ (water level too low) is displayed.
    • The water level indicator will fl ash accompanied with several „beeps“.
  4. Step 4: Fill the device with distilled water
    • Open the water inlet cover and fi ll with water until the water level indicator goes out.
    • The water level indicator may fl ash again when the water fl ows into the heating mat.
    • Keep adding water until the display turns off.
    • Close the lid when the device and the heating mat are full.

Only use distilled water!

OPERATION

Setting the temperature

  • Press the up and down arrow buttons to set the desired temperature.
  • The heating temperature can be set between 10 °C and 45 °C.
  • The current water temperature is shown on the device display. This value approaches the desired temperature when the device circulates the heated water.

Setting the timer mode

  • There are 3 different modes available.
  1. C1 – Normal timer mode
  2. C2 – Cooler timer mode
  3. C3 – Warmer timer mode
  • Press and hold the timer mode button for 2 seconds to tare between modes.
  • To exit timer mode, press one of the arrow buttons for the temperature setting.

Normal mode (8 h)

  • Maintains a temperature of 36 °C for the fi rst 30 minutes.
  • After 30 minutes, the temperature drops to 32 °C, which is maintained for 6 hours.
  • In the last 1.5 hours, the temperature returns to 36 °C.

Cooler mode (14.5 h)

  • Maintains a temperature of 32 °C for the fi rst 30 minutes.
  • After 30 minutes, the temperature drops to 30 °C, which is maintained for 6 hours.
  • After a further 60 minutes, the temperature drops to 28 °C, where it is maintained for 4 hours.
  • In the last 2.5 hours, the temperature rises back to 32 °C.

Warm mode (14 h)

  • Maintains a temperature of 38 °C for the fi rst 30 minutes.
  • After 30 minutes, the temperature drops to 36 °C, which is maintained for 6 hours.
  • After a further 60 minutes, the temperature drops to 34 °C, where it remains for 4 hours.
  • In the last 2.5 hours, the temperature rises back to 38 °C.

Automatic dimming mode

  • If no button is pressed for a period of 30 seconds, the light strip switches to dimming mode.
  • To change the device control again, simply press any button on the device.

SPECIAL SETTINGS

Setting the timer

  • The timer can be set from 1 to 24 hours. Once the preset time has elapsed, the device switches off automatically.
  • After pressing the timer button twice, the display fl ashes.
  • Set the desired time by pressing the up and down arrow buttons.
  •  As soon as the target time is displayed, the setting is active after 5 seconds and the display switches to the temperature display.
  • Press the timer button twice or switch the device off / on to deactivate the timer function.
  • The device does not switch off automatically if the timer is set to „00“.

Setting the childproof lock

  • The childproof lock is used to prevent unwanted operation.
  • With the device switched on, press the childproof lock button to switch the function on or off.
    • Display E5 LOCK ON
    • Display EC LOCK OFF
    • When lock mode is activated, the buttons are deactivated.

Switch between Fahrenheit (°F) and Celsius (°C)

  • Press and hold the switch-on button for 20 seconds until „D1“ (°C) or „D0“ (°F) is  displayed.
  • Restarting the system activates the desired value.

SPARE PARTS

Art. No. Description
E64802001 Warming Mat, small, L 55 x W 55 cm
E64802002 Warming Mat, large, L 55 x W 100 cm

For veterinary use only!
As with all warming devices patients must be checked regularly for pressure points and possible overheating.

GERMANY
EICKEMEYER KG
Eltastraße 8
78532 Tuttlingen
T +49 7461 96 580 0
F +49 7461 96 580 90
info@eickemeyer.de
www.eickemeyer.de

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals